start › cdms › k100 › run_diary › 20150908

2015/09/08 (Tuesday)

previous day

Current setup: DIB1/2/3 → cables → 3.2 MHz filters → HV1/2 → DCRC50/55. Unused LED is grounded.
Overweekend, Cf source in shield activating detector
0930 Fill LN2 cold trap full.
0935 Tower/MC(H35)/MC sensor/Cold Plate = 30.8/25.9/40.0/151 mK, Still = 1.38 K / 0.075 mbar, G1/G2 = 145 mbar, 1K Pot = 1.74 K / 24 mbar, Main bath = 68.5%.
Remove Cf-252 source from fridge.
0945 Mini-bake detector (4 LEDs - two from each DCRC).
1000 Notice that FETs are off on HV2. Power them up.
Realize that HV bias was still applied during mini-bake. Turn HV bias to 0V and mini-bake again.
1035 At 0V HV bias, set up phonon thresholds.
Phonon thresholds
pA pB pC pD
HV1 45 35 75 -
HV2 - 45 45 25
1040 Still seeing ~80Hz noise on HV2pB with amplitude of twice white noise. (This seems to be a DCRC55pB problem, see tomorrow's diary entry)
Apply 0V HV bias to crystal (leakage current of 1.5nA). Start 10 minute data series.
1057 Stop Data. Flash 30s.
1115 Apply +5V to crystal (leakage current of 7.3nA). No noticeable shift in phonon baselines. Start 10 minute data series.
1129 Stop data. Zero HV bias. Flash 30s.
1151 Apply +10V to crystal (leakage current of ~15.4nA). Start 10 minute data series.
1205 Stop data. Zero HV bias. Flash 30s.
1220 At ~12V, HV2pB starts showing upside down pulses, and HV2pC needed a relock.
Set HV=11.5V and allow signals to “settle.”
Adjust driver offsets to place baselines at ~-0.7V.
1225 Upside down pulses still on HV2pB, but other HV2 pulses are essentially gone (breakdown). HV1 still showing good pulses.
Apply +11.5V to crystal (leakage current of 16.1nA). Start 10 minute data series.
1239 Stop data. Zero HV bias. Flash 30s.
1245 Disconnect HV Box and DCRCs.
1255 Set up iZIP mode, DCRC55=DIB1, DCRC33=DIB2, DCRC50=DIB3 in the configuration: EBox → Cable → Gender Swap → 3.2MHz Filter → DCRC
Power DCRCs, zap SQUIDs.
1330 Several channels railed and noise looks larger and spikier than usual. Check cable connections.
1550 Stop DCRC50 and power cycle to see if noise goes away.
1612 No more spiky noise on DIB3. Set up channels with FG=16/5/16 (for DIBs 1/2/3), OG=3. Adjust Driver Offsets to place phonon baselines at ~-0.7V Set phonon thresholds below:
Phonon thresholds
pA pB pC pD
DIB1 50 50 50 55
DIB2 45 50 45 60
DIB3 55 65 50 55
1650 Create 0V MultiDAQ file from above settings. Create -/+2V file as well.
1715 Set DAQDAQ to take 8 cycles of 15 minute data series with 30s flash and 30 minute cooldown in between series. 1000 initial randoms. 2000 events/file. OR trigger across all phonon channels. Start data series. No phonon baselines shifted or railed.
1720 Fill LN2 cold trap full.