UMN CDMS Wiki archive
start › cdms › plasticscintillator › data

SiPM & Plastic Scintillator R&D Data Summary

March 2018

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20180330 20180330_1600V_singlePE.dat channel 2: PMT R1332 @ -1600.V, 1mV scale, -1mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_1650V_singlePE.dat channel 2: PMT R1332 @ -1650.V, 1mV scale, -1mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_1700V_singlePE.dat channel 2: PMT R1332 @ -1700.V, 2mV scale, -1.25mV threshold, bare face. Scope was self-triggered, 20 nsec/div, 0 nsec offset. Measurement of R1332 PMT gain 10000
20180330 20180330_1750V_singlePE.dat channel 2: PMT R1332 @ -1750.V, 2mV scale, -1.25mV threshold, bare face. Scope was self-triggered, 20 nsec/div, 0 nsec offset. Measurement of R1332 PMT gain 10000
20180330 20180330_1800V_singlePE.dat channel 2: PMT R1332 @ -1800.V, 2mV scale, -1.25mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_1850V_singlePE.dat channel 2: PMT R1332 @ -1850.V, 5mV scale, -1.5mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_1900V_singlePE.dat channel 2: PMT R1332 @ -1900.V, 5mV scale, -1.5mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_1950V_singlePE.dat channel 2: PMT R1332 @ -1950.V, 5mV scale, -2mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_2000V_singlePE.dat channel 2: PMT R1332 @ -2000.V, 5mV scale, -6mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_2050V_singlePE.dat channel 2: PMT R1332 @ -2050.V, 10mV scale, -6.5mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000
20180330 20180330_2100V_singlePE.dat channel 2: PMT R1332 @ -2100.V, 10mV scale, -8mV threshold, bare face. Scope was self-triggered Measurement of R1332 PMT gain 10000

February 2018

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
trial2_cf252_1inPolyChannel_gdNewVialCenter_shield6_andTrig_60min.txt
20180222 Na-22 source placed between NaI detector B and C. Trigger: coincidence between Det. B and C. Channel 0: Detector B anode, +1101V bias, 0.5V scale, -0.25V offset, -15mV threshold. Channel 1: Detector C anode, +1099V bias, 0.5V scale, -0.25V offset, -15mV threshold.

January 2018

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20180209 180209_183653_na22_andTrig_1000mVrange_30min.txt, 180209_190653_na22_andTrig_1000mVrange_30min.txt, 180209_193653_na22_andTrig_1000mVrange_30min.txt, 180209_200653_na22_andTrig_1000mVrange_30min.txt, 180209_203653_na22_andTrig_1000mVrange_30min.txt, 180209_210653_na22_andTrig_1000mVrange_30min.txt, 180209_213653_na22_andTrig_1000mVrange_30min.txt, 180209_220653_na22_andTrig_1000mVrange_30min.txt, 180209_223653_na22_andTrig_1000mVrange_30min.txt, 180209_230653_na22_andTrig_1000mVrange_30min.txt Na-22 source placed between NaI detector B and C. Trigger: coincidence between Det. B and C. Channel 0: Detector B anode, +1100V bias, 1V scale, -0.5V offset, -30mV threshold. Channel 1: Detector C anode, +1100V bias, 1V scale, -0.5V offset, -30mV threshold. 10 Data Files, 30 mins each
20180209

180210_191716_na22_andTrig_500mVrange_30min.txt 180210_194716_na22_andTrig_500mVrange_30min.txt 180210_201716_na22_andTrig_500mVrange_30min.txt 180210_204716_na22_andTrig_500mVrange_30min.txt 180210_211716_na22_andTrig_500mVrange_30min.txt 180210_214717_na22_andTrig_500mVrange_30min.txt 180210_221717_na22_andTrig_500mVrange_30min.txt 180210_224717_na22_andTrig_500mVrange_30min.txt | Na-22 source placed between NaI detector B and C. Trigger: coincidence between Det. B and C. Channel 0: Detector B anode, +1100V bias, 500mV scale, -0.25V offset, -30mV threshold. Channel 1: Detector C anode, +1100V bias, 500mV scale, -0.25V offset, -30mV threshold. | | 10 Data Files, 30 mins each | | |

20180209

180211_204059_na22_andTrig_200mVrange_30min.txt 180211_211059_na22_andTrig_200mVrange_30min.txt 180211_214059_na22_andTrig_200mVrange_30min.txt 180211_221059_na22_andTrig_200mVrange_30min.txt 180211_224059_na22_andTrig_200mVrange_30min.txt 180211_231059_na22_andTrig_200mVrange_30min.txt 180211_234059_na22_andTrig_200mVrange_30min.txt 180212_001059_na22_andTrig_200mVrange_30min.txt 180212_004059_na22_andTrig_200mVrange_30min.txt | Na-22 source placed between NaI detector B and C. Trigger: coincidence between Det. B and C. Channel 0: Detector B anode, +1100V bias, 250mV scale, -0.10V offset, -30mV threshold. Channel 1: Detector C anode, +1100V bias, 250mV scale, -0.10V offset, -30mV threshold. | | 10 Data Files, 30 mins each | | |

February 2017

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
Cf-252 run. Cf-252 source positioned on top of scintillator slab. channel 0: PMT 14 @ -1250.V, 0.5V scale, -50.mV threshold, middle optical fiber (masked). channel 1: PMT 51 @ -1250V, 0.2V scale, -200.mV threshold, end optical fiber (masked). channel 3: trigger pulse from coincidence unit, 40ns coincidence window.
20170201 Cf-252 background run. channel 0: PMT 14 @ -1250.V, 0.5V scale, -50.mV threshold, middle optical fiber (masked). channel 1: PMT 51 @ -1250V, 0.2V scale, -200.mV threshold, end optical fiber (masked). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. analyzed
20170201 20170201_na22_2000v_DMR12d_20mm.dat Na-22 run. Na-22 source positioned between DMR12 and generic scintillator inside of collimator. channel 0: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face D). channel 1: PMT 8575B @ -1750V, generic scintillator(1cm x 2 cm x 4cm), 0.2V scale, -200.mV threshold. channel 3: trigger pulse from coincidence unit, 40ns coincidence window. First runs with collimator analyzed
20170201 20170201_na22_2000v_DMR12d_220mm.dat Na-22 run. Na-22 source positioned between large scintillator slabs and a perpendicular generic scintillator at location (22cm,16cm). channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.1V scale, -300.mV threshold. channel 2: PMT R1332 @ -2200.V, 0.05V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. First runs with collimator analyzed
20170201 20170201_na22_2000v_DMR12d_100mm.dat Na-22 run. Na-22 source positioned between DMR12 and generic scintillator inside of collimator. channel 0: PMT R1332 @ -2000.V, 0.1V scale, -50.mV threshold, DMR12 (face D). channel 1: PMT 8575B @ -1750V, generic scintillator(1cm x 2 cm x 4cm), 0.2V scale, -200.mV threshold. channel 3: trigger pulse from coincidence unit, 40ns coincidence window. First runs with collimator analyzed

