UMN CDMS Wiki archive
start › cdms › k100 › data

K100 Data Quality

For data quality diagnostics see Data Quality page

Latest DQ

K100 Data Summary

For a list and info of recent data series in online spreadsheet form: dataserieslist.xlsx

For a table of data series after Run 54 see Data series table from database

For a key between series times and MIDAS ODB run number see MIDAS run number key

For locations of raw data, primary and backup, see this spreadsheet (updated 2016 Aug 13)
Data backup instructions here

For a G106a analysis tasks summary see R54-55-56 analysis

Run 54 (G106a)

Run Run Mode Run Info # of Events/File Comments Analysis Processed
07150827_1512 Am-241
Pb-210
DIB1 -0 V, DIB2 +0 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 20 ADC bins; QETs set from 2015/08/27, Front End Gain 1; Output Gain 1; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150827_1523 Am-241
Pb-210
DIB1 -0 V, DIB2 +0 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 20 ADC bins; QETs set to zero (superconducting), Front End Gain 1; Output Gain 1; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150827_1527 Am-241
Pb-210
DIB1 -0 V, DIB2 +0 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 20 ADC bins; QETs set to 400/-400/400 uA on Dibs 1/2/3, Front End Gain 1; Output Gain 1; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150828_1544 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 20 ADC bins; QETs set from 2015/08/27, Front End Gain 1; Output/Driver Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150828_1547 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 20 ADC bins; QETs set from 2015/08/27, Front End Gain 1; Output/Driver Gain 3; ~7000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150831_1437 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1; Output Gain 3 2000 test data only (AK)
07150831_1519 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150831_1602 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 16/5/16 for Dibs 1/2/3 ; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.1/20 mins. 2000 (AK)
07150831_1642 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150831_1714 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150831_1745 Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150901_1328
07150901_1402
07150901_1433
07150901_1504
Am-241
Pb-210
DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150901_1607
07150901_1638
07150901_1734
Am-241
Pb-210
DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. 2000 (AK)
07150904_1654 Am-241
Pb-210
HV = 0 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition= 10 mins, no flash after. 2000 HV data (AK)
07150904_1708 Am-241
Pb-210
HV = 7.00 V / 11.5 nA.; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition= 11 mins, no flash after. 2000 HV data (AK)
07150904_1724 Am-241
Pb-210
HV = 15.00 V / 24.8 nA; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/08/31; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition= 21 mins, no flash after. 2000 HV data (AK)
07150908_1117 Am-241
Pb-210
HV = +5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition= 10 mins, 30 sec flash after. 2000 HV data (MF)
07150908_1152 Am-241
Pb-210
HV = +10 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition= 10 mins, 30 sec flash after. 2000 HV data (MF)
07150908_1227 Am-241
Pb-210
HV = +11.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 1 ; Output Gain 3; 1000 initial randoms ; acquisition= 10 mins, 30 sec flash after. 2000 HV data (MF)
07150908_1714
07150908_1801
07150908_1935
07150908_2021
07150908_2108
07150908_2155
07150908_2241
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 16/5/16; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (MF/AK)
07150909_1154 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 16/5/16; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 2/0.5/30 mins. 2000 Randoms for Front-end sweeps (AK)
07150909_1217 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 10/3/10; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 2/0.5/30 mins. 2000 Randoms for Front-end sweeps (AK)
07150909_1240 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 2/0.5/30 mins. 2000 Randoms for Front-end sweeps (AK)
07150909_1303 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/08; QETs set from 2015/08/27, Front End Gain 1/1/1; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 2/0.5/30 mins. 2000 Randoms for Front-end sweeps (AK)
07150909_1839
07150909_2012
07150909_2145
07150909_2319
Cf-252
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels(except pB1), thresholds set from 2015/09/09; QETs set from 2015/08/27, Front End Gain 16/5/16; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 Doesn't Process Correctly (AK)
07150909_1925
07150909_2059
07150909_2232
07150910_0005
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels(except pB1), thresholds set from 2015/09/09; QETs set from 2015/08/27, Front End Gain 16/5/16; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 Doesn't Process Correctly (AK)
07150910_1154 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/09; QETs set from 2015/08/27, Front End Gain 16/5/16; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AK)
07150910_1241 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/09; QETs set from 2015/08/27, Front End Gain 16/5/16; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AK)
07150915_1741
07150915_1915
07150915_2048
07150915_2221
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/15; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV/AK)
07150915_1828
07150915_2001
07150915_2134
07150915_2307
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/15; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV/AK)
07150916_1010 Cf-252
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/15; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 Bad data rate, data aborted (autoprocessed)
07150916_1237
07150916_1410
07150916_1737
07150916_1910
Cf-252
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/15; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AK)
-
(AV)
(AV)
07150916_1323
07150916_1458
07150916_1824
07150916_1957
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/15; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 -
=
(AV)
(AV)
07150917_1151 Ba-133
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/15; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AK)
