| 1240 | Trouble with DIB2pB, requires many relocks. Reset DIB2pB driver offset to 0.034 V. |
| 1250 | DIB3pC looks very fuzzy (note high threshold above) |
| 1420 | Troubles with DIB3pC. It either rails or pulses go upside-down: relock fixes it but problem returns. Played with adjusting settings, found ranges at which pulses still present: Lock Point from 0-5 (default 3.02), SQUID bias from 36 to 86 (default 66), QET bias from -90 to -170 (default -108). No help. |
| 1440 | Power cycle DCRC33 and zap SQUIDs. This seems to fix the DIB2pB problem, pulses remain upright, no railing. |
| 1455 | Take data - aborted series. Randoms inadvertently set to 10000. Zero Q bias, flash 30 seconds. |
| 1518 | Take data. -4/+4 V bias. Sample background data after overnight activate/bake, 15 min. Output gains at setting=3, Front-end gains at 1. |
| 1536 | Stop data, zero Q, flash 30 seconds |
| 1555 | Change Front-end Gains to 16/5/16 for DIB1/2/3. |
| 1556 | Adjust driver offset for DIB3pC to 0.024 because channel railed. |
| 1600 | Take data. -4/+4 V bias. Noise test with non-zero front-end gains, 1000 randoms. Output gains at setting=3, Front-end gains at 16/5/16 for DIB1/2/3. |
| 1604 | Data rate bogs down around 800 randoms - presumably phonon thresholds are too low for adjusted gains. Never mind, 800 randoms is enough. Zero Q, flash 30 sec |
| 1610 | Reduce front-end gains to 1 in preparation for a DAQDAQ test, automated multi-series data taking. |
| 1640 | Save four MultiDAQ files, with -4/+4 V versus 0 V and Front-end Gains = 1 versus >1. Output gains at setting=3. Attempt DAQDAQ2 with 1000 initial randoms, 2000 events/file, 10 minute runs, 3 series of 4 DAQ files (all the same file), 6 seconds flashing / 20 min cooldown between series, phonon OR trigger |
| 1642 | Take data, test of DAQDAQ2. No railings or shifted baselines, data rate fine (about 14 Hz) |
| 1715 | Previous test did not involve MultiDAQ restarts because DAQDAQ was doing 3 series/file. Stop data, new test with 1 series/file and 3 cycles, which should close MultiDAQ between series. |
| 1717 | Take data, another test of DAQDAQ2. Again, all looks fine. |
| 1755 | Fill LN cold trap Dewar |
| 1800 | Stop data, start baking detector. |
| 1805 | Place Cf source in shield near detector for overnight activation |