Run 53 (G106a)
| Run | Run Mode | Run Info | # of Events/File | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07150612_1728 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 5; Output Gain 1; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150612_1759 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 5; Output Gain 2; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150612_1831 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 5; Output Gain 4; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150612_1902 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 5; Output Gain 7 (RevC on DIB1 DIB2) or 8 (RevC.1 on DIB3); 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150612_1933 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; Output Gain 1; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150612_2005 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; Output Gain 2; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150612_2036 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; Output Gain 4; 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150612_2108 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; Output Gain 7 (RevC on DIB1 DIB2) or 8 (RevC.1 on DIB3); 1000 initial randoms ; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | |||
| 07150619_1545 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150619_1759 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150619_1923 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150619_2209 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150619_2245 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150619_2311 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150620_0039 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150620_1402 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | N | ||
| 07150620_1527 | Am-241 Pb-210 Ba-133 @ 1m | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150620_1806 | Am-241 Pb-210 Ba-133 @ 1m | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1107 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1200 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1316 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1606 | Am-241 Pb-210 | Junk | 2k | ANV | N | |
| 07150621_1610 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1717 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1749 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1824 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1852 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | ANV | Y | |
| 07150621_1920 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 30/5/20 mins. Manually start/stop | 2k | Y | ||
| 07150622_1219 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | DIB1pA threshold inadvertently set to 1. Run interrupted by pencil sharpener interference. | Y | |
| 07150622_1252 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | DIB1pA threshold inadvertently set to 1. | ||
| 07150622_1340 07150622_1407 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | DIB1pA threshold inadvertently set to 1. | Y | |
| 07150622_1442 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | DIB1pA threshold inadvertently set to 1. Shield under construction during run. | Y | |
| 07150622_1608 07150622_1612 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Shield under construction during run. | Y | |
| 07150622_1641 07150622_1711 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 25 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Shield under construction during run. | Y | |
| 07150622_1738 07150622_1806 07150622_1834 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 25 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Shield in place. LHe transfer 17:50 to 18:30 | Y | |
| 07150623_0802 | Am-241 Pb-210 Ba-131 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 100 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Cf-252 source removed right before data taking after an overnight exposure | Y | |
| 07150623_0843 07150623_0925 07150623_1005 07150623_1043 | Am-241 Pb-210 Ba-131 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 25 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Y | ||
| 07150623_1234 07150623_1307 07150623_1352 07150623_1420 07150623_1448 07150623_1519 07150623_1554 07150623_1624 07150623_1652 07150623_1717 07150623_1743 07150623_1809 07150623_1837 07150623_1902 07150623_1928 07150623_1955 07150623_2020 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 25 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Y | ||
| 07150624_0831 07150624_0859 07150624_0927 07150624_0952 07150624_1031 07150624_1127 07150624_1204 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 25 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Cf-252 source removed right before data taking after an overnight exposure | Y | |
| 07150625_0944 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 25 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Taken with DCRC55 in DIB1, voltage set to -4V Qinner and -4V Qouter | Y | |
| 07150625_1013 | Am-241 Pb-210 Ba-131 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 25 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Taken with DCRC55 in DIB1, voltage set to -4V Qinner and -4V Qouter | Y | |
| 07150625_1041 07150625_1112 07150625_1140 07150625_1209 07150625_1242 07150625_1323 | Am-241 Pb-210 Ba-131 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 2k | Taken with DCRC55 in DIB1, voltage set to -4V Qinner and -4V Qouter | ||
| 07150625_1405 | Am-241 Pb-210 Ba-131 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 90/5/20 mins. Manually start/stop | 2k | Taken with DCRC55 in DIB1, voltage set to -4V Qinner and -4V Qouter | Y | |
| 07150625_1604 07150625_1641 07150625_1731 07150625_1801 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 10k | Taken with DCRC55 in DIB1, voltage set to -4V Qinner and -4V Qouter | C (10000) | |
| 07150629_0911 07150629_0952 07150629_1022 07150629_1055 07150629_1137 07150629_1641 | Am-241 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; OR trigger on phonon channels, threshold of 50 ADC bins; QETs set from 2015/06/09, Front End Gain 50; 1000 initial randoms ; acquisition/flash/cooldown = 15/5/20 mins. Manually start/stop | 10k | Taken with DCRC55 in DIB1, voltage set to -4V Qinner and -4V Qouter | ANV | C (10000) |
Run 46 (G106a)
| Run | Run Mode | Run Info | # of Events/File | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07140325_0647 07140325_0737 07140325_0828 | Am-241 Ba-133 Pb-210 | DIB1 +4.0 V, DIB2 -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=40, Qo=20; DIB 2 Qi=35; Qo=20 ADC bins); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | 3 dib readout. Dibs 1 and 2 processed | AK | Y |
| 07140325_0919 07140325_1009 07140325_1059 | Am-241 Ba-133 Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=40, Qo=20; DIB 2 Qi=35; Qo=20 ADC bins); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | 3 dib readout. Dibs 1 and 2 processed | AK | Y |
| 07140328_0429 07140328_0519 | Am-241 Ba-133 Pb-210 | DIB1 +4.0 V, DIB2 -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=40, Qo=20; DIB 2 Qi=35; Qo=20 ADC bins); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | Dibs 1 and 2 read out and processed | AK | Y |
| 07140328_0610 07140328_0700 | Am-241 Ba-133 Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=40, Qo=20; DIB 2 Qi=35; Qo=20 ADC bins); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | Dibs 1 and 2 read out and processed | AK | Y |
| 07140401_0628 07140401_0658 07140401_1107 | Am-241 Ba-133 Pb-210 | DIB1 0 V, DIB2 0 V; 3.2768 ms trace length; 400 us delay;Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | AK | Y | |
| 07140401_1138 | Am-241 Ba-133 Pb-210 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay;Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | AK | Y | |
| 07140401_1209 | Am-241 Ba-133 Pb-210 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay;Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels; acquisition/flash/cooldown = 10/0.1/20 mins. | 10k | AK | Y | |
| 07140402_2243 07140402_2314 07140402_2344 07140403_0015 07140403_0045 07140403_0115 | Am-241 Ba-133 Pb-210 | DIB1 +4.0 V, DIB2 -4.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140403_0145 07140403_0218 07140403_0251 07140403_0324 07140403_0357 07140403_0430 | Am-241 Ba-133 Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140403_0447 07140403_0519 07140403_0552 07140403_0625 07140403_0658 07140403_0731 | Am-241 Ba-133 Pb-210 | DIB1 +3.0 V, DIB2 +-3.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140403_0748 07140403_0821 07140403_0854 07140403_0925 07140403_0958 07140403_1031 | Am-241 Ba-133 Pb-210 | DIB1 -3.0 V, DIB2 +3.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140404_1521 07140404_1551 07140404_1622 07140404_1652 071404_1723 071404034_1754 | Am-241 Ba-133 Pb-210 | DIB1 +3.0 V, DIB2 -3.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140404_1823 07140404_1853 07140404_1925 07140404_1955 07140404_2025 07140404_2056 | Am-241 Ba-133 Pb-210 | DIB1 -3.0 V, DIB2 +3.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140404_2125 07140404_2155 07140404_2225 07140404_2255 07140404_2325 07140404_2356 | Am-241 Ba-133 Pb-210 | DIB1 0.0 V, DIB2 0.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140405_0802 07140405_0852 071405_0923 07140405_0954 07140405_1026 07140405_1056 | Am-241 Ba-133 Pb-210 | DIB1 +4.0 V, DIB2 +-4.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140405_1122 07140405_1153 07140405_1225 07140405_1256 07140405_1328 07140405_1358 | Am-241 Ba-133 Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 50 ADC bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 2k | AK | Y | |
| 07140407_1422 07140407_1452 07140407_1521 | Am-241 Ba-133 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Data acquired in following order: no changes made to ground, isolate system from IVC pump, remove dewar metal cords. | Y | |
| 07140407_1802 | Am-241 Ba-133 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Fridge frame shorted to earth ground. | Y | |
| 07140408_1729 07140408_1754 07140408_1819 07140409_1723 | Am-241 Ba-133 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Continued to investigate suspected short. Acquired data in following order: no purposeful short between fridge frame and still line, metal braid shorting wall frame to fridge frame, fridge frame shorted to earth ground, no purposeful short and lab electronics powered off. | Y | |
| 07140410_1351 07140410_1423 07140410_1617 | Am-241 Ba-133 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except D3, which was set at 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Continued troubleshooting to find source of phonon noise. Acquired data in the following configurations: Small fan striking 1K Pot needle valve, fan still powered - not striking needle valve, copper shield attached to DCRC33. | Y | |
| 07140411_1503 | Am-241 Ba-133 Pb-210 | Flash detector for 1 minute followed by 20 minute cooldown and 5 minute data series. DIB1 (DCRC33) 0V bias with DIB2 grounded; 3.2768 ms trace length; 400 us delay; 200 uS coincidence time; 2K events per file. Phonon OR trigger thresholds set at 30 ADC for phonon channels A&D and 35 ADC for phonon channels B&C. | 2k | DIB 2 grounded with shorting board | Y | |
| 07140411_1918 | Am-241 Ba-133 Pb-210 | Flash detector for 1 minute followed by 20 minute cooldown and 5 minute data series. DIB1 grounded with DIB2 (DCRC33) 0V bias; 3.2768 ms trace length; 400 us delay; 200 uS coincidence time; 2K events per file. Phonon OR trigger thresholds set at 30 ADC for phonon channels A&D and 35 ADC for phonon channels B&C. | 2k | DIB 2 grounded with shorting board | Y | |
| 07140415_0111 07140415_0559 | Am-241 Ba-133 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0138 | Am-241 Ba-133 Pb-210 | DIB1 -1 V, DIB2 +1 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0204 | Am-241 Ba-133 Pb-210 | DIB1 -2 V, DIB2 +2 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0230 | Am-241 Ba-133 Pb-210 | DIB1 -3 V, DIB2 +3 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0256 | Am-241 Ba-133 Pb-210 | DIB1 -4 V, DIB2 +4 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0322 | Am-241 Ba-133 Pb-210 | DIB1 -5 V, DIB2 +5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0348 | Am-241 Ba-133 Pb-210 | DIB1 -6 V, DIB2 +6 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0415 | Am-241 Ba-133 Pb-210 | DIB1 -7 V, DIB2 +7 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0441 | Am-241 Ba-133 Pb-210 | DIB1 -8 V, DIB2 +8 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0507 | Am-241 Ba-133 Pb-210 | DIB1 -9 V, DIB2 +9 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0533 | Am-241 Ba-133 Pb-210 | DIB1 -10 V, DIB2 +10 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0626 | Am-241 Ba-133 Pb-210 | DIB1 +1 V, DIB2 -1 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0652 | Am-241 Ba-133 Pb-210 | DIB1 +2 V, DIB2 -2 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0718 | Am-241 Ba-133 Pb-210 | DIB1 +3 V, DIB2 -3 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0744 | Am-241 Ba-133 Pb-210 | DIB1 +4 V, DIB2 -4 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0810 | Am-241 Ba-133 Pb-210 | DIB1 +5 V, DIB2 -5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0836 | Am-241 Ba-133 Pb-210 | DIB1 +6 V, DIB2 -6 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0903 | Am-241 Ba-133 Pb-210 | DIB1 +7 V, DIB2 -7 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0929 | Am-241 Ba-133 Pb-210 | DIB1 +8 V, DIB2 -8 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_0955 | Am-241 Ba-133 Pb-210 | DIB1 +9 V, DIB2 -9 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140415_1021 | Am-241 Ba-133 Pb-210 | DIB1 +10 V, DIB2 -10 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, and D3, which were set at 40, 30, and 40 ADC bins, respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140417_0610 07140417_0635 07140417_0701 07140417_0726 | Am-241 Ba-133 Pb-210 | DIB1 +12.0 V, DIB2 -12.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140417_0752 07140417_0817 07140417_0842 07140417_0902 | Am-241 Ba-133 Pb-210 | DIB1 -12.0 V, DIB2 +12.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140417_0933 07140417_0958 07140417_1023 07140417_1049 | Am-241 Ba-133 Pb-210 | DIB1 +10.0 V, DIB2 -10.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140417_1115 07140417_1140 07140417_1205 07140417_1230 | Am-241 Ba-133 Pb-210 | DIB1 -10.0 V, DIB2 +10.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140418_0600 07140418_0625 07140418_0650 07140418_0715 | Am-241 Ba-133 Pb-210 | DIB1 +8.0 V, DIB2 -8.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140418_0741 07140418_0807 07140418_0832 07140418_0857 | Am-241 Ba-133 Pb-210 | DIB1 -8.0 V, DIB2 +8.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140418_0923 07140418_0948 07140418_1013 07140418_1038 | Am-241 Ba-133 Pb-210 | DIB1 +6.0 V, DIB2 -6.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140418_1104 07140418_1129 07140418_1155 07140418_1220 | Am-241 Ba-133 Pb-210 | DIB1 -6.0 V, DIB2 +6.0 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; ionisation data series, although SQUIDs were locked and TESs were biased; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2 and D3, which were set at 40, 30 and 40 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | |||
| 07140419_0244 07140419_0310 07140419_0335 07140419_0400 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140419_0426 07140419_0451 07140419_0516 07140419_0541 | Am-241 Pb-210 | DIB1 -1.0 V, DIB2 +1.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140419_0608 07140419_0633 07140419_0658 07140419_0723 | Am-241 Pb-210 | DIB1 -2.0 V, DIB2 +2.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140419_0749 07140419_0814 07140419_0839 07140419_0904 | Am-241 Pb-210 | DIB1 -3.0 V, DIB2 +3.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140419_0930 07140419_0956 07140419_1021 07140419_1046 | Am-241 Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140419_1112 07140419_1137 07140419_1202 07140419_1227 | Am-241 Pb-210 | DIB1 -5.0 V, DIB2 +5.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140419_1254 07140419_1319 07140419_1344 07140419_1409 | Am-241 Pb-210 | DIB1 -6.0 V, DIB2 +6.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_0359 07140420_0424 07140420_0449 07140420_0514 | Am-241 Pb-210 | DIB1 -7.0 V, DIB2 +7.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_0540 07140420_0605 07140420_0631 07140420_0656 | Am-241 Pb-210 | DIB1 -8.0 V, DIB2 +8.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_0722 07140420_0747 07140420_0812 07140420_0837 | Am-241 Pb-210 | DIB1 -9.0 V, DIB2 +9.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_0904 07140420_0929 07140420_0954 07140420_1019 | Am-241 Pb-210 | DIB1 -10.0 V, DIB2 +10.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_1045 07140420_1110 07140420_1135 07140420_1200 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_1227 07140420_1252 07140420_1317 07140420_1342 | Am-241 Pb-210 | DIB1 +1.0 V, DIB2 -1.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_1408 07140420_1433 07140420_1458 07140420_1523 | Am-241 Pb-210 | DIB1 +2.0 V, DIB2 -2.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_1550 07140420_1615 07140420_1640 07140420_1705 | Am-241 Pb-210 | DIB1 +3.0 V, DIB2 -3.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_1731 07140420_1756 07140420_1821 07140420_1847 | Am-241 Pb-210 | DIB1 +4.0 V, DIB2 -4.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_1913 07140420_1938 07140420_2003 07140420_2028 | Am-241 Pb-210 | DIB1 +5.0 V, DIB2 -5.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_2054 07140420_2119 07140420_2144 07140420_2210 | Am-241 Pb-210 | DIB1 +6.0 V, DIB2 -6.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140420_2236 07140420_2301 07140420_2326 07140420_2351 | Am-241 Pb-210 | DIB1 +7.0 V, DIB2 -7.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140421_0017 07140421_0042 07140421_0108 07140421_0133 | Am-241 Pb-210 | DIB1 +8.0 V, DIB2 -8.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140421_0159 07140421_0224 07140421_0249 07140421_0314 | Am-241 Pb-210 | DIB1 +9.0 V, DIB2 -9.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140421_0340 07140421_0405 07140421_0431 07140421_0456 | Am-241 Pb-210 | DIB1 +10.0 V, DIB2 -10.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140421_2312 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140421_2338 | Am-241 Pb-210 | DIB1 -1.0 V, DIB2 +1.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0004 | Am-241 Pb-210 | DIB1 -2.0 V, DIB2 +2.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0031 | Am-241 Pb-210 | DIB1 -3.0 V, DIB2 +3.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0057 | Am-241 Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0123 | Am-241 Pb-210 | DIB1 -5.0 V, DIB2 +5.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0150 | Am-241 Pb-210 | DIB1 -6.0 V, DIB2 +6.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0216 | Am-241 Pb-210 | DIB1 -7.0 V, DIB2 +7.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0242 | Am-241 Pb-210 | DIB1 -8.0 V, DIB2 +8.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0308 | Am-241 Pb-210 | DIB1 -9.0 V, DIB2 +9.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0335 | Am-241 Pb-210 | DIB1 -10.0 V, DIB2 +10.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0401 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0427 | Am-241 Pb-210 | DIB1 +1.0 V, DIB2 -1.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0453 | Am-241 Pb-210 | DIB1 +2.0 V, DIB2 -2.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0520 | Am-241 Pb-210 | DIB1 +3.0 V, DIB2 -3.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0546 | Am-241 Pb-210 | DIB1 +4.0 V, DIB2 -4.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0612 | Am-241 Pb-210 | DIB1 +5.0 V, DIB2 -5.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0638 | Am-241 Pb-210 | DIB1 +6.0 V, DIB2 -6.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0704 | Am-241 Pb-210 | DIB1 +7.0 V, DIB2 -7.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0731 | Am-241 Pb-210 | DIB1 +8.0 V, DIB2 -8.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0757 | Am-241 Pb-210 | DIB1 +9.0 V, DIB2 -9.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140422_0823 | Am-241 Pb-210 | DIB1 +10.0 V, DIB2 -10.0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; Qi DIB 3 trigger threshold set at 1 ADC bin; QET Bias 0 V for all channels; acquisition/flash/cooldown = 5/0.1/20 mins. | 500 | Y | ||
| 07140502_1638 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, A3, B3 and C3, which were set at 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias for all channels set to 0uA. | 500 | Y | ||
| 07140502_1705 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, A3, B3 and C3, which were set at 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140502_1731 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, A3, B3 and C3, which were set at 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to magnitude of 10uA higher than values on 2014/05/02. | 500 | Y | ||
| 07140502_1757 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, A3, B3 and C3, which were set at 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to magnitude of 200uA. | 500 | Y | ||
| 07140502_1823 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except C2, D2, A3, B3 and C3, which were set at 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to values on 2014/05/02, and polarity set to Inverting rather than Non-Inverting. | 500 | Y | ||
| 07140504_0535 07140504_0625 07140504_0715 | Am-241 Ba-133 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0 V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 minutes. | 2000 | Ionisation-only; neutralisation study. | HC | Y |
| 07140504_0806 07140504_0856 07140504_0946 | Am-241 Ba-133 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0 V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 minutes. | 2000 | Ionisation-only; neutralisation study. | HC | Y |
| 07140508_0457 07140508_0533 07140508_0608 07140508_0646 07140508_0721 07140508_0756 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140509_0534 07140509_0610 07140509_0646 07140509_0722 07140509_0757 07140509_0833 07140509_0910 07140509_0946 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | AK | Y | |