07150917_1238 Ba-133
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/15; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AK)
07150917_1417 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/17; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AK)
07150917_1458 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/17; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AK)
07150917_1654
07150917_1827
07150917_2000
07150917_2134
Cf-252
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/17; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150917_1740
07150917_1914
07150917_2047
07150917_2220
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/17; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150921_1505 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test fails because trace length set incorrectly in MultiDAQ
07150921_1530 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias=0, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, s/c fails because trace length set incorrectly in MultiDAQ
07150921_1602 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias=400 uA, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, resistive fails because trace length set incorrectly in MultiDAQ
07150921_1634 Am-241
Pb-210
DIB1 0V, DIB2 0V;50 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test fails because trace length set incorrectly in MultiDAQ
07150922_1141 Am-241
Pb-210
DIB1 0V, DIB2 0V; 50 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test fails because trace length set incorrectly in MultiDAQ
07150922_1232 Am-241
Pb-210
DIB1 0V, DIB2 0V; 50 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias=0, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, s/c fails because trace length set incorrectly in MultiDAQ
07150922_1336 Am-241
Pb-210
DIB1 0V, DIB2 0V; 50 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias=400 uA, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, resistive fails because trace length set incorrectly in MultiDAQ
07150922_1818
07150922_1951
07150922_2124
07150922_2258
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150922_1905
07150922_2038
07150922_2211
07150922_2344
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150923_1732
07150923_1802
07150923_1810
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 automation test, short series (AK) - failed to process
07150924_1104
07150924_1109
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 automation test, short series (AK) - failed to process
07150924_1205 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test (MF) (AV) - failed to process, empty root files
07150924_1231 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias = 0, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, s/c (MF) (AV) - failed to process, empty root files
07150924_1305 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias = 400 uA, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, resistive (MF) (AV) - failed to process, empty root files
07150924_1415 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, DCRCs stacked (MF) (AV) - failed to process, empty root files
07150924_1703 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, no filters (MF) (AV) - failed to process, empty root files
07150924_1728 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias = 0, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, no filters, s/c (MF) (AV) - failed to process, empty root files
07150924_1752 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27 except QET bias = 400 uA, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms 2000 long-trace noise test, no filters, resistive (MF) (AV) - failed to process, empty root files
07150925_1633 Ba-133
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150925_1736
07150925_1829
07150925_1921
07150925_2014
07150925_2107
07150925_2317
07150926_0010
07150926_0102
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150928_1315
07150928_1403
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/25; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150928_1716
07150928_1803
07150928_1851
07150928_1939
07150928_2027
07150928_2115
07150928_2203
07150928_2251
07150928_2338
07150929_0026
07150929_0114
07150929_0202
07150929_0249
07150929_0337
07150929_0426
07150929_0513
07150929_0602
07150929_0649
07150929_0737
07150929_0825
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/25; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150929_1120
07150929_1125
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/25; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 automation test (AV)
07150929_1131
07150929_1245
07150929_1253
07150929_1300
07150929_1304
07150929_1315
07150929_1550
07150929_1640
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/25; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 automation test (AV)
07150929_1716
07150929_1803
07150929_1851
07150929_1939
07150929_2028
07150929_2116
07150929_2204
07150929_2252
07150929_2340
07150930_0028
07150930_0116
07150930_0204
07150930_0251
07150930_0339
07150930_0428
07150930_0516
07150930_0604
07150930_0652
07150930_0740
07150930_0828
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, all thresholds set to 65 ADC; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07150930_1236 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/25; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 automation test (AV)
07150930_1248
07150930_1336
07150930_1424
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/25; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 automation test (AV)
07150930_1740
07150930_1827
07150930_1915
07150930_2004
07150930_2052
07150930_2140
07150930_2228
07150930_2316
07151001_0004
07151001_0052
07151001_0140
07151001_0228
07151001_0315
07151001_0404
07151001_0452
07151001_0540
07151001_0628
07151001_0716
07151001_0804
07151001_0852
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/09/25; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (MF)
07151002_1711
07151002_1759
07151002_1847
07151002_1935
07151002_2022
07151002_2111
07151002_2159
07151002_2247
07151002_2335
07151003_0023
07151003_0111
07151003_0159
07151003_0247
07151003_0335
07151003_0423
07151003_0512
07151003_0559
07151003_0648
07151003_0736
07151003_0824
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 Lost Side1pF (DIB1pB) during first series, dead throughout entire set. So this is all bad data. (MF)
