UMN CDMS Wiki archive
start › cdms › k100 › run_summary › run_31

Run 31

Summary

Setup

G103a

Figure 1: G103a Packaged at SLAC.
g103a_side1c_packaged.jpg g103a_side2a_packaged.jpg
Figure 2: G103a DIB Layout.

Accelerometer Data

Data available in csv format from here.

Table 1: Vertical red lines in Figures 1, 2 & 3 correspond to the following events from FedEx's website in chronological order (tracking no. 793712429915):
Date Time [PDT] Activity Location
June 22 2012 13:40 Picked up Menlo Park, CA
June 22 2012 17:55 Left origin FedEx location Menlo Park, CA
June 22 2012 22:51 Arrived at FedEx location Memphis, TN
June 23 2012 02:19 Departed FedEx location Memphis, TN
June 23 2012 04:04 At destination sort facility Minneapolis, MN
June 23 2012 04:51 At local FedEx facility St. Paul, MN
June 25 2012 05:00 On FedEx vehicle for delivery St. Paul, MN
June 25 2012 09:23 Delivered Minneapolis, MN
Figure 3: Pressure exerted on puck during its journey across the country as measured by the accelerometer.
Figure 4: Temperature of puck over the course of its voyage across the country as measured by the accelerometer.
Figure 5: G-force sustained by puck during its traumatic odyssey as measured by the accelerometer. The lines between the points have no meaning as the accelerometer recorded the acceleration when its trigger threshold of 2g was satisfied. They are present here to help your eyes.

Tower Assembly & Electrical Setup

Figure 6: Assembled tower.
Figure 7: New adapter plate with space for tower themometer wire.
Figure 8: Tower attached to 30 mK stage.
Table 2: Run 31 setup
Detector G103a
Tower T1
DIB 1 2 3
Face 2 5 6
Side coax ZB1-1-6 ZB1-1-9 ZB1-1-11
SQUET ZC23 ZC28 ZC70
Stripline #19 #20 #21
Readout DCRCv3

Source plate is same as that used in previous 100 mm crystal runs. One Am-241 source facing Side 2 in Q2 position. This means that source is close to the centre of Side 2 Phonon Channel E with reference to Fig. 2.

Electrical Problems

Short to ground on SCUHBA extender plug PCB

After assembling the tower and conducting a tower level continuity check, a short to ground was found. Through a process of elimination, the short was found on the new SCUHBA extender plug PCB. More specifically, pins 7 & 8 were shorted to ground. Running the soldering iron across the masks broke the connection to ground. The results of the tower level continuity checks can be found here (run diary page: 2012/06/29).

Shorts to ground when conducting electrical checks after tower installation

Found a number of shorts to ground after tower installation (QET bias, QET return and Qi & Qo bias). The shorts appeared to capacitive in nature, ie. the resistance increased in time when measuring with the multimeter. When conducting these checks, one appears to be charging a very large capacitor. Additionally, there is a potential drop across the shorted channels.

Stripline 20 and SQUET card ZC-28 have additional shorts between SQUID bias B, C & D to ground, and Feedback B & C to ground. This is probably not surprising, as SQUET card ZC-28 failed SQUID isolation tests as documented in SQUET elog entry 967.

While attempting to determine the source of these shorts, some pins on the ebox feedthroughs were found to have resistances of ∼100 MΩ to ground. Flooding the sockets with methanol, wiping with a paint brush and blow drying with nitrogen gas was found to be an effective way to increase the isolation of these pins from ground. Unfortunately, this did not appear to have an effect on the short.

The details of electrical checks at various temperatures during cool down are detailed here (run diary page: 2012/07/05). The resistances of the shorts are probably sufficiently high enough at liquid helium temperatures not to cause any problems during data acquisition.