| 07140510_0702 07140510_0737 07140510_0812 07140510_0847 07140510_0922 07140510_0958 07140510_1033 07140510_1108 07140510_1143 07140510_1218 07140510_1254 07140510_1329 07140510_1404 07140510_1439 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140511_0545 07140511_0621 07140511_0656 07140511_0731 07140511_0806 07140511_0841 07140511_0917 07140511_0952 07140511_1027 07140511_1102 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140512_0223 07140512_0258 07140512_0333 07140512_0408 07140512_0443 07140512_0519 07140512_0554 07140512_0629 07140512_0704 07140512_0740 07140512_0815 07140512_0850 07140512_0925 07140512_1001 07140512_1036 07140512_1111 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140513_0027 07140513_0102 07140513_0138 07140513_0213 07140513_0249 07140513_0324 07140513_0359 07140513_0435 07140513_0510 07140513_0546 07140513_0621 07140513_0657 07140513_0733 07140513_0808 07140513_0843 07140513_0918 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140513_1457 07140513_1533 07140513_1609 07140513_1646 07140513_1723 07140513_1758 07140513_1835 07140513_1911 07140513_1948 07140513_2025 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140514_0851 07140514_0926 07140514_1002 07140514_1038 07140514_1114 07140514_1150 07140514_1225 07140514_1301 07140514_1336 07140514_1412 07140514_1447 07140514_1523 07140514_1559 07140514_1634 07140514_1710 07140514_1746 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140515_0457 07140515_0528 07140515_0558 07140515_0629 07140515_0659 07140515_0729 07140515_0800 07140515_0830 07140515_0900 07140515_0931 07140515_1001 07140515_1031 07140515_1102 07140515_1132 07140515_1202 07140515_1233 07140515_1303 07140515_1333 07140515_1404 07140515_1434 07140515_1504 07140515_1535 07140515_1605 07140515_1636 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140516_0614 07140516_0645 07140516_0715 07140516_0745 07140516_0816 07140516_0846 07140516_0916 07140516_0947 07140516_1017 07140516_1047 07140516_1118 07140516_1148 07140516_1218 07140516_1249 07140516_1319 07140516_1350 07140516_1420 07140516_1450 07140516_1520 07140516_1551 07140516_1621 07140516_1652 07140516_1722 07140516_1754 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140517_0352 07140517_0422 07140517_0453 07140517_0523 07140517_0554 07140517_0625 07140517_0655 07140517_0726 07140517_0756 07140517_0827 07140517_0858 07140517_0928 07140517_0959 07140517_1030 07140517_1100 07140517_1131 07140517_1201 07140517_1232 07140517_1302 07140517_1333 07140517_1404 07140517_1434 07140517_1505 07140517_1536 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140518_0534 07140518_0605 07140518_0636 07140518_0706 07140518_0737 07140518_0808 07140518_0839 07140518_0909 07140518_0940 07140518_1011 07140518_1041 07140518_1112 07140518_1143 07140518_1213 07140518_1244 07140518_1315 07140518_1345 07140518_1416 07140518_1446 07140518_1517 07140518_1548 07140518_1618 07140518_1649 07140518_1719 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140519_0222 07140519_0253 07140519_0323 07140519_0354 07140519_0424 07140519_0454 07140519_0524 07140519_0555 07140519_0625 07140519_0655 07140519_0725 07140519_0755 07140519_0825 07140519_0856 07140519_0926 07140519_0956 07140519_1028 07140519_1058 07140519_1131 07140519_1201 07140519_1231 07140519_1304 07140519_1334 07140519_1404 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140519_1914 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; charge OR trigger thresholds set at 40 ADC bins for all channels; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias for all channels set to 0uA. | 500 | Y | ||
| 07140519_1940 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; charge OR trigger thresholds set at 40 ADC bins for all channels; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140519_2006 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; charge OR trigger thresholds set at 40 ADC bins for all channels; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to magnitude of 10uA higher than values on 2014/05/02. | 500 | Y | ||
| 07140519_2033 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; charge OR trigger thresholds set at 40 ADC bins for all channels; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to magnitude of 200uA. | 500 | Y | ||
| 07140519_2059 | Am-241 Pb-210 | DIB1 0 V, DIB2 0 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; charge OR trigger thresholds set at 25 ADC bins for all channels; acquisition/flash/cooldown = 5/0.1/20 mins; QET Bias set to values on 2014/05/02, and polarity set to Inverting rather than Non-Inverting. | 500 | Y | ||
| 07140520_0549 07140520_0619 07140520_0649 07140520_0719 07140520_0750 07140520_0820 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140521_0633 07140521_0709 07140521_0744 07140521_0819 07140521_0854 07140521_0929 07140521_1004 07140521_1039 07140521_1114 07140521_1150 07140521_1225 07140521_1300 07140521_1335 07140521_1410 07140521_1445 07140521_1520 | Am-241 Cf-252 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 2000 | Y | ||
| 07140521_1851 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; Channel C2 (D Side 1) SQUID unlocked; Other QET Biases set to values on 2014/05/02. | 500 | Y | ||
| 07140521_1917 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 5/0.1/20 mins; All DIB 2 channels (A2, B2, C2 & D2) SQUIDs unlocked; Other QET Biases set to values on 2014/05/02. | 500 | Y | ||
| 07140522_0543 07140522_0618 07140522_0653 07140522_0729 07140522_0804 07140522_0839 07140522_0914 07140522_0949 07140522_1024 07140522_1059 07140522_1134 07140522_1209 07140522_1244 07140522_1320 07140522_1355 07140522_1430 | Am-241 Cf-252 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140523_0446 07140523_0521 07140523_0556 07140523_0631 07140523_0706 07140523_0742 07140523_0817 07140523_0852 07140523_0927 07140523_1002 07140523_1037 07140523_1112 07140523_1148 07140523_1223 07140523_1258 07140523_1333 | Am-241 Cf-252 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140523_1734 07140523_1809 07140523_1844 07140523_1920 | Am-241 Ba-133 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 15/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140523_1956 07140523_2026 07140523_2056 07140523_2126 | Am-241 Ba-133 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140524_0040 07140524_0110 07140524_0140 07140524_0210 07140524_0240 07140524_0310 07140524_0341 07140524_0411 07140524_0441 07140524_0511 07140524_0541 07140524_0611 07140524_0641 07140524_0711 07140524_0741 07140524_0812 07140524_0842 07140524_0912 07140524_0942 07140524_1012 07140524_1042 07140524_1112 07140524_1142 07140524_1212 | Am-241 Cf-252 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140525_1054 07140525_1124 07140525_1154 07140525_1224 07140525_1255 07140525_1325 07140525_1355 07140525_1425 07140525_1455 07140525_1526 07140525_1556 07140525_1626 | Am-241 Cf-252 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140525_1927 07140525_1957 07140525_2028 07140525_2058 07140525_2128 07140525_2158 07140525_2228 07140525_2258 07140525_2329 07140525_2359 07140526_0029 07140526_0059 07140526_0129 07140526_0159 07140526_0229 07140526_0300 07140526_0330 07140526_0400 07140526_0430 07140526_0500 07140526_0530 07140526_0600 07140526_0630 07140526_0701 | Am-241 Cf-252 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 3.2768 ms trace length; 400 us delay; phonon OR trigger thresholds set at 25 ADC bins for all phonon channels except B1, C2, D2, A3, B3 and C3, which were set at 30, 50, 17, 12, 8, and 8 ADC bins respectively; acquisition/flash/cooldown = 10/0.1/20 mins; QET Bias set to values on 2014/05/02. | 500 | Y | ||
| 07140526_1332 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 75 mK tower temperature. | Y | |
| 07140526_1424 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 75 mK tower temperature. | Y | |
| 07140526_1734 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 105 ± 5 mK tower temperature. | Y | |
| 07140526_1825 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 105 ± 5 mK tower temperature. | Y | |
| 07140527_0209 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 290 ± 5 mK tower temperature. | Y | |
| 07140527_0300 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 290 ± 5 mK tower temperature. | Y | |
| 07140527_0600 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 455 ± 5 mK tower temperature. | Y | |
| 07140527_0651 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 455 ± 5 mK tower temperature. | Y | |
| 07140527_1327 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 630 ± 5 mK tower temperature. | Y | |
| 07140527_1419 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 630 ± 5 mK tower temperature. | Y | |
| 07140527_1640 | Am-241 Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 850 ± 5 mK tower temperature. | Y | |
| 07140527_1731 | Am-241 Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V, DIB3 0V; 0.4096 ms trace length; 100 us delay; charge OR trigger thresholds set at 40 ADC bins for all charge channels; acquisition/flash/cooldown = 30/0.1/20 mins; ionisation-only. | 500 | 850 ± 5 mK tower temperature. | Y |
Run 45 (G101b floated phonon rail)
| Run | Run Mode | Run Info | # of Events/File | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07140203_1522 | Am-241 Ba-133 | DIB1 Qi & Qo 0 V; 3.2768 ms trace length; 400 us delay; trigger on phonon channel A; acquisition =15 mins. | 20k | AK | not | |
| 07140203_2310 07140204_0130 07140204_0350 07140204_0611 | Am-241 Ba-133 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 120/0.1/20 mins. | 10k | HC | Y | |
| 07140205_2210 07140205_2310 07140206_0011 07140206_0111 | Am-241 Ba-133 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 40/0.1/20 mins. | 2k | HC | Y | |
| 07140206_1351 07140206_1416 | Am-241 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140206_1443 07140206_1508 | Am-241 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) +4.8 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140206_1534 07140206_1559 | Am-241 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) +12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140206_1625 07140206_1651 | Am-241 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) -12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140206_1717 07140206_1742 | Am-241 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) -4.8 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140206_1808 07140206_1833 | Am-241 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_0607 07140207_0632 | Am-241 | DIB1 Qi(Vq) -10.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_0659 07140207_0724 | Am-241 | DIB1 Qi(Vq) -10.0 V, Qo(Vp) -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_0750 07140207_0815 | Am-241 | DIB1 Qi(Vq) -10.0 V, Qo(Vp) -10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_0841 07140207_0907 | Am-241 | DIB1 Qi(Vq) +10.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_0933 07140207_0958 | Am-241 | DIB1 Qi(Vq) +10.0 V, Qo(Vp) +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_1024 07140207_1049 | Am-241 | DIB1 Qi(Vq) +10.0 V, Qo(Vp) +10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_2038 07140207_2103 | Am-241 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_2129 07140207_2155 | Am-241 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) -3.2 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_2221 07140207_2247 | Am-241 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) -8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140207_2313 07140207_2338 | Am-241 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0004 07140208_0029 | Am-241 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) +3.2 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0056 07140208_0121 | Am-241 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) +8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0147 07140208_0212 | Am-241 | DIB1 Qi(Vq) -6.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0239 07140208_0304 | Am-241 | DIB1 Qi(Vq) -6.0 V, Qo(Vp) -2.4 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0330 07140208_0355 | Am-241 | DIB1 Qi(Vq) -6.0 V, Qo(Vp) -6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0421 07140208_0447 | Am-241 | DIB1 Qi(Vq) +6.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0513 07140208_0538 | Am-241 | DIB1 Qi(Vq) +6.0 V, Qo(Vp) +2.4 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0604 07140208_0629 | Am-241 | DIB1 Qi(Vq) +6.0 V, Qo(Vp) +6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0656 07140208_0721 | Am-241 | DIB1 Qi(Vq) -5.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0747 07140208_0812 | Am-241 | DIB1 Qi(Vq) -5.0 V, Qo(Vp) -2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0838 07140208_0904 | Am-241 | DIB1 Qi(Vq) -5.0 V, Qo(Vp) -5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_0930 07140208_0955 | Am-241 | DIB1 Qi(Vq) +5.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_1021 07140208_1046 | Am-241 | DIB1 Qi(Vq) +5.0 V, Qo(Vp) +2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_1113 07140208_1138 | Am-241 | DIB1 Qi(Vq) +5.0 V, Qo(Vp) +5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140208_2352 07140209_0017 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0043 07140209_0109 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) -1.6 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0135 07140209_0201 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0227 07140209_0252 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0318 07140209_0343 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) +1.6 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0410 07140209_0435 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0501 07140209_0526 | Am-241 | DIB1 Qi(Vq) -3.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0552 07140209_0618 | Am-241 | DIB1 Qi(Vq) -3.5 V, Qo(Vp) -1.4 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0644 07140209_0709 | Am-241 | DIB1 Qi(Vq) -3.5 V, Qo(Vp) -3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0735 07140209_0800 | Am-241 | DIB1 Qi(Vq) +3.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0827 07140209_0852 | Am-241 | DIB1 Qi(Vq) +3.5 V, Qo(Vp) +1.4 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_0918 07140209_0943 | Am-241 | DIB1 Qi(Vq) +3.5 V, Qo(Vp) +3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140209_2352 07140210_0018 07140210_0043 07140210_0109 | Am-241 | DIB1 Qi(Vq) -3.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_0135 07140210_0200 07140210_0225 07140210_0251 | Am-241 | DIB1 Qi(Vq) -3.0 V, Qo(Vp) -1.2 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_0317 07140210_0342 07140210_0407 07140210_0432 | Am-241 | DIB1 Qi(Vq) -3.0 V, Qo(Vp) -3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_0458 07140210_0523 07140210_0549 07140210_0614 | Am-241 | DIB1 Qi(Vq) +3.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_0640 07140210_0705 07140210_0730 07140210_0755 | Am-241 | DIB1 Qi(Vq) +3.0 V, Qo(Vp) +1.2 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_0821 07140210_0847 07140210_0912 07140210_0937 | Am-241 | DIB1 Qi(Vq) +3.0 V, Qo(Vp) +3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_1003 07140210_1028 07140210_1053 07140210_1118 | Am-241 | DIB1 Qi(Vq) -2.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_1145 07140210_1210 07140210_1235 07140210_1300 | Am-241 | DIB1 Qi(Vq) -2.5 V, Qo(Vp) -1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_1326 07140210_1351 07140210_1417 07140210_1442 | Am-241 | DIB1 Qi(Vq) -2.5 V, Qo(Vp) -2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_1508 07140210_1533 07140210_1558 07140210_1623 | Am-241 | DIB1 Qi(Vq) +2.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_1649 07140210_1715 07140210_1740 07140210_1805 | Am-241 | DIB1 Qi(Vq) +2.5 V, Qo(Vp) +1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140210_1831 07140210_1856 07140210_1921 07140210_1946 | Am-241 | DIB1 Qi(Vq) +2.5 V, Qo(Vp) +2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_0254 07140211_0319 07140211_0344 07140211_0409 | Am-241 | DIB1 Qi(Vq) -2.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_0436 07140211_0501 07140211_0526 07140211_0551 | Am-241 | DIB1 Qi(Vq) -2.0 V, Qo(Vp) -0.8 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_0617 07140211_0642 07140211_0708 07140211_0733 | Am-241 | DIB1 Qi(Vq) -2.0 V, Qo(Vp) -2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_0759 07140211_0824 07140211_0849 07140211_0914 | Am-241 | DIB1 Qi(Vq) +2.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_0940 07140211_1006 07140211_1031 07140211_1056 | Am-241 | DIB1 Qi(Vq) +2.0 V, Qo(Vp) +0.8 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_1122 07140211_1147 07140211_1212 07140211_1237 | Am-241 | DIB1 Qi(Vq) +2.0 V, Qo(Vp) +1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_1303 07140211_1329 07140211_1354 07140211_1419 | Am-241 | DIB1 Qi(Vq) -1.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_1445 07140211_1510 07140211_1535 07140211_1600 | Am-241 | DIB1 Qi(Vq) -1.5 V, Qo(Vp) -0.6 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_1626 07140211_1652 07140211_1717 07140211_1742 | Am-241 | DIB1 Qi(Vq) -1.5 V, Qo(Vp) -1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_1808 07140211_1833 07140211_1858 07140211_1923 | Am-241 | DIB1 Qi(Vq) +1.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_1950 07140211_2015 07140211_2040 07140211_2105 | Am-241 | DIB1 Qi(Vq) +1.5 V, Qo(Vp) +0.6 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_2131 07140211_2156 07140211_2221 07140211_2246 | Am-241 | DIB1 Qi(Vq) +1.5 V, Qo(Vp) +1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140211_2313 07140211_2338 07140212_0003 07140212_0028 | Am-241 | DIB1 Qi(Vq) -1.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_0054 07140212_0119 07140212_0144 07140212_0209 | Am-241 Ba-133 | DIB1 Qi(Vq) -1.0 V, Qo(Vp) -0.4 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_0235 07140212_0301 07140212_0326 07140212_0351 | Am-241 | DIB1 Qi(Vq) -1.0 V, Qo(Vp) -1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_0417 07140212_0442 07140212_0507 07140212_0532 | Am-241 | DIB1 Qi(Vq) +1.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_0559 07140212_0624 07140212_0649 07140212_0714 | Am-241 | DIB1 Qi(Vq) +1.0 V, Qo(Vp) +0.4 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_0740 07140212_0805 07140212_0830 07140212_0855 | Am-241 | DIB1 Qi(Vq) +1.0 V, Qo(Vp) +1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_1813 07140212_1838 07140212_1903 07140212_1929 | Am-241 | DIB1 Qi(Vq) +0.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_1955 07140212_2020 07140212_2045 07140212_2110 | Am-241 | DIB1 Qi(Vq) +0.5 V, Qo(Vp) +0.2 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_2136 07140212_2201 07140212_2227 07140212_2252 | Am-241 | DIB1 Qi(Vq) +0.5 V, Qo(Vp) +0.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140212_2318 07140212_2343 07140213_0008 07140213_0033 | Am-241 | DIB1 Qi(Vq) -0.5 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140213_0059 07140213_0124 07140213_0150 07140213_0215 | Am-241 | DIB1 Qi(Vq) -0.5 V, Qo(Vp) -0.2 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140214_0308 07140214_0334 07140214_0359 07140214_0424 | Am-241 | DIB1 Qi(Vq) -0.5 V, Qo(Vp) -0.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140214_1623 07140214_1650 07140214_1717 07140214_1745 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) 0 V; 3.2768 ms trace length; 400 us delay; SQUIDs A, C & D locked; trigger on DIB1 phonon channel A; pA Trigger Threshold of 1 ADC Bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140214_1804 07140214_1834 07140214_1903 07140214_1933 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) 0 V; 13.1072 ms trace length; 400 us delay; SQUIDs A, C & D locked; trigger on DIB1 phonon channel A; pA Trigger Threshold of 1 ADC Bin; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140215_0513 07140215_0538 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_0604 07140215_0629 | Am-241 | DIB1 Qi(Vq) -4.8 V, Qo(Vp) -12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_0655 07140215_0720 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_0747 07140215_0812 | Am-241 | DIB1 Qi(Vq) +4.8 V, Qo(Vp) +12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_0838 07140215_0903 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_0929 07140215_0954 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) -10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_1021 07140215_1046 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_1112 07140215_1137 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) +10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_1203 07140215_1228 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_1255 07140215_1320 | Am-241 | DIB1 Qi(Vq) -3.2 V, Qo(Vp) -8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_1346 07140215_1411 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140215_1437 07140215_1502 | Am-241 | DIB1 Qi(Vq) +3.2 V, Qo(Vp) +8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_0627 07140216_0652 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_0718 07140216_0743 | Am-241 | DIB1 Qi(Vq) -2.4 V, Qo(Vp) -6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_0810 07140216_0835 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_0901 07140216_0926 | Am-241 | DIB1 Qi(Vq) +2.4 V, Qo(Vp) +6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_0952 07140216_1017 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1044 07140216_1109 | Am-241 | DIB1 Qi(Vq) -2.0 V, Qo(Vp) -5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1135 07140216_1200 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1226 07140216_1251 | Am-241 | DIB1 Qi(Vq) +2.0 V, Qo(Vp) +5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1318 07140216_1343 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1409 07140216_1434 | Am-241 | DIB1 Qi(Vq) -1.6 V, Qo(Vp) -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1500 07140216_1525 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1552 07140216_1617 | Am-241 | DIB1 Qi(Vq) +1.6 V, Qo(Vp) +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1643 07140216_1708 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1734 07140216_1759 | Am-241 | DIB1 Qi(Vq) -1.4 V, Qo(Vp) -3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1826 07140216_1851 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140216_1917 07140216_1942 | Am-241 | DIB1 Qi(Vq) +1.4 V, Qo(Vp) +3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_0559 07140217_0624 07140217_0649 07140217_0714 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_0740 07140217_0806 07140217_0831 07140217_0856 | Am-241 | DIB1 Qi(Vq) -1.2 V, Qo(Vp) -3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_0922 07140217_0947 07140217_1012 07140217_1037 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_1104 07140217_1129 07140217_1154 07140217_1219 | Am-241 | DIB1 Qi(Vq) +1.2 V, Qo(Vp) +3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_1245 07140217_1310 07140217_1335 07140217_1401 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_1427 07140217_1452 07140217_1517 07140217_1542 | Am-241 | DIB1 Qi(Vq) -1.0 V, Qo(Vp) -2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_1608 07140217_1633 07140217_1658 07140217_1724 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140217_1750 07140217_1815 07140217_1840 07140217_1905 | Am-241 | DIB1 Qi(Vq) +1.0 V, Qo(Vp) +2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140218_0333 07140218_0358 07140218_0423 07140218_0449 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140218_0515 07140218_0540 07140218_0605 07140218_0631 | Am-241 | DIB1 Qi(Vq) -0.8 V, Qo(Vp) -2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140218_0657 07140218_0722 07140218_0747 07140218_0812 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140218_0838 07140218_0904 07140218_0929 07140218_0954 | Am-241 | DIB1 Qi(Vq) +0.8 V, Qo(Vp) +2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140218_2246 07140218_2318 07140218_2349 07140219_0020 07140219_0051 07140219_0123 07140219_0154 07140219_0225 07140219_0256 07140219_0328 07140219_0359 | Am-241 | DIB1 Qi(Vq) +0.0 V, Qo(Vp) +0.0 V; QET bias decrease from 0 uA to -200 uA in steps of -20 uA; 13.1072 ms trace length; 400 us delay; trigger on DIB1 phonon channel A Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 500 | |||
| 07140219_0430 07140219_0501 07140218_0532 07140219_0603 07140219_0635 07140219_0706 07140219_0737 07140219_0808 07140219_0840 07140219_0911 07140219_0942 | Am-241 | DIB1 Qi(Vq) -3.0 V, Qo(Vp) -3.0 V; QET bias decrease from 0 uA to -200 uA in steps of -20 uA; 13.1072 ms trace length; 400 us delay; trigger on DIB1 phonon channel A Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 500 | |||
| 07140219_1013 07140219_1044 07140218_1116 07140219_1147 07140219_1218 07140219_1249 07140219_1321 07140219_1352 07140219_1423 07140219_1454 07140219_1525 | Am-241 | DIB1 Qi(Vq) +3.0 V, Qo(Vp) -+3.0 V; QET bias decrease from 0 uA to -200 uA in steps of -20 uA; 13.1072 ms trace length; 400 us delay; trigger on DIB1 phonon channel A Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 10/0.1/20 mins. | 500 | |||