January 2017

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20170103 20170103_na22_1250v_DMR12b_pmt14.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT #14 @ -1250.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window.
20170103 20170103_na22_1400v_DMR12b_pmt14.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT 14 @ -1400.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window.
20170103 20170103_na22_2000v_DMR12a.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window. analyzed
20170103 20170103_na22_2000v_DMR12a_highthresh.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window.
20170103 20170103_na22_2000v_DMR12b.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window. analyzed
20170103 20170103_na22_2000v_DMR12c.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window. analyzed
20170103 20170103_na22_2000v_DMR12c_v2.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window.
20170103 20170103_na22_2000v_DMR12d.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 5.V scale, -50.mV threshold, DMR09. channel 3: trigger pulse from coincidence unit, 40ns coincidence window. analyzed
20170109 20170109_monochromator_DMR12a.txt Monochromator run. Sample: DMR12A. LED: 305nm UV. Step size: 1 nm. Data range: 350nm to 600nm. analyzed
20170109 20170109_monochromator_DMR12b.txt Monochromator run. Sample: DMR12B. LED: 305nm UV. Step size: 1 nm. Data range: 350nm to 600nm. analyzed
20170109 20170109_monochromator_DMR12c.txt Monochromator run. Sample: DMR12C. LED: 305nm UV. Step size: 1 nm. Data range: 350nm to 600nm. analyzed
20170109 20170109_monochromator_DMR12d.txt Monochromator run. Sample: DMR12D. LED: 305nm UV. Step size: 1 nm. Data range: 350nm to 600nm. analyzed
20170110 20170110_na22_2000v_DMR12a_30mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 2V scale, -50.mV threshold, DMR12 (face A). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 30mm away from middle PMT face. Note: 2V scale for middle PMT was used. 15000 analyzed
20170110 20170110_na22_2000v_DMR12a_90mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 2V scale, -50.mV threshold, DMR12 (face A). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 90mm away from middle PMT face. Note: 2V scale for middle PMT was used. 15000 analyzed
20170110 20170110_na22_2000v_DMR12a_150mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face A). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 150mm away from middle PMT face. Note: 0.5V scale for middle PMT was used. 15000 analyzed
20170110 20170110_na22_2000v_DMR12a_210mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face A). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 210mm away from middle PMT face. Note: 0.5V scale for middle PMT used. If a larger scale was used, small pulses would have been indiscernible from noise. 15000 analyzed
20170110 20170110_na22_2000v_DMR12a_30mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 2V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 30mm away from middle PMT face. Note: 2V scale for middle PMT was used. 15000 analyzed
20170110 20170110_na22_2000v_DMR12a_90mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 90mm away from middle PMT face. Note: 1V scale for middle PMT was used. 15000 analyzed
20170110 20170110_na22_2000v_DMR12a_150mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 150mm away from middle PMT face. Note: 0.5V scale for middle PMT was used. 15000 analyzed
20170110 20170110_na22_2000v_DMR12a_210mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. Source placed 210mm away from middle PMT face. Note: 0.5V scale for middle PMT used. If a larger scale was used, small pulses would have been indiscernible from noise. 15000 analyzed
20170112 20170112_na22_2200v_BC400_5_5.dat Na-22 run. Na-22 source positioned between large scintillator slabs and a perpendicular generic scintillator at location (5cm,5cm). channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -300.mV threshold. channel 2: PMT R1332 @ -2200.V, 0.05V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. 15 min run. analyzed
20170112 20170112_na22_2200v_BC400_40_5.dat Na-22 run. Na-22 source positioned between large scintillator slabs and a perpendicular generic scintillator at location (40cm,5cm). channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -300.mV threshold. channel 2: PMT R1332 @ -2200.V, 0.05V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. 15 min run. analyzed
20170112 20170112_na22_2200v_BC400_22_16.dat Na-22 run. Na-22 source positioned between large scintillator slabs and a perpendicular generic scintillator at location (22cm,16cm). channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -300.mV threshold. channel 2: PMT R1332 @ -2200.V, 0.05V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40ns coincidence window. 15 min run. analyzed
20170113 20170113_na22_2000v_DMR12d_20mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170113 20170113_na22_2000v_DMR12d_40mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170113 20170113_na22_2000v_DMR12d_60mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170113 20170113_na22_2000v_DMR12d_80mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. switched to 0.5V scale. analyzed
20170113 20170113_na22_2000v_DMR12d_100mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170113 20170113_na22_2000v_DMR12d_120mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.5V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170113 20170113_na22_2000v_DMR12d_140mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.2V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. switched to 0.2V scale. analyzed
20170113 20170113_na22_2000v_DMR12d_160mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.1V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. switched to 0.1V scale. run time extended to 20 min otherwise no discernible double peak in analysis. analyzed
20170113 20170113_na22_2000v_DMR12d_180mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.2V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170113 20170113_na22_2000v_DMR12d_200mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.1V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170113 20170113_na22_2000v_DMR12d_220mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 0.1V scale, -50.mV threshold, DMR12 (face D). channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170119 20170119_na22_2000v_BC410_20mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, BC410 generic scintillator. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170119 20170119_na22_2000v_BC410_40mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, BC410 generic scintillator. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170119 20170119_na22_2000v_BC410_40mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, BC410 generic scintillator. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170119 20170119_na22_2000v_BC410_40mm.dat Na-22 run. Na-22 source positioned between scintillator sample and a perpendicular generic scintillator. channel 0: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 0.5V scale, -50.mV threshold. channel 2: PMT R1332 @ -2000.V, 1V scale, -50.mV threshold, BC410 generic scintillator. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. analyzed
20170120 20170121_muon_2000v_BC400.dat, 20170123_muon_2000v_BC400.dat Muon run. channel 0: PMT 14 @ -1250V, generic scintillator(1cm x 2 cm x 4cm), 1V scale, -200mV threshold. channel 2: PMT R1332 @ -2200.V, 500mV scale, BC400 generic scintillator slab. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window. not recorded: PMT 8575B @ -1700V, generic scintillator(1cm x 2 cm x 4cm), 1V scale, -200mV threshold. analyzed
20170129 20170129_cf252_BC400.dat Cf-252 run. Cf-252 source placed on top of scintillator slabs. channel 0: PMT 14 @ -1250V, middle fiber of BC400 PS slab, 0.1V scale, -200mV threshold. channel 1: PMT R1332 @ -1700.V, 1V scale, -50.mV threshold, end fiber of BC400 PS slab. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window, coincidence between channels 0 and 1. analyzed
20170129 20170129_cf252_BC400_background.dat Background run. Source was removed. channel 0: PMT 14 @ -1250V, middle fiber of BC400 PS slab, 0.1V scale, -200mV threshold. channel 1: PMT R1332 @ -1700.V, 1V scale, -50.mV threshold, end fiber of BC400 PS slab. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window, coincidence between channels 0 and 1. analyzed
20170129 20170129_na22_BC400.dat Na-22 run. Na-22 source placed on top of scintillator slabs. channel 0: PMT 14 @ -1250V, middle fiber of BC400 PS slab, 0.1V scale, -200mV threshold. channel 1: PMT R1332 @ -1700.V, 1V scale, -50.mV threshold, end fiber of BC400 PS slab. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window, coincidence between channels 0 and 1. analyzed
20170129 20170129_cf252_DMR09.dat Cf-252 run. Cf-252 source placed on top of DMR09 Gd loaded sample. channel 0: PMT 14 @ -1250V, middle fiber of BC400 PS slab, 0.1V scale, -200mV threshold. channel 1: PMT R1332 @ -1700.V, 1V scale, -50.mV threshold, end fiber of BC400 PS slab. channel 3: trigger pulse from coincidence unit, 40 ns coincidence window, coincidence between channels 0 and 1. analyzed

December 2016

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?

November 2016

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?

October 2016

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20161001
20161001 20161001_151646_monochromator_DMR01.txt Monochromator run. Sample: DMR01. LED: 305nm UV. Step size: 1 nm. Data range: 400nm to 600nm. Good runs. Can be used in paper.
20161001 20161001_152650_monochromator_DMR06.txt
20161001 20161001_153748_monochromator_DMR07.txt
20161001 20161001_154612_monochromator_DMR09.txt
20161004
20161010
20161016
20161017
20161022

September 2016

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20160929

August 2016

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20160801 20160801_muon_2000v_DMR09.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR09 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator.
20160801 20160801_115830_monochromator_DMR07.txt Monochromator run. Sample: DMR07. LED: ___UV. Step size: 1 nm. Data range: 400nm to 600nm. It was later determined that MONOCHROMATOR MUST BE POWER CYCLED BETWEEN EACH RUN
20160801 20160801_123310_monochromator_DMR06.txt
20160801 20160801_124216_monochromator_DMR01.txt
20160801 20160801_132513_monochromator_DMR09.txt
20160802 20160802_muon_2000v_DMR09.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR09 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator.
20160808 several monochromator runs taken Distance of sample face from monochromator face varied. Results from runs inconclusive. No meaningful data.
20160812 20160812_na22_1797v_DMR09.dat Na-22 run. Na-22 source positioned between two parallel scintillator pieces. channel 0: PMT R1332 @ -1800.V, 16.V scale, -300.mV threshold, DMR09. PMT #14 @ -1220V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -300.mV threshold. channel 2: trigger pulse from coincidence unit, 40ns coincidence window.
20160815 20160815_muon_1700v_DMR09.dat Muon run. channel 0: middle PMT, R1332 @ -1700V, 16V scale, DMR09 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator.
20160815 20160815_na22_1700v_DMR09.dat Na-22 run. Na-22 source positioned between two parallel scintillator pieces. channel 0: PMT R1332 @ -1800.V, 16.V scale, -300.mV threshold, DMR09. PMT #14 @ -1220V, generic scintillator(1cm x 2 cm x 4cm), 1.6V scale, -300.mV threshold. channel 2: trigger pulse from coincidence unit, 40ns coincidence window.

July 2016

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20160701 20160701_muon_2000v_DMR09.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR09 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. This run can be used in paper.
20160711 20160711_muon_2000v_DMR01.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR01 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. This run can be used in paper.
20160714 20160714_muon_2000v_DMR06.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR06 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. This run can be used in paper.
20160718 20160718_muon_1800v_DMR06.dat Muon run. channel 0: middle PMT, R1332 @ -1800V, 16V scale, DMR09 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator.
20160729 20160729_muon_2000v_DMR09.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR09 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator.

June 2016

Date Run Run Info Run Notes # of Events / Files DAQ configuration file Status?
20160606 20160606_muon_DMR01 Muon run. channel 0: middle PMT, R1332 @ -1720V, 16V scale, DMR01 unloaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator parallel to bottom scintillator. Middle scintillator perpendicular to top scintillator. Test data file. Scope scale set too low so pulse data was cut off. Data run needs to be redone.
20160609 20160609_muon_DMR06.dat Muon run. channel 0: middle PMT, R1332 @ -1726V, 16V scale, DMR06 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1714V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1229V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator parallel to bottom scintillator. Middle scintillator perpendicular to top scintillator.
20160614 20160614_muon_DMR01.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR01 unloaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1709V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1229V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Perpendicular scintillator alignments used. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. Scintillator orientation changed to reduce number of blank triggers.
20160617 20160617_muon_2000v_DMR01.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR01 unloaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1700V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. Middle PMT voltage increased.
20160620 20160620_muon_DMR01.dat Muon run. channel 0: middle PMT, R1332 @ -1720V, 16V scale, DMR01 unloaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. Middle PMT voltage increased.
20160624 20160624_muon_2000v_DMR07.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR07 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. Bad run. Ringing was observed do to incorrect termination of joints throughout signal lines. This problem was local to this run.
20160628 20160628_muon_2000v_DMR07.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, DMR07 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1703V, generic scintillator (2cm x 2 cm x 4cm), 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1220V, generic scintillator(2cm x 2 cm x 4cm), 1.6V scale, -300mV threshold. scope triggers when channel 1 & 2 are above threshold within 40 ns coincidence window. Top scintillator perpendicular to bottom scintillator. Middle scintillator parallel to bottom scintillator. This run can be used in the paper.