Cryogenic Problems

Fixed mixing chamber thermometer, broke tower thermometer

During the upgrade and maintenance work, no fault was seen when conducting room temperature checks on the mixing chamber thermometer. However, as it is known that there is a fault at low temperatures, the connector was resoldered. This thermometer now works at base temperature.

During reassembly of the fridge, the insulation of wire leading to the tower thermometer was severed. As there is no replacement cable, a repair was attempted. Unfortunately, the repair appears to have been unsuccessful, and the tower thermometer is not readible. It is possible that during tower installation, the tower thermometer was not connected sufficiently well. However, it is more than likely that the problem lies further up and will require a new cable.

Therefore, our base temperature during this run is quoted from the mixing chamber thermometer. The tower thermometer is likely to be higher than this by a few mK. The actual difference in temperature is difficult to ascertain, as there has been no run when both of these thermometers were working.

Initial Leak

During the initial cooldown, two cryoleaks were found. A small leak of the order of 1E-6 mbar l/s was found in one of the indium seals towards the top of the IVC. This was slightly surprising as these flanges had been tested for leaks at liquid nitrogen temperatures specifically during Run 30. From past experience, it is possible to run for a short time with this level of a leak.

However, a larger leak of the order of 1E-2 mbar l/s was detected towards the end of nitrogen blow out. This indicated that the bottom IVC flange had a gross leak. This was also a little surprising as it was thought the reasons for the leak lay in the orientation of the flange. After removing the flange, it was found that too much vacuum grease had been used.

After tightening the flanges at the top of the IVC and replacing the indium seal at the bottom, the second cooldown was successful.

Poor Initial Base Temperature

Unfortunately, a pretty high base temperature of 89 mK was reached. Cool down itself was rather slow. As stated above, this is from the mixing chamber thermometer, and the tower temperature is likely to be higher. Locking the DIB 1 SQUIDs and attempting to find the Tcs by increasing power to the mixing chamber heater indicated that the base temperature was above the critical temperature of the phonon channels.

A single shot test was conducted to verify this, and to determine the cooling power of the fridge. The results are shown in Fig. 9. Tcs were found to be lie at around 80 mK, as detailed in 'Tc Measurements' section below.

Figure 9: Dump Pressure (G2) and Mixing Chamber Temperature as a function of Time during Single-Shot.

As expected, the cooling power was pathetic. The mixture was extracted from the fridge in preparation for cleaning the cold traps. Puzzlingly, 592 mbar of mixture was recovered, yet 605 mbar was thrown into the fridge. No leak from the mixing chamber lines to the IVC was found during the throughput test at 77 K. Pressing on, the cold traps were cleaned and the mixture was recondensed. A lower base temperature of 63 mK was achieved.

Whether this had anything to do with cleaning the cold traps is unknown. The traps were dirty, but during this cycle some of the mixture has been left in the dump. The total proportion of mixture left in the dump is unknown, but can be determined during warm up. It is possible that this is the reason for the lower base temperature.

Initial Look at the Phonon Channels

Tc Measurements

Critical temperature measurements of the TESs are shown in Table 3. Measurements for DIB 1 channels were taken when cooling down during single shot test detailed above. This value is quoted as the fridge cooled at a slower rate, thus providing a more accurate reading (please see Fig. 9 for details). The Tcs for phonon channels on DIBs 2 & 3 were measured after a new, lower base temperature had been established. These measurements were performed by supplying power to the mixing chamber heater to gradually increase the temperature.

It should be noted that the Tcs in Table 3 were measured with the mixing chamber thermometer rather than the tower thermometer. As mentioned above, the tower thermometer was unavailable during this run. Therefore, the true Tcs may lie at higher temperatures than those determined.

Table 3: Tc Measurements (using mixing chamber thermometer).
Side & Channel DIB Channel Tc [mK] ± 1 mK
Side 1 A D2 76
Side 1 B A1 83
Side 1 C C1 84
Side 1 D C2 no result*
Side 1 E C3 85
Side 1 F B1 83
Side 2 A B2 86
Side 2 B A3 82
Side 2 C A2 78
Side 2 D D1 79
Side 2 E D3 81
Side 2 F B3 76

* Could not determine transition for Side 1 Channel D.

