Agenda

  • Run 71 status
    • This cooldown: no IVC leak, fridge working fine. Base is 34 mK. All channels working.
    • Had a DAQ glitch a week ago that was applying incorrect detector biases - fixed that.
    • Performing hot LED flashes to restore phonon collection efficiency after running the motor - seems to work.
    • Studying stepper motor behavior indicates that it probably isn't really working right…
      • Heating during stepping - tried smoothing capacitors and changing dwell
      • DC testing shows resistance increase of ~10 Ohm @detector which can explain heating.
      • Solder joint to NbTi may cause local (runaway) heating to drive ~7 cm of NbTi normal
      • Conclusion: may only be moving <4 steps each time we move for a total of <1mm so far.
    • Will try to make a bunch of little moves and search for evidence of Am movement in phonon signals.
    • Probably will end the run within one week.
  • Should we design and build new cold hardware?
  • SQUET recall
    • Martin wants 5 of our 9 SQUETs sent back to Colorado. He wants good SQUIDs.
    • A lot of issues to deal with:
      • there are bad SQUID chips
      • we need to change shunt resistances for iZIPv6 versus HV, but don't want to wirebond too many times
      • some SQUETs have problems at the pins
        • can we get spare SQUETs with the SQUID chips removed? Could theoretically glue good SQUID chips onto good SQUET.
    • Anthony says we may be able to get alternative SQUID chips
    • We have 1.5-ish months to deal with this
  • Analysis status:
    • Run 68 (n,gamma with S101)
      • Migdal analysis pending
      • New (n,γ) simulations, modified signal shape
    • Run 70
      • analysis team includes Pramod, Matt, maybe Zach - should coordinate, have meetings
      • Need to look at the recent PuBe data, work on a low-threshold energy quantity, see if we can see a hint of (n,γ) signal. Also develop cuts and look for low-energy junk
    • Run 67 (USD detector with novel time-dependent gain)
      • Dongming has characterized lots of microscopic processes, wants us to write a paper
      • Matt will work on rough model of detector behavior
    • SiHV event reconstruction
      • Derek working on this
      • Hongguang developing connection to DMC
  • Lab and infrastructure
    • Leak detector: After pump rebuild, pump no longer turned. Jake and Derek fixed this problem. However, it still doesn't seem to be able to pull a good vacuum. Using scroll pump assist in order to use leak detector. | Teleconference link: | https://fnal.zoom.us/j/9082275310 |