UMN CDMS Wiki archive
start › cdms › k100 › run_summary › run_54 › noise_and_dcrc_arrangements_3pc

Noise PSDs at various TES states

DIB1 DIB2 DIB3
pA pB pC pD pA pB pC pD pA pB pC pD
Data taken on 2015 Oct 08
DIB1 and DIB3: Rev C.1, total output gain = 16, front-end gain = 5
DIB2: Rev C, total output gain = 15, front-end gain = 2
0 charge bias

DCRCs stacked


DCRCs spread out

Comments:

  • There was rather prominent ~250 Hz noise during these tests (typically we avoid this pattern when taking data).
  • In all cases the spread-out arrangement improves low frequency noise (including the peak)
  • In most cases the stacked arrangement worsens the mid-frequency noise
  • DIB1 was more affected by stacking (compared to side-by-side) than the other two DIBs
  • Stacking had a weaker effect in these tests than on the September 24 tests. In the previous case the DCRCs were separated only by a couple of pieces of paper; in the current case they were separated by pads of paper.