Problems with Side 1 Phonon Channel D (C2)

Although it was possible to lock the SQUIDs on Side 1 Phonon Channel D (C2), a relatively large QET bias of 570 μA had to be applied to see any pulses. This should be compared with QET biases between 60 and 110 μA for the other phonon channels. Additionally, the pulses are of a different shape than expected. Please see Fig. 10 for an example.

Figure 10: Typical Pulses from DIB 2's Phonon Channels.
Side 2 Channel C (A2) Side 2 Channel A (B2)
pha_dib2.jpg phb_dib2.jpg
Side 1 Channel D (C2) Side 1 Channel A (D2)
phc_dib2.jpg phd_dib2.jpg

Gamma Shield

A full write-up is available: Testing the Poly-Lead Shield with S34 Background Gamma Spectra.

Figure 11: Shield Drawings (left) and Shield installed around HPGe Crystal (right).
Figure 12: Comparing Gamma Spectrum of S34 Background with full shield and with no shield (07/06/2012 and 16/06/2012)

Charge Spectra Inventory

This is an inventory of all the charge spectra collected thus far that would be useful for analyses such as the seagull plot. This also informs us for which biases we are missing good quality data. For series that are boxed together, those series are a half hour long. The series on 2012/07/21 and 2012/07/22 was also recorded over a half hour.

Charge Data Analysis

The data below is ordered by the bias the data series was taken at. If the bias is designated “ -/+ 2V ”, this means that Dib1 is negatively biased at 2V and Dib2 is positively biased at 2V. “ +/- 2V ” indicates that the Dibs have their polarity reversed. To know which board is associated with which Dib, the board configuration is written as well.

The biases shown below are (in order):
-/+ 2V
+/- 2V
-/+ 4V
+/- 4V
-/+ 6V
+/- 6V
+/- 8V
+/- 10V
-/+ 1V
-/+ 3V
The spectra are not calibrated. We scaled up the charge energy by a factor of 1000. QIOFnoXvolts, etc represents the charge energy at the feedback capacitor.

Bias Series Board Configuration Trigger Data Run Time Length
-/+ 2V 07120727_0436 Dib1=DCRC 33, Dib2= DCRC32, Dib3=DCRC31 12 phonon channels 2 hours
Fig 1: First 15 Minutes First 15 Minutes
Bias Series Board Configuration Trigger Data Run Time Length
+/- 2V 07120724_2334 Dib1=DCRC 33, Dib2= DCRC32 12 phonon channels 2 hours
Fig 2: Full Series/ First 15 Minutes Charge vs. Time
Bias Series Board Configuration Trigger Time Length
-/+ 4V 07120728_0737 Dib1=DCRC 33, Dib2= DCRC32, Dib3=DCRC31 Phonon channels 2 hours
Fig 3: First 15 minutes Charge vs. Time
Bias Series Board Configuration Trigger Time Length
+/- 4V 07120727_0658 Dib1=DCRC 33, Dib2= DCRC32, Dib3=DCRC31 12 phonon channels 2 hours
Fig 4: First 15 Minutes Full Series/ First 15 Minutes
Bias Series Board Configuration Trigger Time Length
-/+ 6V 07120725_1134 Dib1=DCRC 33, Dib2= DCRC32 Phonon channels 2 hours
Fig 5: Full Series/ First 15 Minutes Charge vs. Time
Bias Series Board Configuration Trigger Time Length
+/- 6V 07120727_1140 Dib1=DCRC 33, Dib2= DCRC32, Dib3=DCRC31 12 phonon channels 2 hours
Fig 6: First 15 Minutes Full Series/ First 15 Minutes

