start › cdms › k100 › run_diary › 20191011

2019/10/11 (Friday)

Previous Day

Status: Tower Assembly
1415 Measure how far source can move from center before hitting contacts by rotating shaft. Use meter to verify contact. Positions should be good within about 1/4 turn.

Source at position 0 (center). Shaft mount screws face viewer. This is to the left if you are looking from the back of the motor along the shaft.
Near contact -16.5 turns -13.10 mm
Far contact +52.5 turns +41.67 mm
1455 Return source to Position 0 (center) by rotating shaft.
1505 Notice that the nearer indium blocker looks very thin. Is this the same one as before?

Gabe verifies for me that it’s not the same one. We should make a new thicker one.
1510 Mount source plate in source housing in the same orientation as on Sep 17: