| | Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. |
| | Overnight: Tried taking background data but failed after first series. Second series wrote and copied to /chocula but didn't process. A third series directory was created on vuk-01 but it is empty - apparently DAQDAQ got stuck at that point. |
| 0855 | DAQDAQ is still running (arrow is black), and the EventBuilder window is open but not running, showing rate = 0.02 Hz and event number = 10001. The three windows showing traces are not open, so it looks like on the third series DAQDAQ was able to create a directory and open EventBuilder but got stuck there. |
| 0905 | Tower/MC(H35)/MC sensor/Cold Plate = 30.9/25.1/41/158 mK, Still = 1.36 K / 0.11 mbar, G1/G2 = 145/155 mbar, 1K Pot = 1.72 K / 22 mbar, Main bath = 34.8% |
| 0910 | Fill LN cold trap Dewar. |
| 0915 | Stop DAQDAQ, close EventBuilder. Channels look ok on DetectorReadout, no visible low-f noise. Stop data vi's and Putty sessions to release DCRCs for MIDAS tests. |
| 1542 | Commence LHe transfer (43.2%). |
| 1620 | Set up LabVIEW DAQ after MIDAS tests. All channels there, and only slight low-frequency noise. Commence resetting threshold values to optimal (lowest possible) values. |
| 1625 | Set phonon thresholds as low as possible: |