| 07140220_0308 07140220_0334 07140220_0359 07140220_0424 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140220_0450 07140220_0515 07140220_0540 07140220_0605 | Am-241 | DIB1 Qi(Vq) -0.6 V, Qo(Vp) -1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140220_0632 07140220_0657 07140220_0722 07140220_0747 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140220_0813 07140220_0838 07140220_0903 07140220_0928 | Am-241 | DIB1 Qi(Vq) +0.6 V, Qo(Vp) +1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140220_0955 07140220_1020 07140220_1045 07140220_1110 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) -1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140220_1136 07140220_1201 07140220_1226 07140220_1251 | Am-241 | DIB1 Qi(Vq) -0.4 V, Qo(Vp) -1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140220_1317 07140220_1342 07140220_1408 07140220_1433 | Am-241 | DIB1 Qi(Vq) 0 V, Qo(Vp) +1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140220_1459 07140220_1524 07140220_1549 07140220_1614 | Am-241 | DIB1 Qi(Vq) +0.4 V, Qo(Vp) +1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140221_0605 07140221_0631 | Am-241 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) -12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140221_0657 07140221_0722 | Am-241 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) +12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140221_0748 07140221_0813 | Am-241 | DIB1 Qi(Vq) -10.0 V, Qo(Vp) -10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140221_0839 07140221_0904 | Am-241 | DIB1 Qi(Vq) +10.0 V, Qo(Vp) +10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140221_0931 07140221_0956 | Am-241 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) -8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140221_1022 07140221_1047 | Am-241 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) +8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140221_1113 07140221_1138 | Am-241 | DIB1 Qi(Vq) -6.0 V, Qo(Vp) -6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140221_1204 07140221_1230 | Am-241 | DIB1 Qi(Vq) +6.0 V, Qo(Vp) +6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140221_1256 07140221_1321 | Am-241 | DIB1 Qi(Vq) -5.0 V, Qo(Vp) -5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140221_1347 07140221_1412 | Am-241 | DIB1 Qi(Vq) +5.0 V, Qo(Vp) +5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140221_1438 07140221_1503 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140221_1530 07140221_1555 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140221_1621 07140221_1646 | Am-241 | DIB1 Qi(Vq) -3.5 V, Qo(Vp) -3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140221_1712 07140221_1737 | Am-241 | DIB1 Qi(Vq) +3.5 V, Qo(Vp) +3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0146 07140222_0211 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) -12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0238 07140222_0303 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) +12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0329 07140222_0354 | Am-241 Ba-133 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0420 07140222_0445 | Am-241 Ba-133 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0511 07140222_0537 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) -8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0603 07140222_0628 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) +8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0654 07140222_0719 | Am-241 Ba-133 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0745 07140222_0811 | Am-241 Ba-133 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0837 07140222_0902 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) -6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_0928 07140222_0953 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) +6.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_1019 07140222_1044 | Am-241 Ba-133 | DIB1 Qi(Vq) -6.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_1111 07140222_1136 | Am-241 Ba-133 | DIB1 Qi(Vq) +6.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_1202 07140222_1227 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_1253 07140222_1318 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_1345 07140222_1410 | Am-241 Ba-133 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140222_1436 07140222_1501 | Am-241 Ba-133 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Y | ||
| 07140223_0626 07140223_0651 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) -10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140223_0717 07140223_0743 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) +10.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140223_0809 07140223_0834 | Am-241 Ba-133 | DIB1 Qi(Vq) -10.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140223_0900 07140223_0925 | Am-241 Ba-133 | DIB1 Qi(Vq) +10.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140223_0951 07140223_1016 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) -5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140223_1043 07140223_1108 | Am-241 Ba-133 | DIB1 Qi(Vq) 0 V, Qo(Vp) +5.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140223_1134 07140223_1159 | Am-241 Ba-133 | DIB1 Qi(Vq) -5.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | |||
| 07140223_1225 07140223_1250 | Am-241 Ba-133 | DIB1 Qi(Vq) +5.0 V, Qo(Vp) 0 V; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; Qi Trigger Threshold of 27 ADC Bins, Qo Trigger Threshold of 20 ADC Bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2k |
Run 41 (G101b floated phonon rail)
| Run | Run Mode | Run Info | # of Events/File | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07131112_0445 07131112_0516 07131112_0546 07131112_0616 | Am-241 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) -4.8 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131112_0648 07131112_0719 07131112_0749 07131112_0820 | Am-241 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) -3.2 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131112_0851 07131112_0922 07131112_0952 07131112_1023 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) -1.6 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131112_1054 07131112_1125 07131112_1155 07131112_1226 | Am-241 | DIB1 Qi(Vq) -2.0 V, Qo(Vp) -0.8 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131112_1257 07131112_1328 07131112_1358 07131112_1429 | Am-241 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131112_1500 07131112_1531 07131112_1601 07131112_1632 | Am-241 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131112_1703 07131112_1734 07131112_1804 07131112_1835 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131112_1906 07131112_1937 07131112_2007 07131112_2038 | Am-241 | DIB1 Qi(Vq) -2.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131113_0236 07131113_0307 07131113_0337 07131113_0408 | Am-241 | DIB1 Qi(Vq) -12.0 V, Qo(Vp) -12.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131113_0439 07131113_0510 07131113_0540 07131113_0611 | Am-241 | DIB1 Qi(Vq) -8.0 V, Qo(Vp) -8.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131113_0642 07131113_0713 07131113_0743 07131113_0814 | Am-241 | DIB1 Qi(Vq) -4.0 V, Qo(Vp) -4.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131113_0846 07131113_0916 07131113_0947 07131113_1017 | Am-241 | DIB1 Qi(Vq) -2.0 V, Qo(Vp) -2.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131114_0417 07131114_0448 07131114_0518 07131114_0549 | Am-241 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) +4.8 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131114_0620 07131114_0651 07131114_0721 07131114_0752 | Am-241 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) +3.2 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131114_0824 07131114_0854 07131114_0925 07131114_0955 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) +1.6 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131114_1027 07131114_1057 07131114_1128 07131114_1158 | Am-241 | DIB1 Qi(Vq) +2.0 V, Qo(Vp) +0.8 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_0133 07131115_0204 07131115_0234 07131115_0305 | Am-241 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_0336 07131115_0407 07131115_0437 07131115_0508 | Am-241 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_0540 07131115_0610 07131115_0641 07131115_0711 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_0743 07131115_0813 07131115_0844 07131115_0914 | Am-241 | DIB1 Qi(Vq) +2.0 V, Qo(Vp) 0.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_0946 07131115_1016 07131115_1047 07131115_1117 | Am-241 | DIB1 Qi(Vq) +12.0 V, Qo(Vp) +12.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_1149 07131115_1219 07131115_1250 07131115_1320 | Am-241 | DIB1 Qi(Vq) +8.0 V, Qo(Vp) +8.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_1352 07131115_1422 07131115_1453 07131115_1523 | Am-241 | DIB1 Qi(Vq) +4.0 V, Qo(Vp) +4.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131115_1555 07131115_1625 07131115_1656 07131115_1727 | Am-241 | DIB1 Qi(Vq) +2.0 V, Qo(Vp) +2.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_0220 07131116_0250 07131116_0321 07131116_0351 | Am-241 | DIB1 Qi(Vq) -4.8 V, Qo(Vp) -12.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_0423 07131116_0453 07131116_0524 07131116_0554 | Am-241 | DIB1 Qi(Vq) -3.2 V, Qo(Vp) -8.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_0626 07131116_0656 07131116_0727 07131116_0757 | Am-241 | DIB1 Qi(Vq) -1.6 V, Qo(Vp) -4.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_0829 07131116_0900 07131116_0930 07131116_1001 | Am-241 | DIB1 Qi(Vq) -0.8 V, Qo(Vp) -2.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_1032 07131116_1103 07131116_1133 07131116_1204 | Am-241 | DIB1 Qi(Vq) +4.8 V, Qo(Vp) +12.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_1235 07131116_1306 07131116_1336 07131116_1407 | Am-241 | DIB1 Qi(Vq) +3.2 V, Qo(Vp) +8.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_1438 07131116_1509 07131116_1539 07131116_1610 | Am-241 | DIB1 Qi(Vq) +1.6 V, Qo(Vp) +4.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y | |
| 07131116_1641 07131116_1712 07131116_1743 07131116_1813 | Am-241 | DIB1 Qi(Vq) +0.8 V, Qo(Vp) +2.0 V, DIB2 Qi & Qo grounded; 0.4096 ms trace length; 100 us delay; OR trigger on DIB1 charge channels; acquisition/flash/cooldown = 10/0.5/20 mins. | 10k | HC | Y |
Run 39 (G106a)
| Run | Run Mode | Run Info | # of Events/File | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07130825_1450 07130825_1515 07130825_1540 07130825_1605 | Am-241 & Pb-210 | DIB1 +4.0 V, DIB2 -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130825_1631 07130825_1657 07130825_1722 07130825_1747 | Am-241 & Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130825_1813 07130825_1838 07130825_1903 07130825_1929 | Am-241 & Pb-210 | DIB1 +3.0 V, DIB2 -3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130825_1954 07130825_2020 07130825_2045 07130825_2110 | Am-241 & Pb-210 | DIB1 -3.0 V, DIB2 +3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130825_2136 07130825_2201 07130825_2226 07130825_2251 | Am-241 & Pb-210 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130825_2317 07130825_2343 07130826_0008 07130826_0033 | Am-241 & Pb-210 | DIB1 -2.5 V, DIB2 +2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130826_0059 07130826_0124 07130826_0149 07130826_0214 | Am-241 & Pb-210 | DIB1 +2.0 V, DIB2 -2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130826_0240 07130826_0306 07130826_0331 07130826_0356 | Am-241 & Pb-210 | DIB1 -2.0 V, DIB2 +2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130826_0422 07130826_0447 07130826_0512 07130826_0537 | Am-241 & Pb-210 | DIB1 +1.0 V, DIB2 -1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130826_0603 07130826_0628 07130826_0654 07130826_0719 | Am-241 & Pb-210 | DIB1 -1.0 V, DIB2 +1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130826_0745 07130826_0810 07130826_0835 07130826_0900 | Am-241 & Pb-210 | DIB1 +8.0 V, DIB2 -8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130826_0926 07130826_0951 07130826_1017 07130826_1042 | Am-241 & Pb-210 | DIB1 -8.0 V, DIB2 +8.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | HC | Y |
| 07130827_0446 07130827_0511 07130827_0536 07130827_0602 07130827_0627 07130827_0652 07130827_0717 07130827_0742 07130827_0807 07130827_0833 07130827_0858 07130827_0923 07130827_0948 07130827_1013 07130827_1039 07130827_1104 07130827_1129 07130827_1154 07130827_1219 07130827_1244 07130827_1310 07130827_1335 07130827_1400 07130827_1425 | Am-241 & Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | Poor neutralisation - must retake | AK | Y |
| 07130828_0103 07130828_0128 07130828_0154 07130828_0219 | Am-241 & Pb-210 | DIB1 +3.5 V, DIB2 -3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130828_0245 07130828_0310 07130828_0335 07130828_0400 | Am-241 & Pb-210 | DIB1 -3.5 V, DIB2 +3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130828_0426 07130828_0452 07130828_0517 07130828_0542 | Am-241 & Pb-210 | DIB1 +4.5 V, DIB2 -4.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130828_0608 07130828_0633 07130828_0658 07130828_0723 | Am-241 & Pb-210 | DIB1 -4.5 V, DIB2 +4.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130828_0749 07130828_0814 07130828_0840 07130828_0905 | Am-241 & Pb-210 | DIB1 +12.0 V, DIB2 -12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130828_0931 07130828_0956 07130828_1021 07130828_1046 | Am-241 & Pb-210 | DIB1 -12.0 V, DIB2 +12.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130829_1507 07130829_1532 07130829_1558 07130829_1624 | Am-241 & Pb-210 | DIB1 +4.0 V, DIB2 -4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130829_1650 07130829_1716 07130829_1741 07130829_1807 | Am-241 & Pb-210 | DIB1 -4.0 V, DIB2 +4.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130829_1833 07130829_1859 07130829_1924 07130829_1950 | Am-241 & Pb-210 | DIB1 +3.0 V, DIB2 -3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130829_2016 07130829_2042 07130829_2107 07130829_2133 | Am-241 & Pb-210 | DIB1 -3.0 V, DIB2 +3.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130829_2159 07130829_2225 07130829_2250 07130829_2316 | Am-241 & Pb-210 | DIB1 +2.0 V, DIB2 -2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130829_2342 07130830_0008 07130830_0034 07130830_0059 | Am-241 & Pb-210 | DIB1 -2.0 V, DIB2 +2.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130830_0125 07130830_0151 07130830_0216 07130830_0242 | Am-241 & Pb-210 | DIB1 +1.0 V, DIB2 -1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y | |
| 07130830_0308 07130830_0334 07130830_0400 07130830_0425 | Am-241 & Pb-210 | DIB1 -1.0 V, DIB2 +1.0 V; 0.4096 ms trace length; 100 us delay; OR trigger on charge channels (thresholds: DIB 1 Qi=26, Qo=40; DIB 2 Qi=36; Qo=40 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 1k | HC | Y |
Run 37 (G101b)
| Run | Run Mode | Run Info | # of Events/Files | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07130627_0344 | Ba-133 & Pb-210 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 40/0.1/20 mins. | 2k | AK | Y | |
| 07130627_0445 | Ba-133 & Pb-210 & Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 40/0.1/20 mins. | 2k | AK | Y | |
| 07130627_0546 | Ba-133 & Pb-210 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 40/0.1/20 mins. | 2k | AC | Y | |
| 07130627_0647 | Ba-133 & Pb-210 & Am-241 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 40/0.1/20 mins. | 2k | AC | Y | |
| 07130627_0749 | Ba-133 & Pb-210 & Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | CP | Y | |
| 07130627_0850 | Ba-133 & Pb-210 & Am-241 | DIB1 -2.5 V, DIB2 +2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 40/0.1/20 mins. | 2k | CP | Y | |
| 07130627_1900 07130627_1941 07130627_2021 07130627_2102 07130627_2142 07130627_2222 07130627_2302 07130627_2342 07130628_0022 07130628_0102 07130628_0143 07130628_0223 | Pb-210 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 20/0.1/20 mins. | 2k | Y | ||
| 07130628_0303 07130628_0344 07130628_0424 07130628_0504 07130628_0544 07130628_0624 07130628_0704 07130628_0744 07130628_0825 07130628_0905 07130628_0945 07130628_1025 | Pb-210 & Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (32/28 ADC bins for Dib 1/2); acquisition/flash/cooldown = 20/0.1/20 mins. | 2k | Y | ||
| 07130629_1628 07130629_1659 07130629_1729 07130629_1759 07130629_1829 07130629_1859 | Pb-210 & Am-241 | DIB1 +0.5 V, DIB2 -0.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger lowered to 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130629_1931 07130629_2001 07130629_2031 07130629_2101 07130629_2131 07130629_2202 | Pb-210 & Am-241 | DIB1 -0.5 V, DIB2 +0.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger lowered to 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130630_0420 07130630_0500 07130630_0541 07130630_0621 07130630_0701 07130630_0741 07130630_0821 07130630_0901 07130630_0941 07130630_1021 07130630_1102 07130630_1142 07130630_1222 07130630_1302 07130630_1342 07130630_1422 07130630_1502 07130630_1542 07130630_1623 07130630_1703 07130630_1743 07130630_1823 | Pb-210 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130630_1904 07130630_1944 07130630_2024 07130630_2104 07130630_2144 07130630_2224 07130630_2305 07130630_2345 07130701_0025 07130701_0105 07130701_0145 07130701_0225 07130701_0305 07130701_0346 07130701_0426 07130701_0506 07130701_0546 07130701_0626 07130701_0706 07130701_0746 07130701_0827 07130701_0907 | Pb-210 & Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130628_2208 07130628_2239 07130628_2309 07130628_2339 07130629_0009 07130629_0040 07130629_0110 07130629_0140 07130629_0110 07130629_0140 07130629_0210 07130629_0240 07130629_0310 07130629_0340 07130629_0411 07130629_0441 07130629_0511 07130629_0541 | Pb-210 & Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; trigger on Qi channels (DIB 1 trigger threshold increased to 34 ADC bins due to noisy baseline, DIB 2 trigger threshold remains unchanged at 28 ADC bins); 10/0.1/20 mins acquisition/flash/cooldown. | 2k | AC | Y | |
| 07130629_0611 07130629_0641 07130629_0712 07130629_0742 07130629_0812 07130629_0842 07130629_0912 07130629_0942 07130629_1012 07130629_1043 07130629_1113 07130629_1143 07130629_1213 07130629_1243 07130629_1313 07130629_1343 | Pb-210 & Am-241 | DIB1 -2.5 V, DIB2 +2.5 V; trigger on Qi channels (DIB 1 trigger threshold increased to 34 ADC bins due to noisy baseline, DIB 2 trigger threshold remains unchanged at 28 ADC bins); 10/0.1/20 mins acquisition/flash/cooldown. | 2k | AC | Y | |
| 07130701_2032 07130701_2102 07130701_2132 07130701_2202 07130701_2232 07130701_2302 | Pb-210 & Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130701_2334 07130702_0004 07130702_0034 07130702_0104 07130702_0134 07130702_0204 | Pb-210 & Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130702_0235 07130702_0305 07130702_0335 07130702_0406 07130702_0436 07130702_0506 | Pb-210 & Am-241 | DIB1 +1.5 V, DIB2 -1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130702_0537 07130702_0607 07130702_0637 07130702_0707 07130702_0737 07130702_0807 | Pb-210 & Am-241 | DIB1 -1.5 V, DIB2 +1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130702_0839 07130702_0909 07130702_0939 07130702_1009 07130702_1039 07130702_1109 | Pb-210 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130702_1140 07130702_1210 07130702_1241 07130702_1311 07130702_1341 07130702_1411 | Pb-210 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130702_1442 07130702_1512 07130702_1542 07130702_1612 07130702_1642 07130702_1713 | Pb-210 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130702_1744 07130702_1814 07130702_1844 07130702_1914 07130702_1944 07130702_2014 | Pb-210 & Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130703_0129 07130703_0159 07130703_0229 07130703_0259 07130703_0329 07130703_0400 | Pb-210 & Am-241 | DIB1 +3.5 V, DIB2 -3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130703_0431 07130703_0501 07130703_0531 07130703_0602 07130703_0632 07130703_0702 | Pb-210 & Am-241 | DIB1 -3.5 V, DIB2 +3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels only; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130704_0154 07130704_0224 07130704_0254 07130704_0324 07130704_0354 071307014_0425 | Pb-210 & Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 071307014_0456 07130704_0526 07130704_0556 07130704_0626 07130704_0656 07130704_0726 | Pb-210 & Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130704_0757 07130704_0827 07130704_0858 07130704_0928 07130704_0958 071307024_1028 | Pb-210 & Am-241 | DIB1 +1.5 V, DIB2 -1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130704_1059 071307024_1129 07130704_1159 07130704_1229 07130704_1259 07130704_1330 | Pb-210 & Am-241 | DIB1 -1.5 V, DIB2 +1.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | Y | ||
| 07130704_1401 07130704_1431 07130704_1501 07130704_1531 07130704_1601 07130704_1631 | Pb-210 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130704_1702 07130704_1732 07130704_1802 07130704_1833 07130704_1903 07130704_1933 | Pb-210 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130704_2004 071307024_2034 07130704_2104 07130704_2134 07130704_2204 07130704_2235 | Pb-210 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130704_2305 07130704_2336 07130705_0006 07130705_0036 07130705_0106 07130705_0136 | Pb-210 & Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130705_0207 071307035_0237 07130705_0307 07130705_0338 07130705_0408 07130705_0438 | Pb-210 & Am-241 | DIB1 +3.5 V, DIB2 -3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | CP | Y | |
| 07130705_0509 07130705_0539 07130705_0609 07130705_0639 07130705_0710 07130705_0740 | Pb-210 & Am-241 | DIB1 -3.5 V, DIB2 +3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | CP | Y | |
| 07130705_1343 07130705_1423 07130705_1503 | Pb-210 & Am-241 | DIB1 +5 V, DIB2 -5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130705_1545 07130705_1625 07130705_1705 | Pb-210 & Am-241 | DIB1 -5 V, DIB2 +5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130705_2217 07130705_2257 07130705_2337 | Pb-210 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130706_0019 07130706_0059 07130706_0139 | Pb-210 & Am-241 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130706_0220 07130706_0300 07130706_0340 | Pb-210 & Am-241 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130706_0421 07130706_0501 07130706_0542 | Pb-210 & Am-241 | DIB1 -8 V, DIB2 +8 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130706_0623 07130706_0703 07130706_0743 | Pb-210 & Am-241 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130706_0824 07130706_0904 07130706_0944 | Pb-210 & Am-241 | DIB1 -10 V, DIB2 +10 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130706_1025 07130706_1106 07130706_1146 | Pb-210 & Am-241 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130706_1227 07130706_1307 07130706_1347 | Pb-210 & Am-241 | DIB1 -12 V, DIB2 +12 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130710_1659 07130710_1729 07130710_1759 07130710_1830 07130710_1900 07130710_1930 | Pb-210, Am-241 and Ba-133 | DIB1 + 2.5 V, DIB2 - 2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130710_2001 07130710_2031 07130710_2101 07130710_2131 07130710_2201 07130710_2231 | Pb-210, Am-241 and Ba-133 | DIB1 - 2.5 V, DIB2 + 2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130710_2302 07130710_2333 07130711_0003 07130711_0033 07130711_0103 07130711_0133 | Pb-210, Am-241 and Ba-133 | DIB1 + 3 V, DIB2 - 3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_0204 07130711_0234 07130711_0305 07130711_0335 07130711_0405 07130711_0435 | Pb-210, Am-241 and Ba-133 | DIB1 - 3 V, DIB2 + 3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_0506 07130711_0536 07130711_0606 07130711_0636 07130711_0706 07130711_0737 | Pb-210, Am-241 and Ba-133 | DIB1 + 3.5 V, DIB2 - 3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_0808 07130711_0838 07130711_0908 07130711_0938 07130711_1008 07130711_1039 | Pb-210, Am-241 and Ba-133 | DIB1 - 3.5 V, DIB2 + 3.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_1507 07130711_1547 07130711_1628 | Pb-210, Am-241 and Ba-133 | DIB1 + 5 V, DIB2 - 5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_1709 07130711_1749 07130711_1829 | Pb-210, Am-241 and Ba-133 | DIB1 - 5 V, DIB2 + 5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_1911 07130711_1951 07130711_2031 | Pb-210, Am-241 and Ba-133 | DIB1 + 8 V, DIB2 - 8 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_2112 07130711_2152 07130711_2232 | Pb-210, Am-241 and Ba-133 | DIB1 - 8 V, DIB2 + 8 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130711_2313 07130711_2354 07130712_0034 | Pb-210, Am-241 and Ba-133 | DIB1 + 10 V, DIB2 - 10 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130712_0115 07130712_0155 07130712_0235 | Pb-210, Am-241 and Ba-133 | DIB1 - 10 V, DIB2 + 10 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130712_0316 07130712_0356 07130712_0436 | Pb-210, Am-241 and Ba-133 | DIB1 + 12 V, DIB2 - 12 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130712_0518 07130712_0558 07130712_0638 | Pb-210, Am-241 and Ba-133 | DIB1 - 12 V, DIB2 + 12 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 20/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130712_2135 07130712_2205 07130712_2235 07130712_2305 07130712_2335 07130713_0006 | Pb-210, Am-241 | DIB1 + 0.5 V, DIB2 - 0.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130713_0037 07130713_0107 07130713_0137 07130713_0207 07130713_0237 07130713_0307 | Pb-210, Am-241 | DIB1 + 2.5 V, DIB2 - 2.5 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y | |
