2019/11/01 (Friday)
| Status: At 77K discovered IVC leak near bottom of IVC. Warming now to check indium seal. Working toward 2nd attempt at Run 71 cooldown. | |
| 0925 | MC(Cernox) = 282.5 K |
| 0930 | Pump out IVC . |
| 1000 | Turn off MC and Still heaters. |
| 1020 | MC(Cernox) = 282.7 K |
| 1025 | Lower main bath |
| 1030 | Leak detector on IVC = 29 pbarL/s |
| 1030 | Squirt helium around middle IVC seal, bottom stripline tube seal, lower IVC seal. No change in signal. |
| 1045 | Try loosening all the screws on the bottom seal, then squirt helium at it again. Still no detectable leak. Thought about trying to cool it down with LN2, but probably would have to cool it more than I'd like. |
| 1055 | Stop leak detector, erect clean tent, fill IVC with N2. |
| 1120 | After opening IVC, inspect bottom flange. In several places scratches were visible extending from the inner edge to a screw hole. Polished these areas with Scotch Brite, being careful to rub back and forth tangential to the circle, no inward/outward polishing. Cleaned these areas with methanol and a wipe. With just a little polishing the scratches disappeared. |
| 1130 | Also polish the bottom piece flange a little with 1500 grit sandpaper and lightly with Scotch Brite |
| 1200 | Put a new indium o-ring on bottom flange piece and install, tighten screws most of the way |
| 1205 | Start pumping IVC with scroll pump |
| 1210 | Retighten all screws on all three other in-bath IVC flanges - top, middle, and stripline tube. They all seemed quite tight already, couldn't turn any more than a small amount. A few are starting to have stripped heads, but they still seemed to be rather tight. |
| 1215 | Switch to Turbo to pump out IVC |
| 1300 | Turbo on IVC = 400 nbar |
| 1310 | LD on IVC = 18 pbarL/s Squirt helium at top of stripline tube and bottom IVC seal. No leak seen. |
| 1420 | Check fridge pressures: LHe trap/Condenser/Still = 0.1/0.1/0.015 mbar (very low). Pump out with scroll pump. |
| 1450 | Tighten IVC screws one more time, then raise main bath. Adjust fridge insert mount brackets slightly to align fridge/rack/main bath better. |
| 1455 | Turbo on IVC = 1.2 μbar, IVC = 5.4 μbar with Turbo valved out for a few minutes. Stop pumping out IVC to pump out cryostat vacuum jacket. |
| 1505 | Cryostat vacuum jacket = 6.3 mbar. Pump out with scroll pump. |
| 1520 | Start pumping cryostat vacuum jacket with turbo pump |
| 1545 | Turbo on cryostat vacuum jacket = 1.4 μbar, cryostat vacuum jacket = 74 μbar with turbo valved out for a two minutes |
| 1635 | Turbo on cryostat vacuum jacket = 1.0 μbar, cryostat vacuum jacket = 46 μbar with turbo valved out for a two minutes Stop pumping cryostat vacuum jacket. |
| 1645 | IVC = 40 μbar. Start pumping IVC with turbo pump. |