UMN CDMS Wiki archive
start › cdms › k100 › run_summary › run_73 › ibis
Detector Side1 channel map
DCRC channel: A B C D
HVbox1 pE pB pD DCRCX(4)
HVbox2 pA pC pF DCRC50(3)

spreadsheet

at 0-Volts without pi filters

DCRC4pB / Side1pB 2020-07-23
240 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
320 .2131 tower temp [mK] 56.4
335 .1908
305 .2263 V offset: 1.0184
600 -.4317 Prefac: 0.610088922
800 -.9176
700 -.6756 note: no pi filters
290 .2097
280 .1962
260 .1361
240 .0265
230 -.0736
220 -.242
DCRC4pC / Side1pD 2020-07-24
320 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -0.5879 tower temp [mK] 56.6
700 -0.382
600 -0.1748 V offset: 1.0593
360 0.2327 Prefac: 0.499109015
340 0.2094
380 0.2413 note: no pi filters
320 0.107
300 -0.0291
280 -0.209
280 -0.2512
260 Rail
DCRC3pB / Side1pA 2020-07-24
300 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -0.2203 tower temp [mK] 57.0
700 -0.0434
600 0.1314 V offset: 1.1949
400 0.4362 Prefac: 0.52733119
380 0.427
360 0.4121 note: no pi filters
340 0.3744
320 0.3100
300 0.2109
280 0.0612
260 -0.6654
DCRC3pC / Side1pC 2020-07-24
340 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 0.1161 tower temp [mK] 57.1
700 0.2662
600 0.4182 V offset: 1.3169
460 0.6149 Prefac: 0.632544977
440 0.6293
420 0.6343 note: no pi filters
400 0.6308
380 0.6149
360 0.5812
340 0.5346
320 0.4357
300 -0.0389
DCRC3pD / Side1pF 2020-07-24
380 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -0.1132 tower temp [mK] 56.9
700 0.0784
600 0.2710 V offset: 1.4196
440 0.4878 Prefac: 0.45846253
420 0.4531
400 0.3849 note: no pi filters
380 0.2811
360 0.1235
340 -0.7571

At 0-Volts (pi filters back on)

DCRC4pB / Side1pB 2020-07-27
320 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -.4036 tower temp [mK] 52.6
700 -.171
600 .0645 V offset: 1.4572
380 .5628 Prefac: 0.6139724
360 .5716
340 .5412
320 .4606
300 .3383
280 .164
260 RAILED
DCRC4pC / Side1pD 2020-07-27
390 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -.2894 tower temp [mK] 53.2
700 -.0707
600 .1494 V offset: 1.4602
440 .4753 Prefac: 0.666099666
420 .4798
400 .4436
380 .3622 osc (small)
360 .2403 osc (small)
340 .0768
320 RAILED
DCRC1pB / Side1pA 2020-07-27
360 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 .4209 tower temp [mK] 53.5
700 .4203
600 .4176 V offset: ?
500 .4034 Prefac: ?
480 .398
460 .3923 note: level in the normal region
440 .3843
420 .3773
400 .3691
380 .3542
360 .2558
340 .1078
320 -.158 Osc (Med)
300 RAILED
DCRC1pC / Side1pC 2020-07-27
420 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 .0775 tower temp [mK] 53.6
700 .2421
600 .4073 V offset: 1.3943
540 .500 Prefac: 0.653594771
520 .5221
500 .5363 note:
480 .540
460 .5305
440 .5027
420 .4493
400 .365
380 .249
360 .045 Osc
340 RAILED
DCRC1pD / Side1pF 2020-07-27
440 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -.028 tower temp [mK] 53.7
700 .1734
600 .3738 V offset: 1.5832
520 .5165 Prefac: 0.590034965
500 .5356
480 .535 note:
460 .5056
440 .4357 osc
420 .3288 Osc
400 .1825 Osc
380 -.046 Osc
360 RAILED
pA pB pC pD pE pF
QET bias 360 320 420 390 440
prefactor 0.635 0.614 0.6536 0.6661 0.59

8/3/20 New Data (tower a little cooler)

