start › cdms › k100 › run_diary › 20151020

2015/10/20 (Tuesday)

Previous Day

Current setup: iZIP mode, DIB1/2/3 → cables → 3.2 MHz filters → DCRC55/33/50.
Overnight: Taking background data. Last flash at 6:45. Tower temp stable.
0910 Tower/MC(H35)/MC sensor/Cold Plate = 31.1/25.7/41.7/157 mK, Still = 1.34 K / 0.095 mbar, G1/G2 = 105/105 mbar, 1K Pot = 1.63 K / 6 mbar, Main bath = 43.7%
1K Pot temperature and pressure and circulation pressures significantly lower than last week. Probably the pot behavior changed when the bath got very low yesterday.
0920 Start LHe transfer at 43.7%
0950 Finish LHe transfer at 88.9%
1022 Place Ba source behind fridge inside top most part of shield.
1059 Set phonon thresholds to yesterday's values.
1101 Start Qbias sweep Ba data 3 cycles of 30 minute series, FG=5/2/5, OG=3, with 1000 initial randoms, 2000 events/file, 30s flash and 25min cooldown in between. The Qbias was swept in the following order: +/-2V, +/-4V, +/-6V, +/-8V, 0V, -/+6V, -/+8V.
1338 During the third series (+/-6V), the event rate was noticed to oscillate between the typical 15Hz and 1Hz. This was only noticed right at the end of the series, so it's unclear how long this was happening. The data directory for this series only contains 5 files instead of the 12 that are in the directories for the first two series, though. Pull the Ba source slightly higher (directly below the very top of the shield) in hopes of decreasing the strain on the computer.
1410 During fourth data series (+/-8V), the rate drops low (~2Hz). Modify all phonon trigger thresholds to be 10ADC bins higher. This brings the event rate back up to 15Hz. The new phonon thresholds are below:
A B C D
DIB1 46 50 52 50
DIB2 34 28 30 38
DIB3 50 56 68 58
1635 During cooldown, fill LN2 cold trap full.