UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20161208

2016/12/08 (Thursday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters → cables 1,2,3 → HV board inputs 1,3,2 → HV board outputs 1,2 → DCRC55(1)/50(3)
Taking overnight data at HV=15 V.
between 30 and 40 minutes into the first series (1207_1550), DCRC3pC railed. By the beginning of the second series (1207_1650), DCRC1pB, DCRC1pC, DCRC3pC were all railed.
0820 Main bath = 73.6%
0820 HV = 15V, Vtest = 0.7 mV. So Vdetector = 14.7 V
0841 Fill LN2 cold trap full.
1020 Finish microIbIs at 15 V. For some reason the Vbias values are all lower, significantly, than at 0V. Why is this?? Yesterday I showed that at 30V they were all higher which makes sense, additional parasitic power.
The reductions range from 50 to 130 mV at x2 driver gain. The average is 85 mV.
1040 Finish a semi-full IbIs on DCRCpC to compare with the 0V IbIs curve. The transition region is definitely lower.
1046 Lower bias to 0V.
1104 Finish micoIbIs at 0V. The Vbias values are still lower than previous OV values and are even lower than the 15V values taken above. The reductions (from Dec 6th 0V values) range from 60 to 165 mV at 2x driver gain. The average is 112 mV.
1105 Flash detector for 30s.
1146 Finish microIbIs at 0V post-flash. Values seem similar to the 15V measurement taken earlier today with a reduction range of 40 to 131 mV at 2x driver gain, with an average of 84mV.
1358 Tower/MC/Coldplate = 32.9/35.8/226 mK, Still = 1.2 K / 110 μbar, G1/G2 = 187/160 mbar, 1K Pot = 1.69 K / 26 mbar. MB = 65.3%. Boil off rate: a bit faster than slow mode.
Begin testing higher Tower temperature to see how oscillation region changes.
1415 Apply 5 uW to MC heater. After 5 minutes Tower is around 45 mK, not entirely stable
1420 Apply 10 uW to MC heater. After 5 minutes Tower levels out around 60 mK, with some wiggles.
1425 Apply 12 uW to MC heater. Pulses still seen in most channels but getting weaker, probably shouldn't go much higher. After 10 minutes Tower is around 59 mK.
1435 Apply 14 uW to MC heater. After about 5 minutes Tower is around 62 mK
1440 Apply 16 uW to MC. After about 5 minutes Tower is around 67 mK. We'll try it here.
1505 Finish IbIs curve for DCRC1PA. Transition region was small, did not see 10kHz. Tower temp was pretty stable between 66.5 and 67.2 mK.
1600 Filled in a few IbIs points on DRC1pA
1615 Finished IbIs and oscillation measurement for DCRC1pB. Although noise at 10kHz went up a bit there was never any sign of a peak at 10kHz, more just a rising of the flat part of the PSD.
1625 Finish IbIs/oscillation measurement for DCRC1pC, Tower = 66 mK. Again no real sign of 10kHz peak
1635 Finish IbIs/oscillation measurement for DCRC1pD. Tower = 65.8 mK. In this case there was a shallow peak at 10kHz somewhere in the transition region.
1640 Turn off MC heater to let Tower cool to base

The plot above shows the IbIs results for one channel at the two temperatures. The results fit theoretical expectations. Shown here is a quadratic extrapolation of the normal part of the curve back to the vertical axis to find the voltage offset for the normal and transition part of the curve. The four points near the vertical axis represent the superconducting part of the curve. The transition part of the curve should extend from the normal down to a line that has exactly 1/2 the slope of the superconducting part. The data points taken in transition end at the lowest bias before the channel falls into superconducting.

1700 Tower = 35 mK
1700 Flash detector. (Note that HV is at 0V)
1705 Prebias detector to 55 V
1710 Dial HV back down to 45 V
1720 Tower = 34.5 mK and falling. Return all channels to standard QET biases and adjust driver offsets. Set Driver Gains to 4x.
1725 HV = 45 V, Vtest = 2.5 mV. So Vdetector = 43.9 V
1732 Start a sequence of 14 1-hour series of 45V data with Frontend Gain = 5, OG = 4, DCRC3pB intentionally railed, with 1000 beginning randoms, with no flashing in between series, and a 35ADC bin threshold trigger on DCRC3 pD.
1810 Fill LN2 cold trap full.
1905 DCRC1pC is railed (too low) and threshold seems too high, will try to adjust from home…
1943 Adjust DCRC1pC ADCoffset from 0.33 to 0.23
1944 Adjust DCRC3pD threshold from 35 to 30
1955 DCRC1pC still seems railed and threshold still seems high, will try again…
1958 Adjust DCRC3pD threshold from 30 to 25
1959 Adjust DCRC1pC ADCoffset from 0.23 to 0.13
2005 Threshold looks ok now but DCRC1pC still railed
2008 Adjust DCRC1pC ADCoffset from 0.13 to -0.07
2015 OK DCRC1pC is unrailed but probably a little too high, baseline around 11000 in the BatViewer window (max is around 17000). Try tweaking it some…
2017 Adjust DCRC1pC ADCoffset from -0.07 to -0.03
2025 DCRC1pC baseline is around 7500 now. Let's try to get it close to 2500.
2028 Adjust DCRC1pC ADCoffset from -0.03 to 0.0
2028 Adjust DCRC1pC ADCoffset from 0.0 to 0.03

Next Day