start › cdms › k100 › run_diary › 20200725

2020/07/25 (Saturday)

Previous Day

  • Lab in-person: Matt
  • Virtual partner: Nick
0740 Matt Fritts body temperature = 97.4 F
Setup Run 73. DIB 1/3/4 → Stripline 43/48/41 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)-master; DCRC33(2) → HV control. DCRCs on overnight. Lead/poly shield erected (4 lead walls, 2 poly walls).
0655 MB = 30.8%
Yesterday the Tower temperature took an upward excursion for a while after the LN2 fill, maximum 57.0 mK, but it has cooled steadily from 3pm till this morning
0810 Reset QET biases to higher values to avoid oscillations. All are near the top of their transitions:
QET Bias DCRC4/DCRC3 = 0,330,360,0/0,380,400,460
0810 Tower = 53.4 mK. (Temperature went down a little after adjusting the biases, maybe the oscillations were heating it slightly)
0812 Start a 60 min series at HV = 0,
CDMS detector: Driver Gains = 1x, DCRC4/3 QET biases = (0,330,360,0)/(0,380,400,460) uA, Front-End Gains = (3/3/3)/(5/5/5)
1000 BOR randoms, trigger on DCRC1pD threshold = 4
Use sequence SimpleDataLoopNoFlash.msl
0830 Write a sequence Al_flash_DCRC3.msl to do the hot Al flashes like we did in R71 after moving the source
0905 Matt arrives in lab
0910 Tower/MC/Coldplate = 53.4/29.5/188 mK, Still = 1.45 K / 0.15 mbar / 5 mW, G1/G2= 276/257 mbar, 1K Pot = 1.66 K / 30±2 mbar.
0915 Fill LN2 cold trap bath
0927 Start LHe transfer @ MB = 25.0%
0945 Tested source mover control. Plugged the extra motor into the control board: plug into red board, wires from one motor into [a], wires from the other in [b]. Turn on GwINSTEK power supply, plug into BNC on control board, turn voltage up to 7.1 V. Light comes on on the red board at above around 3.5 V (Nick says it should work anywhere between 3 and 15 V).
0950 From the web browser, create.arduino.cc/editor(etc), logged in as Nick, type “help” into Send to see commands. Try mv commands: mv +48 for one rotation forward, mv -96 for two rotations backwards - it works.
0955 Nick says that the sequence we used to run it for a few steps with a pause and many repetitions is Python code, not clear how to send it from the fridge computer running Windows. We should be able to plug it into jiajin and communicate from there. May not be able to do it remotely. Need to figure out how to communicate to one of the COM ports from a terminal (it's not telnet, it's a serial connection). We were sorta able to connect from the fridge computer using PuTTY, using a serial connection to COM4 - we saw the “hello” greeting but we weren't able to type in commands.
1002 End LHe transfer @ MB = 92.4%
1045 Matt leaves lab
1310 Processed and analyzed this morning's data. After raising the QET biases noise is better by about a factor of 2 (3.7 keVee) but still 1 order of magnitude worse than Run72. Gain is much lower than Run72, like 20% or less. Why?
1310 Maybe for some reason the QETs have trapped flux. Let's try a hot Al flash, like we do for the source mover. This should also serve as a detector flash/bake.
1315 Do hot Al flash using DCRC3 LEDs using sequence Al_flash_DCRC3.msl
Tower temperature rises to 94 mK then falls. That's not very high.
The settings for hot Al flash are LEDPulseWidth/LEDRepRate/LED1,2Current = 400/500/0.8. This is in contrast to standard flash settings, 100/5000/0.6
1320 Turn on LED1 for 2 minutes. Tower rises to 87 mK. Turn off LED for a bit to let it cool
1325 Try negative LED current, no heating.
1320 Turn on LED2 for 2 minutes. Tower rises to 88 mK. Turn off LED for a bit to let it cool
1335 Try the standard flash settings, turn on LEDs. Didn't see any temperature rise. The hot settings still cause heating though.
On Feb 25 in Run 72, LEDs heated Tower to 500 mK in 10 seconds. What's wrong this time?? It was a different DCRC but why should that matter??
1525 Tower = 53.6 mK
1530 Try turning on the LEDs with the standard flash settings. After 3 minutes Tower only warms to 55.1 mK. Turn them off again.
1535 Try turning on LEDs again with standard width and rate but increase current 0.6→1.0 mA. After 3 minutes Tower warms to 55.8 mK. Turn them off again.
1540 Try again with 1.0 mA, width 500, rate 100 (this is a mostly-on mode). After 3 minutes reaches 117 mK. Turn them off again.
1545 Try again in this 500/100 mostly-one mode, but increase current 1.0→1.5 mA. After 3 minutes reaches 128 mK. Turn them off again.
1555 Try again with current at 3.0 mA. (Not sure how high it goes). After 3 minutes reaches 130 mK. Turn them off again.
Took another look at pulses, they still seem really weak. I'd like to try warming the detector above 1.2 K and back down, in case there is trapped flux in the QETs for some reason. We could try suspending circulation and turning on the MC heater. I'll try this, and monitor the fridge pressures - if they get too high I'll release some pressure into the dump.
1640 Tower = 54.1 mK
1640 Close V6 completely. After a few minutes Still pressure starts rising. Tower temperature rising a little. Condenser pressure falls.
1645 Tower temperature rises above 100 mK
1650 Tower = 390 mK and rising steadily. Let's give it a boost with the MC heater.
1650 Turn on MC heater → 1 mW. Doesn't do much. Try 10 mW. Looks like it's warming a bit faster.
1655 Tower reaches 1.3 K. Turn off MC heater.
1655 G1/G2 = 144/140 mbar
1655 Restart circulation: V6→12.8%. Condenser peaks at 180 mbar. Tower → 1.28 K
1705 V6→13.2%. Condenser peaks at 220 mbar. Tower → 1.23 K. Temperature flattish, maybe because we're working to get the Al into s/c
1715 V6→13.4%. Condenser peaks at 275 mbar. Tower → 1.19 K.
1730 V6→13.6%. Condenser → 340 mbar without peaking. Still → 55 mbar. Tower rises → 1.46 K.
1740 Turn off Still heater, it was still at 5 mW
1745 Condenser = 360 mbar, Still = 65 mbar, Tower = 1.49 K. Condenser hasn't peaked but let's try opening V6 more anyway
1745 V6→13.8%. Condenser → 430 mbar without peaking. Still → 60 mbar. Tower dips a little then rises again → 1.46 K.
1750 V6→14.0%. Trying to power through, although it's having trouble. Maybe should have let Condenser pressure get lower before trying to start circulation.
1755 OK looking better now, Condenser peaks at 470 mbar and now falling again, Still → 6.5 mbar, Tower finally cooling again → 1.28 K
1800 V6→14.4%. Condenser → 350 mbar, Still → 2.6 mbar, Tower → 1.07 K
1810 V6→15.0%. Condenser peaks at 350 mbar, Still → 1.5 mbar, Tower → 0.73 K and dropping quickly
1820 V6→100%. Condenser peaks at 370 mbar, Still → 0.22 mbar, Tower → 0.48 K and falling
1830 Tower = 200 mK, Condenser = 330 mbar, Still = 0.17 mbar
1950 Tower = 58.6 mK, Condenser = 234 mbar, Still = 0.12 mbar
1950 Switch on Still Heater → 5 mW
1950 Pulses are back but they don't look any better. What gives??? Why is it so different from Run72?

Next Day