| 07130713_0338 07130713_0408 07130713_0439 07130713_0509 07130713_0539 07130713_0609 | Pb-210, Am-241 | DIB1 + 4 V, DIB2 - 4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi and Qo channels; DIB 1 trigger at 32 ADC bins, DIB 2 trigger at 28 ADC bins; 10/0.1/20 minutes acquisition/flash/cool-down cycle. | 2k | AC | Y |
Run 36 (G101b)
| Run | Run Mode | Run Info | # of Events/Files | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07130312_1638 07130312_1703 07130312_1728 07130312_1753 07130312_1818 | Pb-210 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (35 ADC bins); acquisition/flash/cooldown = 5/0.1/20 mins. | 2k | Slight evidence of Pb-210 peak; Need more stats | RR, AC | Y - raw files don't exist |
| 07130313_2024 07130313_2114 07130313_2204 | Ba-133& Pb-210 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (30/35 ADC bins for Dib 1/2); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | Zip 1 has low stats | RR | Y |
| 07130313_2256 07130313_2346 07130314_0036 | Ba-133& Pb-210 & Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (30/35 ADC bins for Dib 1/2); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | Zip 2 has low stats; Files 13 and 14 from *_0036 were removed. | RR | Y |
| 07130315_0337 07130315_0412 07130315_0447 07130315_0523 07130315_0558 07130315_0633 07130315_0708 07130315_0743 07130315_0818 07130315_0853 07130315_0928 07130315_1003 07130315_1038 07130315_1114 | Pb-210 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (30/35 ADC bins for Dib1/2); acquisition/flash/cooldown = 15/0.1/20 mins. | 2k | Slight evidence of Pb-210 peak; dominance of Qo events on electron-collecting side | RR | Y |
| 07130315_1417 07130315_1507 | Pb-210 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (30/35 ADC bins for Dib1/2); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | Low energy/ Qouter event dominance on Dib1 | RR | Y |
| 07130315_1606 07130315_1656 07130315_1746 | Ba-133& Pb-210 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (30/35 ADC bins for Dib1/2); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | AC | Y | |
| 07130319_1344 | Pb-210 & Am-241 | DIB2 -4 V; 0.4096 ms trace length; 100 us delay; trigger on Qi (23 ADC); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | RR | Y | |
| 07130320_1411 | Pb-210 & Am-241 | DIB2 +4 V; 0.4096 ms trace length; 100 us delay; trigger on Qi (23 ADC); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | RR | Y | |
| 07130321_1401 | Pb-210 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; OR trigger on Qi channels (25/32 ADC bins for Dib1/2); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | RR | Y | |
| 07130321_1602 | Pb-210 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; AND trigger on Qi channels (25/32 ADC bins for Dib1/2); acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | RR | Y | |
| 07130321_1908 | Pb-210 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger only on Qi2 at 32 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 2k | RR | Y |
Run 34 (G106a)
| Run | Run Mode | Run Info | # of Events/Files | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07130204_1255 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_1430 | Ba-133 & Am-241 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_1522 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_1613 | Ba-133 & Am-241 | DIB1 +5 V, DIB2 -5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_1704 | Ba-133 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_1755 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_2100 | Ba-133 & Am-241 | DIB1 -8 V, DIB2 +8 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_2151 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_2242 | Ba-133 & Am-241 | DIB1 -5 V, DIB2 +5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130204_2333 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_0024 | Ba-133 & Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_0115 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_1839 | Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_1930 | Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_2022 | Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_2113 | Am-241 | DIB1 +1.5 V, DIB2 -1.5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_2204 | Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130205_2255 | Am-241 | DIB1 +0.5 V, DIB2 -0.5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1117 | Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1208 | Am-241 | DIB1 -2.5 V, DIB2 +2.5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1300 | Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1351 | Am-241 | DIB1 -1.5 V, DIB2 +1.5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1442 | Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1533 | Am-241 | DIB1 -0.5 V, DIB2 +0.5 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1823 | Ba-133 & Am-241 | DIB1 -10 V, DIB2 +10 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130206_1914 | Ba-133 & Am-241 | DIB1 -12 V, DIB2 +12 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 1 at 45/40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130207_1052 | Ba-133 & Am-241 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | Y | |||
| 07130207_1144 | Ba-133 & Am-241 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | Y | |||
| 07130207_1827 | Y-88 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | No Y-88 peaks | RR | Y | |
| 07130207_1918 | Y-88 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | No Y-88 peaks | RR | Y | |
| 07130207_2010 | Y-88 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | Y | |||
| 07130208_1113 | Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130208_1204 | Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130208_1255 | Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130208_2304 | Am-241 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | RR | Y | ||
| 07130208_2355 | Am-241 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | Y | |||
| 07130208_0047 | Am-241 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; trigger on Qi/Qo on Dib 2 at 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | Y | |||
| 07130210_0152 | Cf-252 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger on phonon channels A1, B1, C1, D2, C3 on Side1 at 25 ADC bins each.; acquisition/flash/cooldown = 30/0.1/20 mins. | Bad data | Y | ||
| 07130210_0242 | Cf-252 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger on phonon channels A1, B1, C1, D2, C3 on Side1 at 25 ADC bins each.; acquisition/flash/cooldown = 30/0.1/20 mins. | Bad data | Y | ||
| 07130210_0333 | Cf-252 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger on phonon channels A1, B1, C1, D2, C3 on Side1 at 25 ADC bins each.; acquisition/flash/cooldown = 30/0.1/20 mins. | Bad data | Y | ||
| 07130210_1855 07130210_2254 07130210_2344 07130211_0035 07130211_0125 07130211_0215 07130211_0305 07130211_0355 07130211_0445 07130211_0535 07130211_0625 07130211_0715 07130211_0806 07130211_0856 | Cf-252 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger on phonon channels A1, B1, C1, D2, C3 on Side1 at 25 ADC bins each.; acquisition/flash/cooldown = 30/0.1/20 mins. | Data quality doesn't seem good. Last two files have not been processed. | RR, SM | Y | |
| 07130214_0154 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger on B1, C1, D1, A2, A3, C3 at 25 ADC; acquisition/flash/cooldown = 120/0.1/20 mins. | Ba-133 is visible, though spectrum isn't great | RR | Y | |
| 07130214_0416 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 3.2768 ms trace length; 400 us delay; Trigger on B1, C1, D1, A2, A3, C3 at 25 ADC; acquisition/flash/cooldown = 120/0.1/20 mins. | This and previous phonon data sets in this run have coincidence time of 50 uS. Ba-133 is visible, though spectrum isn't great | RR | Y | |
| 07130217_1650 | Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger on phonon channels (see run summary). 200 uS coincidence time. | A short 20 minute data set to act as a sanity check for charge and phonon collection. Unfortunately, at the end 20 minutes G1 and G2 started to increase dramatically. | RR | Y |
Run 32 (G103a)
| Run | Run Mode | Run Info | # of Events/Files | Comments | Analysis | Processed |
|---|---|---|---|---|---|---|
| 07121012_0136 07121012_0256 07121012_0416 | Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 20 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 5 x 10k 5 x 10k 5 x 10k | Used old SQUID settings. | Y Y Y |
|
| 07121012_0538 07121012_0658 07121012_0818 | Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 20 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 2 x 10k 6 x 10k 6 x 10k | Used old SQUID settings; unsure why 07121012_0538 comprises of only two files. | Y Y Y |
|
| 07121013_0805 07121013_0926 07121013_1046 | Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k | Used new SQUID settings. | Y Y Y |
|
| 07121013_1207 07121013_1327 07121013_1447 | Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k | Used new SQUID settings. | Y Y Y |
|
| 07121014_0317 07121014_0438 07121014_0558 | Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 20 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings. | X Y Y |
|
| 07121014_0719 07121014_0839 07121014_0959 | Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 20 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings. | Y Y Y |
|
| 07121017_0311 07121017_0439 07121017_0611 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 20 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 5 x 10k 5 x 10k 4 x 10k | Used old SQUID settings; file 6 from 07121017_0439 looks empty so removed. | Y Y Y |
|
| 07121017_0713 07121017_0834 07121017_0955 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 20 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 5 x 10k 5 x 10k 5 x 10k | Used old SQUID settings; no flash before 07121017_0713, unsure why. | Y X Y |
|
| 07121018_0823 07121018_0949 07121018_1115 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 4 x 10k 4 x 10k 4 x 10k | Used old SQUID settings; unsure why files are spread out by 86 minutes rather than expected 80 minutes. | X Y Y |
|
| 07121018_1225 07121018_1349 07121018_1513 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 5 x 10k 5 x 10k 5 x 10k | Used old SQUID settings; unsure why files are spread out by 84 minutes rather than expected 80 minutes; no flash before 07121018_1225, unsure why. | Y X X |
|
| 07121019_0305 07121019_0426 07121019_0546 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 6 x 10k 5 x 10k 6 x 10k | Used old SQUID settings. | Y Y Y |
|
| 07121019_0706 07121019_0830 07121019_0954 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 5 x 10k 5 x 10k 5 x 10k | Used old SQUID settings; unsure why files are spread out by 84 minutes rather than expected 80 minutes; restarting DAQDAQ2 vi seems to have solved problem with flashing witnessed in previous data sets. | Y X Y |
|
| 07121020_2322 07121021_0045 07121021_0206 | Ba-133 & Am-241 | DIB1 -3 V, DIB2 +3 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 5 x 10k 5 x 10k 5 x 10k | Used old SQUID settings. | X X X |
|
| 07121021_0324 07121021_0447 07121021_0609 | Ba-133 & Am-241 | DIB1 +3 V, DIB2 -3 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 5 x 10k 5 x 10k 5 x 10k | Used old SQUID settings. | Y Y Y |
|
| 07121022_0228 07121022_0348 07121022_0508 | Am-241 | DIB1 -3 V, DIB2 +3 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings. | X X X |
|
| 07121022_0630 07121022_0750 07121022_0910 | Am-241 | DIB1 +3 V, DIB2 -3 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings. | Y Y Y |
|
| 07121023_0624 07121023_0744 07121023_0904 | Am-241 | DIB1 -3 V, DIB2 +3 V; 3.2768 ms trace length; 400 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings. | Y X X |
|
| 07121023_1026 07121023_1146 07121023_1306 | Am-241 | DIB1 +3 V, DIB2 -3 V; 3.2768 ms trace length; 400 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings. | Y Y Y |
|
| 07121024_0331 07121024_0451 07121024_0611 | Am-241 | DIB1 -3 V, DIB2 +3 V; 3.2768 ms trace length; 400 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs). | Y X Y |
|
| 07121024_0733 07121024_0853 07121024_1013 | Am-241 | DIB1 +3 V, DIB2 -3 V; 3.2768 ms trace length; 400 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 60/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs). | Y Y Y |
|
| 07121027_0556 07121027_0647 07121027_0737 07121027_0827 | Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121027_0918 07121027_1008 07121027_1058 07121027_1148 | Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 8 x 10k 7 x 10k 8 x 10k 8 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121027_1240 07121027_1330 07121027_1420 07121027_1510 | Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121028_0705 07121028_0755 | Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 4 x 10k 4 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active, DAQ crashed 21 minutes into data series 07121028_0755, data quality suspect as signs of trapped flux were visible on phonon pulses. Confirmed bad data. | Y Y |
|
| 07121030_0329 07121030_0419 07121030_0509 07121030_0559 | Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 3 x 10k 3 x 10k 3 x 10k 3 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active, smaller than expected number of files for unknown reason. Confirmed bad data. | Y Y Y Y |
|
| 07121030_0651 07121030_0741 07121030_0831 07121030_0922 | Am-241 | DIB1 -2.5 V, DIB2 +2.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 4 x 10k 4 x 10k 4 x 10k 4 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active, smaller than expected number of files for unknown reason. Confirmed bad data. | Y Y Y Y |
|
| 07121101_0758 07121101_0849 | Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y |
|
| 07121101_0940 07121101_1030 | Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y |
|
| 07121102_0302 07121102_0352 07121102_0442 07121102_0533 | Am-241 | DIB1 -2.5 V, DIB2 +2.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 8 x 10k 8 x 10k 8 x 10k 8 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121102_0624 07121102_0714 07121102_0804 07121102_0854 | Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121103_0105 07121103_0155 07121103_0246 07121103_0336 | Am-241 | DIB1 +1.5 V, DIB2 -1.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121103_0427 07121103_0517 07121103_0607 07121103_0657 | Am-241 | DIB1 -1.5 V, DIB2 +1.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121103_0748 07121103_0838 07121103_0928 07121103_1019 | Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121103_1110 07121103_1200 07121103_1250 07121103_1340 | Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 6 x 10k 6 x 10k 5 x 10k 5 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121103_2332 07121104_0023 07121104_0113 07121104_0103 | Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active, time stamp oddity due to end of daylight savings time. | Y Y Y Y |
|
| 07121104_0154 07121104_0244 07121104_0334 07121104_0425 | Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 8 x 10k 8 x 10k 8 x 10k 8 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y X X |
|
| 07121104_0515 07121104_0606 07121104_0656 07121104_0746 | Am-241 | DIB1 +0.5 V, DIB2 -0.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 5 x 10k 5 x 10k 5 x 10k 5 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121104_0837 07121104_0927 07121104_1017 07121104_1107 | Am-241 | DIB1 -0.5 V, DIB2 +0.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 4 x 10k 4 x 10k 4 x 10k 4 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121105_0621 07121105_0712 07121105_0802 07121105_0852 | Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121105_0943 07121105_1033 07121105_1123 07121105_1213 | Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121106_0642 07121106_0732 07121106_0822 07121106_0912 | Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 8 x 10k 8 x 10k 8 x 10k 8 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121106_1003 07121106_1053 07121106_1144 07121106_1234 | Am-241 | DIB1 +5 V, DIB2 -5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 7 x 10k 7 x 10k 7 x 10k 7 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121106_1325 07121106_1415 07121106_1505 07121106_1555 | Am-241 | DIB1 -5 V, DIB2 +5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 6 x 10k 6 x 10k 6 x 10k 6 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121108_0330 07121108_0420 07121108_0510 07121108_0601 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. Only one file per data series; unsure why. Confirmed bad data. | Y Y Y Y |
|
| 07121108_0652 07121108_0742 07121108_0832 07121108_0922 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. Only one file per data series; unsure why. Confirmed bad data. | Y Y Y Y |
|
| 07121108_1013 07121108_1103 07121108_1154 07121108_1244 | Ba-133 & Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. Only one file per data series; unsure why. Confirmed bad data. | Y Y Y Y |
|
| 07121108_1335 07121108_1425 07121108_1515 07121108_1605 | Ba-133 & Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. Only one file per data series; unsure why. Confirmed bad data. | Y Y Y Y |
|
| 07121109_0518 07121109_0608 07121109_0658 07121109_0748 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. Only one file per data series; unsure why. Confirmed bad data. | Y Y Y Y |
|
| 07121109_0839 07121109_0929 07121109_1019 07121109_1110 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. Only one file per data series; unsure why. Confirmed bad data. | Y Y Y Y |
|
| 07121111_0447 07121111_0537 07121111_0627 07121111_0717 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 8 x 10k 8 x 10k 8 x 10k 8 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121111_0808 07121111_0859 07121111_0949 07121111_1039 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 8 x 10k 8 x 10k 8 x 10k 8 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y |
|
| 07121112_0109 07121112_0135 07121112_0200 07121112_0225 07121112_0250 07121112_0315 07121112_0341 07121112_0406 07121112_0431 07121112_0457 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y Y Y Y Y Y Y |
|
| 07121112_0522 07121112_0547 07121112_0612 07121112_0637 07121112_0702 07121112_0727 07121112_0753 07121112_0818 07121112_0843 07121112_0909 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y Y Y Y Y Y Y |
|
| 07121112_0933 07121112_0959 07121112_1024 07121112_1049 07121112_1114 07121112_1139 07121112_1204 07121112_1230 07121112_1255 07121112_1320 | Ba-133 & Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y Y Y Y Y Y Y |
|
| 07121112_1346 07121112_1411 07121112_1436 07121112_1501 07121112_1526 07121112_1552 07121112_1617 07121112_1642 07121112_1707 07121112_1733 | Ba-133 & Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y Y Y Y Y Y Y |
|
| 07121113_0544 07121113_0609 07121113_0634 07121113_0659 07121113_0725 07121113_0750 07121113_0815 07121113_0840 07121113_0905 07121113_0931 | Ba-133 & Am-241 | DIB1 -1.5 V, DIB2 +1.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y Y Y Y Y Y Y |
|
| 07121113_0956 07121113_1021 07121113_1046 07121113_1112 07121113_1137 07121113_1202 07121113_1228 07121113_1253 07121113_1318 07121113_1343 | Ba-133 & Am-241 | DIB1 +1.5 V, DIB2 -1.5 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k 2 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y Y Y Y Y Y Y |
|
| 07121114_0305 07121114_0355 07121114_0445 07121114_0536 07121114_0626 07121114_0716 07121114_0807 07121114_0857 07121113_0947 07121113_1037 | Ba-133 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; trigger on all DIB 1 & DIB 2 charge channels 40 ADC bins; acquisition/flash/cooldown = 30/0.1/20 mins. | 8 x 10k 8 x 10k 8 x 10k 8 x 10k 8 x 10k 8 x 10k 8 x 10k 8 x 10k 8 x 10k 8 x 10k | Used old SQUID settings, charge channel trigger coincidence time 10 μs (reduced from 200 μs), charge-only data although phonon channels active. | Y Y Y Y Y Y Y Y Y Y |
|
| 07121117_1251 07121117_1318 07121117_1346 07121117_1411 07121117_1437 07121117_1503 07121117_1528 07121117_1554 07121117_1621 07121117_1647 07121117_1715 07121117_1740 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used final SQUID settings, phonon channel trigger coincidence time 10 μs (reduced from 200 μs), actual acquisition times vary from 3-8 minutes for unknown reason; data is probably fine. | Y Y Y Y Y Y Y Y Y Y Y Y |
|
| 07121117_1754 07121117_1824 07121117_1857 07121117_1923 07121117_1952 07121117_2019 07121117_2044 07121117_2111 07121117_2136 07121117_2204 07121117_2232 | Ba-133 & Am-241 | DIB1 +3 V, DIB2 -3 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used final SQUID settings, phonon channel trigger coincidence time 10 μs (reduced from 200 μs), no flash before data series 07121117_1754, 11 data series taken rather than the twelve that were requested, actual acquisition times vary from 5-13 minutes for unknown reason; apart from first series, data is probably fine. | Y Y Y Y Y Y Y Y Y Y Y |
|
| 07121118_0210 07121118_0237 07121118_0303 07121118_0335 07121118_0401 07121118_0434 07121118_0507 07121118_0533 07121118_0606 07121118_0638 07121118_0711 07121118_0743 07121118_0816 07121118_0849 07121118_0922 07121118_0952 07121118_1022 07121118_1055 07121118_1120 07121118_1147 | Cf-252 & Am-241 | DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k 1 x 10k | Used final SQUID settings, phonon channel trigger coincidence time 10 μs (reduced from 200 μs), 20 data series taken rather than the twenty-four that were requested, actual acquisition times vary from 5-14 minutes for unknown reason; data is probably fine. | Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y |
|
| 07121119_0736 07121119_0801 07121119_0827 07121119_0852 07121119_0917 07121119_0942 07121119_1007 07121119_1033 07121119_1058 07121119_1124 07121119_1149 07121119_1215 07121119_1240 07121119_1305 07121119_1330 07121119_1356 07121119_1421 07121119_1446 07121119_1512 07121119_1537 07121119_1602 07121119_1628 07121119_1653 07121119_1718 | Am-241 | DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k | Used final SQUID settings, phonon channel trigger coincidence time 10 μs (reduced from 200 μs). | Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y |
|
| 07121119_2318 07121119_2343 07121120_0009 07121120_0034 07121120_0100 07121120_0125 07121120_0150 07121120_0216 07121120_0241 07121120_0306 07121120_0332 07121120_0357 07121120_0422 07121120_0448 07121120_0513 07121120_0538 07121120_0604 07121120_0629 07121120_0654 07121120_0719 07121120_0745 07121120_0810 07121120_0835 07121120_0901 | Cf-252 & Am-241 | DIB1 +3 V, DIB2 -3 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k 2 x 2k | Used final SQUID settings, phonon channel trigger coincidence time 10 μs (reduced from 200 μs). | Y Y Y Y Y X Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y |
|
| 07121120_2153 07121120_2218 07121120_2243 07121120_2309 07121120_2334 07121120_2359 07121121_0024 07121121_0049 07121121_0115 07121121_0140 07121121_0205 07121121_0230 07121121_0256 07121121_0321 07121121_0346 07121121_0412 07121121_0437 07121121_0502 07121121_0528 07121121_0553 07121121_0618 07121121_0643 07121121_0709 07121121_0734 | Am-241 | DIB1 +3 V, DIB2 -3 V; 3.2768 ms trace length; 400 us delay; trigger on all phonon channels 25 ADC bins; acquisition/flash/cooldown = 5/0.1/20 mins. | 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k 3 x 2k | Used final SQUID settings, phonon channel trigger coincidence time 10 μs (reduced from 200 μs). | Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y Y |
Run 31 (G103a)
Italicized series are currently being re-processed.
