| Status | Transferred LHe Friday. pumping out IVC with turbo pump. |
| 1205 | Lab = 22.7 C, 39% humidity |
| 1209 | MC(Tower)/Sorb/1K Pot = 5.01/8.6/6.6 K, Fridge(G1) = 593 mbar, MB = 40.7%. Turbo pressure = 2.5 μbar, IVC = 16 μbar. |
| 1242 | Start LHe transfer (MB = 40.4%) |
| 1248 | MC(Tower)/Sorb/1K Pot = 5.00/7.9/6.99 K, Fridge(G1) = 570 mbar, MB = 41.5%. Turbo pressure = 2.5 μbar. Recover meter is quite squeaky… |
| 1302 | MC(Tower)/Sorb/1K Pot = 4.97/6.9/5.4 K, Fridge(G1) = 524 mbar, MB = 63.7%. Turbo pressure = 1.4 μbar. |
| 1308 | MC(Tower)/Sorb/1K Pot = 4.94/6.9/5.4 K, Fridge(G1) = 480 mbar, MB = 78.1%. Turbo pressure = 1.0 μbar. Transfer is going quickly, decrease He cylinder pressure slightly. |
| 1320 | MC(Tower)/Sorb/1K Pot = 4.94/7.0/5.4 K, Fridge(G1) = 479 mbar, MB = 84.2%. Turbo pressure = 1.0 μbar. |
| 1333 | End LHe transfer (MB = 92.0%). |
| 1335 | MC(Tower)/Sorb/1K Pot = 4.97/6.9/5.3 K, Fridge(G1) = 530 mbar. Turbo pressure = 2.2 μbar. IVC = 8.6 μbar. |
| 1335 | Fill LN2 cold trap. |