2018/07/17 (Tuesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3)
0943 Tower/MC/Coldplate = 31.3/29.1/184 mK, Still = 1.4 K / 0.150 mbar / 6 mW, G1/G2 = 140/131 mbar, 1K Pot = 1.54 K / 8.5 mbar. MB = 35.0%
0948 Fill LN2 cold trap full.
1000 DCRCs are behaving again (no stuck output) but noise/low energy events observed yesterday are still present.
1007 Turning off the room lights and clean room filters doesn't help.
1010 1K pot has been drifting down over the last several days. Open 1K Pot needle valve 75° → 80°. 1K Pot → 1.6 K / 32 mbar
1115 Tower = 32.8 mK
1131 Start 30-minute 0 V background data series, Driver Gains = 2x, QET biases = (280/260/305)/(140/295/300) uA, Front-End Gains = 10, 1000 initial randoms, trigger on DCRC3pB/C/D threshold = 8/16/16. To be compared to bias scan data from 2018/07/01, but shifted some DCRC3 QET biases slightly to avoid oscillations.
1220 Tower temp wandered up to 34.5 mK and back down to 32.7 mK during that series.
1223 Set HV=40V with 15% prebis for 5 min.
1234 Start 30-minute 40 V background data series, Driver Gains = 2x, QET biases = (280/260/305)/(140/295/300) uA, Front-End Gains = 10, 1000 initial randoms, trigger on DCRC3pB/C/D threshold = 8/16/16. To be compared to bias scan data from 2018/07/01
1236 Tower = 32.6 mK
1318 Tower = 31.0 mK
1453 Start LHe transfer @ MB = 27.1%
1542 End LHe transfer @ MB = 77.1%
1615 Set HV=0V
1625 1K pot has been falling, now 1.55 K / 12 mbar
Open 1K Pot needle valve 80° → 90°. Pot → 1.66 K / 45 mbar. Watched traces, didn't see any obvious change.
1640 Fill LN2 cold trap bath
1129 1K pot failed. G1/G2 = 860/790 mbar, 1K Pot = 4.5 K / 1.3 mbar. Close V6 and open V14 to stop circulation and relieve pressure. G1/G2 begin to fall.

Next Day