2019/04/07 (Sunday)
| Setup | DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRCX(4)/DCRC50(3) DCRC33(2) → 1kV power supply → HV input Lead poly shield is erected, N/E/W walls complete, no poly or lead on S wall. |
|---|---|
| 0615 | Tower = 39.1 mK |
| 0620 | Apply 10 uW of power to MC heater |
| 0630 | tower = 56.5 mK. Looks like it might flatten out at 57 or 58 mK |
| We will see whether a higher base temperature gives us better QET performance. We tried this in Run 61. We will need to adjust the QET biases downward (by maybe 50 uA?) to get to good points in transition. The hope is that the oscillations will be much lower in amplitude. | |
| 0640 | Tower = 56.7±0.1 mK, looks like it's flat here (a little wiggly) |
| 1509 | Tower/MC/Coldplate = 56.5/48.8/189 mK, Still = 1.71±0.15 K / 0.150 mbar, G1/G2 = 310/296 mbar, 1K Pot = 1.78 K / 13 mbar. MB = 24.1% |
| 1518 | Commence LHe transfer, MB = 24.1% |
| 1520 | Fill LN2 cold trap full. |
| 1538 | Tower = 59.2 mK and rising slowly |
| 1543 | Tower = 61.0 mK and still rising. Maybe we pushed a channel into a bad state somehow? Set all QET biases to 0. Heating slows for a minute, then resumes. Probably due to the transfer then. |
| 1550 | Tower starts to flatten out at 61.9 mK |
| 1555 | Cease LHe transfer, MB = 91.7% |
| 1600 | Tower = 60.5 mK and cooling. Snap and re-apply QET biases. |
| 1615 | Tower = 59.4 mK and cooling slowly. |
| 1616 | Got all channels in transition with (mostly) lowered QET biases. DCRC4/3 QET biases = (390/350/390)/(500/410/450) uA. DCRC3 pC is at a higher value than before, but it is now above the oscillation region. There doesn't seem to be any room below it now. Not sure how these values will hold up as the tower continues to cool back down, but it's a start. |