February 2015

Date Run Run Info # of Events / Files DAQ configuration file Status?
2015/02/16 20150216_na-22_wls-fiber.dat Na-22 source positioned between two scintillator pieces, one with WLS fiber. channel 0: PMT R1332 @ -2000.V, 16.V scale, -30.mV threshold, scintillator with WLS fiber. channel 1: PMT 8575B @ -1700.V, 1.6V scale, -180.mV threshold, generic scintillator. channel 2: trigger pulse. 100 samples @ 1.0 GSample/sec 30530 events None Uploaded

January 2015

Date Run Run Info # of Events / Files DAQ configuration file Status?
2015/01/22 20150122_muon_run_gd-scintillator.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, JPB 1-101 Gd-loaded scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1700V, generic scintillator, 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1200V, generic scintillator, 1.6V scale, -180mV threshold. scope triggers when channel 1 & 2 are above threshold within 100 ns coincidence window. 12216 events None Uploaded
2015/01/19 20150119_muon_run.dat Muon run. channel 0: middle PMT, R1332 @ -2000V, 16V scale, JPB 1-74-1 scintillator. channel 1 (not recorded): bottom PMT, 8575B @ -1700V, generic scintillator, 4.0V scale, -300mV discriminator threshold. channel 2: top PMT #14 @ -1200V, generic scintillator, 1.6V scale, -180mV threshold. scope triggers when channel 1 & 2 are above threshold within 100 ns coincidence window. 13000 events None Uploaded
2015/01/19 20150119_na-22_test.dat Na-22 source positioned between two scintillator pieces, no Gd loading. channel 0: PMT R1332 @ -2000.V, 16.V scale, -30.mV threshold, JPB 1-74-2 (non-gadolinium) scintillator. channel 1: PMT 8575B @ -1700.V, 1.6V scale, -180.mV threshold, generic scintillator. channel 2: trigger pulse. 100 samples @ 1.0 GSample/sec 30002 events None Uploaded
2015/01/15 20150115_na-22_test.dat Na-22 source positioned between two scintillator pieces, one Gd-loaded. channel 0: PMT R1332 @ -2000.V, 16.V scale, -30.mV threshold, JPB 1-101 (gadolinium) scintillator. channel 1: PMT 8575B @ -1700.V, 1.6V scale, -180.mV threshold, generic scintillator. channel 2: trigger pulse. 100 samples @ 1.0 GSample/sec 30001 events None Uploaded

October 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/10/13 141013_muon_fiber.dat SiPM 3 on channel 0, placed on end of WLS fiber in scintillator; trigger PMTs – top PMT is #18 at -1200V, -180mV threshold; bottom PMT is 8575B at -1700V, -300mV threshold 8498 141013_muon_fiber.tb4 Uploaded
2014/10/09 141009_muon_telescope_x3_no-fiber.dat SiPM 4 on channel 0, SiPM 2 on channel 1, SiPM 3 on channel 2, SiPM 1 on channel 3 (dead), aligned in 2×2 grid centered at middle of scintillator cross section; trigger PMTs – top PMT is #18 at -1200V, -180mV threshold; bottom PMT is 8575B at -1700V, -300mV threshold 8162 141009_ext-trig.tb4 Uploaded

September 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/09/30 20140930_muon-telescope.dat PMT R1332 @ -2000V (channel 1), scintillator sample with new WLS fiber + foil mask, end of fiber away from PMT is painted with TiO2 paint; trigger PMTs: PMT 14 (channel 4) @ -1200V, -180mV threshold, PMT 8575B (channel 3) @ -1700V, -300mV threshold ~8k None Uploaded
2014/09/12 20140912_muon_fiber_foil-mask.dat PMT R1332 @ -2000V (channel 1), scintillator sample with WLS fiber + foil mask; trigger PMTs: PMT 14 (channel 4) @ -1200V, -180mV threshold, PMT 8575B (channel 3) @ -1700V, -300mV threshold ~8k? None Uploaded
2014/09/09 20140909_muon_fiber.dat PMT R1332 @ -2000V (channel 1), scintillator sample with WLS fiber; trigger PMTs: PMT 14 (channel 4) @ -1200V, -180mV threshold, PMT 8575B (channel 3) @ -1700V, -300mV threshold ~8k? None Uploaded
2014/09/08 20140908_R1332_wls-fiber_am241_foil_0.8.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber, foil separating scintillator and PMT, Am-241 source placed on opposite end of scintillator, 0.8V FS, -30mV self-trigger via discriminator ~10k None Uploaded
2014/09/08 20140908_R1332_wls-fiber_am241_foil_1.6.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber, foil separating scintillator and PMT, Am-241 source placed on opposite end of scintillator, 1.6V FS, -30mV self-trigger via discriminator ~10k None Uploaded
2014/09/05 20140905_R1332_wls-fiber_bkgnd_foil_1.6.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber, foil separating scintillator and PMT, 1.6V FS, -30mV self-trigger via discriminator ~10k None Uploaded
2014/09/05 20140905_R1332_wls-fiber_bkgnd_foil_0.8.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber, foil separating scintillator and PMT, 0.8V FS, -30mV self-trigger via discriminator ~10k None Uploaded
2014/09/04 20140904_R1332_wls-fiber_bkgnd_0.8.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber is optically coupled to PMT, 0.8V FS, -30mV self-trigger via discriminator ~10k None Uploaded
2014/09/04 20140904_R1332_wls-fiber_am241_0.8.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber is optically coupled to PMT, Am-241 source placed against opposite end of scintillator, 0.8V FS, -30mV self-trigger via discriminator ~10k None Uploaded
2014/09/04 20140904_R1332_wls-fiber_am241_1.6.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber is optically coupled to PMT, Am-241 source placed against opposite end of scintillator, 1.6V FS, -30mV self-trigger via discriminator ~10k None Uploaded
2014/09/04 20140904_R1332_wls-fiber_bkgnd_1.6.dat PMT R1332 @ -2000V, ~2" scintillator sample with WLS fiber is optically coupled to PMT, 1.6V FS, -30mV self-trigger via discriminator ~10k None Uploaded

August 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/08/11 20140811_muontelescope_2200V.dat 2" scintilator (sample JPB 1-60) attached to PMT R1332 @ -2200V, 16.V full-scale (FS), channel 0; two-fold coincidence on PMTs 18 (top, -180mV threshold, 1.6V FS, channel 4) and 8575B (bottom, -300mV threshold, 4.0V FS, channel 3) ~8k None Uploaded
2014/08/05 20140805_muontelescope_2000V.dat 2" scintilator (sample JPB 1-60) attached to PMT R1332 @ -2000V, 16.V full-scale (FS), channel 0; two-fold coincidence on PMTs 18 (top, -180mV threshold, 1.6V FS, channel 4) and 8575B (bottom, -300mV threshold, 4.0V FS, channel 3) ~8k None Uploaded
2014/08/01 20140801_muontelescope.dat 2" scintilator (sample JPB 1-60) attached to PMT R1332 @ -1800V, 16.V full-scale (FS), channel 0; two-fold coincidence on PMTs 18 (top, -180mV threshold, 1.6V FS, channel 4) and 8575B (bottom, -300mV threshold, 4.0V FS, channel 3) ~8k None Uploaded
2014/08/01 20140801_R1332_2200V_scint1-60_am241.dat 2" scintilator (sample JPB 1-60) attached to PMT R1332 @ -2200V, -90mV self-trigger threshold, 1.6V full scale, Am-241 source placed >1cm from scintillator ~10k None Uploaded
2014/08/01 20140801_R1332_2200V_scint1-60_background.dat 2" scintilator (sample JPB 1-60) attached to PMT R1332 @ -2200V, -90mV self-trigger threshold, 1.6V full scale, no source ~10k None Uploaded

July 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/07/31 20140731_R1332_0.75inch-sample_2200V_background_90mV-threshold.dat 0.75" sample on R1332 PMT @ -2200V, -90mV self-trigger threshold, 1.6V full scale ~10k None Uploaded
2014/07/31 20140731_R1332_0.75inch-sample_2200V_am241_90mV-threshold.dat 0.75" sample on R1332 PMT @ -2200V, -90mV self-trigger threshold, 1.6V full scale, Am-241 source ~5mm from scintillator ~10k None Uploaded
2014/07/31 20140731_R1332_10mm-sample_2200V_am241_90mV-threshold.dat 10mm sample on R1332 PMT @ -2200V, -90mV self-trigger threshold, 1.6V full scale, Am-241 source ~5mm from scintillator ~10k None Uploaded
2014/07/31 20140731_R1332_10mm-sample_2200V_background_90mV-threshold.dat 10mm sample on R1332 PMT @ -2200V, -90mV self-trigger threshold, 1.6V full scale ~10k None Uploaded
2014/07/30 20140730_R1332_2200V_am241_90mV-threshold.dat 13mm sample on R1332 PMT @ -2200V, -90mV self-trigger threshold, 1.6V full scale, Am-241 source ~5mm from scintillator ~200k None Uploaded
2014/07/29 20140729_R1332_2200V_am241_90mV-threshold.dat 13mm sample on R1332 PMT @ -2200V, -90mV self-trigger threshold, 1.6V full scale, Am-241 source ~5mm from scintillator ~10k None Uploaded
2014/07/29 20140729_R1332_2200V_background_90mV-threshold.dat 13mm sample on R1332 PMT @ -2200V, -90mV self-trigger threshold, 1.6V full scale ~10k None Uploaded
2014/07/29 20140729_R1332_2200V_am241.dat 13mm sample on R1332 PMT @ -2200V, -30mV self-trigger threshold, 0.8V full scale, Am-241 source ~5mm from scintillator ~10k None Uploaded
2014/07/29 20140729_R1332_2000V_am241.dat 13mm sample on R1332 PMT @ -2000V, -30mV self-trigger threshold, 0.8V full scale, Am-241 source ~5mm from scintillator ~10k None Uploaded
2014/07/29 20140729_R1332_2000V_background.dat 13mm sample on R1332 PMT @ -2000V, -30mV self-trigger threshold, 0.8V full scale ~10k None Uploaded
2014/07/28 20140728_R1332_1800V_background2.dat 13mm sample on R1332 PMT @ -1800V, -30mV self-trigger threshold, 0.8V full scale ~10k None Uploaded
2014/07/28 20140728_R1332_1800V_am241.dat 13mm sample on R1332 PMT @ -1800V, -30mV self-trigger threshold, 0.8V full scale, Am-241 source placed ~5mm from scintillator ~10k None Uploaded
2014/07/28 20140728_R1332_1800V_background.dat 13mm sample on R1332 PMT @ -1800V, -30mV self-trigger threshold, 0.8V full scale ~10k None Uploaded

