UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20100729

2010/07/29

Run diary

0933 Stop overnight run 07100728_2249 at file number 75. Save the settings as Preset 3, load Preset 1 and start flashing
Temperature during past 5 hours
IGH at 0936
1056 Fill N2 cold trap full
1337 Stop flashing
1513 We are at a point ready to take data. I saw a lot of noise showing up in the pulses when the heat gun for little blue is running. It's best to take data at night. I'll flash the detector again
1513 flash settings
1800 Stop flashing
1854 Start and stop the run cause B has a large offset, adjust driver offset
1856 Start to take 4“ shield run
07100729_1856 settings
2011 Start series 07100729_2011 with 8” of neutron shield, DCRC settings same as 07100729_1856
2138 Start series 07100729_2138 with 2“ of polyethylene shield, DCRC settings same as 07100729_1856
2305 Start series 07100729_2305 with no polyethylene shield, DCRC settings same as 07100729_1856
2410 Stopped series 07100729_2305

Analysis

Cuts used:

% position cut
cQin = 0.93 <= Qpart & Qpart <=1.07;
cDel = (-50 < Xdel & Xdel <50) & (-50 < Ydel & Ydel <50 );
 
% fitting quality cut
cQchi2 = QSOFchisq(detnum,0) < 9000 + Qsum.^2*0.21;
 
% energy cut
cE = 5 < Qsum & Qsum < 100;
cPt = Pt > 5;
cQsum = Qsum >5;
Plot Cut
Qpart logstep cDel & cPt & cQsum
Qsum linstep cQin & cDel & cQchi2 & Pt
(Qsum, QSOFchisq) cDel & cQin
(Qi, Qo) cDel & cPt
(Xdel, Qi) cDel & cQin & cQchi2 & cPt
Pt overall calibration cDel & cQin & cQchi2 & cQsum & cPt
(Pt, Qsum) cDel & cQin & cQchi2
Phonon relative calibration No cut
(Xdel, Ydel) cQin & cQchi2 & cE(55<Qsum<65)', 'cQin & cQchi2 & cE & cAmDel'
(Xpart, Ypart) 'cc = cDel & cQin & cQchi2 & cE', 'cc = cc & cAmDel'
Yield plot (Pt, Y=Qsum/Pt) cDel & cQin & cQchi2
Neutralization plot (Time, Y) cAmDel & cQin & cQchi2 & cE(30<Qsum<90)
Neutralization plot (Time, Qsum) cAmDel & cQin & cQchi2

Overnight Am-241 run 07100728_2249

Approximately 11 hour run to check neutralisation and obtain background neutron rate.

Polyethylene neutron shielding tests

Neutron rates and fluxes in the following two plots obtained with data taken in 1 hour Cf-252 runs with 0", 2", 4", & 8" of polyethylene shielding. Ambient neutron rate and flux obtained with data from overnight Am-241 run.

1 hour Cf-252 and Am-241 run with 0" polyethylene shield 07100729_2305

1 hour Cf-252 and Am-241 run with 2" polyethylene shield 07100729_2138

1 hour Cf-252 and Am-241 run with 4" polyethylene shield 07100729_1856

1 hour Cf-252 and Am-241 run with 8" polyethylene shield 07100729_2011