start › cdms › k100 › run_diary › 20170703

2017/07/03 (Monday)

Previous Day

Status 4K. Pumping out IVC with turbo pump since Friday. Mixture is open to dump, pump (off), Trap 2, fridge.
0840 Lab = 22.2 C, 42% humidity
0845 MC(Tower)/Sorb/1K Pot = 5.02/8.1/6.25 K, Fridge(G1) = 593 mbar, MB = 72.4%. Turbo pressure = 2.2 μbar, IVC = 16 μbar.
IVC pressure seems to be as low as in Run 62.
0850 Fill LN2 cold trap bath
0850 Pump Trap 2 contents to dump. We'll use Trap 1 for circulation.
0900 Connect 4He pump to 1K Pot pumping line, turn on pump, open the two valves connecting pump exhaust to recovery
0905 Open mixture to dump, pump (off), Trap 1, fridge. Trap 2 is empty.
0905 Open 1K Pot speedivalve. Mechanical gauge rises to about 1/3 bar. Open V4a (note that pumping lines had been evacuated). Didn't have 4He pump isolated so this pumps out quickly to 0.020 mbar. 1K Pot temperature falls to 4K but then rises again to 6.2 K.
0910 Open 1K Pot needle valve about 10 degrees. 1K Pot → 3.3 K / 3.2 mbar
Open needle valve a few more degrees. 1K Pot → 3.1 K / 4.0 mbar
Open needle valve a few more degrees. 1K Pot → 3.0 K / 4.5 mbar
Open needle valve a few more degrees. 1K Pot → 2.7 K / 6.5 mbar, MC temperature falls to 3.5 K
Open needle valve a few more degrees. 1K Pot → 2.6 K / 12.0 mbar
Open needle valve a few more degrees. 1K Pot → 2.3 K / 19 mbar
Open needle valve a few more degrees. 1K Pot → 1.9 K / 28 mbar
0940 Tower = 2.95 K, Fridge (G1/G2) = 280 mbar, both falling
0950 Tower = 2.52↓ K, Fridge (G1/G2) = 150↓ mbar, 1K Pot = 1.69 K / 26 mbar
0955 1K Pot falling to around 20 mbar, open needle valve a few more degrees.
1000 Tower = 2.16↓ K, Fridge (G1/G2) = 75 mbar, 1K Pot = 1.66 K / 18 mbar, open needle valve a few more degrees.
1005 Coldplate/Still = 3.5/2.35 K
1005 1K Pot spikes to 1.9 K, pressure off-scale. Temperature oscillates around 1.85 K, pressure still high
1010 Tower = 2.11 K, Fridge (G1/G2) = 62 mbar, both flat
1015 Close needle valve a bit, 1K Pot settles around 1.75 K / 60 mbar
1020 Tower/Coldplate/Still = 2.11/2.64/2.49 K, Fridge (G1/G2) = 60 mbar, 1K Pot = 1.75 K / 55 mbar
1025 Turn on 3He Pump. Close V9, keeping 58 mbar of mixture in dump.
1025 Start circulation through Trap 1, V6 = 13.0%. Condenser peaks at 68 mbar, Tower → 2.08 K and falling
1030 V6 → 13.3%. Condenser(G1) → 120 mbar (peak), Tower → 1.79 K (and falling). Coldplate/Still = 2.57/2.00 K
1035 V6 → 13.6%. Condenser(G1) → 184 mbar (peak), Tower → 1.52 K (and falling). Coldplate/Still = 1.94/1.59 K. 1K Pot = 1.74 K / 50 mbar. Still pressure falls into readable range.
1045 V6 → 14.0%. Condenser(G1) → 185 mbar (peak), Tower → 1.24 K (and falling). MC/Coldplate = 1.24/1.36 K, Still = 1.16 K / 4.0 mbar.
1055 V6 → 15.5%. Condenser(G1) → 129 mbar (climbing), Tower → 0.75 K (and falling). MC/Coldplate = 0.79/0.62 K, Still = 1.15 K / 0.70 mbar.
1105 Condenser(G1) flat at 134 mbar
1110 V6 → 100%. Condenser(G1) → 135 mbar (peak), Tower → 0.34 K (and falling). MC/Coldplate = 0.27/0.43 K, Still = 1.12 K / 0.12 mbar.
1120 Tower = 135 mK and falling.
1130 Tower = 90 mK and falling. Still = 0.09 mbar.
Open V14 to bring in the rest of the mixture. Still peaks at 400 mbar then falls below 30 mbar. G1/G2 peak around 160 mbar then fall again. Tower temperature rises to 140 mK.
1140 Tower = 63.5 mK. Still = 1.5 mbar and falling (V14 still open)
1150 Tower = 47.7 mK. Still = 0.75 mbar. Close V14. G1/G2 = 85 mbar and rising slowly.
1155 Tower/MC/Coldplate = 36.8/36.8/179 mK, Still = 1.2 K / 0.060 mbar, G1/G2 = 89/87 mbar, 1K Pot = 1.74 K / 50 mbar. Still thermometer a little unstable
1230 Tower = 29.7 mK [image not archived: cool.svg]
1250 Tower/MC/Coldplate = 29.5/30.9/176 mK, Still = 1.2 K / 0.060 mbar, G1/G2 = 90/88 mbar, 1K Pot = 1.73 K / 50 mbar.
1310 Set DCRC1 SQUID settings so that they match the DIB1 SQUET from the last run, the last time we had 2-sided running.
1315 Plug DIB1 into DCRC1 and turn on DCRC1. Tower temperature spikes a couple of mK, then returns.
1320 Zap SQUIDs on DCRC1/DIB1. MC rises to 33.6 mK then falls again, Still pressure spikes then back to normal.
1335 Tried starting vnc session to view pulses but it isn't working…
1345 VNC authentication didn't work for Matt, had to have Allison do it. Don't know why.
1400 See first pulses! DIB1 pA, had to bring QET bias down to 45 uA to see strong pulses. 40 uA knocks it down into s/c. Rate seems pretty high, and pulses have much longer fall times than the recent HV detectors. Fall time around 1.3 ms.
1430 See pulses on DIB1 pA/pC (at 45/55 uA). pD has some “rain” dripping off traces, maybe due to the short to chassis. Tried tuning its SQUID. Can't see pulses, not sure if IbIs makes sense. On pB IbIs seems to make sense but channel appears to be normal all the way down to 0 uA.
1435 Set up to measure Tc in DIB1 pA and pC, the working channels. QET bias = 0, test signal 12 mVpp / 100 Hz connected to QET.
1435 Tower = 29.3 mK
1455 Finish Tc check on DIB1 pA/pC. Heated tower as high as 58 mK by applying up to 11 uW power to heater.
While temperature rising, both seemed to go through full transition by 54 mK. Then while temperature falling we could see approximate beginning and end of transition.
DIB1 pA (Side1 pD) : 53.8 - 54.0 mK
DIB1 pC (Side1 pE) : 54.5 - 55.0 mK
SQUET ZC13 SQUET ZC39 SQUET ZC26
Run 62 DIB1 = DCRC1 Run 62 DIB3 = DCRC2 Run 62 DIB4 = DCRC4
chan SQUID bias Lockpoint chan SQUID bias Lockpoint chan SQUID bias Lockpoint
pA -91.2 -4.3 pA -54.0 -1.8 pA -81.8 -3.3
pB -78.8 -3.3 pB -60.3 -1.8 pB -78.3 -2.9
pC -82.0 -4.8 pC -46.3 -2.0 pC -85.0 -3.5
pD -85.2* -2.7 pD 85.0 5.7 pD -85.0 -5.7
Run 63 DIB1 = DCRC1 Run 63 DIB4 = DCRC2 Run 63 DIB3 = DCRC4
*today after SQUID tuning changed DIB1pD SQUID bias to -100 uA
1520 Estimate event rate: 74 pulses in 0.5 seconds = 148 Hz. Compare with Run 62, had 89 Hz with no shield. Shield blocks out 65 Hz so we should be able to reduce the rate to 83 Hz.
Setup: DIB1 → 3.2 MHz filters 1 → cable 1 → gender changer 1 → DCRC55(1)
1605 Tower/MC/Coldplate = 29.2/30.8/175 mK, Still = 1.2 K / 0.060 mbar, G1/G2 = 91/88 mbar, 1K Pot = 1.73 K / 45 mbar.
1615 Fill LN2 cold trap bath
1620 Filling the LN2 trap bath brought Tower temp even lower, 29.0 mK. Impressive.

Next Day