| 07120912_2213 07120912_2303 07120912_2353 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | 10 x 10k 10 x 10k 10 x 10k | Y | |||
| Run | Run Mode | Run Info | # of Events/Files | Comments | Analysis | Processed | |
|---|---|---|---|---|---|---|---|
| 07120719_2328 | Am-241 | DIB1 0 V, DIB2 0 V; 3.2768 ms trace length; 400 us delay; Trigger on all phonon channels but C2 | 1 file | Dib1=DCRC 32, Dib2= DCRC33 | |||
| 07120720_1419 | Am-241 | DIB1 0 V, DIB2 0 V; 3.2768 ms trace length; 400 us delay; Trigger on all phonon channels but C2 (?) | 1 file | ||||
| 07120721_0128 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Dib1=DCRC 32, Dib2= DCRC33 It seems no flashing occurred for the series taken through *22_0327… error in LED selection for DAQ files. Flashing only occured manually after DAQ crashed | Y | ||
| 07120721_0219 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 10 files | Y | |||
| 07120721_0441 | Ba-133 & Am-241 | DIB1 -2.5 V, DIB2 +2.5 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_0534 | Ba-133 & Am-241 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_0626 | Ba-133 & Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_0718 | Ba-133 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_0809 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 10 files | Y | |||
| 07120721_0901 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 10 files | Y | |||
| 07120721_0953 | Ba-133 & Am-241 | DIB1 -5 V, DIB2 +5 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_1044 | Ba-133 & Am-241 | DIB1 +5 V, DIB2 -5 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_1135 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_1853 | Ba-133 & Am-241 | DIB1 0 V, DIB2 0 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | Zero bias data | ||||
| 07120721_2039 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120721_2308 | Ba-133 & Am-241 | DIB1 -8 V, DIB2 +8 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 10 files | Y | |||
| 07120722_0001 | Ba-133 & Am-241 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120722_0053 | Ba-133 & Am-241 | DIB1 -10 V, DIB2 +10 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120722_0144 | Ba-133 & Am-241 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120722_0236 | Ba-133 & Am-241 | DIB1 -12 V, DIB2 +12 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120722_0327 | Ba-133 & Am-241 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; Trigger 25 ADC bins on phonon channels; data/flashing/cooldown = 30/0.1/20 min. | 9 files | Y | |||
| 07120723_2148 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Dib1=DCRC 33, Dib2= DCRC32 Not good spectra | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) \\ | Y | ||
| 07120724_0011 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 18 files | Not good spectra | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) | Y | |
| 07120724_0232 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 20 files | Not good spectra | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) | Y | |
| 07120724_0454 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 17 files | Not good spectra | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) | Y | |
| 07120724_0716 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 13 files | Not good spectra | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) | Y | |
| 07120724_0938 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 12 files | Not good spectra No charge data for zip2 | AK (qsum spectra, qpart, Qi, Qo) | Y | |
| 07120724_2334 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 11 files | Dib1=DCRC 33, Dib2= DCRC32 | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |
| 07120725_0156 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 12 files | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120725_0420 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 10 files | x | |||
| 07120725_0644 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 10 files | x | |||
| 07120725_0911 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 11 files | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120725_1134 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 11 files | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120726_0115 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Dib1=DCRC 33, Dib2= DCRC32, Dib3=DCRC31 These files seem too small processed. | x | |||
| 07120726_0338 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Detector not flashed before this series These files seem too small processed. | x | |||
| 07120726_0602 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Detector not flashed before this series These files seem too small processed. | x | |||
| 07120726_0827 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Detector not flashed before this series These files seem too small processed. | x | |||
| 07120726_1053 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 5 files instead of 9-10 | Detector not flashed before this series These files seem too small processed. | x | ||
| 07120726_1316 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Detector not flashed before this series These files seem too small processed. | x | |||
| 07120727_0215 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Dib1=DCRC 33, Dib2= DCRC32, Dib3=DCRC31 | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) RR (max qsum spectra w/ cuts) | Y | ||
| 07120727_0436 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) RR (max qsum spectra w/ cuts) | Y | |||
| 07120727_0658 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) RR (max qsum spectra w/ cuts) | Y | |||
| 07120727_0919 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Only 2 files | x | |||
| 07120727_1140 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | AK (qsum spectra, chi2 vs. qsum, Qi, Qo, qsum vs. time) RR (max qsum spectra w/ cuts) | Y | |||
| 07120727_1401 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Only 1 file | x | |||
| 07120728_0254 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Dib1=DCRC 33, Dib2= DCRC32, Dib3=DCRC31 | Y | |||
| 07120728_0516 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | Y | ||||
| 07120728_0737 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | RR (Qsum1 and 2, Max(qsum), Qi/Qo, chimax vs. max qsum) | Y | |||
| 07120728_0959 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | RR (Qsum1 and 2, Max(qsum), Qi/Qo, chimax vs. max qsum) | Y | |||
| 07120728_1220 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | 1 file | X | |||
| 07120728_1442 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 3.2768 ms trace length; 400 us delay; Trigger 20 ADC bins on 12 phonon channels; data/flashing/cooldown = 120/0.1/20 min. | RR (Qsum1 and 2, Max(qsum), Qi/Qo, Qpart, chimax vs. max qsum) SM(chi2_vs_qi(s1_and_s2), qi(s2)_vs_time, qsum(s1)_vs_time(high_energy), qsum(s1)_high_energy, qsum(s2)_high_energy)) | Y | |||
| 07120729_1929 07120729_2019 07120729_2109 | Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; Trigger on 4 Q channels at 40 ADC bins; Data/flash/cooldown = 30/0.1/20 min | Dib1=DCRC 33, Dib2= DCRC32 - charge only data | Y | |||
| 07120729_2200 07120729_2250 07120729_2340 | Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger on 4 Q channels at 40 ADC bins; Data/flash/cooldown = 30/0.1/20 min | Charge pulses are okay | Y | |||
| 07120730_0031 07120730_0122 07120730_0212 | Am-241 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; Trigger on 4 Q channels at 40 ADC bins; Data/flash/cooldown = 30/0.1/20 min | Charge pulses not good - spectrum looks noisy | Y | |||
| 07120730_0304 07120730_0355 07120730_0446 | Am-241 | DIB1 -8 V, DIB2 +8 V; 0.4096 ms trace length; 100 us delay; Trigger on 4 Q channels at 40 ADC bins; Data/flash/cooldown = 30/0.1/20 min | 0304 and 0446 have especially small files- not sure why Charge pulses not good | Y | |||
| 07120806_1925 07120806_2015 07120806_2105 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Dib1=DCRC 33, Dib2= DCRC32 | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120806_2157 07120806_2247 07120806_2337 | Ba-133 & Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120807_0028 07120807_0118 07120807_0208 | Ba-133 & Am-241 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | RR (Qi/Qo, Q vs. T, Qsum, Max Qsum) | Y | |||
| 07120807_0300 07120807_0350 07120807_0440 | Ba-133 & Am-241 | DIB1 -8 V, DIB2 +8 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Files unusually small | data quality not good | AK (qsum spectra, Qi and Qo vs. time) | Y | |
| 07120807_0531 07120807_0625 07120807_0718 | Ba-133 & Am-241 | DIB1 -10 V, DIB2 +10 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Files unusually small | data quality not good | AK (qsum spectra, Qi and Qo vs. time) | Y | |
| 07120807_0803 07120807_0857 07120807_0952 | Ba-133 & Am-241 | DIB1 -12 V, DIB2 +12 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Files unusually small | data quality not good | Y | ||
| 07120807_1035 07120807_1125 07120807_1215 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120807_1306 07120807_1356 07120807_1447 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120807_1538 07120807_1628 07120807_1718 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120807_1937 07120807_2027 07120807_2117 | Ba-133 & Am-241 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Dib1=DCRC 33, Dib2= DCRC32 Detector may not have been flashed before this series | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120807_2210 07120807_2300 07120807_2350 | Ba-133 & Am-241 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not very good | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120808_0041 07120808_0131 07120808_0221 | Ba-133 & Am-241 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not good | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120808_0313 07120808_0403 07120808_0453 | Ba-133 & Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120808_0544 07120808_0635 07120808_0725 | Ba-133 & Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120808_0816 07120808_0906 07120808_0956 | Ba-133 & Am-241 | DIB1 -5 V, DIB2 +5 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120808_1048 07120808_1138 07120808_1228 | Ba-133 & Am-241 | DIB1 -7 V, DIB2 +7 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120808_1408 07120808_1459 07120808_1549 | Ba-133 & Am-241 | DIB1 -7 V, DIB2 +7 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | |||
| 07120808_1640 07120808_1731 07120808_1821 | Ba-133 & Am-241 | DIB1 -9 V, DIB2 +9 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not good | AK (qsum spectra, chi2 vs. qsum, Qi and Qo vs. time) | Y | ||
| 07120808_1912 07120808_2004 07120808_2057 | Ba-133 & Am-241 | DIB1 -13 V, DIB2 +13 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not good | AK (qsum spectra, Qi and Qo vs. time) | Y | ||
| 07120808_2143 07120808_2234 07120808_2324 | Ba-133 & Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not good | AK (qsum spectra, Qi and Qo vs. time) | Y | ||
| 07120809_0015 07120809_0105 07120809_0155 | Ba-133 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120809_0247 07120809_0337 07120809_0427 | Ba-133 & Am-241 | DIB1 +5 V, DIB2 -5 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120809_0518 07120809_0608 07120809_0658 | Ba-133 & Am-241 | DIB1 +7 V, DIB2 -7 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not good | Y | |||
| 07120809_0750 07120809_0840 07120809_0930 | Ba-133 & Am-241 | DIB1 +9 V, DIB2 -9 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not good | Y | |||
| 07120809_1021 07120809_1111 07120809_1202 | Ba-133 & Am-241 | DIB1 +13 V, DIB2 -13 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120809_1253 07120809_1343 07120809_1433 | Ba-133 & Am-241 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at (56, 40, 32, 32) ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Data quality not good | Y | |||
| 07120910_1702 07120910_1752 07120910_1843 | Ba-133 & Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120910_1934 07120910_2024 07120910_2115 | Ba-133 & Am-241 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120910_2205 07120910_2256 07120910_2346 | Ba-133 & Am-241 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120911_0037 07120911_0127 07120911_0217 | Ba-133 & Am-241 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120911_0308 07120911_0359 07120911_0449 | Ba-133 & Am-241 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | Y | ||||
| 07120912_1709 07120912_1800 07120912_1850 | Ba-133 & Am-241 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | 9 x 10k 9 x 10k 9 x 10k | Y | |||
| 07120912_1941 07120912_2031 07120912_2122 | Ba-133 & Am-241 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | 10 x 10k 10 x 10k 10 x 10k | Y | |||
| 07120912_2213 07120912_2303 07120912_2353 | Ba-133 & Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | 10 x 10k 10 x 10k 10 x 10k | Y | |||
| 07120913_0044 07120913_0134 07120913_0225 | Ba-133 & Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | 10 x 10k 10 x 10k 10 x 10k | Y | |||
| 07120913_0316 07120913_0406 07120913_0456 | Ba-133 & Am-241 | DIB1 +1 V, DIB2 -1 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | 10 x 10k 10 x 10k 10 x 10k | Y | |||
| 07120913_0547 07120913_0637 07120913_0728 | Ba-133 & Am-241 | DIB1 -1 V, DIB2 +1 V; 0.4096 ms trace length; 100 us delay; Trigger on (Qi/DIB1, Qo/DIB1, Qi/DIB2, Qo/DIB2) at 40 ADC bin. Data/flash/cooldown = 30/0.1/20/min. | 10 x 10k 10 x 10k 10 x 10k | Y | |||
Run 28 (G101b)
| Run | Run Mode | Run Info | # of Events/Files | Comments | |
|---|---|---|---|---|---|
| 07120513_0059 | Am-241 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 27x10k | Bad data? | |
| 07120514_2137 | Am-241 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 22x10k | DAQ crashed 1 hour 45 minutes into run. Bad data? | |
| 07120515_0007 | Am-241 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 26x10k | Bad data? | |
| 07120516_0120 07120516_0230 07120516_0341 07120516_0451 07120516_0602 07120516_0712 07120516_0823 07120516_0934 | Am-241 & Ba-133 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 11x10k 11x10k 11x10k 11x10k 11x10k 11x10k 11x10k 11x10k | ||
| 07120516_1715 07120516_1826 07120516_1936 07120516_2046 07120516_2157 07120516_2307 | Am-241 & Ba-133 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k | ||
| 07120517_0018 07120517_0129 07120517_0239 07120517_0349 07120517_0459 07120517_0610 07120517_0720 07120517_0830 07120517_0940 07120517_1051 07120517_1201 | Am-241 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k | ||
| 07120518_0024 07120518_0135 07120518_0245 07120518_0357 | Am-241 & Ba-133 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 11x10k 11x10k 11x10k 11x10k | ||
| 07120518_0505 07120518_0616 07120518_0726 07120518_0836 | Am-241 & Ba-133 | DIB1 +5 V, DIB2 -5 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120518_0947 07120518_1057 07120518_1207 07120518_1318 | Am-241 & Ba-133 | DIB1 +3 V, DIB2 -3 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 15x10k 16x10k 16x10k 16x10k | ||
| 07120518_1428 07120518_1538 07120518_1648 07120518_1759 | Am-241 & Ba-133 | DIB1 +2 V, DIB2 -2 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120519_0011 07120519_0122 | Am-241 & Ba-133 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 12x10k 7x10k | Power cut means data series 0712059_0112 is only 30 minutes long. | |
| 07120519_2024 07120519_2135 07120519_2245 07120519_2356 | Am-241 & Ba-133 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 12x10k 12x10k 12x10k 12x10k | ||
| 07120520_0106 07120520_0216 07120520_0327 07120520_0437 | Am-241 & Ba-133 | DIB1 -5 V, DIB2 +5 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120520_0548 07120520_0658 07120520_0808 07120520_0918 | Am-241 & Ba-133 | DIB1 -3 V, DIB2 +3 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120520_1029 07120520_1139 07120520_1249 07120520_1400 | Am-241 & Ba-133 | DIB1 -2 V, DIB2 +2 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 15x10k 16x10k 16x10k 16x10k | ||
| 07120520_2146 07120520_2257 07120521_0007 07120521_0118 07120521_0228 07120521_0338 | Am-241 & Ba-133 | Ba-133 source 23“ underneath detector. DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 11x10k 11x10k 11x10k 11x10k 11x10k 11x10k | χ² distributions rather wide for both sides, flashing doesn't seem to restore neutralisation for some reason, questionable quality. May have to retake. | |
| 07120521_0448 07120521_0559 07120521_0709 07120521_0819 07120521_0924 07120521_1040 | Am-241 & Ba-133 | Ba-133 source 23” underneath detector. DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 15x10k 16x10k 16x10k 15x10k 16x10k | χ² distributions rather wide for both sides, flashing doesn't seem to restore neutralisation for some reason, questionable quality. May have to retake. | |
| 07120521_2209 07120521_2320 07120522_0030 07120522_0140 | Am-241 & Ba-133 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 11x10k 11x10k 11x10k 11x10k | ||
| 07120522_0251 07120522_0402 07120522_0512 07120522_0622 | Am-241 & Ba-133 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120522_0733 07120522_0843 07120522_0953 07120522_1104 | Am-241 & Ba-133 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120522_2224 07120522_2335 07120523_0046 07120523_0157 | Am-241 & Ba-133 | DIB1 -12 V, DIB2 +12 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 12x10k 12x10k 12x10k 12x10k | ||
| 07120523_0306 07120523_0416 07120523_0526 07120523_0637 | Am-241 & Ba-133 | DIB1 -10 V, DIB2 +10 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120523_0747 07120523_0857 07120523_1008 07120523_1118 | Am-241 & Ba-133 | DIB1 -8 V, DIB2 +8 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120523_2312 07120524_0023 07120524_0133 07120524_0244 | Am-241 & Ba-133 | DIB1 +2.5 V, DIB2 -2.5 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 11x10k 12x10k 12x10k 12x10k | ||
| 07120524_0354 07120524_0504 07120524_0615 07120524_0725 | Am-241 & Ba-133 | DIB1 +3.5 V, DIB2 -3.5 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 50/0.1/20 min. | 16x10k 16x10k 16x10k 16x10k | ||
| 07120524_1738 07120524_1749 07120524_1800 07120524_1811 07120524_1822 07120524_1833 07120524_1844 07120524_1855 07120524_1906 07120524_1917 07120524_1928 07120524_1940 | Am-241 & Ba-133 | +/-4 V -/+4 V +/-4 V -/+4 V +/-4 V -/+4 V +/-4 V -/+4 V +/-4 V -/+4 V +/-4 V -/+4 V | 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; bias flipping, no flash, 10 minutes per series. | 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k | Fine Corrupt last file Fine Fine Fine Corrupt last file Fine Fine Full 10 minutes? Fine Corrupt last file Fine |
| 07120524_2158 | Am-241 & Ba-133 | DIB1 +12 V, DIB2 -12 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 35x10k | ||
| 07120525_0020 | Am-241 & Ba-133 | DIB1 -12 V, DIB2 +12 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 37x10k | ||
| 07120525_0242 | Am-241 & Ba-133 | DIB1 +10 V, DIB2 -10 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 37x10k | ||
| 07120525_0503 | Am-241 & Ba-133 | DIB1 -10 V, DIB2 +10 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 37x10k | ||
| 07120525_0725 | Am-241 & Ba-133 | DIB1 +8 V, DIB2 -8 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 37x10k | ||
| 07120525_0947 | Am-241 & Ba-133 | DIB1 -8 V, DIB2 +8 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 36x10k | ||
| 07120526_2209 | Am-241 & Ba-133 | DIB1 +6 V, DIB2 -6 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 36x10k | ||
| 07120527_0031 | Am-241 & Ba-133 | DIB1 -6 V, DIB2 +6 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 37x10k | ||
| 07120527_0253 | Am-241 & Ba-133 | DIB1 +4 V, DIB2 -4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 38x10k | ||
| 07120527_0515 | Am-241 & Ba-133 | DIB1 -4 V, DIB2 +4 V; 0.4096 ms trace length; 100 us delay; Trigger 40 ADC bins all Q; data/flashing/cooldown = 120/0.1/20 min. | 37x10k | ||
Run 23(G102)
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07120131_0029 07120131_0149 07120131_0310 07120131_0430 07120131_0550 07120131_0710 07120131_0830 07120131_0951 07120131_1111 07120131_1231 | Am-241 | -7V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at (48 36 48 36) ADC bins on Q1-4; data/flashing/cooldown = 60/0.1/20 min. | 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k 12x10k 11x10k 11x10k 11x10k | comment |
| 07120201_0232 07120201_0452 07120201_0712 | Ba-133 & Am-241 | -1V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 21x10k 23x10k 22x10k | peaks aren't very clear |
| 07120201_0937 07120201_1157 07120201_1418 | Ba-133 & Am-241 | -2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 27x10k 26x10k 21x10k | peaks aren't very clear |
| 07120201_1643 | 5x10k | Bias/Config Unknown | ||
| 07120201_2147 07120201_2237 07120201_2327 | Ba-133 & Am-241 | -3V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 6x10k 6x10k | peaks aren't very clear |
| 07120202_0022 07120202_0112 07120202_0203 | Ba-133 & Am-241 | -4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k | peaks are noticeable but not very distinct |
| 07120202_0258 07120202_0348 07120202_0438 | Ba-133 & Am-241 | -5V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 8x10k | peaks are noticeable but not very distinct |
| 07120202_0533 07120202_0623 07120202_0713 | Ba-133 & Am-241 | -6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k | peaks are noticeable but not very distinct |
| 07120202_2235 07120202_2325 07120203_0015 07120203_0106 | Ba-133 & Am-241 | -4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 5x10k 5x10k 6x10k | comment |
| 07120203_0200 | Ba-133 & Am-241 | -8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 32 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 1x10k | DAQ crashed |
| 07120203_1631 07120203_1721 07120203_1811 | Ba-133 & Am-241 | -4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k | comment |
| 07120203_2033 07120203_2253 07120204_0113 07120204_0333 07120204_0553 07120204_0813 07120204_1034 07120204_1254 | Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 23x10k 23x10k 23x10k 23x10k 23x10k 23x10k 23x10k 22x10k | comment |
| 07120205_0010 07120205_0100 07120205_0150 07120205_0240 | Ba-133 & Am-241 | -2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 6x10k 6x10k 6x10k | comment |
| 07120205_0335 07120205_0426 07120205_0516 07120205_0606 | Ba-133 & Am-241 | -3V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k 7x10k | comment |
| 07120205_0701 07120205_0751 07120205_0841 07120205_0931 | Ba-133 & Am-241 | -5V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k 7x10k | comment |
| 07120205_1026 07120205_1116 07120205_1207 07120205_1257 | Ba-133 & Am-241 | -6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k 7x10k | comment |
| 07120205_2145 07120205_2235 07120205_2325 07120206_0015 | Ba-133 & Am-241 | -7V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 5x10k 5x10k 5x10k | comment |
| 07120206_0110 07120206_0200 07120206_0250 07120206_0341 | Ba-133 & Am-241 | -8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k 7x10k | comment |
| 07120206_0436 07120206_0526 07120206_0616 07120206_0706 | Ba-133 & Am-241 | -9V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k 7x10k | comment |
| 07120206_0801 07120206_0851 07120206_0941 07120206_1031 | Ba-133 & Am-241 | -1V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k 7x10k | comment |
| 07120206_2227 07120206_2317 07120207_0007 07120207_0057 | Ba-133 & Am-241 | -2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 5x10k 5x10k 5x10k | Not triggering on Q4 |
| 07120207_0152 07120207_0242 07120207_0332 07120207_0422 | Ba-133 & Am-241 | -4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | Not triggering on Q4 |
| 07120207_0517 07120207_0608 07120207_0658 07120207_0748 | Ba-133 & Am-241 | -6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | Not triggering on Q4 |