07151003_1550
07151003_1637
07151003_1726
07151003_1814
07151003_1902
07151003_1949
07151003_2037
07151003_2126
07151003_2214
07151003_2302
07151003_2350
07151004_0038
07151004_0126
07151004_0215
07151004_0302
07151004_0350
07151004_0438
07151004_0527
07151004_0615
07151004_0703
07151004_0751
07151004_0839
07151004_0927
07151004_1016
07151004_1103
07151004_1151
07151004_1239
07151004_1328
07151004_1416
07151004_1504
07151004_1552
07151004_1640
07151004_1728
07151004_1817
07151004_1904
07151004_1952
07151004_2040
07151004_2129
07151004_2217
07151004_2304
07151004_2353
07151005_0041
07151005_0129
07151005_0217
07151005_0305
07151005_0353
07151005_0441
07151005_0529
07151005_0618
07151005_0705
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 Lost Side1pF (DIB1pB) between 1004_1552 and 1004_1640. See the Oct 4 PSD trends. See also 2015/10/05 (Monday) run diary entry for mentions of problems with this channel. So the last 19 or maybe 20 of these 50 series are bad. (AV)
07151005_1330 Ba-133
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151005_1418 Ba-133
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151005_1527 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 aborted, DIB1 pB problem (AV)
07151005_1550
07151005_1637
07151005_1726
07151005_1814
07151005_1902
07151005_1950
07151005_2037
07151005_2126
07151005_2214
07151005_2302
07151005_2350
07151006_0038
07151006_0127
07151006_0215
07151006_0303
07151006_0350
07151006_0438
07151006_0527
07151006_0616
07151006_0703
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 Side1pF railed by the beginning of _1814
possible during _1726
Remains railed for all subsequent series
(AV)
07151006_1505 Am-241
Pb-210
DIB1 0V, DIB2 0V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 randoms 2000 front-end-gain scan, randoms only
07151006_1559 Am-241
Pb-210
DIB1 0V, DIB2 0V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 15/5/15; Output Gain 3; 1000 randoms 2000 front-end-gain scan, randoms only. Series aborted, problems with railing. (AV)
07151007_1142 Am-241
Pb-210
DIB1 0V, DIB2 0V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/07; QETs set from 2015/08/27, Front End Gain 15/5/15; Output Gain 3; 1000 randoms 2000 front-end-gain scan, randoms only (AV)
07151007_1146 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/07; QETs set from 2015/08/27, Front End Gain 15/5/15; Output Gain 3; 1000 randoms 2000 front-end-gain scan, randoms only (AV)
07151007_1305 Am-241
Pb-210
DIB1 0V, DIB2 0V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/07; QETs set from 2015/08/27, Front End Gain 30/10/30; Output Gain 3; 1000 randoms 2000 front-end-gain scan, randoms only (AV)
07151007_1641
07151007_1729
07151007_1817
07151007_1905
07151007_1953
07151007_2041
07151007_2130
07151007_2217
07151007_2306
07151007_2354
07151008_0042
07151008_0130
07151008_0218
07151008_0306
07151008_0354
07151008_0442
07151008_0531
07151008_0619
07151008_0707
07151008_0755
07151008_0843
Cf-252
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151008_1058
07151008_1145
Ba-133
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151008_1429 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 2000 initial randoms 2000 long-trace randoms (AV) failed to process, empty Root files
07151008_1515 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 2000 initial randoms 2000 long-trace randoms, DCRCs stacked (AV) failed to process, empty Root files
07151008_1549
07151008_1608
Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 2000 initial randoms 2000 long-trace randoms, DCRCs spread out (AV) failed to process, empty Root files
07151008_1657
07151008_1745
07151008_1833
07151008_1921
07151008_2009
07151008_2056
07151008_2145
07151008_2234
07151008_2322
07151009_0010
07151009_0058
07151009_0146
07151009_0235
07151009_0322
07151009_0410
07151009_0458
07151009_0547
07151009_0636
07151009_0723
07151009_0812
07151009_0900
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 DCRCs are still spread out, like the previous test
started with noticeable 250Hz noise pattern
(AV)
07151009_1029
07151009_1117
07151009_1205
07151009_1253
07151009_1342
Ba-133
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151009_1654 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 Series aborted early because DIB1pB was lost (AV)
07151010_1641
07151010_1728
07151010_1817
07151010_1905
07151010_1953
07151010_2041
07151010_2129
07151010_2218
07151010_2306
07151010_2354
07151011_0043
07151011_0130
07151011_0219
07151011_0307
07151011_0355
07151011_0443
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 DIB1pB lost during or after first series, so this is all bad data (AV) second series failed, empty Root files
07151011_1148
07151011_1236
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/02; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 DAQDAQ problem, first series OK, second failed to process, then vi locked up (AV)
07151012_1648
07151012_1756
07151012_1848
07151012_1939
07151012_2028
07151012_2119
07151012_2209
07151012_2259
07151012_2350
07151013_0040
07151013_0129
07151013_0219
07151013_0311
07151013_0401
07151013_0452
07151013_0542
07151013_0632
07151013_0722
07151013_0813
07151013_0902
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151013_1450
07151013_1534
07151013_1620
07151013_1704
07151013_1749
07151013_1834
07151013_1918
07151013_2006
07151013_2051
07151013_2136
07151013_2221
07151013_2306
07151013_2352
07151014_0037
07151014_0122
07151014_0207
07151014_0252
07151014_0336
07151014_0423
07151014_0508
07151014_0554
07151014_0639
Cf-252
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/25 mins. 2000 cooldown time reduced to compensate for vi modification; in the new version a 25 minute setting corresponds to 30 minutes of actual cooldown (AV)
07151014_1156
07151014_1242
07151014_1651
07151014_1742
07151014_1826
07151014_1911
07151014_1956
07151014_2040
07151014_2127
07151014_2212
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 It's possible that the last series in this list had a problem with DIB1pB (Side1pF) (AV)
07151014_2258
07151014_2343
07151015_0028
07151015_0113
07151015_0158
07151015_0243
07151015_0327
07151015_0412
07151015_0459
07151015_0545
07151015_0629
07151015_0714
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (Continuation of set from above.)