June 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/06/11 20140611_muon-test_3ch.dat 3 day run with muon scope PMTs; PMT 18 @ -1200V (top) on channel 3 with 0.4V full scale (284mV trigger threshold), PMT 8575B @ -1700V (bottom) on channel 4 with 4.0V full scale (214mV trigger threshold), PMT R1332 @ -2000V (middle) on channel 1 with 16V full scale ~4k None Uploaded
2014/06/06 20140606_muon-test_2ch.dat 4 day run with muon scope; PMT 18 @ -1200V (top) with 0.8V full scale and 402mV trigger threshold; PMT 8575B @ -1700V (bottom) with 4.0V full scale and 116mV trigger threshold; middle PMT removed ~19k None Uploaded

May 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/05/27 20140527_muon_test.dat ~3 day run with muon telescope, same setup as on 2014/05/22 (PMT 18 @ -1200V & 8575B @ -1700V for trigger, PMT R1332 @ -2000V for signal), full-scale set to 8V & +3.5V offset, so saturation should occur at -7.5V ~6k None Uploaded
2014/05/22 20140522_muon-test.dat 5 day run with muon telescope (PMT 18 @ -1200V & 8575B @ -1700V for trigger, PMT R1332 @ -2000V for signal), 16V full-scale & +7.5V offset, so saturation should occur at -15.5V ~12k None Uploaded
2014/05/07 20140507_R1332_Am241_2000V_12.5mV-threshold.dat R1332 PMT @ -2000V, Am-241 source, non-Gd scintillator, -12.5mV self-trigger, full scale increased ~100k None Uploaded
2014/05/07 20140507_R1332_background_2000V_12.5mV-threshold.dat R1332 PMT @ -2000V, no source, non-Gd scintillator, -12.5mV self-trigger, full scale increased ~100k None Uploaded
2014/05/05 20140505_R1332_no-Gd-Am241_2000V_10mV-threshold.dat R1332 PMT @ -2000V, Am-241 source, non-Gd scintillator, -10.0mV self-trigger ~100k None Uploaded
2014/05/05 20140505_R1332_no-Gd-background_2000V_10mV-threshold.dat R1332 PMT @ -2000V, no source, non-Gd scintillator, -10.0mV self-trigger ~100k None Uploaded
2014/05/02 20140502_R1332_Gd-Am241_2000V_10mV-threshold.dat R1332 PMT @ -2000V, Am-241 source, Gd-doped scintillator, -10.0mV self-trigger ~100k None Uploaded
2014/05/02 20140502_R1332_Gd-background_2000V_10mV-threshold.dat R1332 PMT @ -2000V, background, Gd-doped scintillator, -10.0mV self-trigger ~100k None Uploaded
2014/05/01 20140501_R1332_SPE_2200V_2.5mV-threshold.dat Hamamatsu R1332 PMT @ -2200V, SPE test with -2.5mV self-trigger ~100k None Uploaded
2014/05/01 20140501_R1332_SPE_1800V_2.5mV-threshold.dat Hamamatsu R1332 PMT @ -1800V, SPE test with -2.5mV self-trigger ~100k None Uploaded

April 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/04/30 20140430_R1332_SPE_2000V_2.5mV-threshold.dat Hamamatsu R1332 PMT @ -2000V, SPE test with -2.5mV self-trigger ~100k None Uploaded
2014/04/30 20140430_PMT8575B_Am241_2000V_50mV-threshold.dat Burle PMT @ -2000V, self trigger @ -50mV, Am-241 source, Gd-doped scintillator ~100k None Uploaded
2014/04/30 20140430_PMT8575B_Am241_1800V_50mV-threshold.dat Burle PMT @ -2000V, self trigger @ -50mV, Am-241 source, Gd-doped scintillator ~100k None Uploaded
2014/04/29 20140429_PMT8575B_background_2000V_50mV-threshold.dat Burle PMT @ -2000V, self trigger @ -50mV, no source, Gd-doped scintillator ~100k None Uploaded
2014/04/29 20140429_PMT8575B_background_1800V_50mV-threshold.dat Burle PMT @ -2000V, self trigger @ -50mV, no source, Gd-doped scintillator ~100k None Uploaded
2014/04/28 20140428_PMT8575B_Am-241_2000V_50mV-threshold.dat Burle PMT @ -2000V, self trigger @ -50mV, Am-241 source, Bertan HV supply ~100k None Uploaded
2014/04/28 20140428_PMT8575B_background_2000V_50mV-threshold.dat Burle PMT @ -2000V, self trigger @ -50mV, no source, Bertan HV supply ~100k None Uploaded
2014/04/25 20140425_PMT8575B_Am-241_1800V_50mV-threshold.dat Burle PMT @ -1800V, self trigger @ -50mV, Am-241 source, Bertan HV supply ~100k None Uploaded
2014/04/25 20140425_PMT8575B_background_1800V_50mV-threshold.dat Burle PMT @ -1800V, self trigger @ -50mV, no source, Bertan HV supply ~100k None Uploaded
2014/04/23 20140423_PMT8575B_SPE_1300V_0.4mV-threshold.dat, 20140423_PMT8575B_SPE_1400V_0.4mV-threshold.dat, 20140423_PMT8575B_SPE_1500V_0.4mV-threshold.dat, 20140423_PMT8575B_SPE_1600V_0.4mV-threshold.dat, 20140423_PMT8575B_SPE_1700V_0.4mV-threshold.dat, 20140423_PMT8575B_SPE_1800V_0.4mV-threshold.dat Measurement of single photoelectron peak; Burle PMT @ voltages from -1300V to -1800V in 100V increments, threshold at -0.4mV, Bertan HV supply ~10k per voltage None Uploaded
2014/04/18 20140418_PMT8575B_SPE_1310V_1.2mV-threshold.dat Measurement of single photoelectron peak; Burle PMT @ -1310V, threshold at -1.2mV, Bertan HV supply ~100k None Uploaded
2014/04/18 20140418_PMT8575B_SPE_1460V_1mV-threshold.dat Measurement of single photoelectron peak; Burle PMT @ -1460V, threshold at -1.0mV, Bertan HV supply ~100k None Uploaded
2014/04/18 20140418_PMT8575B_SPE_1310V_0.4mV-threshold.dat Measurement of single photoelectron peak; Burle PMT @ -1310V, threshold at -0.400mV, switched back to old power supply (Bertan) 106071 None Uploaded
2014/04/17 20140417_PMT14_SPE_1100V_1mV-threshold.dat Measurement of single photoelectron peak; PMT 14 @ -1100V, threshold at -1.00mV, switched to new power supply 10447 None Uploaded
2014/04/08 20140408_PMT14_SPE_1100V_1mV-threshold.dat Measurement of single photoelectron peak; PMT 14 @ -1100V, threshold at -1.00mV ~20k events None Uploaded
2014/04/07 20140407_PMT14_SPE_1100V_2mV-threshold.dat Measurement of single photoelectron peak for PMT 14; PMT 14 @ -1100V, threshold set to -2.00mV 10547 events None Uploaded
2014/04/07 20140407_PMT14_SPE_1100V_1mV-threshold.dat Measurement of single photoelectron peak using new oscilloscope (LeCroy 434); PMT 14 @ -1100V, threshold set to -1.00mV 10462 events None Uploaded

February 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/02/05 20140205_pmt14_self-trig_background_1100v_Gd-scint.dat PMT 14 @ -1100V, no source; Gd-doped 1-101 scintillator; 200mV per division, -30mV self trigger 1 file, unknown event count None Uploaded
2014/02/04 20140204_pmt14_self-trig_cf252_1100v_Gd-scint.dat PMT 14 @ -1100V, Cf-252 source (unmoderated); Gd-doped 1-101 scintillator; 200mV per division, -30mV self trigger 1 file, unknown event count None Uploaded

