2012/07/05
| 1140 | Filled LN2 cold trap. |
|---|---|
| 1320 | Conduct electrical check at 5.4 K: |
| Detector: | G103a DIB1 | BH: | Stripline: | 19 | ||||||||
| SC: | ZB1-1-6 | 300 K | LN | LHe | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 2 | Date | 2012-07-03 | Date | 2012-07-04 | Date | 2012-07-05 | |||||
| SQUET: | ZC23 | Who | AK & RR & DS | Who | HC & AK | Who | DS & AK | |||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis | |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | ||||||||||
| Nominal: | Nominal: | Nominal: | ||||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | ||||||||||
| Phonon | ||||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.4 | OK | 1 - 2 k | 1.7 | OK | 80 - 140 | 106 | OK | |
| SQUID bias B/Ret | 19 | 34 | 5 | OK | 2.1 | OK | 80 - 140 | 98 | OK | |||
| SQUID bias C/Ret | 8 | 24 | 4.6 | OK | 1.8 | OK | 80 - 140 | 104 | OK | |||
| SQUID bias D/Ret | 23 | 39 | 4.3 | OK | 1.6 | OK | 80 - 140 | 94 | OK | |||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 18 | OK | 9 - 13 k | 7.5 | OK | 1.0 - 1.3 k | 1.26 | OK | |
| Feedback B/Ret | 2 | 18 | 20 | OK | 7.5 | OK | 1.0 - 1.3 k | 1.28 | OK | |||
| Feedback C/Ret | 41 | 25 | 19 | OK | 7.9 | OK | 1.0 - 1.3 k | 1.26 | OK | |||
| Feedback D/Ret | 22 | 6 | 18 | OK | 7.3 | OK | 1.0 - 1.3 k | 1.25 | OK | |||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 120, 103, 92, 96, 93 | 15 | 30 - 40 | 80,63,52,56,53 | 10 | 20 - 30 | 73, 58, 44, 47, 44 | 2.5 | |
| QET B/Ret | 35 | 29, 33, 47, 48 | 92, 91, 106, 92 | 13 | 52,51,67,52 | 10 | 20 - 30 | 38, 37, 53, 37 | OK | |||
| QET C/Ret | 7 | 5, 9, 10, 28 | 97, 86, 75, 85 | 12 | 58,47,36,46 | 10 | 20 - 30 | 45, 33, 22, 32 | 2.5 | |||
| QET D/Ret | 38 | 36, 40, 43, 44 | 99,85, 85, 85 | 12 | 60,47,46,47 | 10 | 20 - 30 | 47, 34, 33, 33 | 3 | |||
| Charge | ||||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.068 | OK | 1.0 - 1.1 k | 1.045 | OK | 1.0 - 1.1 k | 1.26 | OK | |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 27 | 12 | < 1 nS | 27 | 7 | <1 nS | 1.8 | OK | ||
| Qo bias/QET Ret | 11 | 27 | 12 | 27 | 7 | <1 nS | 0.5 | OK | ||||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 24, 24 | OK | 6 - 8 nS | 21,21 | OK | <1 nS | 19, 19 | OK | |
| Qo feedback/Drain/Source | 30 | 12, 27 | 24, 24 | OK | 21,21 | OK | <1 nS | 20, 20 | OK | |||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK | |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | |||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | 0.725 | OK | 0.9 - 1.0 V | 1.092 | OK | — | OK | ||
| LED 1 | chassis | 42 | 1.0 - 1.1 V | OPEN | – | 1.2 - 1.5 V | open | – | — | – | ||
| LED 2 | chassis | 26 | 1.11 | – | 1.379 | – | — | – | ||||
| Detector: | G103a DIB2 | BH: | Stripline: | 20 | |||||||
| SC: | ZB1-1-9 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 5 | Date | 2012-07-03 | Date | 2012-07-04 | Date | 2012-07-05 | ||||
| SQUET: | ZC28 | Who | AK & RR & DS | Who | HC & AK | Who | DS & AK | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.0 k | OK | 1 - 2 k | 1.6 | OK | 80 - 140 | 113 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.3 k | 12 nS | 1.6 | 9 nS | 80 - 140 | 97 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.9 k | 12 nS | 1.9 | OK | 80 - 140 | 101 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 4.3 k | 10 nS | 1.6 | OK | 80 - 140 | 88 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 23.7 k | OK | 9 - 13 k | 8.7 | OK | 1.0 - 1.3 k | 1.25 | OK |