DIB1pB (Side1pF) is bad for all these series. For the first 8 it exhibited a test pattern; for the remaining 4 it was flatline
(AV); first 8 failed to process (empty Root files) probably because of the DIB1pB behavior.
07151016_1117
07151016_1128
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 45/0.5/30 mins. 2000 problems with channels, aborted early (AV)
07151016_1134 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 45/0.5/30 mins. 2000 long series test for neutralization study (AV)
07151016_1324
07151016_1409
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151016_1704
07151016_1712
07151016_1757
07151016_1842
07151016_1927
07151016_2011
07151016_2058
07151016_2143
07151016_2228
07151016_2313
07151016_2359
07151017_0044
07151017_0129
07151017_0214
07151017_0259
07151017_0344
07151017_0428
07151017_0513
07151017_0601
07151017_0646
07151017_0731
07151017_0816
07151017_0902
07151017_0946
07151017_1032
07151017_1117
07151017_1202
07151017_1246
07151017_1331
07151017_1418
07151017_1504
07151017_1548
07151017_1634
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151018_1018 Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 series stopped early (started automatically during transfer) (AV)
07151018_1136
07151018_1222
07151018_1307
07151018_1352
07151018_1437
07151018_1521
07151018_1609
07151018_1654
07151018_1739
07151018_1824
07151018_1909
07151018_1955
07151018_2040
07151018_2124
07151018_2209
07151018_2256
07151018_2340
07151019_0025
07151019_0110
07151019_0154
07151019_0239
07151019_0327
07151019_0412
07151019_0458
07151019_0542
07151019_0628
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/12; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 (AV)
07151019_1159
07151019_1244
07151019_1329
07151019_1414
07151019_1500
07151019_1544
07151019_1630
Am-241
Pb-210
see Comments for biases; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 15/0.5/30 mins. 2000 DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
DIB1 0V, DIB2 0V
DIB1 -6V, DIB2 +6V
DIB1 -8V, DIB2 +8V
(AV)
07151019_1710
07151019_1810
07151019_1912
07151019_2011
07151019_2112
07151019_2212
07151019_2313
Am-241
Pb-210
see Comments for biases; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
DIB1 0V, DIB2 0V
DIB1 -6V, DIB2 +6V
DIB1 -8V, DIB2 +8V
(AV)
07151020_0013
07151020_0113
07151020_0213
07151020_0313
07151020_0414
07151020_0514
07151020_0615
Am-241
Pb-210
see Comments for biases; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
DIB1 0V, DIB2 0V
DIB1 -6V, DIB2 +6V
DIB1 -8V, DIB2 +8V
(AV)
07151020_1102
07151020_1202
07151020_1301
07151020_1406
07151020_1506
07151020_1606
07151020_1706
07151020_1807
07151020_1907
07151020_2007
07151020_2107
07151020_2208
07151020_2308
07151021_0009
07151021_0109
07151021_0209
07151021_0309
07151021_0409
07151021_0509
07151021_0611
07151021_0711
Ba-133
Am-241
Pb-210
see Comments for biases; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/20; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
DIB1 0V, DIB2 0V
DIB1 -6V, DIB2 +6V
DIB1 -8V, DIB2 +8V
DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
DIB1 0V, DIB2 0V
DIB1 -6V, DIB2 +6V
DIB1 -8V, DIB2 +8V
DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
DIB1 0V, DIB2 0V
DIB1 -6V, DIB2 +6V
DIB1 -8V, DIB2 +8V
(AV)
Strong Pu-Be source neutron activation, 65 minutes, 2015/10/21 9:40 - 10:45
07151021_1202 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 45/0.5/30 mins. 2000 long series neutralization test (AV)
07151021_1650 aborted early
07151021_1655
07151021_1755
07151021_1855
07151021_1956
07151021_2055
07151021_2155
07151021_2256
07151021_2357
07151022_0057
07151022_0157
07151022_0257
07151022_0358
07151022_0457
07151022_0557
07151022_0658
07151022_0758
07151022_0859
07151022_0959
Am-241
Pb-210
see Comments for biases; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 DIB1 -8V, DIB2 +8V
DIB1 -6V, DIB2 +6V
DIB1 -4V, DIB2 +4V
DIB1 -2V, DIB2 +2V
DIB1 0V, DIB2 0V
DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
DIB1 -8V, DIB2 +8V
DIB1 -6V, DIB2 +6V
DIB1 -4V, DIB2 +4V
DIB1 -2V, DIB2 +2V
DIB1 0V, DIB2 0V
DIB1 +2V, DIB2 -2V
DIB1 +4V, DIB2 -4V
DIB1 +6V, DIB2 -6V
DIB1 +8V, DIB2 -8V
(AV)
07151022_1137 aborted early
07151022_1139 Am-241
Pb-210
DIB1 -8V, DIB2 +8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

