UMN CDMS Wiki archive
start › cdms › plasticscintillator › ebook › 20170119na22

Na-22 Measurements of Generic Scintillator BC410

Experimental Set Up

Identical setup as was used here. Runs were performed on BC410 generic plastic scintillator (1cm x 1.3cm x 10cm). Runs were done in 20mm intervals for 10 minutes each. This was done as an extrapolation of runs done on 20170110 to further investigate the attenuation coefficient of an unloaded scintillator.

Data

DMR-12D, 20mm

Figure 1: DMR-12 Measurements

Analysis

BC410 Generic Scintillator
Distance of Source from PMT (mm) Peak Location (nVs) Error (+/- nVs)
20 14.25 3.75
60 6.9 2.1
100 4.4 1.6
140 2.2 0.7
Figure 17: Data Summary, Generic Scinitllator
[image not archived: 20170119_na22_bc410_summary.png]

Conclusion