January 2014

Date Run Run Info # of Events / Files DAQ configuration file Status?
2014/02/05 20140205_pmt14_self-trig_background_1100v_Gd-scint.dat PMT 14 @ -1100V, no source; Gd-doped 1-101 scintillator; 200mV per division, -30mV self trigger 1 file, unknown event count None Uploaded
2014/02/04 20140204_pmt14_self-trig_cf252_1100v_Gd-scint.dat PMT 14 @ -1100V, Cf-252 source (unmoderated); Gd-doped 1-101 scintillator; 200mV per division, -30mV self trigger 1 file, unknown event count None Uploaded
2014/01/31 20140131_self-trig_cf252-no-scint_1100v_PMT14_200mV-scale.dat PMT 14 @ -1100V, Cf-252 source, unmoderated; no scintillator; 200mV per division, -30mV self trigger 1 File, unknown event count None Uploaded
2014/01/31 20140131_self-trig_background-no-scint_1100v_PMT14_200mV-scale.dat PMT 14 @ -1100V, no source, no scintillator (dark count test); 200mV per division, -30mV self trigger 1 File, unknown event count None Uploaded
2014/01/31 20140131_self-trig_background_1100v_PMT14_200mV-scale.dat PMT 14 @ -100V, no source, non-Gadolinium 1-66-x scintillator, 200mV per division on scope, self-trigger on -30mV or greater pulse 1 File, unknown event count None Uploaded
2014/01/28 20140128_self-trig_cf251_1100v_PMT14_200mV-scale.dat PMT 14 @ -1100V with Cf-252 source (unmoderated) attached to end of (non-Gadolinium) scintillator, full-scale on scope set to 50mV per division (8 divisions total), -30mV self-trigger 1 File, unknown event count None Uploaded
2014/01/24 20140124_self-trig_am241_1100v_PMT14_200mV-scale.dat PMT 14 @ -1100V with am-241 source attached to end of scintillator, full-scale on scope set to 200mV per division (8 divisions total) 1 File, unknown event count None Uploaded
2014/01/23 20140123_self-trig_am241_1100v_PMT14_50mV-scale.dat same as 2014/01/21 50mV per division run but with more events 1 File, unknown event count None Uploaded
2014/01/21 20140121_self-trig_am241_1100v_PMT14_500mV-scale.dat PMT 14 @ -1100V with am-241 source attached to end of scintillator, full-scale on scope set to 500mV per division (8 divisions total) 1 File, unknown event count None Uploaded
2014/01/21 20140121_self-trig_am241_1100v_PMT14_50mV-scale.dat PMT 14 @ -1100V with am-241 source attached to end of scintillator, full-scale on scope set to 50mV per division (8 divisions total) 1 File, unknown event count None Uploaded
2014/01/1920140119_vertical_4.5cm_1100v_PMT14(middle).dat, 20140119_vertical_4.5cm_1100v_PMT18(top).dat, 20140119_vertical_4.5cm_1100v_PMT3(bottom).dat 4.5 vertical separation of PMTs; same as 2014/01/13 run 3 files None Uploaded
2014/01/1720140117_vertical_1.5cm_1100v_PMT14(middle).dat, 20140117_vertical_1.5cm_1100v_PMT18(top).dat, 20140117_vertical_1.5cm_1100v_PMT3(bottom).dat 1.5cm vertical separation of PMTs; otherwise the same configuration as 2014/01/13 run 3 files None Uploaded
2014/01/1520140115_vertical_1100v_PMT14(middle).dat, 20140115_vertical_1100v_PMT18(top).dat, 20140115_vertical_1100v_PMT3(bottom).dat same configuration as previous run 3 files None Uploaded
2014/01/1320140113_vertical_1100v_PMT14(middle).dat, 20140113_vertical_1100v_PMT18(top).dat, 20140113_vertical_1100v_PMT3(bottom).dat all -1100V, top and bottom PMTs coaxially aligned in a vertical configuration, with middle PMT placed horizontally between the two; 3.5cm vertical separation 3 files None Uploaded

December 2013

Date Run Run Info # of Events / Files DAQ configuration file Status?
2013/12/17 - 2013/12/201217-1220_atten_1100v_PMT3(bottom)_3.5cm.dat, 1217-1220_atten_1100v_PMT14(middle)_3.5cm.dat, 1217-1220_atten_1100v_PMT18(top)_3.5cm.dat see 12/10 - 12/11 run; all -1100V; discriminator threshold set to 300mV, each PMT recorded in separate file, 3.5cm attenuation distance 3 Files None Uploaded
2013/12/12 - 2013/12/171212-1217_attenuation_PMT3(bottom)_1100v_1.5cm-dist.dat, 1212-1217_attenuation_PMT14(middle)_1100v_1.5cm-dist.dat, 1212-1217_attenuation_PMT18(top)_1100v_1.5cm-dist.dat see 12/10 - 12/11 run; discriminator threshold set to 300mV, each PMT recorded in separate file, 1.5cm attenuation distance 3 Files None Uploaded
2013/12/10 - 2013/12/11attenuation_1100v_distance-3.5cm.dat, attenuation_1100v_distance-1.5cm.datPMTs 3 and 18 biased to -1100V, aligned parallel to one another (5.5cm vertical separation); PMT 14 biased to -1100V, placed between top and bottom PMTs, with plane of top of bottom PMTs moved to varying distances from the face of the middle PMT (1.5cm and 3.5cm); 30mV discriminator threshold, 70ns coincidence window; middle PMT recorded2 Files None Analyzed
2013/12/09 - 2013/12/10PMT3(bottom)_1100v_cross6.5cm.dat, PMT18(top)_1100v_cross6.5cm.dat Switched back to old discriminator (Lecroy 623B) with -30mV threshold, PMTs 3, 14, 18 biased to -1100V. PMTs 3 and 18 perpendicular to one another, with a vertical separation of 6.5cm, with their waveforms collected upon coincident trigger of both PMTs (70ns coincidence window) 2 Files None Analyzed
2013/12/04 - 2013/12/06new-descrim_1100v_cross_6.5cm.datTesting Ortec 935 CFD at -25mV discrimination threshold. PMTs 3, 14 and 18 biased to -1100V; all additional run info identical to 11/15 - 11/18 runUnknownNoneAnalyzed

November

Date Run Run Info # of Events / Files DAQ configuration file Status?
2013/11/24 - 2013/11/261124-1126_1200v-all_cross_6.5cm.dat PMT 0 and 3 biased to -1200v, 6.5 cm vertical separation of scintillators, aligned perpendicular to one another; trigger: -25mV threshold for each, 70ns coincidence window; PMT 14 @ -1200v positioned between trigger PMTs, waveform collected upon coincident trigger unknown None Yes
2013/11/22 - 2013/11/241122-1124_sig1100v_trig1200v_cross_6.5cm.dat PMT 0 and 3 biased to -1200v, 6.5 cm vertical separation of scintillators, aligned perpendicular to one another; trigger: -25mV threshold for each, 70ns coincidence window; PMT 14 @ -1100v positioned between trigger PMTs, waveform collected upon coincident trigger unknown None Yes
2013/11/18 - 2013/11/191118-1119_parallel_6.5cm_1100v.dat PMT 0 and 3 biased to -1100v, 6.5 cm vertical separation of scintillators, aligned parallel to one another; trigger: -25mV threshold for each, 70ns coincidence window; PMT 14 @ -1100v positioned between trigger PMTs, waveform collected upon coincident trigger unknown None Yes
2013/11/15 - 2013/11/181115-1118_cross_1100v_6.5cm.dat PMT 0 and 3 biased to -1100v, 6.5 cm vertical separation of scintillators, aligned perpendicular to one another; trigger: -25mV threshold for each, 70ns coincidence window; PMT 14 positioned between trigger PMTs, biased to -1100v, waveform collected upon coincident trigger unknown None Yes
2013/11/14TEST_1000V.DAT PMT 0 and 3 biased to -1100v, 6.5 cm vertical separation of scintillators, aligned parallel to one another; trigger: -25mV threshold for each, 70ns coincidence window; PMT 14 positioned between trigger PMTs, biased to -1000v, waveform collected upon coincident trigger of first two PMTs unknown None Uploaded
2013/11/14TEST.DAT Test run with muon telescope: PMT 0 and 3 biased to -1100v, 6.5 cm vertical separation of scintillators, aligned parallel to one another; trigger: -25mV threshold for each, 70ns coincidence window; PMT 14 positioned between trigger PMTs, biased to -1100v, waveform collected upon coincident trigger of first two PMTs unknown None Yes

October 2013

Date Run Run Info # of Events / Files DAQ configuration file Analysed?
2013/10/09131009_pmt14_background_8280_1100v_x10.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, partial lead castle surrounding scintillator, self-trig @ 8280 threshold, background w/ copper collimator 100k 131009_self-trig.tb4 In Progress
2013/10/09131009_pmt14_am241_8280_1100v_x10.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, partial lead castle surrounding scintillator, self-trig @ 8280 threshold, Am-241 source w/ copper collimator 100k 131009_self-trig.tb4 In Progress

