start › cdms › k100 › run_diary › 20180719

2018/07/19 (Thursday)

Previous Day

Status: circulation and 1K Pot suspended yesterday. Mixture in dump, MC and Still heaters on at 20 mW each
overnight: Tower steady at 5.4 K, Sorb around 7K and slowly increasing, 1K Pot around 5.5 K and slowly increasing
0940 Tower/Sorb/1K Pot = 5.4/7.3/5.7 K, IVC = 74 nbar
The fact that we're stuck at 5.4 K even with the heaters on all night suggests a thermal connection to the 4K bath, e.g. helium in the IVC. Pressure in the IVC doesn't seem high but maybe I shouldn't trust the gauge?
We should try pumping the IVC but first I'll try turning off the heaters like yesterday.
0945 Turn off MC and Still heaters
1020 Tower has fallen to 4.5 K, Pot and Sorb have gone up. 1K Pot >7K, Sorb = 10.6 K. Saw this yesterday too. It's as if having the heater on is boiling He into the IVC which cool the upper stages. Also IVC fell → 40 nbar (possibly lower, forgot to check it again)
1020 Turn MC and Still heaters back on → 20/20 mW
1035 Tower/Sorb/1K Pot = 5.5/7.5/5.9, IVC = 74 nbar, close to before
1040 Begin pumping IVC with turbo
Tower/Sorb/1K Pot temperatures are all starting to rise
1150 Tower/Sorb/1K Pot = 7.9/11.1/>7 K
1228 Tower/Sorb/1K Pot = 10.2/14.1/>7 K. Turn MC and Still heaters down to 10/10 mW
1310 Tower/Sorb/1K Pot = 11.2/15.8/>7 K. Turn MC and Still heaters down to 5/5 mW
1500 Tower/Sorb/1K Pot = 12.2/16.1/>7 K. Turn MC and Still heaters down to 3/3 mW
1645 Tower/Sorb/1K Pot = 12.5/15.9/>7 K.
1648 Fill LN2 cold trap.

Next Day