| 0935 | Perform mixture throughput test at MC = 300 K using above procedure with 3He pump on. Let G1 rise to 530 mbar before Step 3, then it falls to 38 mbar and starts to rise. Throughput = 5.6 mbar/min. |
| 0950 | With 1K Pot speedivalve and needle valve closed, pump on Main bath with scroll pump.
Once P2 < 5 mbar, close pump valve then open 1K pot speedivalve and needle valve (5 turns), notice P2 rise a few mbar. Now pump out bath and 1K pot volume to < 1 mbar on P2. |
| 0955 | Close off scroll pump, then pressurize bath and 1K Pot (both sides) to about 0.5 bar with helium. Then pump it out again with scroll pump to < 1 mbar. |
| 1005 | Pressurize main bath and 1K Pot (both sides) with helium to 380 mbar +1atm on G3, then close 1K Pot speedivalve and needle valve. |
| 1010 | Let 5.4 mbar He exchange gas into IVC |
| 1010 | As on 2019 Oct 29 during the last attempted Run 71 cooldown, we will flow mixture through the fridge during LN2 cooldown. Let mixture from dump through cold LN2 Trap 2 to fridge, both sides. Then we'll isolate the sides and circulate with V6 open partially and mix open to dump. V9, V13b, V12b+speedi, V1, Condenser speedi, Still gate valves open, 3He pump on. Now open V6 to 12.8%. |
| 1015 | Circulation pressures settling to G2/G1/P1 = 650/719/0.24 mbar |
| 1015 | MC(Cernox) = 295.8 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1/P1 (circulation) = 650/719/0.14 mbar, Level resistor = 0.982 kΩ |
| 1020 | Begin LN2 transfer |
| | looks like it might be a slow one. I filled the LN2 trap bath an hour ago, maybe Dewar pressure is low |
| 1030 | Try pressurizing the LN2 Dewar through the vent with N2 from the wall supply. Looks like it might be going a little faster, blowing out of the top bath port stronger. |
| 1035 | MC(Cernox) = 295.0 K, Sorb = >249 K (max), IVC = 5.4 mbar, G2/G1/P1 (circulation) = 655/721/0.26 mbar, Level resistor = 0.987 kΩ |
| 1035 | Close V12b speedivalve for 2 minutes to assess throughput. (Condenser pressure G1 should fall.) G1 falls at 10.6 mbar/min. Open V12b speedivalve again. |
| 1045 | MC(Cernox) = 289.5 K, Sorb = >249 K (max), IVC = 4.6 mbar, G2/G1/P1 (circulation) = 654/722/0.32 mbar, Level resistor = 1.002 kΩ |
| 1045 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 11.0 mbar/min. Open V12b speedivalve again. |
| 1100 | MC(Cernox) = 270.1 K, Sorb = >249 K (max), IVC = 4.0 mbar, G2/G1/P1 (circulation) = 655/722/0.40 mbar, Level resistor = 1.008 kΩ |
| 1100 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 12.0 mbar/min. Open V12b speedivalve again. |
| 1120 | MC(Cernox) = 236.2 K, Sorb = >249 K (max), IVC = 3.0 mbar, G2/G1/P1 (circulation) = 656/722/0.70 mbar, Level resistor = 1.046 kΩ |
| 1120 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 15.3 mbar/min. Open V12b speedivalve again. |
| 1140 | MC(Cernox) = 193.4 K, Sorb = >249 K (max), IVC = 2.5 mbar, G2/G1/P1 (circulation) = 657/722/1.80 mbar, Level resistor = 1.110 kΩ |
| 1140 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 20.1 mbar/min. Open V12b speedivalve again. |
| 1205 | MC(Cernox) = 151.8 K, Sorb = >249 K (max), IVC = 2.2 mbar, G2/G1/P1 (circulation) = 657/722/5.00 mbar, Level resistor = 1.223 kΩ |
| 1205 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 31.1 mbar/min. Open V12b speedivalve again. |
| 1220 | MC(Cernox) = 135.7 K, Sorb = 209.4 K, IVC = 2.2 mbar, G2/G1/P1 (circulation) = 657/721/8.50 mbar, Level resistor = 1.263 kΩ |
| 1220 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 38.6 mbar/min. Open V12b speedivalve again. |
| 1245 | MC(Cernox) = 117.3 K, Sorb = 160.5 K, IVC = 1.9 mbar, G2/G1/P1 (circulation) = 657/721/24.0 mbar, Level resistor = 1.299 kΩ |
| 1245 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 48.6 mbar/min. Open V12b speedivalve again. |
| 1300 | MC(Cernox) = 109.2 K, Sorb = 144.0 K, IVC = 1.9 mbar, G2/G1/P1 (circulation) = 657/720/36.0 mbar, Level resistor = 1.311 kΩ |
| 1300 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 52.8 mbar/min. Open V12b speedivalve again. |
| 1330 | MC(Cernox) = 97.2 K, Sorb = 128.2 K, IVC = 1.6 mbar, G2/G1/P1 (circulation) = 657/720/120 mbar, Level resistor = 1.323 kΩ |
| 1330 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 57.5 mbar/min. Open V12b speedivalve again. |
| 1400 | MC(Cernox) = 90.5 K, Sorb = 123.5 K, IVC = 1.6 mbar, G2/G1/P1 (circulation) = 657/719/650 mbar, Level resistor = 1.328 kΩ |
| 1400 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 57.9 mbar/min. Open V12b speedivalve again. |
| 1405 | MC(Cernox) = 88.3 K, Sorb = 122.6 K, IVC = 1.6 mbar, G2/G1/P1 (circulation) = 657/719/offscale mbar, Level resistor = 1.330 kΩ |
| 1405 | Stop LN2 transfer. Put ping pong ball over bath exhaust port, remove level resistor and put stopper on port. |
| 1530 | MC(Cernox) = 79.6 K, Sorb = 121.8 K, IVC = 1.6 mbar, G2/G1/P1 (circulation) = 655/719/offscale mbar |
| 1530 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 61.0 mbar/min. Open V12b speedivalve again. |
| 1645 | MC(Cernox) = 77.7 K, Sorb = 122.3 K, IVC = 1.6 mbar, G2/G1/P1 (circulation) = 655/719/offscale mbar |
| 1645 | Close V12b speedivalve for 2 minutes to assess throughput. G1 falls at 62.0 mbar/min. Open V12b speedivalve again. |
| 1650 | Stop circulation by closing V6. Turn off 3He pump. |