September 2013

Date Run Run Info # of Events / Files DAQ configuration file Analysed?
2013/09/29130929_pmt14_led_1.85kohm_8350_800v.dat PMT 14 (ch0) @ 800v in darkroom w/ felt box, self-trig @ 8350 threshold, PMT flashed with 100ns blue LED pulse 10k 130929_self-trig.tb4 In Progress
2013/09/29130929_pmt14_led_1.85kohm_8350_850v.dat PMT 14 (ch0) @ 850v in darkroom w/ felt box, self-trig @ 8350 threshold, PMT flashed with 100ns blue LED pulse 10k 130929_self-trig.tb4 In Progress
2013/09/29130929_pmt14_led_1.85kohm_8350_900v.dat PMT 14 (ch0) @ 900v in darkroom w/ felt box, self-trig @ 8350 threshold, PMT flashed with 100ns blue LED pulse 10k 130929_self-trig.tb4 In Progress
2013/09/29130929_pmt14_led_1.85kohm_8350_950v.dat PMT 14 (ch0) @ 950v in darkroom w/ felt box, self-trig @ 8350 threshold, PMT flashed with 100ns blue LED pulse 10k 130929_self-trig.tb4 In Progress
2013/09/29130929_pmt14_led_1.85kohm_8350_1000v.dat PMT 14 (ch0) @ 1000v in darkroom w/ felt box, self-trig @ 8350 threshold, PMT flashed with 100ns blue LED pulse 10k 130929_self-trig.tb4 In Progress
2013/09/29130929_pmt14_led_1.85kohm_8350_1050v.dat PMT 14 (ch0) @ 1050v in darkroom w/ felt box, self-trig @ 8350 threshold, PMT flashed with 100ns blue LED pulse 10k 130929_self-trig.tb4 In Progress
2013/09/29130929_pmt14_led_1.85kohm_8350_1100v.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8350 threshold, PMT flashed with 100ns blue LED pulse 10k 130929_self-trig.tb4 In Progress
2013/09/24130924_sipm124_am241_8320.dat SiPM 1 (ch0), SiPM 2 (ch1), SiPM 4 (ch2) in darkroom w/ felt box, self-trig @ 8320 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Am-241 source 10k 130923_sipm124_self-trig.tb4 In Progress
2013/09/23130923_sipm124_background_8320.dat SiPM 1 (ch0), SiPM 2 (ch1), SiPM 4 (ch2) in darkroom w/ felt box, self-trig @ 8320 threshold, 1-66-x scintillator, no source 10k 130923_sipm124_self-trig.tb4 In Progress
2013/09/21130921_pmt14_am241_8280_1100v_x10_no-scint.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, no scintillator, Am-241 source attached to collimator100k130905_self-trig_long.tb4 In Progress
2013/09/20130920_pmt14_background_8280_1100v_x10.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, no scintillator, no source 100k 130905_self-trig_long.tb4 Yes
2013/09/17130917_pmt14_background_8230_1100v_x10_x10atten.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, x10 PMT signal attenuation 100k 130906_self-trig_long_x10atten.tb4 Yes
2013/09/11130911_pmt14_background_8280_1100v_x10.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator 100k 130905_self-trig_long.tb4 Yes
2013/09/10130910_pmt14_cs137_8230_1100v_x10_x10atten.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Cs-137 source attached to collimator, x10 PMT signal attenuation 100k 130906_self-trig_long_x10atten.tb4 Yes
2013/09/10130910_pmt14_am241_8230_1100v_x10_x10atten.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Am-241 source attached to collimator, x10 PMT signal attenuation 100k 130906_self-trig_long_x10atten.tb4 Yes
2013/09/09130909_pmt14_am241_8280_1100v_x10.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Am-241 source attached to collimator 100k 130905_self-trig_long.tb4 Yes
2013/09/09130909_pmt14_am241_8230_1100v_x10_x5atten.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Am-241 source attached to collimator, x5 PMT signal attenuation 100k 130906_self-trig_long_x5atten.tb4 Yes
2013/09/06130906_pmt14_ba133_8230_1100v_x10_x5atten.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Ba-133 source attached to collimator, x5 PMT signal attenuation 100k 130906_self-trig_long_x5atten.tb4 Yes
2013/09/06130906_pmt14_ba133_8280_1100v_x10.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Ba-133 source attached to collimator 100k 130905_self-trig_long.tb4 Yes
2013/09/05130905_pmt14_ba133_8280_1100v_x10.dat PMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x scintillator, copper collimator coaxial to scintillator, Ba-133 source attached to collimator; saw significantly higher noise than usual 100k 130905_self-trig_long.tb4 Yes
2013/09/03130903_ext-trig_long.datTB4 board noise test, external triggering on no signal 100k 130903_ext-trig_long.tb4 In progress
2013/09/03130903_ext-trig.datTB4 board noise test, external triggering on no signal 10k 130903_ext-trig.tb4 In progress

August 2013

Date Run Run Info # of Events / Files DAQ configuration file Analysed?
2013/08/30130830_pmt14_cs137_8230_1100v_x10_x10atten.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x plastic scintillator & Cs-137 source against scintillator, x10 in-line signal attenuator 100k 130829_pmt_self-trig_long_x10atten.tb4 Yes
2013/08/30130830_pmt14_cs137_8230_1100v_x10_x5atten.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x plastic scintillator & Cs-137 source against scintillator, x5 in-line signal attenuator 100k 130829_pmt_self-trig_long_x10atten.tb4 Yes
2013/08/30130830_pmt14_ba133_8230_1100v_x10_x5atten.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8230 threshold, 1-66-x plastic scintillator & Ba-133 source 2cm from scintillator, x5 in-line signal attenuator 100k 130829_pmt_self-trig_long_x5atten.tb4 Yes
2013/08/27130827_pmt14_co57_8280_1100v_x10.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x plastic scintillator & Co-57 source placed against scintillator 100k 130822_pmt_self-trig_long.tb4 Yes
2013/08/27130827_pmt14_ba133_8280_1100v_x10.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x plastic scintillator & Ba-133 source 2cm from scintillator 100k 130822_pmt_self-trig_long.tb4 Yes
2013/08/26130826_pmt14_am241_8280_1100v_x10.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x plastic scintillator & Am-241 source 2cm from scintillator 100k 130822_pmt_self-trig_long.tb4 Yes
2013/08/26130826_pmt14_background_8280_1100v_x10.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x plastic scintillator, no source 100k 130822_pmt_self-trig_long.tb4 Yes
2013/08/23130823_pmt14_am241_8280_1100v_x10.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x plastic scintillator & Am-241 source 10cm from scintillator 100k 130822_pmt_self-trig_long.tb4 Yes
2013/08/23130823_pmt14_ba133_8280_1100v_x10.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x plastic scintillator & Ba-133 source outside box 100k 130822_pmt_self-trig_long.tb4 Yes
2013/08/22130822_pmt14_ba133_8280_1100v.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold, 1-66-x plastic scintillator & Ba-133 source outside box 10k 130822_pmt_self-trig.tb4 No
2013/08/22130822_pmt3_spe_8280_1200v.datPMT 3 (ch0) @ 1200v in darkroom w/ felt box, self-trig @ 8280 threshold10k130821_pmt_self-trig.tb4Yes
2013/08/21130821_pmt3_spe_8280_1150v.datPMT 3 (ch0) @ 1150v in darkroom w/ felt box, self-trig @ 8280 threshold10k130821_pmt_self-trig.tb4Yes
2013/08/21130821_pmt3_spe_8280_1100v.datPMT 3 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold10k130821_pmt_self-trig.tb4Yes
2013/08/21130821_pmt0_spe_8280_1200v.datPMT 0 (ch0) @ 1200v in darkroom w/ felt box, self-trig @ 8280 threshold10k130820_pmt_self-trig2.tb4Yes
2013/08/20130820_pmt0_spe_8280_1150v.datPMT 0 (ch0) @ 1150v in darkroom w/ felt box, self-trig @ 8280 threshold10k130820_pmt_self-trig2.tb4Yes
2013/08/20130820_pmt0_spe_8280_1100v.datPMT 0 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold10k130820_pmt_self-trig2.tb4Yes
2013/08/20130820_pmt0_spe_8270_1100v_h.datPMT 0 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8270 threshold10k130820_pmt_self-trig.tb4Yes
2013/08/20130820_pmt0_spe_8280_1100v_h.datPMT 0 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8280 threshold10k130820_pmt_self-trig2.tb4Yes
2013/08/20130820_pmt0_spe_8270_1100v.datPMT 0 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8270 threshold10k130820_pmt_self-trig.tb4Yes
2013/08/20130820_pmt18_spe_8270_1150v.datPMT 18 (ch0) @ 1150v in darkroom w/ felt box, self-trig @ 8270 threshold10k130820_pmt_self-trig.tb4Yes
2013/08/20130820_pmt18_spe_8270_1100v.datPMT 18 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8270 threshold10k130820_pmt_self-trig.tb4Yes
2013/08/20130820_pmt14_spe_8270_1150v.datPMT 14 (ch0) @ 1150v in darkroom w/ felt box, self-trig @ 8270 threshold10k130820_pmt_self-trig.tb4Yes
2013/08/20130820_pmt14_spe_8270_1100v.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8270 threshold10k130820_pmt_self-trig.tb4Yes
2013/08/15130815_pmt14_spe_8240_low-light.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8240 threshold10k130815_pmt14_self-trig.tb4Yes
2013/08/15130815_pmt14_spe_8240_low-high.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, self-trig @ 8240 threshold10k130815_pmt14_self-trig.tb4Yes
2013/08/14130814_pmt14_no-trig.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, external trigger disconnected, LED shorted by wire10k130813_pmt14_LED_delay-trig.tb4Yes
2013/08/14130814_delay-trig_pmt14.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, 400ns delay on external trigger, LED shorted by wire10k130813_pmt14_LED_delay-trig.tb4Yes
2013/08/14130814_delay-trig_no-pmt.datno PMT in darkroom w/ felt box, 400ns delay on external trigger, LED shorted by wire10k130813_pmt14_LED_delay-trig.tb4Yes
2013/08/14130814_pmt14_blueled_delayed-trig_2.24kOhm.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, 100ns blue LED pulse w/ 2.24 kOhm inline resistor, PMT window within viewing angle of LED, 15cm separation10k130813_pmt14_LED_delay-trig.tb4Yes
2013/08/14130814_pmt14_blueled_delayed-trig_2.22kOhm.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, 100ns blue LED pulse w/ 2.22 kOhm inline resistor, PMT window within viewing angle of LED, 15cm separation10k130813_pmt14_LED_delay-trig.tb4Yes
2013/08/14130814_pmt14_blueled_delayed-trig_2.13kOhm.datPMT 14 (ch0) @ 1100v in darkroom w/ felt box, 100ns blue LED pulse w/ 2.13 kOhm inline resistor, PMT window within viewing angle of LED, 15cm separation10k130813_pmt14_LED_delay-trig.tb4Yes
2013/08/13130813_pmt14_spe_8262_1.datPMT 14 (ch0) @ 1100v covered in felt in darkroom, self-trig @ 8262 threshold10k130812_pmt3_self-trig.tb4Yes
2013/08/13130813_pmt14_spe_8265_1.datPMT 14 (ch0) @ 1100v covered in felt in darkroom, self-trig @ 8265 threshold10k130812_pmt3_self-trig.tb4Yes
2013/08/12130812_pmt14_spe_8250.datPMT 14 (ch0) @ 1100v covered in felt in darkroom, self-trig @ 8250 threshold10k130812_pmt3_self-trig.tb4Yes
2013/08/12130812_pmt14_spe_8260.datPMT 14 (ch0) @ 1100v covered in felt in darkroom, self-trig @ 8260 threshold10k130812_pmt3_self-trig.tb4Yes
2013/08/08130808_pmt14_spe_8270.datPMT 14 (ch0) @ 1100v covered in felt in darkroom, self-trig @ 8270 threshold 10k130808_pmt3_self-trig.tb4Yes
2013/08/08130808_pmt3_spe_8270.datPMT 3 (ch0) @ 1100v covered in felt in darkroom, self-trig @ 8270 threshold10k130808_pmt3_self-trig.tb4No

