2019/03/05 (Tuesday)
| Status: | Mixture has been cryocleaned several times a few days ago. Tower has been installed and IVC turbo pumping overnight. Need to clean mixture again and leak check IVC. |
|---|---|
| 0955 | Tighten IVC bottom screws one more time. |
| 1000 | Turbo on IVC = 40 nbar. IVC = 2200 nbar with pump valved out. Kinda high maybe. Compare e.g. 20190304 in Run 68, overnight reached 16 nbar pumping, 340 nbar not pumping. Could be slow pumpout due to taping up more holes. |
| 1015 | Put leak detector on IVC. Background = 3.4 pbarL/s. That's nice and low. |
| 1025 | Leak detector = 4.1 pbarL/s. |
| 1030 | Leak check top of fridge and IVC. No leaks seen. |
| 1035 | Start cryocleaning mixture through LN2 Cold Trap 1, V6 = 12.7% |
| 1040 | Increase V6→13.0% to restart flow |
| 1110 | Raise main bath |
| 1040 | Increase V6→13.1% to speed up flow |
| 1135 | Stop cleaning through Trap 1 after 60 min. G2 flow pressure was 570-860 mbar. Pump mixture to dump. Removed 7.0 mbar |
| 1205 | Start cryocleaning mixture through LN2 Cold Trap 2, V6 = 12.4% |
| 1210 | Increase V6→12.9% to restart flow |
| 1250 | Increase V6→13.0% to speed up flow |
| 1305 | Stop cleaning through Trap 2 after 60 min. G2 flow pressure was 570-660 mbar. Pump mixture to dump. Removed 16.2 mbar |
| 1320 | Trap 1 is warm. Check contents: 8.5-1.5 = 7.0 mbar on G1. Pump out with scroll pump. |
| 1335 | Check pressure in condenser line (includes LHe cold trap) at G1 by opening V1. 206 mbar. That's pretty high. Had this been looked at since the end of the last run? Check that all the KF flanges are tight. Pump out with scroll pump. |
| 1340 | Check pressure in fridge at condenser side on G1 by opening condenser valve and V1. About 1 mbar. Check pressure in fridge at still side on P1 by opening still gate valve: about 0.83 mbar. Pump out both sides with scroll pump. |
| 1355 | Open dump to both sides of fridge lines via LN2 cold trap 1. |
| 1400 | With 1K Pot speedivalve and needle valve closed, pump on Main bath with scroll pump. Once P2 < 100 mbar, open 1K pot speedivalve and needle valve (5 turns) to pump out 1K pot volume. |
| 1410 | P2 < 1 mbar, close V1A, 1K Pot speedivalve. (Don't have to pump out lines up to 4He gas line.) Close off scroll pump. Flush MB and 1K pot with 0.4 bar of 4He gas. |
| 1415 | Do another pump and flush. Again, with 1K Pot speedivalve and needle valve closed, pump on Main bath with scroll pump. Once P2 < 100 mbar, open 1K pot speedivalve and needle valve (5 turns) to pump out 1K pot volume. |
| 1420 | P2 < 1 mbar, close off scroll pump. Pressurize MB and 1K pot (both speedivalve and needle valve open) with 175 mbar (above 1 atm) of 4He gas. Close 1K Pot needle valve and speedivalve. |
| 1440 | Turbo on IVC = 34 nbar. IVC = 1600 nbar with pump valved out. Improved since this morning. |
| 1450 | Let 6.3 mbar He exchange gas into IVC (a little high, usual target is 5.4 mbar). |
| 1450 | MC(Cernox) = 297.1 K, Sorb = >249 K (max), IVC = 6.3 mbar, G1 (fridge) = 606 mbar, Level resistor = 0.984 kΩ |
| 1505 | Begin LN2 transfer |
| 1515 | MC(Cernox) = 296.8 K, Sorb = >249 K (max), IVC = 6.3 mbar, G1 (fridge) = 605 mbar, Level resistor = 0.985 kΩ |
| 1530 | MC(Cernox) = 291.2 K, Sorb = >249 K (max), IVC = 5.4 mbar, G1 (fridge) = 604 mbar, Level resistor = 1.001 kΩ |
| 1545 | MC(Cernox) = 274.8 K, Sorb = >249 K (max), IVC = 4.6 mbar, G1 (fridge) = 604 mbar, Level resistor = 1.004 kΩ |
| 1600 | MC(Cernox) = 255.2 K, Sorb = >249 K (max), IVC = 4.0 mbar, G1 (fridge) = 603.8 mbar, Level resistor = 1.018 kΩ |
| 1615 | MC(Cernox) = 228.7 K, Sorb = >249 K (max), IVC = 3.4 mbar, G1 (fridge) = 603.0 mbar, Level resistor = 1.056 kΩ |
| 1730 | MC(Cernox) = 128.6 K, Sorb = 171.3 K, IVC = 2.2 mbar, G1 (fridge) = 599.0 mbar, Level resistor = 1.270 kΩ |
| 1740 | Cease LN2 transfer and put ping pong ball on main bath port. |
HPGe Measurements
| 1100 | Setting up to make measurements of HPGe sensitivity as a function of source position. The point of this measurement is to characterize the active detector volume. Built a lead castle with a slot to produce a thin sheet of flux. The source will go in the lead slot and we'll roll the detector around on a wooden cart to illuminate different regions of it. |
|---|
| 1300 | Set up associated HPGe electronics as described here. Tried slowly turning up the HV bias to the recommended 3.5kV, but the current drawn from the HV supply spikes when approaching 2.2kV and the supply disables itself. Tried waiting ~30 min and repeating, but that did not help. Even so, we are able to see pulses and take spectra using the EasyMCA when biased at 2kV. |
|---|