DCRC4pB / Side1pB 8/3/20
320 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -0.2673 tower temp [mK] 51.7
700 -0.0288
600 0.2148 V offset: 1.6407
400 0.6912 Prefac: 0.559461986
380 0.7161
360 0.7131
340 0.6634
320 0.5634
300 0.4227
280 0.2122 osc
260 rail
DCRC4pC / Side1pD 8/3/20
370 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 -0.1822 tower temp [mK] 51.1
700 0.024
600 0.2307 V offset: 1.4674
440 0.5276 Prefac: 0.533052234
420 0.5172
400 0.4655
380 0.3693 Osc
360 0.252 osc
340 0.061 osc
320 rail
DCRC1pB / Side1pA 8/3/20
360 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 0.6632 tower temp [mK] 51.5
700 0.664
600 0.6602 V offset: 1.57
500 0.6463 Prefac: 0.483941506
480 0.6411
460 0.6347 note:
440 0.6272
420 0.619
400 0.6092
380 0.5977
360 0.5077
340 0.3538
320 -0.8147 random burst of osc
300 -0.5986 superconducting
Voffset is a guess based on comparing IbIs max to offset in other channels
DCRC1pC / Side1pC 8/3/20
420 ← standard driver gain: 1
QET bias [uA] Baseline [V] det bias [V]: 0
800 0.2913 tower temp [mK] 51.7
700 0.4449
600 0.5992 V offset: 1.5201
540 0.6819 Prefac: 0.558455146
520 0.6998
500 0.7090 note:
480 0.7072
460 0.6901
440 0.6518
420 0.5860
400 0.4925
380 0.3611
360 -0.2977
340 -0.148
DCRC1pD / Side1pF 8/3/20
420 ← standard driver gain: 1
QET bias [uA] det bias [V]: 0
800 0.3481 tower temp [mK] 51.9
700 0.5359
600 0.7269 V offset: 1.8505
520 0.8597 Prefac: 0.454882342
500 0.8728
480 0.8623 note:
460 0.8174
440 0.7358 Osc
420 0.623 Osc
400 0.4788
380 -0.1831
360 -0.0077
pA pB pC pD pE pF
QET bias 360 320 420 370 420
prefactor 0.484 0.559 0.558 0.533 0.455

2020 Aug 21: Adjusted DCRC1pD

After the partial warmup, we decided to adjust the standard setting of DCRC1pD (Side1pF) from 420 uA to 410 uA. Here are the new QET biases and prefactors, appropriate for relatively low temperatures, around 52 mK and below.

pA pB pC pD pE pF
QET bias 360 320 420 370 410
prefactor 0.484 0.559 0.558 0.533 0.313

2020 Sep 21: New IbIs and QET biases

New IbIs was done at a warmer temperature: 64.6 mK. Adjusted QET biases such that each channel has close to the same dP/dI

pA pB pC pD pE pF
QET bias 265 210 240 240 275
prefactor 0.392 1.368 0.466 0.718 0.439

2020 Sep 22: New IbIs and QET biases

New IbIs was done at a warmer temperature: 64.6 mK. Adjusted QET biases such that each channel has close to the same dP/dI

pA pB pC pD pE pF
QET bias 320 240 290 290 325
prefactor 0.621 0.882 0.549 0.689 0.631

2020 Oct 10: Begin using standard dP/dI

Beginning on this date, we changed QET biases frequently to account for temperature variations, always setting all channels so that their QET bias / prefactor product was equal to a standard value of 200. Therefore their dP/dI values (and gains) should all be the same.

Standardized QET bias × Prefactor 200

Here's a timeline of different variations in the method:

Oct 10 Using IbIs at several temperatures, make curves of QET bias versus temperature for biases that target the standardized dP/dI.
Oct 12 Start manually adjusting biases during series to correct for temperature variations
Oct 15 Start automatically adjusting biases during series (using Root sequencer)
Oct 18 Start implementing an offset to Tower Thermometer temperature, because IbIs suggests that it differs from detector temperature
Oct 19 Start implementing a linear correction to Tower Thermometer temperature, because variation in detector temperature seems to be much smaller than variation in thermometer. First series we did this with was at 21:34