| 1300 | Matt Fritts body temperature = 96.8 F |
| Setup | Run 73. DIB 1/3/4 → Stripline 43/48/41 → Cable 3/1/2 → nothing/nothing/accessory_interposer → HV_Conn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3)-master; DCRC33(2) → HV control. DCRCs on overnight. Lead/poly shield erected (4 lead walls, 2 poly walls). |
| 1230 | MB 55.2% |
| 1300 | Tower = 61.8 mK, G1/G2= 259/240 mbar, Still = 0.160 mbar / 5 mW, 1K Pot = 1.68 K / 38±2 mbar. |
| | Tower has wavered, fell as low as 60.4 mK at 3am then wandered back up. More steady for the last 4 hours. |
| 1400 | Tower/MC/Coldplate = 61.4/31.1/193 mK, Still = 1.50 K / 0.16 mbar / 5 mW, G1/G2= 259/240 mbar, 1K Pot = 1.67 K / 45±3 mbar. |
| 1410 | Fill LN2 cold trap bath |
| 1415 | Tape plastic sheet over end of E-box, so we can purge it with N2 to reduce HV current leakage |
| 1435 | Hook up HV box test resistor to Keithley voltage meter, start reading this out on OmniLogger vi on fridge computer. |
| 1445 | Start purging E-box with N2 from the LN2 Dewar vent. Opened the valve just barely enough to feel some gas pressure. |
| 1445 | Matt leaves lab |
| 1745 | Matt Fritts body temperature = 97.4 F |