UMN CDMS Wiki archive
start › cdms › k100 › meetings › 20190325
Teleconference link: https://fnal.zoom.us/j/9082275310
Domestic: 1-646-558-8656 (US toll)
Domestic: 1-408-638-0968 (US toll)
Access code: 908-227-5310

Agenda

  • Run 69 (GeHV n,gamma NR calibration) status
    • Tower has cooled to around 42 mK
    • Sten will send back DCRC55(1) (we've been doing OK without it though. Should ask him again in a couple weeks if he has time to look at it.)
    • Data:
      • HV tests: seems ok at least to +440V - at 440V there was an uptick in leakage current, so we are limiting it to 400V
      • At constant HV gain falls over time as counter-bias builds up in detector
      • Continuously increasing bias produces a constant gain. Using sequences to drift HV up from 300V to 400V.
      • Counterbias growth rate depends on event rate (and average event energy). Event rate 99 Hz with no shield, 46 Hz with lead/poly shield (3 walls), 98 Hz with shield and PuBe source
        • With PuBe can run 18-minute data series with HV drift 300V→400V. (Considering counter bias, leakage, and voltage drop across gap, detector bias is probably a little over 200V)
      • 60 keV and 10 keV lines seen clearly.
      • Noise width looks like ~60 eV using unbalanced PTOF and 60keV line as calibration. With balanced PsumOF and calibrating on the 10keV, noise sigma maybe as low as 26 eV. (This is still probably too high.)
    • Need to return PuBe source today
    • Will continue data tests. Try to gather more statistics to see lower Am lines. Understand calibration, threshold, noise.
      • superconducting and normal noise data taken
    • Helium recovery:
      • Last run it was determined that we weren't recovering any Helium from the 4He pump because the oil mist filter was saturated
      • With a new filter element, found that recovery was much worse (it must be leaking to atmosphere). Between-transfer recovery actually negative
      • Removing oil mist filter makes recovery slightly better than before.
        • Between transfers: boiloff 18.3 L/day, recovery with/without pump = 13.3/12.1 L/day (efficiency 73%/66%)
  • Run 68 (n,gamma with S101) analysis:
  • Future 150mm runs
    • 150mm w/phonons - need to design 600mK rod to accommodate 150mm cans
    • Received 4/3-inch DIBs from TAMU - now we can house either detector for phonon readout (would need to machine faces on 4/3-inch housing for DIBs)
    • S1501 and S1502 sent back to TAMU
    • We have decided we can run 8 phonon channels
      • Preferred DIB-pad layout: two DIBs separated by 120 degrees
      • Himangshu's mask design: PDF download
      • Will use existing vacuum-interface electrode, mounted on side opposite the phonon side
      • Will connect directly to two DCRCs, with one diverter board to route one tower wire to HV connection
      • We'll want LEDs - are there some mounted on the DIBs? We can mount external LEDs too - existing electrode has mounts.
    • First 150mm phonon detector will be monolithic, but “donut+hole” arrangement (active veto) proposed as well.
    • Aiming for May to finish fabrication and start run.
  • Other future runs
    • Future HVeV runs?
      • Berkeley can help us with a Soudan-style housing mount (3“ or 4”?)
      • then use original tower, can try different heatsinking for colder SQUID card
      • would need to install optical fiber
        • through top or through stripline tube? need vacuum feedthrough
        • Look at Prisca's deuterium lamp/monochromator/filters, glass UV fiber
      • Is PuBe NR calibration feasible?
        • 1×1×0.4 cm^3 is 630× smaller than SNOLAB detector. Anthony estimates 25 events/day (over full spectrum) for Si, 250 events/day for Ge. any chance we could get multiple chips?
        • Nick's idea: sandwich between two 4“ Ge detectors for gamma coincidence (large solid angle coverage)
    • Other ideas:
      • run SiHV with Nick's movable Am source (or with multiple Am sources)
      • run SiHV with no 4K heatsinking
      • re-run USD1? do we have any additional ideas to test it?
      • anything else?