| 07120208_2205 07120208_2255 07120208_2346 07120209_0036 | Ba-133 & Am-241 | +2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 5x10k 5x10k 5x10k | I think we went back to triggering on Q4 |
| 07120209_0131 07120209_0221 07120209_0311 07120209_0401 | Ba-133 & Am-241 | +4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | comment |
| 07120209_0456 07120209_0546 07120209_0636 07120209_0727 | Ba-133 & Am-241 | +6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | comment |
| 07120209_0822 07120209_0912 07120209_1002 07120209_1052 | Ba-133 & Am-241 | +8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 5x10k 5x10k | comment |
| 07120209_1754 07120209_1845 07120209_1935 07120209_2025 | Ba-133 & Am-241 | +2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 6x10k 6x10k 6x10k | comment |
| 07120209_2120 07120209_2210 07120209_2300 07120209_2350 | Ba-133 & Am-241 | +4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | comment |
| 07120210_0045 07120210_0135 07120210_0226 07120210_0316 | Ba-133 & Am-241 | +6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | comment |
| 07120210_0411 07120210_0501 07120210_0551 07120210_0641 | Ba-133 & Am-241 | +8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | comment |
| 07120212_0042 07120212_0132 07120212_0222 07120212_0312 | Ba-133 & Am-241 | +1V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | Low energy events |
| 07120212_0408 07120212_0458 07120212_0548 07120212_0638 | Ba-133 & Am-241 | +3V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | Low energy events |
| 07120213_1223 07120213_1313 07120213_1403 | Ba-133 & Am-241 | +10V is applied to the back side of crystal. DCRC33 only, 0V, trigger on Qi OR Qo at 40 ADC bin; data/flashing/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k | one HV board in place |
| 07120214_0944 07120214_1034 07120214_1124 | Ba-133 & Am-241 | +1V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k | mixed polarities- faultiness of DCRC |
| 07120214_1426 07120214_1516 07120214_1606 07120214_1656 07120214_1751 | Ba-133 & Am-241 | +1V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 6x10k 6x10k 6x10k 6x10k | Flashed for an hour beforehand, but one DCRC wasn't flashing properly. Might be getting both polarities for pulses. I think this is all +1V - implied by run diary entry |
| 07120214_1846 07120214_1936 07120214_2026 | Ba-133 & Am-241 | +1V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k | comment |
| 07120214_2121 07120214_2211 07120214_2302 | Ba-133 & Am-241 | +3V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k | comment |
| 07120214_2357 07120215_0047 07120215_0137 | Ba-133 & Am-241 | +5V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k | comment |
| 07120215_0232 07120215_0322 07120215_0412 | Ba-133 & Am-241 | +7V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k | comment |
| 07120215_0507 07120215_0557 07120215_0648 | Ba-133 & Am-241 | +9V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k | comment |
| 07120215_0743 07120215_0833 07120215_0923 | Ba-133 & Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 6x10k 7x10k 6x10k | comment |
| 07120215_2344 | Ba-133 & Am-241 | +10V external bias. 0V DCRC bias; Trigger on Q1-4 at 56ADC bin; 60 min data taking. | 11x10k | both HV boards in place |
| 07120216_1218 | Ba-133 & Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 180/0.1/20 min. | 29x10k | |
| 07120216_1835 | Ba-133 & Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 24x10k | 200 mK |
| 07120216_2110 | Ba-133 & Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 25x10k | 200 mK |
| 07120217_0056 07120217_0146 | Ba-133 & Am-241 | -10V Q1-3; Q4 at 0V; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 30/0.1/20 min. | 4x10k 4x10k | comment |
| 07120217_0241 07120217_0331 | Ba-133 & Am-241 | -6V Q1-3; Q4 at 0V; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 6x10k | comment |
| 07120217_0951 07120217_1041 | Ba-133 & Am-241 | +10V Q1-3; Q4 at 0V; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 6x10k | comment |
| 07120217_1151 07120217_1241 | Ba-133 & Am-241 | +6V Q1-3; Q4 at 0V; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 30/0.1/20 min. | 5x10k 5x10k | comment |
| 07120217_1336 | Ba-133 & Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-3; data/flashing/cooldown = 180/0.1/20 min. | 32x10k | comment |
| 07120217_2247 | Am-241 | -7V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 27x10k | comment |
| 07120218_0048 | Am-241 | 0V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 31x10k | comment |
| 07120218_0313 | Am-241 | +7V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 29x10k | comment |
| 07120218_0514 | Am-241 | 0V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 31x10k | comment |
| 07120218_0739 | Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 29x10k | comment |
| 07120218_1112 | Am-241 | 0V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 31x10k | comment |
| 07120218_1337 | Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 29x10k | comment |
| 07120218_1538 | Am-241 | 0V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/20 min. | 31x10k | comment |
| 07120219_0751 07120219_0809 07120219_0826 07120219_0844 | Ba-133 & Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120219_0901 07120219_0918 07120219_0935 07120219_0953 | Ba-133 & Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120219_1010 | Ba-133 & Am-241 | -8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 3x10K | 400 mK - bad set- couldn't fully process |
| 07120219_1030 07120219_1047 07120219_1104 | Ba-133 & Am-241 | -8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k | 400 mK |
| 07120219_1120 07120219_1137 07120219_1154 07120219_1211 | Ba-133 & Am-241 | +8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120219_1229 07120219_1246 07120219_1304 07120219_1321 | Ba-133 & Am-241 | -6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120219_1338 07120219_1356 07120219_1413 07120219_1431 | Ba-133 & Am-241 | +6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120219_1448 07120219_1505 07120219_1522 07120219_1539 | Ba-133 & Am-241 | -4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120219_1557 07120219_1615 07120219_1632 07120219_1649 | Ba-133 & Am-241 | +4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120219_1707 07120219_1724 07120219_1741 07120219_1758 | Ba-133 & Am-241 | -2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 4x10k 5x10k | 400 mK |
| 07120219_2030 | Ba-133 & Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/2 min. | 37x10k | 2 hours at 400 mK |
| 07120219_2233 | Ba-133 & Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/0.1/2 min | 37x10k | 2 hours at 400 mK |
| 07120220_0224 07120220_0244 07120220_0304 07120220_0324 | Ba-133 & Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_0345 07120220_0405 07120220_0425 07120220_0445 | Ba-133 & Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_0506 07120220_0526 07120220_0546 07120220_0606 | Ba-133 & Am-241 | -8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_0627 07120220_0647 07120220_0707 07120220_0728 | Ba-133 & Am-241 | +8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_0748 07120220_0808 07120220_0828 07120220_0848 | Ba-133 & Am-241 | -6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_0909 07120220_0929 07120220_0949 07120220_1010 | Ba-133 & Am-241 | +6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_1030 07120220_1051 07120220_1111 07120220_1131 | Ba-133 & Am-241 | -4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_1151 07120220_1213 07120220_1233 07120220_1253 | Ba-133 & Am-241 | +4V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 4x10k 5x10k 5x10k | 200 mK |
| 07120220_1312 07120220_1332 07120220_1352 07120220_1413 | Ba-133 & Am-241 | -2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 5x10k 5x10k 5x10k 5x10k | 200 mK |
| 07120220_1527 07120220_1547 07120220_1607 07120220_1627 | Ba-133 & Am-241 | +2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/1/4 min. | 4x10k 4x10k 5x10k 5x10k | 200 mK |
| 07120220_1715 07120220_1735 07120220_1755 07120220_1815 | Ba-133 & Am-241 | +2V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 15/0.1/2 min. | 5x10k 5x10k 5x10k 5x10k | 400 mK |
| 07120221_0040 07120221_0130 07120221_0220 07120221_0310 | Am-241 | +7V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 11x10k 11x10k 11x10k 11x10k | 1 hour flashing before this set; base temp |
| 07120221_0401 07120221_0452 07120221_0542 07120221_0632 | Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 11x10k 11x10k 11x10k 11x10k | base temp |
| 07120221_0723 07120221_0813 | Am-241 | +5V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 11x10k 11x10k | base temp |
| 07120221_0904 07120221_0954 | Am-241 | -5V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 11x10k 10x10k | base temp |
| 07120221_1212 | Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/0.1/20 min. | 19x10k | 600 mK |
| 07120221_2206 07120221_2244 07120221_2323 07120222_0001 | Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/3/5 min. | 7x10k 7x10k 7x10k 7x10k | Temperature 2K |
| 07120222_0040 07120222_0118 07120222_0156 07120222_0234 | Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/3/5 min. | 10x10k 10x10k 10x10k 10x10k | Temperature 2K |
| 07120222_0313 07120222_0351 07120222_0429 07120222_0507 | Am-241 | -8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 30/3/5 min. | 10x10k 10x10k 10x10k 10x10k | Temperature 2K |
| 07120222_2303 | Ba-133 & Am-241 | -10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/3/5 min. | 25x10k | T=2K. Flashed detector 35min before data. |
| 07120223_0112 | Ba-133 & Am-241 | +10V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/3/5 min. | 37x10k | T=2K |
| 07120223_0321 07120223_0400 07120223_0438 07120223_0516 | Ba-133 & Am-241 | -8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/3/5 min. | 9x10k 9x10k 9x10k 9x10k | T=2K |
| 07120223_0554 07120223_0633 07120223_0712 07120223_0750 | Ba-133 & Am-241 | +8V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/3/5 min. | 9x10k 9x10k 9x10k 9x10k | T=2K |
| 07120223_0828 07120223_0906 07120223_0944 07120223_1023 | Ba-133 & Am-241 | -6V; Q1-4; 0.4096 ms trace length; 100 us delay; Trigger at 40 ADC bins on Q1-4; data/flashing/cooldown = 120/3/5 min. | 9x10k 9x10k 9x10k 9x10k | T=2K |
| 07120224_0052 | Ba-133 & Am-241 | +35V (DCRCs at +4 V, “ground” electrode at -31 V) | 28x10k | base temp |
| 07120224_0353 | Ba-133 & Am-241 | +35V (DCRCs at +4 V, “ground” electrode at -31 V) | 28x10k | 400 mK |
Run 22(G102)
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07111203_1800 07111203_1920 07111203_2040 | Am-241 | Q1-4 -4 V; 0.4096 ms trace length; 100 us delay; Trigger at (48 36 48 36) ADC bins on Q1-4; data/flashing/cooldown = 60/0.1/20 min. | 10x10k 10x10k 10x10k | comment |
| 07111203_2205 07111203_2325 07111204_0046 | Am-241 | Q1-4 -6 V; 0.4096 ms trace length; 100 us delay; Trigger at (48 36 48 36) ADC bins on Q1-4; data/flashing/cooldown = 60/0.1/20 min. | 12x10k 12x10k 12x10k | comment |
| 07111204_0211 07111204_0331 07111204_0451 | Am-241 | Q1-4 -8 V; 0.4096 ms trace length; 100 us delay; Trigger at (48 36 48 36) ADC bins on Q1-4; data/flashing/cooldown = 60/0.1/20 min. | 12x10k 11x10k 11x10k | comment |
| 07111204_0616 07111204_0736 07111204_0856 | Am-241 | Q1-4 -10 V; 0.4096 ms trace length; 100 us delay; Trigger at (48 36 48 36) ADC bins on Q1-4; data/flashing/cooldown = 60/0.1/20 min. | 10x10k 9x10k 8x10k | comment |
| 07111204_2111 07111204_2232 07111204_2352 | Ba-133 & Am-241 | Q1-4 -4V; Trigger on Q1-4 at 40ADC bin. Trace length 0.4096ms, look back time 100us. Data/flashing/cooldown = 60/0.1/20 min | 10x10k 10x10k 10x10k | comment |
| 07111205_0117 07111205_0237 07111205_0357 | Ba-133 & Am-241 | Q1-4 -8V; Trigger on Q1-4 at 40ADC bin. Trace length 0.4096ms, look back time 100us. Data/flashing/cooldown = 60/0.1/20 min | 11x10k 11x10k 11x10k | comment |
| 07111205_0522 07111205_0642 07111205_0802 | Ba-133 & Am-241 | Q1-4 -13V; Trigger on Q1-4 at 40ADC bin. Trace length 0.4096ms, look back time 100us. Data/flashing/cooldown = 60/0.1/20 min | 11x10k 11x10k 11x10k | comment |
| 07111205_1050 07111205_1210 07111205_1330 | Am-241 | -13V on Q1-4. Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 10x10k 11x10k 11x10k | comment |
| 07111205_1456 07111205_1616 07111205_1736 | Am-241 | +13V on Q1-4. Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 13x10k 13x10k 13x10k | comment |
| 07111206_1030 07111206_1150 07111206_1310 | Ba-133 & Am-241 | -2V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 9x10k 9x10k 9x10k | comment |
| 07111205_1951 07111205_2111 07111205_2232 | Ba-133 & Am-241 | +2V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 10x10k 10x10k 10x10k | comment |
| 07111206_2115 07111206_2235 | Ba-133 & Am-241 | +4V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 11x10k 11x10k | comment |
| 07111207_1149 07111207_1309 | Ba-133 & Am-241 | +6V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 13x10k 12x10k | comment |
| 07111208_1023 07111208_1144 | Am-241 | -13V; Triggering at 40ADC bin. Data/flash/cooldown = 60/0.1/20 min | 15x10k 8x10k | 10 hours flashing |
| 07111209_2009 | Ba-133 & Am-241 | -10V; Triggering at 40ADC bin. Data/flash/cooldown = 60/0.1/20 min | 164x10k | 12 hours of data taking |
| 07111210_1937 | Ba-133 & Am-241 | -10V; Triggering at 40ADC bin. Data/flash/cooldown = 60/0.1/20 min | TBD | Continuous data acquisition to see when data goes bad |
| 07111214_2248 07111215_0010 07111215_0132 | Ba-133 & Am-241 | -4V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 14x10k 14x10k 14x10k | comment |
| 07111215_0255 07111215_0416 07111215_0536 | Ba-133 & Am-241 | -8V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 18x10k 18x10k 18x10k | comment |
| 07111215_0701 07111215_0821 07111215_0941 | Ba-133 & Am-241 | -13V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 18x10k 18x10k 18x10k | comment |
| 07111215_2311 07111216_0031 | source | config | 14x10k 14x10k | comment |
| 07111216_0156 07111216_0316 07111216_0437 | Ba-133 & Am-241 | V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 18x10k 18x10k 18x10k | comment |
| 07111216_0602 07111216_0722 07111216_0843 | Ba-133 & Am-241 | -10V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 18x10k 18x10k 18x10k | comment |
| 07111216_2125 07111216_2248 07111217_0011 | Ba-133 & Am-241 | -4V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 11x10k 11x10k 11x10k | iZIP |
| 07111217_0132 07111217_0252 07111217_0412 | Ba-133 & Am-241 | -6V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 13x10k 13x10k 13x10k | iZIP |
| 07111217_0537 07111217_0659 07111217_0819 | Ba-133 & Am-241 | -8V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 13x10k 13x10k 13x10k | iZIP |
| 07111217_0943 07111217_1104 07111217_1225 | Ba-133 & Am-241 | -10V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 13x10k 13x10k 13x10k | iZIP |
| 07111217_1349 07111217_1510 07111217_1649 | Ba-133 & Am-241 | -13 V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 13x10k 8x10k 11x10k | iZIP |
| 07111223_1628 07111223_1748 07111223_1908 | Am-241 | 8V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 16x10k 17x10k 17x10k | Q3,Q4 not very good |
| 07111223_2033 07111223_2153 07111223_2314 | Am-241 | 10V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 18x10k 18x10k 18x10k | Q3,Q4 not very good |
| 07111224_0039 07111224_0159 07111224_0319 | Am-241 | 13V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 18x10k 18x10k 18x10k | Q3,Q4 not very good |
| 07111224_0444 07111224_0604 07111224_0725 | Am-241 | -3V on Q1-4; Trigger at 40 ADC bins on Q1-4. Trace length 0.4096ms, lookback time 100us. Data/flash/cooldown = 60/0.1/20 min | 17x10k 17x10k 17x10k | Q1 is best |
Run 20 (G102)
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07111026_1111 07111026_1231 07111026_1352 | Am-241 | Q1-4 -4 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; data/flashing/cooldown = 60/0.1/20 min. | 16 x 10k 16 x 10k 16 x 10k | |
| 07111026_1517 07111026_1637 07111026_1758 | Am-241 | Q1-4 +4 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; data/flashing/cooldown = 60/0.1/20 min. | 19 x 10k 16 x 10k 15 x 10k | |
| 07111026_2054 07111026_2214 07111026_2334 | Am-241 | Q1-4 -2 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; data/flashing/cooldown = 60/0.1/20 min. | 15 x 10k 15 x 10k 15 x 10k | |
| 07111027_0100 07111027_1157 07111027_1317 07111027_1437 | Am-241 | Q1-4 +2 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; data/flashing/cooldown = 60/0.1/20 min. | 19 x 10k 19 x 10k 16 x 10k 16 x 10k | |
| 07111027_1835 07111027_1955 07111027_2115 | Am-241 | Q1-4 -6 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; data/flashing/cooldown = 60/0.1/20 min. | 15 x 10k 15 x 10k 15 x 10k | |
| 07111027_2241 07111028_0001 07111028_0121 | Am-241 | Q1-4 +6 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; data/flashing/cooldown = 60/0.1/20 min. | 19 x 10k 19 x 10k 19 x 10k | |
| 07111028_1923 07111028_2044 07111028_2204 | Am-241 | Q1-4 -13 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; data/flashing/cooldown = 60/0.1/20 min. | 14 x 10k 14 x 10k 8 x 10k | Final data series incomplete due to DAQ error. |
| 07111030_1309 | Am-241 & Ba-133 | Q1-4 -13 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins. | 14 x 10k | Series incomplete due to DAQ error. |
| 07111030_1456 | Am-241 & Ba-133 | Q1-4 +13 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; flash for 6 s. | 16 x 10k | |
| 07111030_1814 | Am-241 & Ba-133 | Q1-4 -13 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; flash for 6 s. | 14 x 10k | Tower temperature 400 ± 5 mK. |
| 07111030_1924 | Am-241 & Ba-133 | Q1-4 +13 V; 0.4096 ms trace length; 100 us delay; trigger on Q 40 ADC bins; flash for 6 s. | 13 x 10k | Tower temperature 400 ± 5 mK. |
Run 17, G53
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07110913_1652 07110913_1912 | Ba-133 | -2.5 V DIB1; +2.5 V DIB2; 0.4096 ms trace length; 100 us trigger delay; trigger on Q 40 ADC bins; charge-only; data/flashing/cooldown = 120/0.1/20 min. | 21 x 10k 21 x 10k | |
| 07110914_1939 07110914_2159 07110915_0019 | Ba-133 | +2.5 V DIB1; -2.5 V DIB2; 0.4096 ms trace length; 100 us trigger delay; trigger on Q 40 ADC bins; charge-only; data/flashing/cooldown = 120/0.1/20 min. | 25 x 10k 24 x 10k 3 x 10k | Last series incomplete possibly due to trigger problem or DAQ error. |
| 07110915_0245 07110915_0505 07110915_0725 | Ba-133 | -2.5 V DIB1; +2.5 V DIB2; 0.4096 ms trace length; 100 us trigger delay; trigger on Q 40 ADC bins; charge-only; data/flashing/cooldown = 120/0.1/20 min. | 22 x 10k 22 x 10k 21 x 10k | DIB1 signal lost due to DAQ error in the last run. Rq data seems bad–no trees written. |
| 07110915_1916 07110915_2148 07110916_0020 | Ba-133 | -2.5 V DIB1; +2.5 V DIB2; 3.2768 ms trace length; 400 us trigger delay; trigger on P 20 ADC bins; data/flashing/cooldown = 120/0.1/20 min. | 8 x 10k 8 x 10k 7 x 10k | |
| 07110916_0221 07110916_0453 07110916_0722 | Ba-133 | +2.5 V DIB1; -2.5 V DIB2; 3.2768 ms trace length; 400 us trigger delay; trigger on P 20 ADC bins; data/flashing/cooldown = 120/0.1/20 min. | 10 x 10k 10 x 10k 9 x 10k | |
| 07110916_2109 | Ba-133 | -2.5 V DIB1; +2.5 V DIB2; 3.2768 ms trace length; 400 us trigger delay; trigger on P 20 ADC bins; 6-hour run. | 23 x 10k | Long run: 6 hours. |
| 07110917_0354 | Ba-133 | +2.5 V DIB1; -2.5 V DIB2; 3.2768 ms trace length; 400 us trigger delay; trigger on P 20 ADC bins; 6-hour run. | 28 x 10k | Long run: 6 hours. |
| 07110917_1039 | Ba-133 | -2.5 V DIB1; +2.5 V DIB2; 0.4096 ms trace length; 100 us trigger delay; trigger on P 20 ADC bins; 6-hour run; charge-only. | 58 x 10k | Long run: 6 hours; charge-only. |
Run 16, G102
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07110829_1849 07110829_2109 07110829_2329 07110830_0151 | Am-241 | -4V on Q1-4. Trigger on Q1-4 at 48 ADC bin. Data/flashing/cooldown = 120/0.1/20 min. | 22x10k 22x10k 24x10k 27x10k | Test run |
| 07110830_0415 | Am-241 | +4V on Q1-4. Trigger on Q1-4 at 48 ADC bin. Data/flashing/cooldown = 120/0.1/20 min. | 17x10k | Test run |
| 07110831_0028 07110831_0118 07110831_0208 07110831_0258 | Am-241 | -4V on Q1-4. Trigger on Q1-3 at 48 ADC bin. Data/flash/cooldown = 30/0.1/20min. | 8x10k 10x10k 10x10k 10x10k | Test run |
| 07110831_2039 07110831_2129 07110831_2220 07110831_2310 | Am-241 | -4V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 9x10k 9x10k 9x10k 9x10k | 20min flashing before data. |
| 07110901_0005 07110901_0055 07110901_0145 07110901_0235 | Am-241 | -6V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 9x10k 9x10k 9x10k 9x10k | comment |
| 07110901_0330 07110901_0421 07110901_0511 07110901_0601 | Am-241 | -8V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 9x10k 9x10k 9x10k 9x10k | comment |
| 07110901_0929 07110901_1019 07110901_1110 07110901_1201 | source | -0.5V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | 10min flashing before data. |
| 07110901_1255 07110901_1345 07110901_1435 07110901_1525 | Am-241 | -1V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 7x10k 7x10k 7x10k 7x10k | comment |
| 07110901_1620 07110901_1711 07110901_1801 07110901_1851 | Am-241 | -2V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
| 07110901_1946 07110901_2037 07110901_2127 07110901_2217 | Am-241 | -3V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
| 07110901_2312 07110902_0002 07110902_0052 07110902_0143 | Am-241 | -5V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 6x10k | Q2&4 lost charge amp power at the last series. Not sure when this happened. |
| 07110902_1929 07110902_2019 07110902_2109 07110902_2159 | Am-241 | +1V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 9x10k 8x10k 8x10k 8x10k | 10min flashing before data. |
| 07110902_2254 07110902_2345 07110903_0035 07110903_0125 | Am-241 | +2V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
| 07110903_0220 07110903_0310 07110903_0401 07110903_0451 | Am-241 | +4V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
| 07110903_0546 07110903_0636 07110903_0726 07110903_0816 | Am-241 | +6V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
| 07110903_1644 | Am-241 | -6V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. 6 hour long run, 1 hour flashing before data. | 76x10k | |
| 07110904_0120 | Am-241 | +6V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. 6 hour long run, 1 hour flashing before data. | 96x10k | |
| 07110904_1908 07110904_1959 07110904_2050 07110904_2141 | Am241 & Ba133 | -1V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 6x10k | 1 hour flashing before data |
| 07110904_2236 07110904_2327 07110905_0017 07110905_0108 | Am241 & Ba133 | -2V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | |
| 07110905_0204 07110905_0255 07110905_0345 07110905_0436 | Am241 & Ba133 | -4V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | |
| 07110905_1547 07110905_1640 07110905_1731 07110905_1822 | Am241 & Ba133 | -6V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 6x10k 6x10k 6x10k 7x10k | comment |
| 07110905_1916 07110905_2007 07110905_2058 07110905_2148 | Am241 & Ba133 | +2V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
| 07110905_2244 07110905_2335 07110906_0025 07110906_0116 | Am241 & Ba133 | +4V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
| 07110906_0212 07110906_0302 07110906_0353 07110906_0444 | Am241 & Ba133 | +6V on Q1-3 and 0V on Q4. Trigger on Q1-3 at 32 ADC bin. Data/flash/cooldown = 30/0.1/20 min. | 8x10k 8x10k 8x10k 8x10k | comment |
Run 15, G16K Data
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07110718_2110 07110718_2330 07110719_0150 07110719_0410 | Ba-133 | DIB1 -2.5V ZIP1, DIB2 2.5V ZIP2. Trigger on all 4 charge channels at 64 ADC bin. Ba-133 source. Data/flash/cooldown=120min/6s/20min. | 28x10k 28x10k 28x10k 28x10k | Last series 8 minutes short due to DAQ error. |
| 07110719_2024 07110719_2255 | Ba-133 | -2.5V on DIB1, 2.5V on DIB2. Trigger on 8 phonon channels at (12, 15, 27, 20, 14, 8, 15, 17) ADC bin for A1-D2. Trace length 3.2768ms, lookback time 400us. Data/flash/cooldown = 120min/0.6s/20min. | 10x10k 9x10k | |
