  • Lab in-person: Matt
  • Virtual partner: Michael
0915 Matt Fritts body temperature = 97.5 F
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)-master; DCRC33(2) → HV control. DCRCs on overnight. Lead/poly shield erected (4 lead walls, 2 poly walls).
0855 MB = 64.2%
0920 Tower/MC/Coldplate = 54.6/29.5/183 mK, Still = 1.39 K / 0.15 mbar / 5 mW, G1/G2= 280/257 mbar, 1K Pot = 1.66 K / 30±2 mbar.
Overnight the tower cooled to a minimum of 54.8 mK at 2:00 am, then warmed a little to 55.0 mK, then at 8:00 am started cooling again steadily.
1100 Fill LN2 cold trap bath using the new long copper tube, no rubber hose (we have a new one but I'll just use the copper for now, if it works).
1140 Matt leaves lab
1235 Matt Fritts body temperature = 98.4 F
1430 Finish quick manual IbIs on all working channels. Choose the following transition points:
QET Bias DCRC4/DCRC3 = 0,240,320,0/0,300,340,380
These biases are chosen as reasonably in the middle of the transition somewhere. But pulses aren't very pretty. There seems to be thermoelectric oscillations in all channels.
Here's an alternative set toward the high end of the transitions:
QET Bias DCRC4/DCRC3 = 0,280,340,0/0,340,380,420
It's not obvious that these higher biases give better results. Definitely can see that at the lower bias, pF has higher amplitude pulses (checked this by doing a 100 running average on the pulse viewer). Also requires higher threshold, so lower bias makes stronger pulses and stronger noise. Hard to tell whether S/N is better one way or the other. We'll stick to the lower biases, because this will give more reliable dP/dI calculations from IbIs.
1545 Tower = 55.5 mK
1546 Start a 60 min series at HV = 0,
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (0,240,320,0)/(0,300,340,380) uA, Front-End Gains = (3/3/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC1pD threshold = 40
Use sequence SimpleDataLoopNoFlash.msl
1625 Create a sequence to take data at 5V, drifted, using the python code to generate it:

from directory:

pysequencer

command:

python HVscan.py -iDCRC 2 -HVs 5 -HVcalFile HV_calibrations/HV_DCRC2_control_cals_500V_200721.txt -HVdriftRate 0.031 -HVpreBias 10/2 -tSeries 60 -DCRCs2Flash 3 >> ../../sequencer/HVdrift_5V_31mVps_pre10-2_60min.msl

ANALYSIS OF TODAY'S DATASET (here's the Root code):

  • After processing today's data, I tried a quick calibration and a noise measurement
    • Noise sigma (PTOF) = 6.3 keV
    • This is 20X worse than Run72!
      • probably shouldn't be too surprised given how nasty our pulses look
    • Gain is a little lower than in Run72, about 36% lower
  • The Run 72 0-Volt data from 2020 Mar 24 was taken at a very similar Tower temperature
  • but at significantly higher QET biases
    • biases that according to our IbIs are near the top, or over the top of the transition
    • and we do see much better pulse shapes in this run at those much higher QET biases (in some channels at least)