July 2013

Date Run Run Info # of Events / Files DAQ configuration file Analysed?
2013/07/29130729_sipm2+4_co-60_8320threshold_wlsfiber_x10.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on Co-60 source 10 cm from plastic, 8320 threshold 100k 130729_sipm2+4_self-trig_8320_long.tb4 No
2013/07/29130729_sipm2+4_co-60_8320threshold_wlsfiber.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on Co-60 source 10 cm from plastic, 8320 threshold 10k 130729_sipm2+4_self-trig_8320.tb4 No
2013/07/29130729_100ns_square-wave_XXmV.datFunction generator attached to DAQ; 100ns logic pulses sent to DAQ in order to correlate DAQ pulse height with voltage 100 x 10 130729_100ns_square-wave.tb4 Yes
2013/07/26130726_sipm2+4_y-88_8320threshold_wlsfiber_x10.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on Y-88 source 46 cm from plastic, 8320 threshold 100k 130726_sipm2+4_y-88_8320threshold_wlsfiber_x10.tb4 Yes
2013/07/26130726_sipm2+4_background_8320threshold_wlsfiber.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on background, 8320 threshold 10k 130726_sipm2+4_self-trig_8320.tb4 Yes
2013/07/25130725_sipm2+4_co-60_8320threshold_wlsfiber_x10.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on Co-60 source 10cm from plastic, 8320 threshold 100k 130725_sipm2+4_self-trig_8320_long.tb4 Yes
2013/07/25130725_sipm2+4_co-60_8320threshold_WLSfiber.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on Co-60 source 10cm from plastic, 8320 threshold 10k 130724_sipm2+4_self-trig_8320.tb4 Yes
2013/07/25130725_sipm2+4_background_8320threshold_WLSfiber.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on background, 8320 threshold 10k 130724_sipm2+4_self-trig_8320.tb4 Yes
2013/07/24130724_sipm2+4_cs-136_8320threshold_WLSfiber-2.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber & no fiber-scintillator coupling grease, self-trigger on Cs-137 source 10cm from plastic, 8320 threshold 10k 130724_sipm2+4_self-trig_8320.tb4 Yes
2013/07/24130724_sipm2+4_cs-136_8320threshold_WLSfiber.datSiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, self-trigger on Cs-137 source 10cm from plastic, 8320 threshold 10k 130724_sipm2+4_self-trig_8320.tb4 Yes
2013/07/23130723_sipm2+4_cs-137_8320threshold.dat SiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-66-? scintillator, self-trigger on Cs-137 source 10cm from plastic, 8320 threshold 10k 130723_sipm2+4_self-trig_8320.tb4 Yes
2013/07/22130722_sipm2+4_background_8320threshold.dat SiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-66-? scintillator, self-trigger on background, 8320 threshold 10k 130722_sipm2+4_self-trig_8320.tb4 Yes
2013/07/22130722_sipm3+4_ba-133_8320threshold_short.dat SiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, self-trigger on background, 8320 threshold 10k 130722_sipm3+4_self-trig_8320.tb4 Yes
2013/07/22130722_sipm3+4_ba-133_8320threshold.dat SiPM 2 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, self-trigger on Ba-133 source 3cm from scintillator, 8320 threshold 10k 130722_sipm3+4_self-trig_8320.tb4 Yes
2013/07/22130722_sipm3+4_background_8320threshold.dat SiPM 3 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, self-trigger on background, 8320 threshold 100k 130721_sipm3+4_self-trig_8320.tb4 Yes
2013/07/21130721_sipm3+4_ba-133_8320threshold.dat SiPM 3 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, Ba-133 source 3cm from scintillator, self trigger at 8320 threshold 100k 130721_sipm3+4_self-trig_8320.tb4 Yes
2013/07/19130719_sipm3+4_ba-133_8320threshold.dat SiPM 3 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, Ba-133 source 3cm from scintillator, self trigger at 8320 threshold 10k 130718_sipm3+4_self-trig_8320.tb4 Yes
2013/07/19130719_sipm3+4_ba-133_8400threshold.dat SiPM 3 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, Ba-133 source 1cm from scintillator, self trigger at 8400 threshold 10k 130718_sipm3+4_self-trig_8400.tb4 Yes
2013/07/18130718_sipm3+4_ba-133_8400threshold.dat SiPM 3 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, Ba-133 source 3cm from scintillator, self trigger at 8400 threshold 10k 130718_sipm3+4_self-trig_8400.tb4 Yes
2013/07/18130718_sipm3+4_background_8400threshold.datSiPM 3 (ch0) and SiPM 4 (ch1) @ 20.6V, each in delrin mount, opposite ends of 1-74-1 scintillator w/ WLS fiber, no source, self trigger at 8400 threshold 6.9k 130718_sipm3+4_self-trig_8400.tb4 Yes
2013/07/17130717_sipm3_ba-133_8350threshold_x10.datSiPM 3 @ 20.6V in delrin mount, 1-74-1 scintillator w/ WLS fiber & Ba-133 source opposite to delrin mount, self trigger at 8350 threshold 100k 130717_sipm3_self-trig_8350_long.tb4Yes
2013/07/17130717_sipm3_ba-133_8350threshold.datSiPM 3 @ 20.6V in delrin mount, 1-74-1 scintillator w/ WLS fiber & Ba-133 source opposite to delrin mount, self trigger at 8350 threshold 10k 130717_sipm3_self-trig_8350.tb4Yes
2013/07/17130717_sipm3_am-241_8400threshold.datSiPM 3 @ 20.6V in delrin mount, 1-66-? scintillator w/ Am-241 source opposite to delrin mount, self trigger at 8400 threshold 10k 130717_sipm3_self-trig_8400threshold.tb4 Yes
2013/07/17130717_sipm4_am-241_8400threshold.datSiPM 4 @ 20.6V in delrin mount, 1-66-? scintillator w/ Am-241 source opposite to delrin mount, self trigger at 8400 threshold 10k 130717_sipm4_self-trig_8400threshold.tb4 Yes
2013/07/17130717_sipm4_am-241_8275threshold.datSiPM 4 @ 20.6V in delrin mount, 1-66-? scintillator w/ Am-241 source opposite to delrin mount, self trigger at 8275 threshold 10k 130717_sipm4_self-trig_8275threshold.tb4 Yes
2013/07/17130717_sipm4_am-241.datSiPM 4 @ 20.6V in delrin mount, 1-66-? scintillator w/ Am-241 source opposite to delrin mount, self trigger at 8300 threshold 10k 130717_sipm4_self-trig.tb4 Yes
2013/07/16130716_sipm4_am-241_x10.datSiPM 4 @ 20.6V in delrin mount, 1-66-? scintillator w/ Am-241 source opposite to delrin mount, self trigger at 8350 threshold 100k 130716_sipm4_self-trig_long.tb4 Yes
2013/07/15130715_sipm4_blueled_delayed-trig_2.00kohm.datSiPM 4 @ 20.6V (darkbox), 400ns delayed trigger on blue LED, 2.00kOhm attenuation10k130711_sipm4_LED_delay-trig_variable-resistance.tb4Yes
2013/07/12130712_sipm4_darkcount.datSiPM 4 @ 20.6V (darkbox), self-trigger on dark count/light leak at 8250 threshold, DAQ set to record 1024 samples per triggering (as opposed to default 100) 10k 130712_sipm4_self-trig.tb4 Yes
2013/07/11130711_sipm4_blueled_delayed-trig_1.693kohm.datSiPM 4 @ 20.6V (darkbox), 400ns delayed trigger on blue LED, 1.693kOhm attenuation10k130711_sipm4_LED_delay-trig_variable-resistance.tb4 Yes
2013/07/11130711_sipm4_blueled_delayed-trig_1.750kohm.datSiPM 4 @ 20.6V (darkbox), 400ns delayed trigger on blue LED, 1.750kOhm attenuation10k130711_sipm4_LED_delay-trig_variable-resistance.tb4 Yes
2013/07/11130711_sipm4_blueled_delayed-trig_1.807kohm.datSiPM 4 @ 20.6V (darkbox), 400ns delayed trigger on blue LED, 1.807kOhm attenuation10k130711_sipm4_LED_delay-trig_variable-resistance.tb4 Yes
2013/07/11130711_sipm4_blueled_delayed-trig_1.838kohm.datSiPM 4 @ 20.6V (darkbox), 400ns delayed trigger on blue LED, 1.838kOhm attenuation10k130711_sipm4_LED_delay-trig_variable-resistance.tb4 Yes
2013/07/11130711_sipm4_blueled_delayed-trig_1.906kohm.datSiPM 4 @ 20.6V (darkbox), 400ns delayed trigger on blue LED, 1.906kOhm attenuation10k130711_sipm4_LED_delay-trig_variable-resistance.tb4 Yes
2013/07/11130711_sipm4_blueled_delayed-trig_1.963kohm.datSiPM 4 @ 20.6V (darkbox), 400ns delayed trigger on blue LED, 1.963kOhm attenuation10k130711_sipm4_LED_delay-trig_variable-resistance.tb4 Yes
2013/07/10130711_sipm4_darkcount.datSiPM 4 @ 20.6V in darkbox, self-triggering on dark counts/light leaks at 8250 threshold 10k 130710_sipm4_self-trig.tb4 Yes
2013/07/10130710_sipm4_blueled_delayed-trig_10hz_100ns_x10_1.806kohm.datSiPM 4 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation100k130708_sipm1_LED_delay-trig_long.tb4 Yes
2013/07/09130709_sipm3_blueled_delayed-trig_10hz_100ns_x10_1.806kohm.datSiPM 3 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation100k130708_sipm1_LED_delay-trig_long.tb4 Yes
2013/07/09130709_sipm2_blueled_delayed-trig_10hz_100ns_x10_1.806kohm.dat SiPM 2 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation100k130708_sipm1_LED_delay-trig_long.tb4 Yes
2013/07/08130708_sipm1_blueled_delayed-trig_10hz_100ns_x10_1.806kohm.datSiPM 1 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation100k130708_sipm1_LED_delay-trig_long.tb4 Yes
2013/07/08130708_sipm1_blueled_delayed-trig_10hz_100ns_1.806kOhm.datSiPM 1 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation10k130708_sipm4_LED_delay-trig.tb4Yes
2013/07/08130708_sipm2_blueled_delayed-trig_10hz_100ns_1.806kOhm.datSiPM 2 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation10k130708_sipm4_LED_delay-trig.tb4Yes
2013/07/08130708_sipm4_blueled_delayed-trig-only.datSiPM 4 @ 20.6V in darkbox, external trigger w/ 200ns delay, no LED; short run to collect trigger cross-talk data 300 130708_sipm4_LED_delay-trig.tb4 Yes
2013/07/08130708_sipm3_blueled_delayed-trig_10hz_100ns_1.806kOhm.datSiPM 3 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation10k130708_sipm4_LED_delay-trig.tb4Yes
2013/07/08130708_sipm4_blueled_delayed-trig_10hz_100ns_1.806kOhm.datSiPM 4 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.806kOhm attenuation10k130708_sipm4_LED_delay-trig.tb4Yes
2013/07/08 SiPM header in darkbox is held in place with a spacer, allowing for more consistent positioning of SiPM sensor going forward
2013/07/08130708_sipm4_blueled_delayed-trig_10hz_100ns-2.datSiPM 4 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.623kOhm attenuation10k130708_sipm4_LED_delay-trig.tb4 Yes
2013/07/08130708_sipm4_blueled_delayed-trig_10hz_100ns.datSiPM 4 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.568kOhm attenuation10k130708_sipm4_LED_delay-trig.tb4 Yes
2013/07/05130705_sipm4_blueled_delayed-trig_10hz_100ns_x10.dat SiPM 4 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.493kOhm attenuation100k130705_sipm4_LED_delay-trig_long.tb4 Yes
2013/07/03130704_sipm1_blueled_delayed-trig_10hz_100ns_x10.dat SiPM 1 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.493kOhm attenuation100k130704_sipm1_LED_delay-trig_long.tb4 Yes
2013/07/03130703_sipm1_blueled_delayed-trig_10hz_100ns_1.dat SiPM 1 @ 20.6V in darkbox, external trigger w/ 200ns delay on blue LED pulse, 1.493kOhm attenuation10k130703_sipm1_LED_delay-trig.tb4 Yes
2013/07/02130702_blueled_ext-trig_10hz_100ns_x10_2.dat SiPM 2 @ 20.6V in darkbox, external trigger, blue LED, 1.487kOhm attenuation 100k 130702_sipm2_LED_ext-trigger_long.tb4 Yes

