UMN CDMS Wiki archive
start › cdms › plasticscintillator › ebook › balowthreshold

Attempt to find full Ba-133 peak

Multiple runs with two SiPM at opposite ends of scintillator 1-74-7 with the wavelength-shifting fiber were conducted in the darkroom. Each event was self-triggered when both SiPMs were above the trigger threshold of 8320. When the Ba-133 source is present, it was placed 3 cm from the scintillator. Details of each run are shown below:

Table 1: Details about each low threshold run.
Run Channel 0 SiPM Channel 1 SiPM Source Number of events Data file
1 3 4 Ba-133 10k 130719_sipm3+4_ba-133_8320threshold.dat
2 3 4 Ba-133 100k 130721_sipm3+4_ba-133_8320threshold.dat
3 3 4 No source 100k 130722_sipm3+4_background_8320threshold.dat
4 2 4 Ba-133 10k 130722_sipm3+4_ba-133_8320threshold.dat
5 2 4 No source 10k 130722_sipm3+4_background_8320threshold_short.dat

Why SiPM 3 was changed to SiPM 2

Sometime between run 1 (2013/07/19) and run 2 (2013/07/21) something happened to SiPM 3.

Figure 1: Example of typical pulse from SiPM3 run 3.
Figure 2: Comparison of baseline standard deviations for example runs.

The baseline standard deviations from run 1 (Figure 2a) show SiPM 3 and 4 before SiPM went crazy. Figure 2b shows the baseline standard devations of SiPM 3 and 4 after the extra sinusoidal pattern was visible in the pulses of SiPM 3 (Figure 1). Figure 2c shows the baseline standard deviation of SiPM 2 and 4 after SiPM 3 was removed from the set-up.

Figure 3: Comparison of area under the pulse for SiPM 3 (run 1), SiPM 3 (run 2), and SiPM 2 (run 4).

The histogram of Run 2 in Figure 3 has been scaled by 1/10 to match the total number of events of Runs 1 and 4.

Ba-133 and Background runs with SiPM 3 and SiPM 4

Figure 4: Area under the pulse for SiPM 3 only, SiPM 3 only, and SiPM 3 and SiPM 4 combined.
Figure 5: Comparison of background and source spectra for SiPM 3 and SiPM 4.

All histograms involving SiPM 3 have been gain-matched to SiPM 4. For SiPM 3, pulses with a baseline standard deviation greater than 15 are excluded from the histograms. For SiPM 4, pulses with a baseline standard deviation greater than 7 are excluded. All histograms exclude data with secondary pulses present.

Ba-133 and Background runs with SiPM 2 and SiPM 4

Figure 6: Area under the pulse for SiPM 2 only, SiPM 4 only, and SiPM 2 and SiPM 4 combined.
Figures 7: Comparison of background and source spectra for SiPM 2 and SiPM 4.

All histograms above exclude data with a baseline standard deviation greater than 7 or with secondary peaks are present.