2010/03/12
Run Diary
| Time | |
|---|---|
| 14:26 | Stop LED flashing started 17:41 last night. Keep the detector grounded and cool it down |
| Time | |
|---|---|
| 14:48 | Tower temperature falls to 35mK from ~490mK. It took 22min to cool down. |
| 14:54 | Load preset 2, set triggers and LEDs, start to take data |
| Time | |
|---|---|
| 15:31 | When flashing with LED1 I=164, ON=100us, OFF=2500us, temperature increase from 30mK to 470mK in 1min. |
| 15:32 | Flashing stopped. |
| 15:53 | Tower T drops to 35 mK |
| 21:13 | Readout vi stopped because of the connection error in ReadAll vi, leaving detector being baked. Data stops at 21:10 |
| 22:25 | Stop LED flashing |
| 22:57 | Start to take new data |
Data
| Data Start | Data End | Comment |
|---|---|---|
| 07100312_1454_F0001 | 07100312_2110_F0008 | 100 Files. DCRC: Preset 2; LED1: , I=400uA, ON=100us, OFF=2500us; Data taking: 3min, Flash:3min; Settings see above |
| 07100312_2258_F0001 | 07100313_0108_F0035 | 137 Files. DCRC: Preset 2;LED1: , I=400uA, ON=100us, OFF=2500us; Data taking: 15min, Flash:3min; Settings see above |