0/0 V → high bias setting
(AV)
07151022_1144 Am-241
Pb-210
DIB1 -6V, DIB2 +6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

0/0 V → high bias setting
(AV)
07151022_1150 Am-241
Pb-210
DIB1 +8V, DIB2 -8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

0/0 V → high bias setting
(AV)
07151022_1155 Am-241
Pb-210
DIB1 +6V, DIB2 -6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

0/0 V → high bias setting
(AV)
07151022_1201 Am-241
Pb-210
DIB1 -8V, DIB2 +8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

after setting + side to +14V, then down to high bias setting
(AV)
07151022_1206 Am-241
Pb-210
DIB1 +8V, DIB2 -8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

after setting + side to +14V, then down to high bias setting
(AV)
07151022_1212 Am-241
Pb-210
DIB1 -6V, DIB2 +6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

after setting + side to +14V, then down to high bias setting
(AV)
07151022_1217 Am-241
Pb-210
DIB1 +6V, DIB2 -6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

after setting + side to +14V, then down to high bias setting
(AV)
07151022_1502 Am-241
Pb-210
DIB1 -6V, DIB2 +6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

0/0 V and flash → high bias setting
(AV)
07151022_1530 Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

0/0 V and flash → high bias setting
(AV)
07151022_1644
07151022_1744
07151022_1845
07151022_1945
07151022_2045
07151022_2144
07151022_2246
07151022_2346
07151023_0046
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 (AV)
07151023_1057 Am-241
Pb-210
DIB1 -8V, DIB2 +8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 set + side to +14 V first before starting series (AV)
07151023_1158 Am-241
Pb-210
DIB1 +8V, DIB2 -8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 set + side to +14 V first before starting series (AV)
07151023_1303 Am-241
Pb-210
DIB1 -6V, DIB2 +6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 set + side to +14 V first before starting series (AV)
07151023_1409 Am-241
Pb-210
DIB1 +6V, DIB2 -6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 set + side to +14 V first before starting series (AV)
07151023_1522 Am-241
Pb-210
DIB1 -10V, DIB2 +10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 set + side to +14 V first before starting series (AV)
07151023_1624 Am-241
Pb-210
DIB1 +10V, DIB2 -10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/19; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 set + side to +14 V first before starting series (AV)
07151023_1747
07151023_1751
07151023_1752
07151023_1754
07151023_1756
aborted early
07151023_1757
07151023_1857
07151023_1957
07151023_2056
07151023_2156
07151023_2256
07151023_2356
07151024_0056
07151024_0156
07151024_0256
07151024_0355
07151024_0455
07151024_0555
07151024_0655
07151024_0755
07151024_0855
07151024_0954
07151024_1054
07151024_1154
07151024_1254
07151024_1353
07151024_1454
07151024_1553
07151024_1653
07151024_1753
07151024_1853
07151024_1953
07151024_2053
07151024_2152
07151024_2253
07151024_2353
07151025_0052
07151025_0152
07151025_0252
07151025_0351
07151025_0452
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 partial
07151025_1201 aborted early
07151025_1202
07151025_1302
07151025_1401
07151025_1501
07151025_1600
07151025_1700
07151025_1800
07151025_1859
07151025_1959
07151025_2059
07151025_2158
07151025_2258
07151025_2358
07151026_0057
07151026_0157
07151026_0257
Am-241
Pb-210
DIB1 +2V, DIB2 -2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 partial
07151026_1047 aborted early
07151026_1051 Ba-133
Am-241
Pb-210
DIB1 -8V, DIB2 +8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 noise test, randoms only

