UMN CDMS Wiki archive
start › cdms › plasticscintillator › ebook › muontelescope

Muon Telescope

PMTs 0 and 3 were placed 6.5 cm apart vertically against the telescope scintillator with PMT 14 between them against scintillator 1-66-?. All PMTs were biased to -1100V. PMT 14 was triggered by a coincident event from PMTs 0 and 3. The coincident event was defined by a threshold of -25 mV with a 70 ns coincident window. The configuration of the telescope scintillator, the total time run, and the number of events varied by trial.

Table 1: Data runs
Run Date Configuration Number of events Total time (s) Data File
1 2013/11/14 parallel 754 57699 TEST.DAT
2 2013/11/15 perpendicular 3261 250863 1115-1118_cross_1100v_6.5cm.dat
3 2013/11/18 parallel 320 39469 1118-1119_parallel_6.5cm_1100v.dat
4 2013/11/22 perpendicular 1901 157812 1122-1124_sig1100v_trig1200v_cross_6.5cm.dat

Muon Energy loss calculation

The equation used for calculating the energy loss of a muon is (this equation was taken from the source listed below):

,

where is the energy of the muon and is the distance the muon has travelled.

R. Abbasi, et. al., An improved method for measuring muon energy using the truncated mean of dE/dx, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Volume 703, 1 March 2013, Pages 190-198, ISSN 0168-9002, http://dx.doi.org/10.1016/j.nima.2012.11.081.

The average energy of a muon at sea level was taken to be 4 GeV. The density of polystyrene is . The diameter of the plastic is 2.10 cm. The equation used to calculate the energy lost as the muon passes through the diameter of our polystyrene scintillator is

Energy Spectrum of the (long) Parallel Run

Figure 1: Area under the pulse fit to a gaussian.

Energy conversion: ,

where is energy in MeV and is pulse area in mVs.

Figure 2: Muon Energy Spectrum

Energy Spectrum of the (long) Perpendicular Run

Figure 3: Area under the pulse fit to a gaussian.

Energy conversion: ,

where is energy in MeV and is pulse area in mVs.

Figure 4: Muon Energy Spectrum

Energy Spectrum of the (short) Parallel Run

Figure 5: Area under the pulse fit to a gaussian.

Energy conversion: ,

where is energy in MeV and is pulse area in mVs.

Figure 6: Muon Energy Spectrum

Energy Spectrum of the (short) Perpendicular Run

Figure 7: Area under the pulse fit to a gaussian.

Energy conversion: ,

where is energy in MeV and is pulse area in mVs.

Figure 8: Muon Energy Spectrum

Comparison of Zero Energy Events

Figure 9:Calculation of Empty Events for the parallel (long) configuration.
Figure 10: Calculation of Empty Events for the perpendicular (long) configuration.
Figure 11: Calculation of Empty Events for the parallel (short) configuration.
Figure 12: Calculation of Empty Events for the perpendicular (short) configuration.

The number of empty events was calculated by taking the integral from the far left side of the histogram to the red line shown.

Table 2: Percentage of empty events
Run Configuration Empty Events Total Events Percentage
Parallel (long) 100 754 13.3%
Perpendicular (long) 679 3261 20.8%
Parallel (short) 24 320 7.5%
Perpendicular (short) 503 1901 26.5%

Rate Calculation

Table 3: Event Rate for each Run
Run Configuration Total Events Events less empties Total Time Total Rate Rate less empties
Parallel (long) 754 654 57699 s 0.0131 Hz 0.0113 Hz
Perpendicular (long) 3261 2582 250863 s 0.01300 Hz 0.01029 Hz
Parallel (short) 320 296 39469 s 0.00811 Hz 0.00750 Hz
Perpendicular (short) 1901 1398 157812 s 0.01204 Hz 0.008859 Hz