June 2013

Date Run Run Info # of Events / Files DAQ configuration file Analysed?
2013/06/28130628_blueLED_ext-trig_10hz_100ns_1.487kOhm_x10.dat SiPM 3 @ 20.6V in darkbox, external trigger, blue LED, 1.487kOhm attenuation 100k 130628_sipm3_LED_ext-trigger_long.tb4 Yes
2013/06/27130627_blueLED_ext-trig_10hz_100ns_x10_bias3800.dat SiPM 4 @ 21.5V in darkbox, external trigger, blue LED, 1.487kOhm attenuation 100k 130627_sipm4_LED_ext-trigger_long_newbias.tb4 Yes
2013/06/26130626_blueLED_ext-trig_10Hz_100ns_1.487kOhm_x10.dat SiPM 4 @ 20.6V in darkbox, external trigger, blue LED, 1.487kOhm attenuation 100k 130626_sipm4_LED_ext-trigger_long.tb4 Yes
2013/06/26130626_blueLED_ext-trig_10Hz_100ns_1.487kOhm.dat SiPM 4 @ 20.6V in darkbox, external trigger, blue LED, 1.487kOhm attenuation 10k 130625_sipm4_LED_ext-trigger.tb4 Yes
2013/06/26130626_led_ch1-trig_10Hz_100ns_1.323kOhm_offset-7e8.dat SiPM 4 @ 20.6V in darkbox, channel 1 trigger, blue LED, 1.323kOhm attenuation, adjusted DAQ offset to collect more pre-trigger baseline samples, 7500 threshold 10k 130626_sipm4_LED_ch1-trigger_threshold7500.tb4 Yes
2013/06/26130626_blueLED_ext-trig_10Hz_100ns_1.323kOhm.dat SiPM 4 @ 20.6V in darkbox, external trigger, blue LED, 1.323kOhm attenuation 10k 130625_sipm4_LED_ext-trigger.tb4 Yes
2013/06/25130625_blueLED_ext-trig_10Hz_100ns_1.323kOhm.dat SiPM 4 @ 20.6V in darkbox, external trigger, blue LED, 1.323kOhm attenuation 10k 130625_sipm4_LED_ext-trigger.tb4 Yes
2013/06/25130625_blueLED_ch1-trig_10Hz_100ns_1.323kOhm.dat SiPM 4 @ 20.6V in darkbox, channel 1 trigger, blue LED, 1.323kOhm attenuation, 7500 threshold 10k 130625_sipm4_LED_ch1-trigger-7500.tb4 Yes
2013/06/24130624_greenLED_threshold9000_10Hz_100ns_3.23kOhm[-run#].dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, green LED, 3.23kOhm attenuation 10k x 3 130624_sipm4_LED.tb4 Yes
2013/06/21130621_greenLED_threshold9000_10Hz_100ns_3.23kOhm.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, green LED, 3.23kOhm attenuation 10k 130621_sipm4_LED.tb4 Yes
2013/06/21130621_blueLED_threshold9000_10Hz_100ns_1.328kOhm.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, blue LED, 1.328kOhm attenuation 10k 130619_sipm4_blueLEDTest.tb4 Yes
2013/06/20130620_redLED_threshold9000_10Hz_100ns_4.68kOhm.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, red LED, 4.68kOhm attenuation 10k 130619_sipm4_redLEDTest.tb4 Yes
2013/06/20130620_blueLED_threshold9000_10Hz_100ns_1.66kOhm.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, blue LED, 1.66kOhm attenuation 10k 130619_sipm4_blueLEDTest.tb4 Yes
2013/06/20130620_greenLED_threshold9000_10Hz_100ns_3.42kOhm.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, green LED, 3.42kOhm attenuation 10k 130619_sipm4_yellowLEDTest.tb4 Yes
2013/06/19130619_greenLED_threshold9000_10Hz_100ns_3.88kOhm.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, green LED, 3.88kOhm attenuation 10k 130619_sipm4_yellowLEDTest.tb4 Yes
2013/06/19130619_yellowLED_threshold9000_10Hz_100ns_4.68kOhm.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 9000 threshold, yellow LED, 4.68kOhm attenuation 10k 130619_sipm4_yellowLEDTest.tb4 Yes
2013/06/19130619_yellowledtest_1_thesh8400.dat SiPM 4 @ 20.6V in darkbox, self-trigger @ 8400 threshold, yellow LED, 80 Ohm attenuation 10k 130618_sipm4_yellowLEDTest-2.tb4 Yes
2013/06/18130618_yellowledtest3_100ns-10Hz.dat SiPM 4 @ 20.6V in darkbox, self trigger @ 8900 threshold, yellow LED, 80 Ohm attenuation 10k 130618_sipm4_yellowLEDTest-100ns-10Hz.tb4 Yes
2013/06/18130618_yellowledtest2_1.dat SiPM 4 @ 20.6V in darkbox, self trigger @ 8400 threshold, yellow LED, 80 Ohm attenuation; This file is not analysed, but a run with identical settings is performed the next day 10k 130618_sipm4_yellowLEDTest-2.tb4 No
2013/06/17130617_yellowledtest_1.dat SiPM 4 @ 20.6V in darkbox, self trigger @ 8450 threshold, yellow LED, 80 Ohm attenuation 10k 130617_sipm4_yellowLEDTest.tb4 Yes
2013/06/16130616_greenfotest_1.dat SiPM 4 @ 20.6V in darkbox, self trigger @ 8200 threshold, fibre 10k 130516_sipm4_SelfTrigger.tb4 Yes

May 2013

Date Run Run Info # of Events / Files DAQ configuration file Analysed?
2013/05/17130517_sipm4_greentest_1.dat SiPM 4 @ 20.6V in darkbox, self-trigger on lightleak 10k 130516_sipm4_SelfTrigger.tb4 Yes