| Feedback B/Ret | 2 | 18 | 17.4 k | 12 nS | 7.4 | 9 nS | 1.0 - 1.3 k | 1.27 | OK | ||
| Feedback C/Ret | 41 | 25 | 20.7 k | 11 nS | 8.8 | OK | 1.0 - 1.3 k | 1.38 | OK | ||
| Feedback D/Ret | 22 | 6 | 18.3 k | OK | 7.4 | OK | 1.0 - 1.3 k | 1.25 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 115, 115, 81,85,82 | 13 nS | 30 - 40 | 72,60,43,46,43 | 9 nS | 20 - 30 | 65, 50, 35, 39, 35 | 7 |
| QET B/Ret | 35 | 29, 33, 47, 48 | 76, 75, 108, 76 | 11 nS | 43,42,57,42 | 9 nS | 20 - 30 | 31, 30, 46, 30 | 5 | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 85, 73, 65, 72 | 11 nS | 49,37,30,37 | 9 nS | 20 - 30 | 37, 24, 16, 24 | OK | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 93, 72, 71, 71 | 15 nS | 51,39,38,38 | 9 nS | 20 - 30 | 40, 27, 26, 26 | OK | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.06 | OK | 1.0 - 1.1 k | 1.051 | OK | 1.0 - 1.1 k | 1.23 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | 26 | 12 | < 1 nS | 26 | 7 nS | <1 nS | 3 | 6 | |
| Qo bias/QET Ret | 11 | 25 | 12 | 26 | 7 nS | <1 nS | 0.4 | OK | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23.4, 23.5 | 13 nS | 6 - 8 nS | 21,21 | OK | <1 nS | 18, 18 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23.2, 23,2 | 12 nS | 20,20 | OK | <1 nS | 19, 19 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | .71 V | OK | 0.9 - 1.0 V | 1.083 | OK | — | OK | |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.11 V | – | 1.2 - 1.5 V | 1.385 | – | — | – | |
| LED 2 | chassis | 26 | 1.11 V | – | 1.376 | – | — | – | |||
| Detector: | G103a DIB3 | BH: | Stripline: | 21 | |||||||
| SC: | ZB1-1-11 | 300 K | LN | LHe | |||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Face: | 6 | Date | 2012-07-03 | Date | 2012-07-04 | Date | 2012-07-05 | ||||
| SQUET: | ZC70 | Who | AK & RR & DS | Who | HC & AK | Who | DS&AK | ||||
| Signal | Pos. lead | Gnd lead | Nominal | Actual | Chassis | Nominal | Actual | Chassis | Nominal | Actual | Chassis |
| (Gnd lead) | (Gnd lead) | (Gnd lead) | |||||||||
| Nominal: | Nominal: | Nominal: | |||||||||
| < 0.1 nS | < 0.1 nS | < 0.1 nS | |||||||||
| Phonon | |||||||||||
| SQUID bias A/Ret | 4 | 20 | 3 - 5 k | 4.8k | OK | 1 - 2 k | 2 | OK | 80 - 140 | 100 | OK |
| SQUID bias B/Ret | 19 | 34 | 4.4k | OK | 1.8 | OK | 80 - 140 | 114 | OK | ||
| SQUID bias C/Ret | 8 | 24 | 4.6k | OK | 1.9 | OK | 80 - 140 | 98 | OK | ||
| SQUID bias D/Ret | 23 | 39 | 5.2k | OK | 2.1 | OK | 80 - 140 | 96 | OK | ||
| Feedback A/Ret | 37 | 21 | 15 - 25 k | 29k | OK | 9 - 13 k | 15 | OK | 1.0 - 1.3 k | 1.25 | OK |
| Feedback B/Ret | 2 | 18 | 25k | OK | 9.5 | OK | 1.0 - 1.3 k | 1.27 | OK | ||
| Feedback C/Ret | 41 | 25 | 28k | OK | 14 | OK | 1.0 - 1.3 k | 1.26 | OK | ||
| Feedback D/Ret | 22 | 6 | 34k | OK | 18 | OK | 1.0 - 1.3 k | 1.24 | OK | ||
| QET A/Ret | 3 | 1, 13, 14, 17, 32 | 60 - 80 | 120, 108, 94, 98, 95 | 20 nS | 30 - 40 | 81,67,54,57,54 | 10 nS | 20 - 30 | 70, 54, 38, 42, 39 | 10nS |
| QET B/Ret | 35 | 29, 33, 47, 48 | 88, 87, 103, 87 | 15nS | 47,46,62,47 | 10 nS | 20 - 30 | 33, 32, 48, 32 | 11nS | ||