+14V overshoot on + side
07151026_1131
07151026_1148
aborted early
07151026_1210 Ba-133
Am-241
Pb-210
DIB1 -8V, DIB2 +8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/26; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151026_1311 Ba-133
Am-241
Pb-210
DIB1 +8V, DIB2 -8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/26; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151026_1412 Ba-133
Am-241
Pb-210
DIB1 -6V, DIB2 +6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/26; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151026_1613 Ba-133
Am-241
Pb-210
DIB1 +6V, DIB2 -6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/26; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151026_1714
07151026_1817
07151026_1916
07151026_2016
07151026_2117
07151026_2217
07151026_2317
07151027_0018
07151027_0118
07151027_0218
07151027_0318
07151027_0418
07151027_0519
07151027_0618
07151027_0718
Cf-252
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/26; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151027_1119 Am-241
Pb-210
DIB1 -8V, DIB2 +8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151027_1223 Am-241
Pb-210
DIB1 +8V, DIB2 -8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151027_1326 Am-241
Pb-210
DIB1 -6V, DIB2 +6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151027_1428 Am-241
Pb-210
DIB1 +6V, DIB2 -6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151027_1535
07151027_1639
07151027_1739
07151027_1838
07151027_1938
07151027_2039
07151027_2139
07151027_2238
07151027_2339
07151028_0039
07151028_0138
07151028_0238
07151028_0338
07151028_0440
07151028_0540
07151028_0640
07151028_0740
Cf-252
Am-241
Pb-210
DIB1 +2V, DIB2 -2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151028_1228 Ba-133
Am-241
Pb-210
DIB1 -10V, DIB2 +10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151028_1431 Ba-133
Am-241
Pb-210
DIB1 +10V, DIB2 -10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151028_1640
07151028_1740
07151028_1840
07151028_1945
07151028_2044
07151028_2144
07151028_2244
07151028_2344
07151029_0044
07151029_0144
07151029_0244
07151029_0345
07151029_0445
07151029_0544
07151029_0646
07151029_0745
Cf-252
Am-241
Pb-210
DIB1 -2V, DIB2 +2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151029_1000 Ba-133
Am-241
Pb-210
DIB1 -8V, DIB2 +8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151029_1101 Ba-133
Am-241
Pb-210
DIB1 +8V, DIB2 -8V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151029_1218 Ba-133
Am-241
Pb-210
DIB1 -6V, DIB2 +6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151029_1318 Ba-133
Am-241
Pb-210
DIB1 +6V, DIB2 -6V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151029_1430 Ba-133
Am-241
Pb-210
DIB1 -10V, DIB2 +10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151029_1547 Ba-133
Am-241
Pb-210
DIB1 +10V, DIB2 -10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151029_1653
07151029_2152
07151030_0253
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151029_1752
07151029_1852
07151029_2252
07151029_2351
07151030_0352
07151030_0452
Cf-252
Am-241
Pb-210
DIB1 +2V, DIB2 -2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151029_1952
07151029_2052
07151030_0051
07151030_0153
07151030_0552
07151030_0652
Cf-252
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151030_1141 Am-241
Pb-210
DIB1 -10V, DIB2 +10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151030_1254 Am-241
Pb-210
DIB1 +10V, DIB2 -10V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 +14V overshoot on + side
07151030_1505
07151030_1605
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151030_1709
07151030_1910
07151030_2111
07151030_2311
07151031_0112
07151031_0313
07151031_0513
07151031_0714
07151031_0914
07151031_1115
07151031_1316
07151031_1516
07151031_1716
07151031_1917
07151031_2118
Cf-252
Am-241
Pb-210
DIB1 +2V, DIB2 -2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151030_1809
07151030_2011
07151030_2211
07151031_0012
07151031_0211
07151031_0413
07151031_0613
07151031_0814
07151031_1013
07151031_1215
07151031_1416
07151031_1616
07151031_1816
07151031_2018
07151031_2218
Cf-252
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151101_0851
07151101_0950
07151101_1050
07151101_1150
07151101_1251
07151101_1351
07151101_1452
07151101_1552
07151101_1652
Am-241
Pb-210
DIB1 +2V, DIB2 -2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151101_1752 Am-241
Pb-210
aborted early (DAQ glitch)
07151102_1155 Am-241
Pb-210
DIB1 +1V, DIB2 -1V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151102_1255 Am-241
Pb-210
DIB1 -1V, DIB2 +1V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151102_1357 Am-241
Pb-210
DIB1 +3V, DIB2 -3V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151102_1457 Am-241
Pb-210
DIB1 -3V, DIB2 +3V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151102_1723
07151102_2123
Ba-133
Am-241
Pb-210
DIB1 +1V, DIB2 -1V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151102_1823
07151102_2224
Ba-133
Am-241
Pb-210
DIB1 -1V, DIB2 +1V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151102_1923
07151102_2324
Ba-133
Am-241
Pb-210
DIB1 +3V, DIB2 -3V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151102_2023
07151103_0025
Ba-133
Am-241
Pb-210
DIB1 -3V, DIB2 +3V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/28; QETs set from 2015/08/27, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151103_1009 Am-241
Pb-210
DIB1 +1V, DIB2 -1V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151103_1109 Am-241
Pb-210
DIB1 -1V, DIB2 +1V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151103_1209 Am-241
Pb-210
DIB1 +3V, DIB2 -3V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151103_1308 Am-241
Pb-210
DIB1 -3V, DIB2 +3V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151103_1529 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20.0 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/11/03; QETs set from 2015/08/21 except QETbias = 400 uA, Front End Gain 5/2/5; Output Gain 0/1/0; 2000 randoms 2000 long-trace noise test

normal state, low output gain
07151103_1542 Am-241
Pb-210
aborted early
07151103_1544 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20.0 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/11/03; QETs set from 2015/08/21 except QETbias = 400 uA, Front End Gain 5/2/5; Output Gain 0/1/0; 2000 randoms 2000 long-trace noise test

