2019/03/06 (Wednesday)

Previous Day

Status: Transferred LN2 yesterday
0830 MC(Cernox) = 78.1 K, Sorb = 137 K, IVC = 1.6 mbar, G1 (fridge) = 602 mbar, Level resistor removed last night.
0835 Start pumping out IVC
0845 Turbo on IVC: 10 nbar
0900 Perform electrical check, looks OK. Some issues that weren't seen at 300 K:
1) DIB3 Feedback D/return is too high, 700 kΩ (s/b 8 kΩ). Lots of problems with this SQUET, luckily pC is the only active channel and it's OK
2) DIB4 qI, short to chassis: feedback→chassis 22 nS, chassis→source-or-drain 2 MΩ. This was seen last run too.
No problems with any relevant channels.
0910 Something is wrong with the leak detector. With port blanked, background signal is >100 pbarL/s, usually less than 1. When trying to calibrate it doesn't seem to pump down all the way. Tried opening it to IVC while it was running - this was a mistake, IVC pressure spiked up a couple of orders of magnitude briefly. Hopefully not too much contamination. After a couple of minutes IVC is back down to 40 nbar, but it had been below 10 nbar.
Try turning leak detector off and back on.
0925 Turbo on IVC: 8.6 nbar. So it has recovered it seems.
0930 Reboot of leak detector failed. Display says:

Error (41)
TMP frequency is too low!
Possible causes:
* Frequency converter is faulty
* Turbomolecular pump is faulty

so maybe the turbo pump in the LD has died.
0935 Hit the Restart button and get another error:

Error (56)
P1 > 10 mbar after runup
Possible causes:
* Forepump is faulty
* Valve V1 didn't open