| 07110720_2220 07110721_0043 07110721_0304 07110721_0527 07110721_0750 07110721_1010 07110721_1235 | Ba-133 | -2.5V on DIB1, 2.5V on DIB2. Trigger on 8 phonon channels at (12, 15, 27, 20, 14, 8, 15, 17) ADC bin for A1-D2. Trace length 3.2768ms, lookback time 400us. Data/flash/cooldown = 120min/0.6s/20min. | 9x10k 9x10k 9x10k 9x10k 9x10k 10x10k 10x10k | |
| 07110721_2147 07110722_0016 07110722_0237 07110722_0457 07110722_0722 | Ba-133 | +2.5V on DIB1, -2.5V on DIB2. Trigger on 8 phonon channels at (12, 15, 27, 20, 14, 8, 15, 17) ADC bin for A1-D2. Trace length 3.2768ms, lookback time 400us. Data/flash/cooldown = 120min/0.6s/20min. | 9x10k 7x10k 5x10k 6x10k 9x10k | Possible trigger problem during data sets 07110722_0016, 07110722_0237 and 07110722_0457 |
| 07110724_1400 07110724_1717 07110724_1937 07110724_2158 07110725_0019 | Ba-133 | -2.5V on DIB1, 2.5V on DIB2. Trigger on 8 phonon channels at (12, 15, 27, 20, 14, 8, 15, 17) ADC bin for A1-D2. Trace length 3.2768ms, lookback time 400us. Data/flash/cooldown = 120min/0.6s/20min. | 11x10k 11x10k 11x10k 12x10k 6x10k | |
| 07110725_2026 07110725_2042 07110725_2043 07110725_2058 07110725_2059 07110725_2114 07110725_2116 07110725_2131 07110725_2132 07110725_2147 07110725_2148 07110725_2203 07110725_2204 07110725_2219 07110725_2221 07110725_2236 07110725_2237 07110725_2252 07110725_2253 07110725_2308 07110725_2309 07110725_2324 07110725_2326 07110725_2341 07110725_2342 07110725_2357 07110725_2358 07110726_0013 07110726_0014 07110726_0029 07110726_0031 07110726_0046 07110726_0047 07110726_0102 07110726_0103 07110726_0118 07110726_0119 07110726_0135 07110726_0136 07110726_0151 07110726_0152 07110726_0207 07110726_0208 07110726_0223 07110726_0225 07110726_0240 07110726_0241 07110726_0256 07110726_0257 07110726_0312 07110726_0314 07110726_0329 07110726_0330 07110726_0345 07110726_0346 07110726_0401 07110726_0403 07110726_0418 07110726_0419 07110726_0434 07110726_0435 07110726_0450 07110726_0452 07110726_0507 07110726_0508 07110726_0523 07110726_0524 07110726_0540 07110726_0541 07110726_0556 07110726_0557 07110726_0612 07110726_0613 07110726_0629 07110726_0630 07110726_0645 | Ba-133 | S1/S2=(-2.5/2.5, 2.5/-2.5)V data. Data duration 15 min, ground 1min in between. Trace length 3.2768ms, lookback time 100us. Trigger on phonon channels at (15, 20, 15, 25, 25, 10, 18, 15) ADC bin on A1-D2. | 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k 2x10k 1x10k |
Run 14, G19N
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07110707_0028 07110707_0258 07110707_0528 07110707_0758 | Am241 & Ba133 | DIB1 is zip1. Bias voltages: -2.5 V DIB1; +2.5 V DIB2. Tracelength 0.4096 ms, lookback time 100 us. | 24x10k 25x10k 25x10k 25x10k | 2-hour stability run |
| 07110708_0122 07110708_0352 07110708_0622 07110708_0852 | Am-241 & Ba-133 | DIB1 is zip1. Bias voltages: +2.5 V DIB1; -2.5 V DIB2. Trace length 0.4096 ms, trigger delay 100 μs. | 25 x 10k 25 x 10k 25 x 10k 28 x 10k | 2-hour stability runs. |
| 07110709_0206 07110709_0438 07110709_0710 07110709_0934 07110709_1158 07110709_1429 | Am241 & Ba133 | DIB1 is zip1. Bias voltages: -2.5 V DIB1; +2.5 V DIB2. Trace length 3.2768 ms, lookback time 400 μs. Trigger on 8 phonon channels, threshold: 0x(1C, 12, E, C) ADC bins on DIB1 and 0x(A, 18, C, 12) ADC bins on DIB2. | 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k | |
| 07110710_0021 07110710_0241 07110710_0501 07110710_0721 07110710_0941 07110710_1201 07110710_1421 07110710_1641 | Am241 & Ba133 | DIB1 is zip1. Bias voltages: -2.5 V DIB1; +2.5 V DIB2. Trace length 0.4096 ms, lookback time 100 μs. Trigger on 8 phonon channels, threshold: 0x(1C, 12, E, C) ADC bins on DIB1 and 0x(A, 18, C, 12) ADC bins on DIB2. | 3x10k 4x10k 4x10k 4x10k 13x10k 16x10k 16x10k 13x10k | Phonon-triggered charge data |
| 07110710_1916 07110710_2136 07110710_2356 07110711_0216 07110711_0436 07110711_0656 | Am241 & Ba133 | DIB1 -2.5V. DIB2 0V. Trace length 0.4096ms, lookback time 100us. Trigger on phonon channels. Trigger threshold: (24, 20, 14, 14) on DIB1. 0x(A, 18, C, 12) ADC bins on DIB2. | 15x10k 17x10k 18x10k 17x10k 16x10k 15x10k | One-side charge data. |
| 07110712_0041 07110712_0308 07110712_0537 07110712_0800 07110712_1029 | Am241 & Ba133 | DIB1 +2.5 V, DIB2 -2.5 V; trace length 3.2768 ms; lookback time 400 μs; trigger on 8 phonon channels, threshold: (24, 20, 14, 14) on DIB1; 0x(A, 18, C, 12) ADC bins on DIB2. | 7x10k 9x10k 9x10k 9x10k 7x10k | Opposite polarity |
Run 13
100 mm Detector G102 Ionisation Test
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07110610_0149 | Am-241 | -4 V on Q1-4. Trigger: Q1 OR Q2 OR Q3 OR Q4. Threshold: 60 ADC bins. DIB2→Detector 1, DIB1→Detector 2. | 31x10k | |
| 07110611_1825 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at 60ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 2h/6s/30min. Tower temperature 68.3mK. Trigger baseline/sample = 64/2. | 7x10k | Increase trigger baseline |
| 07110611_2019 07110611_2109 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at 60ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 67.1mK. Trigger baseline/sample = 64/16. Timing link wasn't set for the first 1300 events. | 7x10k 8x10k | Increase trigger sample number |
| 07110611_2210 07110611_2300 07110611_2350 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at 0x(24, 40, 40, 40)ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 68.6mK. Trigger baseline/sample = 64/16. | 8x10k 8x10k 7x10k | See how the data look like if keeping the trigger rate about the same on all 4 channels. |
| 07110612_0041 07110612_0131 07110612_0221 07110612_0311 07110612_0401 07110612_0451 07110612_0542 07110612_0632 07110612_0722 07110612_0812 07110612_0902 07110612_0953 07110612_1043 07110612_1133 07110612_1223 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at 0x(40, 40, 40, 40)ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 68.6mK. Trigger baseline/sample = 64/16. | 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k 9x10k | Regular data |
| 07110612_1359 | Am-241 | True -4V on Q1-4, Trigger on Q2 at (1-8)*16 ADC for 2min each. Trace length 0.4096ms (1024bins), lookback time 100us. 17min data after 6s flashing | 4x10k | Trigger threshold study |
| 07110612_1436 | Am-241 | True -4V on Q1-4, Trigger on Q4 at (1-8)*16 ADC for 2min each. Trace length 0.4096ms (1024bins), lookback time 100us. 17min data after 6s flashing | 4x10k | Trigger threshold study |
| 07110612_1513 07110612_1550 07110612_1628 | Am-241 | True -4V on Q1-4, Trigger on Q1 at 16 ADC. Trace length 0.4096ms (1024bins), lookback time 100us. 17min data after 6s flashing | 4x10k 4x10k 4x10k | Forgot to step trigger level. |
| 07110612_1713 | Am-241 | True -4V on Q1-4, Trigger on Q1 at (1-8)*16 ADC for 2min each. Trace length 0.4096ms (1024bins), lookback time 100us. 17min data after 6s flashing | 4x10k | Trigger threshold study |
| 07110612_1751 | Am-241 | True -4V on Q1-4, Trigger on Q3 at (1-8)*16 ADC for 2min each. Trace length 0.4096ms (1024bins), lookback time 100us. 17min data after 6s flashing | 5x10k | Trigger threshold study |
| 07110612_1919 07110612_2009 07110612_2059 07110612_2150 07110612_2240 07110612_2331 07110613_0021 07110613_0111 07110613_0201 07110613_0252 07110613_0342 07110613_0432 07110613_0522 07110613_0613 07110613_0703 07110613_0754 07110613_0844 07110613_0934 07110613_1025 07110613_1115 07110613_1206 07110613_1256 07110613_1346 07110613_1436 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. Tower temperature 66.3mK. | 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 7x10k 8x10k 8x10k 8x10k 8x10k 8x10k 7x10k 8x10k | Lower the trigger threshold so we don't lose primary events on any channels. |
| 07110613_2348 07110614_0038 07110614_0128 07110614_0218 07110614_0309 07110614_0359 07110614_0449 07110614_0540 07110614_0630 07110614_0721 | Am-241 | True -8V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 7x10k 7x10k 7x10k 7x10k 7x10k 7x10k 7x10k 7x10k 7x10k 7x10k | True -8V |
| 07110614_0813 07110614_0904 07110614_0954 07110614_1044 07110614_1134 07110614_1225 07110614_1315 07110614_1405 07110614_1456 07110614_1546 | Am-241 | True -6V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k 8x10k | True -6V |
| 07110614_1802 07110614_1853 07110614_1943 07110614_2034 07110614_2124 07110614_2214 07110614_2305 07110614_2355 07110615_0045 07110615_0135 | Am-241 | True -2V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 8x10k 7x10k 7x10k 8x10k 8x10k 8x10k 7x10k 8x10k 8x10k 8x10k | True -2V |
| 07110615_0228 07110615_0318 07110615_0408 07110615_0458 07110615_0548 07110615_0639 07110615_0729 07110615_0819 07110615_0909 07110615_0959 | Am-241 | True -1V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 9x10k 10x10k 10x10k 10x10k | True -1V |
| 07110615_1053 07110615_1144 07110615_1234 07110615_1324 07110615_1414 07110615_1504 07110615_1555 07110615_1645 07110615_1735 07110615_1826 | Am-241 | True +1V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 9x10k 10x10k | True +1V |
| 07110615_1919 07110615_2009 07110615_2059 07110615_2150 07110615_2240 07110615_2330 07110616_0020 07110616_0110 07110616_0200 07110616_0251 | Am-241 | True +2V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k | True +2V |
| 07110616_1030 07110616_1120 07110616_1210 07110616_1300 07110616_1351 07110616_1441 07110616_1532 07110616_1906 07110616_1957 07110616_2047 | Am-241 | True +4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 7x10k 7x10k 7x10k 7x10k 7x10k 7x10k 7x10k 8x10k 8x10k 8x10k | True +4V |
| 07110616_2142 07110616_2232 07110616_2322 07110617_0012 07110617_0102 07110617_0152 07110617_0243 07110617_0333 07110617_0424 07110617_0514 | Am-241 | True +6V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k | True +6V |
| 07110617_0607 07110617_0637 07110617_0707 07110617_0738 07110617_0808 07110617_0838 07110617_0908 07110617_0938 | Am-241 | Calibrated biases on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 9x10k 9x10k 10x10k | -4V +4V -4V +4V -4V +4V -4V +4V |
| 07110617_1845 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Flashing 10min before 30min data. | 8x10k | Check flashing time |
| 07110617_1935 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Flashing 10min before 30min data. | 8x10k | Check flashing time |
| 07110618_1334 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. 6s flashing before data | 75x10k | Long neutralization run at -4V |
| 07110618_1949 | Am-241 | True 0V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Detector biased at -4V before data. | 58x10k | Long 0V run after long -4V bias |
| 07110619_0016 | Am-241 | True +4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. 6s flashing before data | 116x10k | +4V long neutralization |
| 07110619_0616 | Am-241 | True 0V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. +4V bias before data | 80x10k | 0V bias long run after +4V bias |
| 07110619_1838 07110619_1928 07110619_2019 07110619_2109 07110619_2159 07110619_2249 | Am-241 | True -7V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k | Bias sweep |
| 07110619_2344 07110620_0034 07110620_0124 07110620_0215 07110620_0305 07110620_0355 | Am-241 | True -5V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k | Bias sweep |
| 07110620_0449 07110620_0539 07110620_0629 07110620_0720 07110620_0810 07110620_0900 | Am-241 | True -3V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k | Bias sweep |
| 07110620_0955 07110620_1045 07110620_1135 07110620_1225 07110620_1315 07110620_1406 | Am-241 | True +3V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k | Bias sweep |
| 07110620_1500 07110620_1550 07110620_1641 07110620_1731 07110620_1822 07110620_1912 | Am-241 | True +5V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k 10x10k 10x10k | Bias sweep |
| 07110620_2118 07110620_2210 07110620_2301 07110620_2352 | Am-241 & Ba-133 | True -4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 6x10k 7x10k 7x10k 7x10k | Ba data |
| 07110621_0044 07110621_0135 07110621_0226 07110621_0317 | Am-241 & Ba-133 | True -2V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 9x10k 9x10k 9x10k 9x10k | Ba data |
| 07110621_0409 07110621_0500 07110621_0550 07110621_0641 | Am-241 & Ba-133 | True +2V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 9x10k 9x10k 9x10k 9x10k | Ba data |
| 07110621_0735 07110621_0825 07110621_0916 07110621_1006 | Am-241 & Ba-133 | True +4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 8x10k 8x10k 8x10k 8x10k | Ba data |
| 07110621_1825 07110621_1915 07110621_2005 07110621_2056 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 8x10k 8x10k 8x10k 8x10k | -3.5V |
| 07110621_2151 07110621_2241 07110621_2331 07110622_0021 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -2.5V |
| 07110622_0117 07110622_0207 07110622_0257 07110622_0347 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -1.5V |
| 07110622_0442 07110622_0532 07110622_0623 07110622_0713 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -0.5V |
| 07110622_0808 07110622_0858 07110622_0948 07110622_1039 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | 0.5V |
| 07110622_1134 07110622_1224 07110622_1314 07110622_1404 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | 1.5V |
| 07110622_1953 07110622_2043 07110622_2133 07110622_2224 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 7x10k 8x10k 8x10k 8x10k | -3.5V |
| 07110622_2319 07110623_0009 07110623_0100 07110623_0150 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -2.5V |
| 07110623_0245 07110623_0335 07110623_0426 07110623_0516 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -1.5V |
| 07110623_0611 07110623_0701 07110623_0752 07110623_0842 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -0.5V |
| 07110623_0937 07110623_1027 07110623_1117 07110623_1207 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 9x10k 9x10k 10x10k | 0.5V |
| 07110623_1303 07110623_1353 07110623_1443 07110623_1533 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | 1.5V |
| 07110624_1336 07110624_1427 07110624_1517 07110624_1607 07110624_1657 | Am-241 | Calibrated -7V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 7x10k 7x10k 7x10k 7x10k 7x10k | Test run after detector baking. Data not good. |
| 07110625_1134 07110625_1224 | Am-241 | True -4V on Q1-4, Trigger on Q1-4 at (22,14,22,14) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 7x10k 8x10k | -4V test run to check the 60keV line after baking |
| 07110626_0319 07110626_0409 07110626_0459 07110626_0549 | Am-241 | True -4V on Q1-4; triggered on Q1 and Q3 at 22ADC bin. | 14x10k 14x10k 13x10k 14x10k | noise is stable. |
| 07110626_1204 | Am-241 | -7V trigger test | 3x10k | |
| 07110626_1414 | Am-241 | -7V trigger test; thresholds are at (26, 32, 38, 30) for Q1-4. | 3x10k | |
| 07110626_1450 07110626_1541 07110626_1632 07110626_1722 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 8x10k 8x10k 8x10k 8x10k | -7V |
| 07110626_1816 07110626_1906 07110626_1956 07110626_2047 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 11x10k | -5V |
| 07110626_2141 07110626_2231 07110626_2322 07110627_0012 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -3.5V |
| 07110627_0107 07110627_0157 07110627_0247 07110627_1008 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 9x10k | -3V |
| 07110627_1103 07110627_1154 07110627_1244 07110627_1334 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 8x10k 8x10k 8x10k 8x10k | -2.5V |
| 07110627_1429 07110627_1520 07110627_1610 07110627_1701 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | -1.5V |
| 07110627_1755 07110627_1845 07110627_1935 07110627_2043 07110627_2133 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 4x10k 10x10k 8x10k | -0.5V |
| 07110627_2228 07110627_2318 07110628_0009 07110628_0059 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | 0.5V |
| 07110628_0154 07110628_0244 07110628_0334 07110628_0424 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | 1.5V |
| 07110628_0519 07110628_0609 07110628_0659 07110628_0750 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 10x10k 10x10k | 3V |
| 07110628_0845 07110628_0935 07110628_1212 07110628_1303 | Am-241 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 10x10k 10x10k 9x10k 8x10k | 5V |
| 07110628_1717 07110628_1807 | Am-241 & Ba-133 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 9x10k 9x10k | -4V |
| 07110628_1902 07110628_1953 | Am-241 & Ba-133 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 7x10k 7x10k | -2V |
| 07110628_2048 07110628_2138 | Am-241 & Ba-133 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 9x10k 9x10k | +2V |
| 07110628_2233 07110628_2323 | Am-241 & Ba-133 | Calibrated bias on Q1-4, Trigger on Q1-4 at (26, 32, 38, 30) ADC. Trace length 0.4096ms (1024bins), lookback time 100us. Data/flash/cooldown = 30min/6s/20min. | 8x10k 7x10k | +4V |
Run 12
iZIPv4 G42 Test.
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| DCRC30 on DIB1; DIB1 is zip1 in oZIP format data | ||||
| 07110428_1946 | Am-241 | -2 V bias DIB1; +2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 14 x 10k | File 14 not written out properly; one phonon channel railed on DIB1; oZIP data format. |
| 07110430_2312 | Am-241 | -2 V bias DIB1; +2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 9 x 10k | iZIPv4 data format. |
| 07110502_0044 | Am-241 & Ba-133 | -2 V bias DIB1; +2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 51 x 10k | oZIP data format, corrupted? |
| 07110502_2009 | Am-241 | -2 V bias DIB1; +2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 10 x 10k | iZIPv4 data format. |
| 07110502_2259 | Am-241 & Ba-133 | -2 V bias DIB1; +2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 9 x 10k | iZIPv4 data format. |
| 07110504_1701 | Am-241 & Ba-133 | -3 V bias DIB1; +3 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 11 x 10k files | iZIPv4 data format. |
| 07110504_2201 | Am-241 & Ba-133 | -3 V bias DIB1; +3 V bias DIB2; phonon channels locked but no QET bias applied; charge-only; Ba-133 source 20“ below top of mu-metal shield | 9 x 10k | iZIPv4 data format. |
| 07110505_1551 | Am-241 & Ba-133 | -3 V bias DIB1; +2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 23 x 10k | iZIPv3 data format. |
| 07110505_2149 | Am-241 & Ba-133 | -2 V bias DIB1; +2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 19 x 10k | iZIPv4 data format. |
| 07110508_2236 | Am-241 & Ba-133 | +2 V bias DIB1; -2 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 13 x 10k | iZIPv4 data format. |
| Trace length and trigger delay divided by 4 in files below (ie. trace length = 409.6 ns and trigger delay = 100 μs unless otherwise specified. | ||||
| 07110510_0009 | Am-241 & Ba-133 | +3 V bias DIB1; -3 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 12 x 10k. | iZIPv4 data format, no timing link between boards. |
| 07110510_1834 07110510_2046 07110510_2307 | Am-241 & Ba-133 | -2.5 V bias DIB1; +2.5 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 7 x 10k 9 x 10k 7 x 10k | iZIPv4 data format. |
| 07110511_1821 | Am-241 & Ba-133 | -2.5 V bias DIB1; +2.5 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 14 x 10k | iZIPv4 data format. |
| 07110511_2353 07110512_0105 07110512_0218 07110512_0330 07110512_0443 | Am-241 & Ba-133 | -2.5 V bias DIB1; +2.5 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 13 x 10k 12 x 10k 11 x 10k 11 x 10k 8 x 10k | iZIPv4 data format; last series 07110512_0443 incomplete due to DAQ error: 44 minutes long; 60 min/6 s/12 min. |
| 07110513_2327 07110514_0043 07110514_0158 07110514_0313 07110514_0428 | Am-241 & Ba-133 | -2.5 V bias DIB1; +2.5 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 13 x 10k 12 x 10k 11 x 10k 11 x 10k 8 x 10k | iZIPv4 data format; last series 07110514_0428 incomplete due to DAQ error: 46 minutes long; 60 min/6 s/15 min. |
| 07110517_0001 07110517_0221 07110517_0441 07110517_0702 07110517_0922 | Am-241 & Ba-133 | -2.5 V bias DIB1; +2.5 V bias DIB2; phonon channels locked but no QET bias applied; charge-only | 13 x 10k 12 x 10k 11 x 10k 11 x 10k 8 x 10k | iZIPv4 data format; last series 07110517_0922 incomplete due to DAQ error: 46 minutes long; 60 min/20 min/60 min. |
| 07110519_0120 07110519_0340 07110519_0600 07110519_0820 07110519_1040 07110519_1300 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge-only | 15 x10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k | oZIP data format; 60 min/20 min/60 min. |
| 07110521_2058 07110521_2318 07110522_0138 07110522_0358 07110522_0618 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge-only | 13 x 10k 12 x 10k 12 x 10k 11 x 10k 7 x 10k | iZIPv4 data format; last series 07110522_0618 incomplete due to DAQ error: 40 minutes long; 60 min/20 min/60 min. |
| 07110522_2054 07110522_2209 07110522_2324 07110523_0039 07110523_0154 07110523_0310 07110523_0425 07110523_0540 07110523_0655 07110523_0810 07110523_0926 07110523_1041 07110523_1156 07110523_1311 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge-only | 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k 15 x 10k | oZIP data format; 60 min/6 s/15 min. |
| 07110524_0118 07110524_0339 07110524_0600 07110524_0820 07110524_1040 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge-only | 13 x 10k 12 x 10k 12 x 10k 11 x 10k 7 x 10k | iZIPv4 data format; last series 07110524_1040 incomplete due to DAQ error; 60 min/6 s/15 min. |
| 07110525_1928 07110525_2147 07110526_0008 07110526_0238 07110526_0453 07110526_0723 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons | 8 x 10k 8 x 10k 8 x 10k 8 x 10k 8 x 10k | oZIP data format; 3.2768 ms trace length; 400 μs trigger delay; 120 min/6 s/15 min. |
| 07110526_1003 07110526_1227 07110526_1450 07110526_1718 07110526_1943 07110526_2205 07110527_0029 07110527_0249 07110527_0506 07110527_0724 07110527_0945 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons | 10 x 10k 10 x 10k 10 x 10k 10 x 10k 6 x 10k 10 x 10k 10 x 10k 10 x 10k 10 x 10k 10 x 10k 10 x 10k | iZIPv4 data format; 3.2768 ms trace length; 400 μs trigger delay; 120 min/6 s/15 min. |
| 07110527_2245 07110528_0000 07110528_0116 07110528_0231 07110528_0346 07110528_0501 07110528_0617 07110528_0732 07110528_0847 07110528_1002 07110528_1117 07110528_1233 07110528_1348 07110528_1503 07110528_1619 07110528_1734 | Am-241 & Ba-133 | -2.5 V DIB1; + 2.5V DIB2; charge and phonons | 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k | iZIPv4 data format; 0.4096 ms trace length; 100 μs trigger delay; bias was not set to zero during flashing and cool down, so junky; 60 min/6 s/15 min. |
| 07110528_1934 07110528_2049 07110528_2204 07110528_2320 07110529_0035 07110529_0150 07110529_0305 07110529_0420 07110529_0536 07110529_0651 07110529_0807 07110529_0922 07110529_1037 07110529_1153 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons | 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 11 x 10k 12 x 10k 12 x 10k 12 x 10k 12 x 10k 12 x 10k 12 x 10k 12 x 10k 13 x 10k | iZIPv4 data format; 0.4096 ms trace length; 100 μs trigger delay; bias set to zero during flashing and cool down, so good; 60 min/6 s/15 min. |
| 07110529_2326 07110530_0014 07110530_0100 07110530_0145 07110530_0232 07110530_0318 07110530_0405 07110530_0452 07110530_0539 07110530_0626 07110530_0715 07110530_0804 07110530_0853 07110530_0944 07110530_1034 07110530_1124 07110530_1210 07110530_1259 07110530_1349 07110530_1438 07110530_1528 | Am-241 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k | 30min data, 6s flashing, 15min cooldown. |
| 07110530_1900 07110530_1948 07110530_2035 07110530_2123 07110530_2208 07110530_2254 07110530_2341 07110531_0028 07110531_0115 07110531_0202 07110531_0250 07110531_0337 07110531_0424 07110531_0512 07110531_0559 07110531_0647 07110531_0734 07110531_0821 07110531_0908 07110531_0955 07110531_1043 | Am-241 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k | Continuation of the data above after squid zapping |
