| | Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50. |
| | tower temperature normal |
| 0920 | DAQ stopped, LabVIEW error message on screen: “Fatal Internal Error: “objheap.cpp” line 523. You will lose any unsaved work.” Caused LabVIEW crash. Turns out Disk I: is completely full so maybe this caused the error. Note: it took 36 series (30 min series, 1 hour cycle) to fill I:. |
| 0930 | Fill LN cold trap Dewar |
| 1000 | After opening up crashed LabVIEW programs, notice that detector has been at +/-2V since the crash. Flash for 30s to neutralize crystal. |
| 1040 | Place Ba source at back of fridge, inside top most part of shield for high-voltage Ba data. |
| 1047 | Start 1000 randoms run of Ba data at -/+8V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 2000 events/file, 30s flash and 25min cooldown after. Use Wednesday's phonon thresholds. |
| | MultiDAQ isn't triggering because we didn't specify a trigger file… Stop data |
| 1051 | Start 1000 randoms run of Ba data at -/+8V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 2000 events/file, 30s flash and 25min cooldown after. Use Wednesday's phonon thresholds. |
| | Rate drops when getting to triggered data. Try raising Ba source as high as possible. That doesn't help. Raise all thresholds by 10 ADC bins. That also doesn't help. |
| 1059 | Stop data after 1000 randoms and 500 triggered events. |
| 1103 | Flash for 30s. |
| | Modify MultiDAQEventBuilder VI to include a 0ms wait in the loop of the state-machine. This will hopefully help to stop the processor from getting eaten up by a constantly running loop. |
| 1120 | Hear large hissing noise from back room. It's the pressure-relief port on the exhaust line of the He4 pump. Oil has spewed from it too and is on the floor. Change exhaust port from Recovery to Exhaust, and a big release of pressure comes out. Close He recovery valve leading from He4 pump exhaust line. He meter on wall is ticking faster than normal for non-transfer times. Stops ticking after ~1minute. |
| 1125 | Still = 0.11 mbar, G1/G2 = 120/105 mbar, 1K Pot = 1.68 K / 13 mbar, Main bath = 54.3% |
| 1130 | Start 30min run of Ba data at -/+8V overshooting +Qbias (side 2) to 14V before starting data, FG=5/2/5, OG=3, 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown after. Use original thresholds ( Wednesday's phonon thresholds.) |
| 1130 | We reconnected the He4 pump exhaust to recovery. The overpressure valve opened again, and we saw the recovery pressure was elevated to about 3 PSI. It happened a couple of times, and temperature spikes were seen (not dissimilar to temperature spikes at the beginning and end of transfers). Also interestingly 250 Hz noise was seen in the phonon traces - there have been a couple of times that we have correlated this noise to high pressure in the bath, so it's likely that at least one source of this noise is gas oscillations, probably producing microphonic noise. |
| 1134 | Stop data because the computer is bogging down again. |
| | Set new phonon thresholds as below. Also add a 0ms wait (as with EventBuilder) to MultiDAQBoard VI. |