normal state, low output gain
07151103_1703
07151103_2001
07151103_2302
07151104_0201
07151104_0500
Cf-252
Am-241
Pb-210
DIB1 -4V, DIB2 +4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151103_1802
07151103_2101
07151104_0001
07151104_0301
07151104_0600
Cf-252
Am-241
Pb-210
DIB1 +2V, DIB2 -2V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151103_1902
07151103_2201
07151104_0102
07151104_0401
07151104_0700
Cf-252
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151104_1110 Y-88
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 short series
07151104_1300
07151104_1400
07151104_1500
07151104_1559
Y-88
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
07151105_1413
07151105_1514
07151105_1624
07151105_1908
07151105_2008
07151105_2108
07151105_2208
07151105_2308
07151106_0008
07151106_0108
07151106_0207
07151106_0307
07151106_0407
07151106_0507
07151106_0607
07151106_0707
Y-88
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 at the end of Dump 5 of _1514 the timing link to DIB2 was temporarily broken
06_0707 series has very high low-frequency noise
07151106_1128
07151106_1224
Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; QETs set from 2015/08/21 except QET bias 400/-400/400 uA (DIB1pA 500 uA), Front End Gain 5/2/5 except pD's =1/1/1; Output Gain 3; 1000 initial randoms. 2000 long-trace noise test
normal state, front-end gain scan
07151106_1254
07151106_1256
Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; QETs set from 2015/08/21 except QET bias 400/-400/400 uA (DIB1pA 500 uA), Front End Gain 5/2/5 except pD's = 15/5/15; Output Gain 3; 1000 initial randoms. 2000 long-trace noise test
normal state, front-end gain scan
07151106_1356
07151106_1359
Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; QETs set from 2015/08/21 except QET bias 400/-400/400 uA (DIB1pA 500 uA), Front End Gain 5/2/5 except pD's = 25/5/25; Output Gain 3; 1000 initial randoms. 2000 long-trace noise test
normal state, front-end gain scan
07151106_1455 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; QETs set from 2015/08/21 except QET bias 400/-400/400 uA (DIB1pA 500 uA), Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms. 2000 long-trace noise test
normal state, front-end gain scan
07151106_1552 Am-241
Pb-210
2000 aborted early
07151106_1610 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; QETs set from 2015/08/21 except pC QET biases 400/-400/400 uA, pD QET biases 0, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms. 2000 long-trace noise test
various TES states
07151110_1110 Am-241
Pb-210
aborted early
07151110_1120 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; Open Loop output, QETs set from 2015/08/21 except QET bias=400 uA for pCs, 0 uA for pDs, Front End Gain 5/2/5; Output Gain 3; randoms only long-trace noise test
07151110_1446 Am-241
Pb-210
aborted early
07151110_1448 Am-241
Pb-210
DIB1 0V, DIB2 0V; 20 ms trace length; Open Loop output, QETs set from 2015/08/21 except QET bias=400 uA for pCs, 0 uA for pDs, DIB1 SQUID bias/lockpts reduced to 1/2 modulation depth, Front End Gain 5/2/5; Output Gain 3; randoms only long-trace noise test
07151110_1705 Am-241
Pb-210
aborted early
07151110_1710 Am-241
Pb-210
aborted early
07151110_1712
07151110_1812
07151110_1911
07151110_2011
07151110_2111
07151110_2210
07151110_2310
07151111_0010
07151111_0109
07151111_0209
07151111_0309
07151111_0408
07151111_0508
07151111_0608
07151111_0708
07151111_0807
Y-88
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set from 2015/10/21; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 DIB2pB aka Side2pA was railed some time before the beginning of the second series. (It's an outer channel so arguably not the most important, but you can't construct a proper total phonon energy so generally this is bad data.)
07151112_1736 Y-88
Am-241
Pb-210
aborted early
07151112_1806
07151112_1907
07151112_2007
07151112_2107
07151112_2207
07151112_2307
07151113_0008
07151113_0107
07151113_0207
07151113_0307
07151113_0406
07151113_0506
07151113_0606
07151113_0705
Y-88
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set to 65 ADC; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 DIB2pB aka Side2pA was railed for all these series. (It's an outer channel so arguably not the most important, but you can't construct a proper total phonon energy so generally this is bad data.)
07151113_1234
07151113_1334
07151113_1523
07151113_1625
07151113_1725
07151113_1826
07151113_1926
07151113_2026
07151113_2126
07151113_2226
07151113_2326
07151114_0027
07151114_0127
07151114_0228
07151114_0328
07151114_0428
07151114_0528
07151114_0628
07151114_0728
07151114_0829
07151114_0930
07151114_1030
07151114_1130
07151114_1230
07151114_1330
07151114_1430
07151114_1530
07151114_1631
07151114_1732
07151114_1832
07151114_1932
07151114_2032
07151114_2133
07151114_2232
07151114_2332
07151115_0033
07151115_0134
07151115_0234
07151115_0334
07151115_0434
07151115_0535
07151115_0634
07151115_0734
07151115_0835
07151115_1023
07151115_1123
07151115_1224
07151115_1324
07151115_1424
07151115_1523
07151115_1625
07151115_1724
07151115_1824
07151115_1924
07151115_2024
07151115_2123
07151115_2223
07151115_2327
07151116_0026
07151116_0127
07151116_0227
07151116_0327
07151116_0428
07151116_0527
07151116_0627
07151116_0727
Y-88
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set to 65 ADC; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000 ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
high low-frequency noise
high low-frequency noise
high low-frequency noise
ok
ok
ok
high low-frequency noise
high low-frequency noise
high low-frequency noise
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
ok
railed channels
railed channels
railed channels
railed channels
07151116_1105
07151116_1205
07151116_1305
07151116_1504
07151116_1605
07151116_1705
07151116_1805
07151116_1906
07151116_2007
07151116_2107
07151116_2207
07151116_2307
07151117_0008
07151117_0107
07151117_0207
07151117_0308
07151117_0409
07151117_0509
07151117_0609
07151117_0710
Y-88
Am-241
Pb-210
DIB1 +4V, DIB2 -4V; 3.2768 ms trace length; 400us delay; OR trigger on phonon channels, thresholds set to 65 ADC; QETs set from 2015/08/21, Front End Gain 5/2/5; Output Gain 3; 1000 initial randoms ; acquisition/flash/cooldown = 30/0.5/30 mins. 2000
Not-identified-as-bad data amounts, hours (2015 November 3)
Biases (V): -10/+10 -8/+8 -6/+6 -4/+4 -3/+3 -2/+2 -1/+1 0/0 +1/-1 +2/-2 +3/-3 +4/-4 +6/-6 +8/-8 +10/-10
Background 1.0 1.0 1.0 28.8 1.0 31.0 1.0 2.3 1.0 14.8 1.0 20.3 1.0 1.0 1.0
Cf 0.0 0.0 0.0 20.0 0.0 20.8 0.0 0.0 0.0 21.5 0.0 20.5 0.0 0.0 0.0
Ba 1.0 1.0 1.0 2.0 1.0 1.0 1.0 1.5 1.0 1.5 1.0 1.5 1.0 1.0 1.0