| 07110531_2205 07110531_2254 07110531_2343 07110601_0032 07110601_0120 07110601_0209 07110601_0258 07110601_0347 07110601_0436 07110601_0525 07110601_0614 07110601_0703 07110601_0751 07110601_0840 07110601_0929 07110601_1017 | Am-241 & Cf-252 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k | 30min data, 6s flashing, 15min cooldown |
| 07110601_1928 07110601_2016 | Am-241 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 3x10k 3x10k | 30min data, 6s flashing, 15min cooldown |
| 07110601_2104 07110601_2153 07110601_2241 07110601_2330 07110602_0019 07110602_0107 07110602_0153 07110602_0243 07110602_0332 07110602_0421 07110602_0511 07110602_0600 07110602_0647 07110602_0734 07110602_0822 | Am-241 & Cf-252 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 3x10k 3x10k 3x10k 3x10k 3x10k 2x10k 3x10k 3x10k 3x10k 3x10k 3x10k 2x10k 2x10k 3x10k 3x10k | 30min data, 6s flashing, 15min cooldown |
| 07110602_0911 07110602_0959 07110602_1045 07110602_1133 07110602_1220 07110602_1308 07110602_1356 07110602_1443 07110602_1531 07110602_1619 07110602_1706 07110602_1752 07110602_1840 07110602_1929 07110602_2017 07110602_2105 07110602_2154 07110602_2242 07110602_2330 07110603_0019 07110603_0107 07110603_0155 07110603_0242 07110603_0330 | Am-241 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k | 30min data, 6s flashing, 15min cooldown |
| 07110603_0418 07110603_0507 07110603_0557 07110603_0646 07110603_0736 07110603_0825 07110603_0915 07110603_1004 | Am-241 & Cf-252 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k | 30min data, 6s flashing, 15min cooldown |
| 07110603_1053 07110603_1142 07110603_1230 07110603_1319 07110603_1407 07110603_1455 07110603_1544 07110603_1632 07110603_1721 | Am-241 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k | 30min data, 6s flashing, 15min cooldown |
| 07110603_2047 07110603_2139 07110603_2231 07110603_2323 07110604_0015 07110604_0107 07110604_0157 07110604_0247 07110604_0336 07110604_0427 07110604_0518 07110604_0608 07110604_0658 07110604_0749 07110604_0840 07110604_0931 07110604_1022 07110604_1113 07110604_1203 07110604_1253 07110604_1343 | Cf-252 w/ n shield | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k | 30min data, 6s flashing, 15min cooldown |
| 07110604_1708 07110604_1823 07110604_1938 07110604_2053 | Am-241 & Ba-133 | +2.5 V DIB1; -2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 0.4096ms (1024 samples), lookback time 100us; ozip format. | 8x10k 8x10k 8x10k 8x10k | 30min data, 6s flashing, 15min cooldown; check polarity effect |
| 07110604_2234 07110604_2321 07110605_0008 07110605_0057 07110605_0148 07110605_0238 07110605_0327 07110605_0414 07110605_0459 07110605_0545 07110605_0631 07110605_0717 07110605_0803 07110605_0850 07110605_0937 07110605_1023 07110605_1110 07110605_1156 07110605_1241 07110605_1327 07110605_1413 07110605_1459 07110605_1544 07110605_1630 | Cf-252 with n-shield | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768ms (8192 samples), lookback time 400us; ozip format. | 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 3x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k 2x10k | 30min data, 6s flashing, 15min cooldown |
| 07110605_1721 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2; charge and phonons, trigger on 8 phonon channels at 18ADC bins. Trace length 3.2768 ms (8192 samples), lookback time 400 μs; ozip format. | 21x10k | Long Ba-133 run, no flashing |
| 07110605_2353 07110606_0013 07110606_0033 07110606_0053 07110606_0113 07110606_0133 07110606_0153 07110606_0214 07110606_0234 07110606_0254 07110606_0314 07110606_0334 07110606_0354 07110606_0414 07110606_0434 07110606_0454 07110606_0514 07110606_0534 07110606_0554 07110606_0615 07110606_0635 07110606_0655 07110606_0715 07110606_0735 07110606_0755 07110606_0815 | Am-241 & Ba-133 | -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 -2.5 V DIB1; +2.5 V DIB2 +2.5 V DIB1; -2.5 V DIB2 | 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k | Flipping biases; trace length 0.4096 ms (1024 samples); lookback time 400 μs; ozip format; no flashing. File 3 of sereis 07110605_2353, 07110606_0113, 07110606_0153, 07110606_0414, 07110606_0454 could not be processed because they were not closed properly. After deleting the last block starting with a980, the files can be processed successfully. I guess specifying an event number smaller the the last one should also work. |
Run 11
100mm detector g101 ionization test
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| DCRC33 on DIB2 | ||||
| 07110317_1126 | Am-241 | Bias -3V on Q1-4. Trigger: Q1 OR Q2 OR Q3 OR Q4. Threshold: 40 ADC bins. DIB2→Detector 1, DIB1→Detector 2. | 18x10k | Couldn't see 60keV line |
| 07110318_1103 | Am-241 | -3V on Q1-4. Trigger: 32 ADC on Q1 | 5x10k | Couldn't see 60keV line |
| 07110318_1405 | Am-241 | -4V on Q1-4. Trigger: 40ADC on Q1 | 5x10K | Couldn't see 60keV line |
| 07110318_1853 | Am-241 & Ba-133 | -4V on Q1, 0V on Q2-4. Trigger, 40ADC on Q1 | 10x10k | Couldn't see 356keV line? |
| 07110319_1130 | Am-241 & Ba-133 | -6V on Q1, 0V on Q2-4. Trigger, 64ADC on Q1 | 14x10k | Couldn't see 356keV line? |
| 07110319_1924 | Am-241 & Ba-133 | -4V on Q1, 0V on Q2-4. Trigger, 64ADC on Q1 | 9x10k | Couldn't see 356keV line? |
| Reduced trace length and used OptimalFilterCharge in processing from the series 07110321_1323. | ||||
| 07110321_1323 | Am-241 | Trace length 0.4096ms, lookback time 100us. Trigger coincidence time 50us. -4V on Q1, 0V on Q2-4. Trigger: 64ADC on Q1. | 10x10k | 60keV line ? |
| 07110321_1943 | Am-241 & Ba-133 | Trace length 0.4096ms, lookback time 100us. Trigger coincidence time 50us. -4V on Q1-4. Trigger: 64ADC on Q1. | 11x10k | |
| 07110322_0020 | Am-241 & Ba-133 | Trace length 0.4096ms, lookback time 100us. Trigger coincidence time 200us. -4V on Q1-4. Trigger: 64ADC on Q1-4. | 182x10k | |
| 07110323_0058 | Am-241 | -4V on Q1-4. Trigger: Q1=48ADC, Q2=51, Q3=75, Q4=59, coincidence time 200us. Am-241. Q samples 1024, lookback time 100us. | 208x10k | |
| DCRC30 on DIB2 | ||||
| 07110324_0118 | Am241 | -4V on Q1 & Q3. Trigger: Q1=17ADC, Q3=15. Trace length 1024, lookback time 100us. DCRC30 on DIB2. DIB1 connected to nothing. | 90x10k | |
| 07110325_0225 07110325_0425 07110325_0625 07110325_0825 | Am241 | -4V on Q1-4. Trigger: Q1(28ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. Auto run enabled, take data 60min, flash 20min, cooldown 40min. Flash settings: DCRC30 on DIB2 (Q1&3), LED 1&2; DCRC33 on DIB1 (Q2&4), LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~57mK. | (22x10k)x4 | Q1&3 were not quite responsive. |
| 07110326_1152 07110326_1352 07110326_1552 | Am241 | -4V on Q1-4. Trigger: Q1(28ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. Auto run enabled, take data 60min, flash 20min, cooldown 40min. Flash settings: DCRC30 on DIB2 (Q1&3), LED 1&2; DCRC33 on DIB1 (Q2&4), LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~57mK. | (18x10k)x3 | Q1&3 were not quite responsive. |
| 07110327_1326 | Am241 | -4V on Q1-4. Trigger: Q1(28ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. | 2x1k | Short run to check neutralization. |
| 07110327_1612 | Am241 | -4V on Q1-4. Trigger: Q1(28ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. | 78x1k | 30min run to check neutralization. |
| DCRC33 on DIB2 | ||||
| 07110328_1916 | Am-241 | -4V on Q1-4. Trigger: Q1(30ADC) OR Q2(28ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. | 10x10k | swapping the two boards |
| 07110329_0954 | Am-241 | -4V on Q1-4. Trigger: Q1(30ADC) OR Q2(28ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC33 was on DIB2 and was mapped to ZIP1. DCRC30 was the master in timing. | 8x10k | reference to ZIP1/2 and master/slave symmetry test |
| 07110329_1119 | Am-241 | -4V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC33 was on DIB2 and was mapped to ZIP2. | 8x10k | zip1/2 symmetry |
| 07110329_1503 | Am-241 | -4V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC33 was on DIB2 and was mapped to ZIP1. DCRC33 was the master in timing. | 8x10k | master/slave test |
| 07110329_2325 | Am-241 | -4V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC33 was on DIB2 and was mapped to ZIP1, and was master. Tower temperature 58mK. | 162x10k | long neutralization run |
| 07110331_0006 07110331_0206 07110331_0406 07110331_0607 | Am-241 | :4 -4V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 60min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~62mK. | (16,19,19,9)x10k | 1h neutralization run, DAQ crashed in the run. |
| 07110331_1329 07110331_1444 07110331_1559 07110331_1714 | Am-241 | -6V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~60.5mK. | (5x10k)x4 | -6V 15min bias sweep |
| 07110331_1830 07110331_1945 07110331_2100 07110331_2216 | Am-241 | -8V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~65mK. | (5x10k)x4 | -8V 15min bias sweep |
| 07110331_2334 07110401_0049 07110401_0204 07110401_0319 07110401_0434 07110401_0549 07110401_0704 07110401_0819 | Am-241 | -2V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~63.4mK. | (5x10k)x8 | -2V 15min bias sweep |
| 07110401_0941 07110401_1056 07110401_1212 07110401_1327 | Am-241 | -1V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower temperature ~63.4mK. | (5x10k)x4 | -1V 15min bias sweep |
| 07110402_1805 07110402_1936 07110402_2106 07110402_2236 07110403_0007 07110403_0137 07110403_0307 07110403_0437 07110403_0608 07110403_0738 07110403_0908 07110403_1039 | Am-241 | Calibrated -4V on Q1-Q4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 30min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. | 9x10k 9x10k 10x10k 11x10k 11x10k 10x10k 10x10k 10x10k 9x10k 9x10k 9x10k 7x10k | Calibrated -4V bias data |
| 07110403_1928 07110403_2058 07110403_2229 07110404_0000 07110404_0131 07110404_0302 07110404_0433 07110404_0603 07110404_0733 07110404_0905 07110404_1035 07110404_1205 07110404_1336 07110404_1507 07110404_1813 07110404_1944 | Am-241 | -5 V bias on Q1 & Q3; -4 V bias on Q2 & Q4 | 7x10k 8x10k 8x10k 8x10k 9x10k 9x10k 9x10k 8x10k 8x10k 8x10k 8x10k 7x10k 5x10k 6x10k 5x10k 5x10k | Check bias effects |
| 07110404_2119 07110404_2320 07110405_0121 07110405_0321 07110405_0522 07110405_0724 07110405_0924 07110405_1125 07110405_1325 | Am-241 | Biases. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 60min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 62.1mK. | 12x10k 11x10k 14x10k 14x10k 14x10k 13x10k 12x10k 12x10k 11x10k | Set biases to mimic the effects of swapping the boards. |
| 07110405_2059 07110405_2216 07110405_2336 07110406_0052 07110406_0207 07110406_0323 07110406_0438 07110406_0553 | Am-241 | Calibrated -6V on Q1-Q4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. | (3x10k)x8 | Calibrated -6V 15min sweep |
| 07110406_0708 07110406_0823 07110406_0940 07110406_1056 07110406_1212 07110406_1327 07110406_1442 | Am-241 | Calibrated -8V on Q1-Q4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. | 4x10k 4x10k 3x10k 3x10k 3x10k 3x10k 3x10k | Calibrated -8V 15min sweep |
| 07110406_1558 07110406_1716 07110406_1836 07110406_1951 07110406_2109 07110406_2229 | Am-241 | Calibrated -2V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 63.5mK. | 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k | Calibrated -2V 15min sweep. |
| 07110406_2345 07110407_0103 07110407_0219 07110407_0335 07110407_0451 07110407_0606 07110407_0722 07110407_0838 | Am-241 | Calibrated -1V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 60min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 61.2mK. | (4x10k)x8 | Calibrated -1V 15min sweep. |
| 07110407_0956 07110407_1111 07110407_1227 07110407_1343 07110407_1500 07110407_1616 07110407_1731 07110407_1846 | Am-241 | Calibrated +1V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 30min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 62.8mK. | 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 3x10k 2x10k | +1V 15min bias sweep. Inverted pulses. |
| 07110407_2002 07110407_2121 07110407_2236 07110407_2355 07110408_0112 07110408_0231 07110408_0347 07110408_0502 07110408_0619 07110408_0735 | Am-241 | Calibrated +2V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 66.3mK. | 3x10k 3x10k 3x10k 3x10k 4x10k 4x10k 4x10k 4x10k 4x10k 3x10k | +2V 15min bias sweep. Inverted pulses. |
| 07110408_0859 07110408_1016 07110408_1131 07110408_1247 07110408_1403 07110408_1520 07110408_1635 07110408_1750 | Am-241 | Calibrated +4V on Q1-4. Trigger: Q1(30ADC) OR Q2(30ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 62.8mK. | 3x10k 2x10k 2x10k 2x10k 3x10k 2x10k 2x10k 2x10k | +4V 15min bias sweep. Inverted pulses. |
| 07110409_0154 07110409_0309 07110409_0425 07110409_0540 07110409_0655 | Am-241 | Calibrated +6V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Auto run enabled, take data 15min, flash 20min, cooldown 40min. Flash settings: DIB2, LED 1&2; DIB1, LED1; LED current 0.6mA, 100us ON, 2500us OFF. Tower T 60.8mK. | 5x10k 5x10k 5x10k 5x10k 5x10k | +6V 15min bias sweep. |
| Timing link might not be set successfully on the following series. | ||||
| 07110409_2335 | Am-241 | Calibrated -1V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. Tower T 61.5mK. | 11x10k | Inverted pulses. |
| 07110410_1205 07110410_1406 | Am-241 | Calibrated -2V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | (9x10k)x2 | |
| 07110410_1611 | Am-241 | Calibrated -3V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 10x10k | |
| 07110410_1814 | Am-241 | Calibrated -4V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 10x10k | |
| 07110410_2043 | Am-241 | Calibrated -5V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 9x10k | |
| 07110410_2244 | Am-241 | Calibrated -6V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 12x10k | |
| 07110411_0044 | Am-241 | Calibrated -7V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 13x10k | |
| 07110411_0245 | Am-241 | Calibrated -8V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 13x10k | |
| 07110411_0445 | Am-241 | Calibrated -1V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 13x10k | Retake with non-inverted pulses |
| 07110411_0646 | Am-241 | Calibrated -(1,3,2,4)V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 4x10k | |
| 07110411_0846 | Am-241 | Calibrated -(4,2,3,1)V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 5x10k | |
| 07110411_1142 07110411_1304 07110411_1420 07110411_1536 07110411_1652 | Am-241 | Calibrated -4V on Q1-4. Trigger: Q1(60ADC) OR Q2(36ADC) OR Q3(30ADC) OR Q4(30ADC); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 15min after 20min flashing. | 2x10k 3x10k 3x10k 3x10k 3x10k | |
| Timing link might not be set successfully on the series above. | ||||
| 07110411_2053 | Am-241 | Calibrated -8V on Q1-4. Trigger: Q1=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 15min after 20min flashing. | 4x10k | 15min short run to check trigger settings. |
| 07110412_0040 07110412_0241 07110412_0441 07110412_0641 07110412_0842 07110412_1042 07110412_1243 07110412_1443 07110412_1644 | Am-241 | Calibrated biases on Q1-4. Trigger: Q1=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Take data 60min after 20min flashing. | 15x10k 20x10k 20x10k 20x10k 20x10k 20x10k 18x10k 20x10k 4x10k | -8V -7V -6V -5V -4V -3V -2V -1V -1324V |
| 07110412_2138 07110412_2155 07110412_2215 07110412_2225 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Flash for 20 min, 10 min -4 V bias across all electrodes, 20 min 0 V bias, 10 min -4 V bias across all electrodes, 20 min 0 V bias. | 3 x 10k 5 x 10k 4 x 10k 7 x 10k | -4 V 0 V -4 V 0 V |
| 07110413_0116 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Ground Q1 & Q3 to chassis and flash for 20 min. | 16 x 10k | -4 V |
| 07110413_0357 07110413_0418 07110413_0439 07110413_0449 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Flash for 20 min, 10 min -4 V bias on Q1, 20 min 0 V bias, 10 min -4 V bias on Q1, 20 min 0 V bias. | 3 x 10k 6 x 10k 3 x 10k 7 x 10k | -4 V (Q1 only) 0 V -4 V (Q1 only) 0 V |
| 07110413_1536 07110413_1546 07110413_1606 07110413_1616 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Flash for 20 min, 10 min -4 V bias on Q2, 20 min 0 V bias, 10 min -4 V bias on Q2, 20 min 0 V bias. | 3 x 10k 5 x 10k 3 x 10k 6 x 10k | -4 V (Q2 only) 0 V -4 V (Q2 only) 0 V |
| 07110413_1745 07110413_1756 07110413_1816 07110413_1826 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Flash for 20 min, 10 min -4 V bias on Q3, 20 min 0 V bias, 10 min -4 V bias on Q3, 20 min 0 V bias. | 1 x 10k 6 x 10k 2 x 10k 6 x 10k | -4 V (Q3 only) 0 V -4 V (Q3 only) 0 V |
| 07110413_2008 07110413_2018 07110413_2038 07110413_2049 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Flash for 20 min, 10 min -4 V bias on Q4, 20 min 0 V bias, 10 min -4 V bias on Q4, 20 min 0 V bias. | 1 x 10k 6 x 10k 2 x 10k 6 x 10k | -4 V (Q4 only) 0 V -4 V (Q4 only) 0 V |
| 07110413_2239 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Flash 20 min, Q1 -4 V, Q2-4 +4 V. | 2 x 10k | -4 V (Q1) +4 V (Q2-4) |
| 07110414_0407 | Am-241 | Calibrated biases on Q1-4. Trigger: Q=46(0x2E), Q2=34(0x22), Q3=24(0x18), Q4=29(0x1D); coincidence time 50us. Trace length 1024, lookback time 100us. DCRC30 on DIB1, ZIP2, master; DCRC33 on DIB2, ZIP1, slave. Flash 20 min, fruit bowl: Q1<Q2<Q3<Q4. | 9 x 10k | -4 V (Q1) -3 V (Q2) -2 V (Q3) -1 V (Q4) |
Corrupted files
Below is a incomplete list of files which could not be processed successfully by BatRoot. It's likely the rq files produced won't be able to be read by root/matlab without problems
07110412_2215_F0004 07110413_0357 3 07110413_0418 6 07110413_0449 7 07110413_2018 6 07110413_1745 1 07110413_1816 2 07110413_1606 3 07110413_1536 3 07110413_2239 2
Run 10
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07110304_1358 | 2 × Am-241 on G9; neutron shield | 82 10k-files | Trigger on phonon channels: threshold 8 ADC bins | |
| 07110305_0350 | 2 × Am-241 on G9 & Cf-252; neutron shield | 61 10k-files | Trigger on phonon channels: threshold 10 ADC bins; Cf-252 source 4’ from metal shield roughly at height of detectors | |
| 07110305_1427 | 2 × Am-241 on G9 & Ba-133; neutron shield | 29 10k-files | Trigger on phonon channels; threshold 10 ADC bins; Ba-133 source 13” below top of metal shield | |
| 07110305_2101 | 2 x Am-241 on G9 & Cf-252; unshielded | 24 10k-files | Trigger on phonon channels; threshold 10 ADC bins; Cf-252 source 4’ from metal shield roughly at height of detectors | |
| 07110306_0028 | 2 x Am-241 on G9; unshielded | 66 10k-files | Trigger on phonon channels; threshold 10 ADC bins |
Run 8
| Data | Detector | Comment |
|---|---|---|
| 07101221_2209 | G9 and G8, 2 Am-241 source on G9 | Data taken: 9 files; event rate 15-20 Hz; DAQ settings and Trigger settings |
| 07101222_2105 | G9 and G8, 2 Am-241 source on G9 | Trigger on phonons. Trigger threshold: DCRC30 G8 3mV; DCRC33 G9 8mV. DAQ setting and Trigger setting |
| 07101222_2138 | G9 and G8, 2 Am-241 source on G9 | 11 files, background run. PhD of one detector railed during the run. DCRC setting and Trigger settings. Trigger threshold: DCRC30 G8 3mV; DCRC33 G9 8mV. |
| 07101222_2351 | G9 and G8, 2 Am-241 source on G9, Cf-252 ~40” from fridge | DCRC setting and Trigger settings. Trigger threshold: DCRC30 G8 3mV; DCRC33 G9 8mV. vi stopped at file 2 |
| 07101223_0122 | Same as 07101222_2351 | vi stopped at file 2 |
| 07101223_0150 | Same as 07101222_2351 | vi stopped at file 2 |
| 07101223_0401 | G9 and G8, 2 Am-241 source on G9, Cf-252 ~40“ from fridge | Trigger threshold: DCRC30 G8=3.5mV (0x17); DCRC33 G9=9mV (0x3b). vi stopped at file 5. |
| 07101223_0503 | G9 and G8, 2 Am-241 source on G9 | Trigger threshold: DCRC30 G8=3.5mV (0x17); DCRC33 G9=9mV (0x3b). 5 files. DCRC30, DCRC33 |
| 07101223_0853 | G9 and G8, 2 Am-241 source on G9, with neutron shield | Trigger threshold: DCRC30 G8=3.5mV (0x17); DCRC33 G9=9mV (0x3b). 6 files. DCRC30, DCRC33 |
| 07101223_1052 | G9 and G8, 2 Am-241 source on G9, with neutron shield, Cf-252 ~40” from fridge | Trigger threshold: DCRC30 G8=3.5mV (0x17); DCRC33 G9=9mV (0x3b). 6 files. DCRC30, DCRC33 |
| 07101223_1224 | G9 and G8, 2 Am-241 source on G9, Ba-133 | 4 files. DCRC30, DCRC33 |
| 07101223_1256 | Same as 07101223_1224, 2 files. Lost G9 phonon B during run. | |
Run 7
| Data | Comment |
|---|---|
| 07101121_1733 | DIB1 -2V, DIB2 +2V. Test series |
| 07101123_2223 | DIB1 -1V, DIB2 +1V. Trigger 400ADC both sides. Tower T=118mK, 50K events |
| 07101123_2252 | DIB1 -2V, DIB2 +2V. Trigger 400ADC both sides. Tower T=118mK |
| 07101123_2320 | DIB1 -0.5V, DIB2 +0.5V, Trigger 400ADC both sides. Tower T=119mK |
| 07101124_1352 | DIB1 -3 V on DCRC33, DIB2 +3 V on DCRC30. Stopped by error. |
Run 6
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07100817_0102 | Am-241 | DAQ settings | 44 10k-files | IZIP G41 DIB2 Am-241 overnight run |
Run 5
| Run | Run Mode | Run Info | # of Events/Files | Comments |
|---|---|---|---|---|
| 07100727_1819 | Am-241 | 2 10k-files | All driver gain are 1. Stopped the run to adjust them | |
| 07100727_2111 | Am-241 | DAQ settings | 10 10k-files, 99958 | Good run, large pulse of channel B(?) had a noise blob on falling edge |
| 07100728_1940 | Am-241 & Cf-252 | DAQ settings | 11 10k-files | Good run, last file has only 1 event. |
| 07100728_2249 | Am-241 | DAQ settings | 75 10k-files | Good run, test neutralization |
| 07100729_1856 | Am-241 & Cf-252 | DAQ settings | 10 10k-files | Good run, 4“ polyethylene shield detector |
| 07100729_2011 | Am-241 & Cf-252 | DAQ settings | 11 10k-files | Good run, 8” polyethylene shield detector |
| 07100729_2138 | Am-241 & Cf-252 | DAQ settings | 12 10k-files | Good run, 2“ polyethylene shield detector |
| 07100729_2305 | Am-241 & Cf-252 | DAQ settings | 10 10k-files | Good run, no polyethylene shield detector |
| 07100730_1700 | Am-241 & Ba-133 | DAQ settings | 11 10k-files | Small 356 keV peak as Ba-133 source too far from detector |
| 07100730_2024 | Am-241 & Ba-133 | DAQ settings | 11 10k-files | Small 356 keV peak, Ba-133 source 4” below top of metal shield |
| 07100730_2239 | Am-241 & Ba-133 | DAQ settings | 24 10k-files | Small 60 keV peak, Ba-133 source 22“ below top of metal shield |
| 07100731_0111 | Am-241 & Ba-133 | DAQ settings | 86 10k-files | Good run, overnight, Ba-133 source 13” below top of metal shield |
Run 3
| Data start | Data end | Comment |
|---|---|---|
| 07100506_1140_F0001 | 07100506_1140_F0004 | 9 Files with 10k events. DCRCv2: Q only, +3V bias, -1V Q Offset, Qthresh 1000 ADCs, Qi enabled in trigger. 137Cs Source. ~2 hours |
| 07100506_1517_F0001 | 07100506_1517_F0004 | 4 Files with 10k events. DCRCv2: Q only, +3V bias, -1V Q Offset, QThresh 30 ADCs, Qi and Qo enabled in trigger. No source. ~60 minutes |
| 07100506_1653_F0001 | 07100506_1653_F0006 | 6 Files with 10k events. DCRCv2: Q only, +3V bias, -0V Q Offset, QThresh 30 ADCs, Qi and Qo enabled in trigger. 133Ba Source. ~60 minutes |
| 07100506_1953_F0001 | 07100506_1953_F0003 | 3 Files with 10k events. DCRCv1: All channels, +3V bias, QThresh 30 ADCs, Qi and Qo enabled in trigger. 133Ba Source. ~20 minutes |
| 07100507_1209 | 07100507_1209 | 19 Files, DCRC v2 charge, 133Ba source. |
| 07100516_0115 | 07100516_0115 | 55 10K files. DCRC v1, 133Ba source. DAQ settings |
Run 2
| Data Start | Data End | Comment |
|---|---|---|
| 07100312_1454_F0001 | 07100312_2110_F0008 | 100 Files. DCRC: Preset 2; LED1: , I=400uA, ON=100us, OFF=2500us; Data taking: 3min, Flash:3min; DAQSettings |
| 07100312_2258_F0001 | 07100313_0108_F0035 | 137 Files. DCRC: Preset 2;LED1: , I=400uA, ON=100us, OFF=2500us; Data taking: 15min, Flash:3min; DAQSettings |
| 07100313_1619_F0001 | 07100313_1817_F0070 | 211 files. DCRC: Preset 2. LED1: , I=400uA, ON=100us, OFF=2500us; Data taking: 30min, Flash:3min; DAQSettings |
| 07100314_1819_F0001 | 07100314_1819_F0142 | 142 Files. DCRC: Preset 2. LED1: , I=400uA, ON=100us, OFF=2500us; Data taking: 60min, Flash:3min; DAQSettings |
| 07100315_1212_F0001 | 07100315_1329_F0013 | 78 files. DCRC : Preset 2. LED1: , I=1000uA, ON=100us, OFF=2500us; Data taking: 5min, Flash: 0.2min; DAQSettings |
| 07100317_1014_F0001 | 07100317_1124_F0056 | 200 Files. DCRC: Preset 2. LED1: , I=1000uA, ON=100us, OFF=2500us; Data taking: 60min, Flash: 0.2min; DAQSettings |