| QET C/Ret | 7 | 5, 9, 10, 28 | 102, 85, 76, 84 | 12 nS | 56,46,37,46 | 10 nS | 20 - 30 | 41, 27, 18, 27 | 3nS | ||
| QET D/Ret | 38 | 36, 40, 43, 44 | 102, 87, 86, 87 | 13 nS | 61,48,48,48 | 10 nS | 20 - 30 | 43, 29, 28, 28 | 5nS | ||
| Charge | |||||||||||
| FET heater +/- | 45 | 46 | 1.0 - 1.1 k | 1.065 | OK | 1.0 - 1.1 k | 1.042 | OK | 1.0 - 1.1 k | 1.24 | OK |
| Qi bias/QET Ret | 15 | 24 - 26 nS | OPEN | OK | < 1 nS | open | OK | <1 nS | 0.51 | OK | |
| Qo bias/QET Ret | 11 | OPEN | OK | open | OK | <1 nS | 2 | 3nS | |||
| Qi feedback/Drain/Source | 50 | 16, 31 | 15 - 17 nS | 23, 23 | OK | 6 - 8 nS | 20,20 | OK | <1 nS | 18.5, 18.4 | OK |
| Qo feedback/Drain/Source | 30 | 12, 27 | 23, 23 | OK | 20,20 | OK | <1 nS | 19.3, 19.1 | OK | ||
| Qi drain | 16 | chassis | < 0.1 nS | – | OK | < 0.1 nS | – | OK | <0.1 nS | – | OK |
| Qo drain | 12 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qi source | 31 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| Qo source | 27 | chassis | – | OK | – | OK | <0.1 nS | – | OK | ||
| FET Tsense +/FET heater- | 49 | 46 | 0.6 - 0.7 V | OPEN | OK | 0.9 - 1.0 V | open | OK | — | – | OK |
| LED 1 | chassis | 42 | 1.0 - 1.1 V | 1.11V | – | 1.2 - 1.5 V | 1.384 | – | — | – | – |
| LED 2 | chassis | 26 | OPEN | – | open | – | — | – | – | ||
| 1500 | Start 1 K pot. Start condensing mixture through nitrogen cold trap 1. |
|---|---|
| 1630 | Majority of mixture has condensed. Start circulating mixture. |
| Quick survey on base temperature: |
| Table 1: What base temperature will we reach? | |
|---|---|
| Chagani | 90 mK |
| Kennedy | 81 mK |
| Radpour | 70 mK |
| Monin | 79 mK |
| Strandberg | 80 mK |
| 1730 | Valve 6 fully open and dump has been bled dry. Circulation procedure complete. Tower is cooling very slowly. The base temperature is still decreasing though. It may be best to wait until it reaches a stable temperature before making any futher comment. |
|---|---|
| 2155 | Base temperature currently 90.6 mK. It had been stuck at 91.3 mK for about an hour, but appears to have started to drop again, albeit very slowly. There is a significantly large amount of mass to cool down, so we could simply wait until tomorrow morning to see what temperature we've reached. No harm in taking a look at the other temperature stages though. |
| Pressures of G1 & G2 are pretty low at 49 mbar & 58 mbar respectively. Pressure of P1 is 0.065 mbar. It may be a good idea to start with the still. | |
| 2220 | Resistances of temperature stage thermometers: |
| Table 2: Resistance of temperature stage thermometers. | |||
|---|---|---|---|
| Stage | Measured Resistance [kΩ] | Resistance during Run 29 [kΩ] | Nominal Resistance from manual [kΩ] |
| Still | 4.52 | 4.5 | 4.53 |
| Cold Plate | 6.1 | 15.5 | 23.4 |
| Mixing Chamber | 22.5 | 25 | 134 |
| 2230 | Cold plate appears to be very warm, which is very surprising considering that we have reached a similar base temperature as that in Run 29 (90 mK vs. 84 mK). Temperature of still is where it should be. |
|---|---|
| 2305 | Still heater has been on at power 1 mW for the last 20 minutes. Mixing chamber temperature has risen to 99 mK. It was unlikely to work as this doesn't appear to be the problem. Turn still heater off. |