UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20180718

2018/07/18 (Wednesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
overnight: 1K Pot failure around 10pm last night (we'll look at the logs for more detail of the failure and conditions preceeding it). Nick caught it at 11:30 pm and suspended circulation; at that time the Condenser pressure had risen above 800 mbar.
0810 V6 closed, V14 open, so pump is pumping on nothing, dump-trap1-condenser path open. Tower = 2.4 K, 1K Pot = 2.5 K / 1 mbar. G1/G2 = 207/184 mbar.
0810 Close 1K Pot needle valve, then close 1K Pot speedivalve. 1K Pot → 3.5 K. Sorb rises above 4 K.
0815 Turbo on IVC = 2.5 nbar. Close valve to turbo pump, if we decide to warm above 4K we can see how much helium comes out.
0930 Tower = 2.8 K, IVC = 30 nbar.
0931 Fill LN2 cold trap
1125 1K Pot >7K. Sorb = 11.5 K
1125 Tower = 3.1 K and rising, IVC = 30 nbar. G1/G2 = 414/375 mbar and rising.
1125 Turn on MC heater → 20 mW, increase Still heater 6→20 mW. Tower temperature rises more quickly. Sorb temperature rises too.
1130 Turn off MIDAS DAQ programs and unplug DCRCs 1&3.
1145 Tower = 3.5 K and flattening a bit, G1/G2 =495/448 mbar and rising quickly than before, Sorb now falling, peaked at 12.5 K now 9.6 K
1155 Tower temperature suddenly starts rising quickly, 3.6 → 5.4 K, IVC 30→63 nbar. 1K Pot falls to 6 K
1200 Tower = 5.4 K and plateauing, G1/G2 = 548/497 mbar and rising steadily, IVC = 63 nbar, Sorb = 7.7 K and falling. 1K Pot = 6K
1250 Tower = 5.5 K, G1/G2 = 576/523 mbar, Sorb = 7.5 K, all flat. IVC = 74 nbar
1250 Close V14 and V13A, open V9 and V5. This pumps on the Condenser and LN2 Trap 1 to dump. Sorb temp falls a little, Tower/1K Pot flat.
1315 Tower = 5.5 K, Sorb = 7.6 K, flat. G2 (dump) = 695 mbar and rising, G1 (condenser + Trap1) = 9.7 mbar and falling. IVC = 63 nbar.
1320 Open V3 to pump on Still side too. G2 peaks to 60 mbar then falls.
1350 Tower = 5.5 K, Sorb = 7.8 K, flat, G2 (dump) = 730 mbar, looks flat. G1 (condsenser+trap1) = 1.5 mbar, Still = 0.6 mbar. IVC = 63 nbar.
On July 6 we warmed the tower to 7.5 K under I think the same conditions, here we're stuck at 5.5 K. Is something anchoring the temperature?
1400 Try turning off both MC heater and Still heater, see where it settles
1430 Tower settled to 4.4 K. So it does seem anchored. Turning off heaters also seems to have caused Sorb/1K Pot temperature to rise, now Sorb = 10.9 K, 1K Pot > 7 K. IVC has fallen to 40 nbar.
1435 Turn Still heater back on to 20 mW
this caused Tower to rise slightly to 4.8 K and caused Sorb to fall, 10.9→9.5 K
1505 Turn MC heater back on to 20 mW
Tower → 5.5 K, Sorb falls again, → 7.7 K, 1K Pot falls → 6 K
1525 IVC = 74 nbar
1536 Start LHe transfer @ MB = 49.6%
Starting transfer caused Sorb/1K Pot to peak to 9.0/6.9 K then fall to 5.4/4.2 K
1610 End LHe transfer @ MB = 93.6%
After end of transfer Sorb rises to 5.9 K
1705 Fill LN2 cold trap full.
1725 Close V12a+speedivalve, V1, V3, V5, no longer pumping to dump. Dump(G2) = 732 mbar. Close V9, turn off 3He pump.
1725 Tower/Sorb/1K Pot = 5.7/5.4/4.2 K, all flat. MC/Still = 20/20 mW. IVC = 63 nbar.

Next Day