36 series of Y-88 data at +4/-4 V = 18 hours

MIDAS data

Data series were also taken for Run 54 using the MIDAS DAQ. They are not included in the list above but they are included in the database:

Data series table from database

Note that this list also contains data taken with the LabVIEW DAQ and currently the table does not distinguish between the two DAQs.

How close are we to the data goals for the run?

Not-identified-as-bad data amounts, hours (2015 October 27)
Biases (V) -10/10 -8/8 -6/6 -4/4 -2/2 0/0 2/-2 4/-4 6/-6 8/-8 10/-10
Background 0.5 1 (3.3) 1 (3.3) 28.8 31 2.3 10.3 20.3 1 (3.3) 1 (3.3) 0.5
Cf 0 0 0 15 12.8 0 8.5 7.5 0 0 0
Ba 0 0.5 (2) 0.5 (2) 2 1 1.5 1.5 1.5 0.5 (2) 0.5 (2) 0
Numbers in (parantheses) include high-bias data without overshooting

Surface rejection

  • general goal: enough neutron events to calculate a NR band (Cf data), enough low-yield surface events to characterize discrimination (bg data), at a few different biases
  • Four biases chosen: -4/+4, -2/+2, +2/-2, +4/-4 V
  • Desired neutron counts: 1000 events in 5-20 keV range, 1000 events in 20-100 keV range
  • Estimated neutron rates: with optimized thresholds, 50/hour in low energy range, 135/hour in high energy range (note: lowering thresholds improved low-E neutron rate by 3×)
    • thus need 20 hours of Cf data at each bias
  • Desired surface event counts: 5000 events in 5-100 keV range
  • We don't have Pb-210 so main source of surface events is Am gammas. Only a small fraction of these will have yields dipping as low as NR. These events are also localized. So it's not likely we will really reach the desired surface event goal.
  • Haven't fully defined surface event goal (what yield range?), so haven't estimated rates. Maybe data-in-hand should be considered sufficient.

Current neutron status:

  • can reach goal in 3 days of data-taking. (Add to this times for transfers and any other pauses)
  • can reach 50% of goal in 0.33 days of data-taking.

Resolution and Charge collection efficiency

  • general goal: enough data to resolve gamma lines (bg and Ba data) over a wide range of biases
  • chosen biases: -8 to +8 V side 1 (opposite on side 2) in 2 V steps, possibly also include -10 and +10 V
  • lines 10 keV (activation), 60 keV (Am), 356 keV (Ba). Others will probably be too weak, but can look
  • For the 356 keV line (at -4/+4 V), tried Side1-Qinner peak fits for various amounts of Ba data; for 15/30/60 minutes of data, error in mean is 20/12/9 %. Probably want at least 30 minutes at each bias, more is preferable
  • For the 10 keV line (at -4/+4 V), tried Side1-Qinner peak fits for various amounts of background data; for 15/30/60 minutes of data, error in mean is 10/9/6 %. Probably want at least 30 minutes at each bias, more is preferable
  • 60 keV peak is stronger than 10 keV peak so it is not a limiting factor
  • biases greater than 4 V must be done during working hours

Current status:

  • for 60 minutes of each type and bias: 0.4 days of working-hours data-taking
  • for 30 minutes of each type and bias: almost done

Other optional data goals

  1. Y-Be source for NR-scale calibration: Anthony thinks we can maybe do better with this than what was done at Soudan because of less material between mono-energetic neutron source and detector. But logistics of borrowing Soudan source and thinking through the required shielding setup would take time, so think about this for the next run.
  2. Take MIDAS data: Anthony suggests taking Cf data at a standard bias, enough to see a NR band, with MIDAS. If good plots can be reproduced it will ensure nothing is wrong with MIDAS-taken data.
  3. Noise tests: looking at our phonon noise PSDs, Sunil and Matt Pyle suggested some followup tests. These could be done in a day.
  4. Y-88 data to search for MISS events: initial calculations from Hannah suggest we could get a pretty good rate (over 10/hour) of MISS-like events with the source placed within about half a meter of the detector, above or below detector. We can try taking some with the source above shining through gaps in the internal lead shielding – or after removing shield with source below cryostat – but better if shield is actually present, so maybe partially dismantled shield – we also brainstormed on ideas for getting the source in and out of the shield under the cryostat.
  5. HV mode temperature study: study more closely temperature increases seen at high HV biases, look at temperatures at different stages, compare to controlled heating of still, MC, detector (via flashing). May give insight into where heat is being generated in HV mode. Could be done in a day.
  6. No-shield data: Suggestion came up today to take some background data without the shield to try to get a cosmogenic neutron rate, compare to with-shield data. Would need to think about how long we'd need to do this to see some neutrons.

Older Runs