so maybe it's the forepump instead
0940 The shared Alcatel leak detector is available - bring it into the lab and warm it up.
1015 Alcatel leak detector seems not to be working either. Won't pump down, stays in “roughing” mode.
1030 After a restart the Alcatel LD seems to be working.
1030 Alcatel LD on IVC = 120 nbar (10 times higher than when our turbo is pumping on it.
1030 Alcatel LD = 5400 pbarL/s, slowly rising to some stable value.
1035 Mini-throughput check: close off Trap 1 and V3. Only V1, condenser speedivalve, still gate valve, and V9 open. Turn on 3He pump, then open V6 to 12.8%. This pumps mix out of the still side and we can see if the pressure goes down on the condenser side. P1 (still) falls to 6 mbar after a few minutes. G1 (condenser) falls, from 603 mbar to 440 mbar in 5 minutes, and continues to fall. Good, looks like we have throughput.
1040 Close V6, turn off 3He pump. Open trap 1 valves and V3 to let mix into fridge lines through trap 1.
1045 Begin blowing out LN2. Main bath (G3) = 440 mbar above 1atm.
Alcatel Leak detector on IVC = 4500 pbarL/s
1050 Alcatel LD = 4200 pbarL/s. Main bath = 451 mbar.
1055 Alcatel LD = 4000 pbarL/s. Main bath = 444 mbar.
1100 Alcatel LD = 4000 pbarL/s. Main bath = 320 mbar and falling.
1105 Alcatel LD = 3800 pbarL/s. Main bath = 155 mbar.
Stop pressurizing main bath.
1110 Pump mixture to dump, so we can clean out warm LN2 Trap 2, still dirty from the last cryocleaning.
1110 Start pumping out main bath with scroll pump. LD = 3600 pbarL/s. Takes 11 minutes to fall to 1 mbar.
1125 Trap 2 is warm. Check contents: 17.7-1.5 = 16.2 mbar on G1 (surprisingly high). Pump out with scroll pump.
1135 Let mixture back in to both sides of the fridge through LN2 Cold Trap 1
1135 Alcatel LD = 3100 pbarL/s.
1145 Pressurize MB with helium to 380 mbar above 1 atm. Open 1K Pot speedivalve and needle valve (2 full turns) to repressurize pot as well. Notice MB pressure falls about 5 mbar. Close 1K Pot valves again.
1145 Alcatel LD = 3000 pbarL/s. IVC pressure while pumping = 86 nbar.
1150 Let 63 μbar He exchange gas into IVC.
Throughput test procedure, Cold Trap 1:
  1. All mixture in dump, LN2 trap 1 cold, condenser & still valves open (3He pump on - optional)
  2. Open V9, V13a, V12a+speedivalve, V3. Wait for G1 to be close to G2.
  3. Close V12a+speedivalve and V3, open V1
  4. monitor G1
1200 Perform mixture throughput test at MC = 77 K using above procedure with 3He pump on. Let G1 rise to 565 mbar before Step 3, then it falls to 68 mbar and starts to rise. Throughput = 35.7 mbar/min.
1215 Let mixture back into fridge, both sides open to LN2 Trap 1 and dump, 3He pump off.
1230 Begin initial LHe transfer
1240 MC(Cernox) = 81.9 K, Sorb = 156 K, IVC = 74 μbar, G1(fridge) = 603.8 mbar. G3(MB) = 460 mbar.
1245 MC(Cernox) = 82.3 K, Sorb = 154 K, IVC = 46 μbar, G1(fridge) = 602.7 mbar. G3(MB) = 430 mbar.
1250 Transfer pressure on G3 started around 460 mbar, then gradually fell to around 360 mbar. Turn the Dewar Heater on to 4 PSIg. Transfer pressure climbed to 530 mbar after a few minutes and recovery meter was spinning fast, so turn it off again.
1255 Transfer pressure on G3 has fallen to 340 mbar (around 2 PSI on Main Bath gauge). Turn Dewar Heater on again to 4 PSIg. Pressure is not rising as quickly this time so we'll leave it on for a while.
1300 MC(Cernox) = 77.3 K, Sorb = 129 K, IVC = 19 μbar, G1(fridge) = 591.9 mbar. G3(MB) = 385 mbar.
1320 MC(Cernox) = 62.3 K, Sorb = 72 K, IVC = 14 μbar, G1(fridge) = 520.2 mbar. G3(MB) = 457 mbar. (MB gauge = 2.9 PSI, Storage Dewar gauge = 4.2 PSI.)
1330 MC(Cernox) = 56.6 K, Sorb = 38 K, IVC = 12 μbar, G1(fridge) = 472.9 mbar. G3(MB) = 418 mbar. (Storage Dewar gauge = 4.5 PSI.)
1350 MC(Cernox) = 47.5 K, Sorb = 7.4 K, IVC = 10 μbar, G1(fridge) = 459.2 mbar. G3(MB) = 376 mbar. MB = 55.7%
this is one of those times when the IVC pressure seems to be falling too fast. I think we concluded before that this was because the gauge got dirty.
1350 Increase resistance range on bridge to 2.00 kΩ, so that the Cernox thermometer doesn't saturate and make us miss the pumpout point (10 K)
1400 MC(Cernox) = 40.7 K, Sorb = 7.0 K, IVC = 8.6 μbar, G1(fridge) = 456.3 mbar. G3(MB) = 363 mbar. MB = 79.3%
1405 MC(Cernox) = 38.9 K, Sorb = 7.0 K, IVC = 8.6 μbar, G1(fridge) = 455.2 mbar. G3(MB) = 360 mbar.
1410 End LHe transfer @ MB = 88.0%
1435 MC(Cernox) = 23.9 K, Sorb = 6.2 K, IVC = 7.4 μbar, G1(fridge) = 501.4 mbar.
1450 MC(Cernox) = 19.8 K, Sorb = 6.0 K, IVC = 7.4 μbar
1500 MC(Cernox) = 17.5 K, Sorb = 5.9 K, IVC = 6.3 μbar
1510 MC(Cernox) = 14.8 K, Sorb = 5.7 K, IVC = 6.3 μbar
1522 MC(Cernox) = 11.9 K, Sorb = 5.6 K, IVC = 1.6 μbar
IVC falling fast, let's start pumping
1523 Start pumping out IVC with turbo
1615 MC(Cernox) = 10.4 K, Sorb/1K Pot = 6.4/5.1 K, Turbo on IVC = 2.5 nbar, G1(fridge) = 538 mbar.
Note temperature is creeping up. Last time it fell to 4K. So maybe I started pumping out too soon. Tomorrow it may be hard to cool to 1K.

HPGe Measurement

1100 Set up HPGe detector with HVbias=1.85 kV. Bias was unstable above this level. Took 1000s background spectrum using EasyMCA in this manner.
1125 Remove 133Ba source from safe and place in lead castle.
1130 Take 1000s Ba data with source slot aimed at z=8“ and source at r=16” from the center of the HPGe detector. Hint of 356 keV line.
1150 Take 1000s Ba data with source slot aimed at z=6“ and source at r=16” from the center of the HPGe detector. Strong 356 keV line, maybe 80keV also.
1150 Take 1000s Ba data with source slot aimed at z=4“ and source at r=16” from the center of the HPGe detector. No clear Ba lines.
1240 Take 1000s Ba data with source slot aimed at z=5“ and source at r=16” from the center of the HPGe detector. No clear Ba lines. Detector may have been moved. Should repeat.
1315 Take 1000s Ba data with source slot aimed at z=7“ and source at r=16” from the center of the HPGe detector. Lines are back.
1345 Take 1000s Ba data with source slot aimed at z=7.5“ and source at r=16” from the center of the HPGe detector. Lines present.
1430 Take 1000s Ba data with source slot aimed at z=5.5“ and source at r=16” from the center of the HPGe detector. Some lines.
1510 Take 1000s Ba data with source slot aimed at z=5“ and source at r=16” from the center of the HPGe detector.

Next Day