2020/07/23 (Thursday)
- Lab in-person: Matt
- Virtual partner: Pramod
| 0810 | Matt Fritts body temperature = 97.5 F |
|---|---|
| Setup | Run 73. DIB 1/3/4 → Stripline 43/48/41 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)-master; DCRC33(2) → HV control. DCRCs on overnight. Lead/poly shield erected (4 lead walls, 2 poly walls). |
| 0730 | MB 27.1% |
| 0810 | Tower = 56.6 mK, G1/G2= 262/240 mbar, Still = 0.150 mbar / 5 mW, 1K Pot = 1.67 K / 36±2 mbar. |
| Tower has wavered, fell as low as 60.4 mK at 3am then wandered back up. More steady for the last 4 hours. | |
| 0920 | Tower/MC/Coldplate = 56.5/30.4/188 mK, Still = 1.45 K / 0.15 mbar / 5 mW, G1/G2= 263/240 mbar, 1K Pot = 1.67 K / 36±2 mbar. |
| 0946 | Begin LHe transfer @ MB = 21.0% |
| 1005 | Try adjusting phonon channels. Lowered a number of QET biases, now see strong pulses on every working channel (DCRC4pA/Side1pE is not operational, like last run it behaves like an open TES). But there is ugly noise. Let's try zapping SQUIDs. |
| 1005 | ZAP all SQUIDs on DCRCs 3&4 |
| 1010 | Fill LN2 cold trap bath |
| 1023 | End LHe transfer @ MB = 92.6% |
quick SQUID Tuning notes:
- Trigger on Test Signal
- Connect Test Signal to feedback, 400 Hz, maybe 100 mVpp
- Set Polarity to OPEN and ADCsource to PREAMP
- Start with QET bias = 0, Lockpoint = 0, SQUID bias = 0
- Increase SQUID bias (negative for RevC.1, positive for RevC) to maximized modulation depth (Vpp) while avoiding cusps
- Increase Lockpoint (negative for RevC.1, positive for RevC) so that x-axis intersects at point with high slope
- then set back to normal: disconnect test signal, set ADCsource to FEEDBACK, set Polarity to NON-INVERTING
| 1100 | Tune SQUID settings on DCRC3pC (we did DCRC3pB yesterday) |
|---|---|
| Seeing some ugly noise on most channels. Let's try powercycling the DCRCs | |
| 1105 | Turn off DAQ programs, unplug DCRCs |
| 1110 | Power DCRCs, restart DAQ programs. ZAP all SQUIDs on DCRCs 3&4 |
| 1150 | Try to find good QET biases for all working channels. Some of them have very ugly pulses shapes, others OK |
| 1155 | MB = 91.1% |
| 1230 | The tube for transferring LN2 burst today. Trying to order another from the chem stockroom. |
| 1325 | Rig up a long copper pipe to attach to the LN2 Dewar which extends all the way to the LN2 Trap Bath Dewar |
| 1400 | Pick up new tubing from Chem Stockroom |
| 1400 | Matt finally leaves lab |
| 1705 | Do quick IbIs on DCRC4pB. It turns out that the QET bias we chose to use earlier is not fully in the transition region. so we chose a lower value. But the pulse shapes at this lower value are ugly. |
| 2145 | Matt Fritts body temperature = 98.6 F |