Anything below is tentative

Bias Series Board Configuration Trigger Time Length
+/- 8V 07120807_1937
07120807_2027
07120807_2117
Dib1=DCRC 33, Dib2= DCRC32 Charge channels 30 min/series
Fig 7: Full Series/ First 15 Minutes Charge vs. Time
Bias Series Board Configuration Trigger Time Length
+/- 10V 07120807_2210
07120807_2300
07120807_2350
Dib1=DCRC 33, Dib2= DCRC32 Charge channels 30 min/series
Fig 8: Full Series/ First 15 Minutes Charge vs. Time
Bias Series Board Configuration Trigger Time Length
-/+ 1V 07120808_0313
07120808_0403
07120808_0453
Dib1=DCRC 33, Dib2= DCRC32 Charge channels 30 min/series
Fig 9: Full Series/ First 15 Minutes Charge vs. Time
Bias Series Board Configuration Trigger Time Length
-/+ 3V 07120808_0544
07120808_0635
07120808_0725
Dib1=DCRC 33, Dib2= DCRC32 Charge channels 30 min/series
Fig 10: Full Series/ First 15 Minutes Charge vs. Time

Phonon Pulses and Analysis

Tc Measurements as of 2012/08/08

Tc of Phonon channels on DIB1/DCRC33 when MC T rising at 2mK/min
A B C D
81.46 81.84 82.3 74.3
82.1 83 82.9 75
Temperature dropping at 2mK/min
82.1 83.7 83 74.5
81.6 82 82.4
Tc of Phonon channels on DIB2/DCRC32 when MC T rising at 2mK/min
A B C D
77 86.2 79.5 73.6
78.3 87.1 74.7
Temperature dropping at 2mK/min
78.7 87.2 80.1 74.5
76.9 86.4 79.9 73.5
Tc of Phonon channels on DIB3/DCRC31 when MC T rising at 2mK/min
A B C D
79.7 75.6 84.5 79.7
81 76.5 85.2 81
Temperature dropping at 2mK/min
81 76.7 85.2 81.1
80.2 75.7 84.8 79.8

IbIs

Figure 1: IbIs measurement for C1 (DCRC 33 plugged into Dib 1)

Figure 2: IbIs Measurements for Channel C2
Figure 3: IbIs Measurements for Channel D2 (right).

Phonon Pulses from 12 Channels

First thing to do is to plot raw pulses and fit a double exponential function to them. F=A * { Exp [ - ( t - t0 ) / t2 ] - Exp [ - ( t - t0 ) / t3 ] } + B.

Here A is the amplitude, t is time, t0 is the starting point of the pulse, t2 is the fall time constant, t3 is the rise time constant, B is the offset.

Series:07120727_1140
Blue = channel A; Green = channel B; Red = channel C; Light Blue = channel D; ms=microseconds

Chi^2 cuts

The next step is to select only good pulses. We look at Chi^2 distribution for all 12 channels. Below is an example of such distribution for one of the channels. (the only difference in the pictures is the scale)

Different scale

Histograms

ms=microseconds ms=microseconds
ms=microseconds ms=microseconds
ms=microseconds ms=microseconds
ms=microseconds ms=microseconds

Fall Time Constant vs Phonon energy

Below are the plots of the Fall Time Constant vs Phonon energy deposited into the channel. There are two plots for channel B1(second picture clearly indicates independence of the FTC).

Not all channels are included. Others don't have enough statistics. The conclusion one can make is that FTC is independent of the energy.

DIB1
[image not archived: ftc_vs_energy_d1_.png]
ms=microseconds
ms=microseconds
DIB2
ms=microseconds
DIB3
ms=microseconds

Rise Time Constant vs Phonon energy

It's harder to make conclusions about the dependence of RTC on energy. Some of the plots show independence(D1,D2) while others show inverse relation. More investigation is needed.

DIB1
DIB2
DIB3

As a comparison here is a similar G42 analysis