| 1420 | Perform mixture throughput test at MC = 300 K using above procedure with 3He pump on. Let G1 rise to 550 mbar before Step 3, then it falls to 41 mbar and starts to rise. Throughput = 5.9 mbar/min. |
| 1430 | Check pressure in 1K Pot: 30 mbar on P2, 0.03 atm on mechanical gauge. Pump out with Scroll Pump |
| 1435 | With 1K Pot speedivalve and needle valve closed, pump on Main Bath with scroll pump. Once P2 < 2 mbar, open 1K Pot speedivalve and needle valve (5 turns) to pump out 1K pot volume to < 1 mbar on P2. |
| 1445 | Close off scroll pump, then pressurize bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar. |
| 1450 | Close off scroll pump, then pressurize bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar. |
| 1455 | Pressurize main bath and 1K Pot (both sides) with helium to 475 mbar+1atm on G3, then close 1K Pot speedivalve and needle valve. |
| 1515 | As on 2021 Nov 30, we will flow mixture through the fridge during LN2 cooldown. Let mixture from dump through cold LN2 Trap 1 to fridge, both sides. Then isolate the sides and circulate with V6 open partially and mix open to dump. V9, V13b, V12b+speedi, V1, Condenser speedi, Still gate valves open, 3He pump on. Open V6 to 12.6%. |
| 1520 | P1(Condenser) → 0.12 mbar. |
| 1525 | Try closing V12a speedivalve. G1 falls as it should, showing we have throughput. Falls at 10.5 mbar/min . Open V12a speedivalve again. |
| 1535 | Let 5.4 mbar He exchange gas into IVC. |
| 1550 | MC(Cernox) = 298.4 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1/P1 (circulation) = 664/735/0.16 mbar, Level resistor = 0.975 kΩ. |
| 1550 | Begin LN2 transfer |
| 1550 | Goldman thermometer = 2.043 kOhm. |
| 1600 | MC(Cernox) = 296.8 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1/P1 (circulation) = 664/735/0.19 mbar, Level resistor = 1.089 kΩ. |
| 1600 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 24.1 mbar/min. Open V12a speedivalve again. |
| 1630 | MC(Cernox) = 227.6 K, Sorb = >249 K (max), IVC = 2.5 mbar, G2/G1/P1 (circulation) = 665/734/0.45 mbar, Level resistor = 1.325 kΩ. |
| 1630 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 23.1 mbar/min. Open V12a speedivalve again. |
| 1700 | MC(Cernox) = 165.4 K, Sorb = 195.1 K, IVC = 2.2 mbar, G2/G1/P1 (circulation) = 664/733/1.5 mbar, Level resistor = 1.327 kΩ. |
| 1700 | Close V12a speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 41.8 mbar/min. Open V12a speedivalve again. |
| 1723 | MC(Cernox) = 135.0 K, Sorb = 145.4 K, IVC = 1.9 mbar, G2/G1/P1 (circulation) = 663.8/732.1/off_scale mbar, Level resistor = 1.327 kΩ. |
| 1730 | Stop LN2 transfer. Close LN2 Dewar valve. Pull out Level Resistor and put stopper in port. Put ping-pong ball on MB exhaust port. |
| 1730 | Close V6 and turn off 3He pump. |