start › cdms › k100 › run_diary › 20150821

2015/08/21 (Friday)

previous day

0836 Filled LN2 cold trap full
0840 Tower/Cold Plate = 31.5/151 mK, Still = 1.33K/0.11 mbar, G1/G2 = 155/150 mbar, 1K pot = 1.74 K/24 mbar, Main Bath = 72.5%
0926 Variac was sitting on equipment rack near Lakeshore bridge. Found that while the variac was powered, Still temp was reading artificially high ar around 1.33K. When the variac was switched off, the Still temp reading fell to a more normal value of 1.1K
1200 Transfer stick was cold on previous transfers. Pumped transfer stick down to < 1e-1 mbar using scroll pump.
1330 Look for Tc transitions in HV1 pA,B,C. QET bias = 0, send 10mVpp 100Hz signal down QET lines. base = 31.4mK.
HV (V) I_leak (nA) Tower temp (mK) TES status
12.42 17.8 31.4 small phonon signal = normal
10.93 14.7 31.4 large phonon = superconducting
11.31 15.5 31.4 increasing HV, A then B then C go normal

Connecting a single probe to I_leak test point produces 60Hz noise in charge signals and drives TESs superconducting (cooling!?!?) Tower temp unaffected.

HV (V) I_leak (nA) TES status
11.29 15.8 all SC
13.03 19.3 A then B then C go normal. Return to SC after a minute.
13.94 21.4 normal. Probes drive them SC, but return to normal after removal. Eventually go SC again after a minute.

Switch to increasing HV until TESs go normal, waiting for them to return to SC, then measuring I_leak

HV (V) TES status I_leak (nA)
14.29 normal, 1min, then SC 22.3
14.89 “ -
15.49 ” 24.9
16.16 “ 26.3
17.00 normal, not returning to SC after 5 min. Probe drives them SC 28
17.51 normal, few minutes, SC 29.3
18.48 normal, few min, SC 30.7
19.01 normal, didn't go SC after ~20 min even when measuring I_leak 32.2
1800 Filled LN2 trap full.