2018/07/06 (Friday)
| Setup: DIB1/3/4 → 3.2 MHz filters 2/3/1 → cable 3/1/2 → gender changer 3/?/2 → HVConn 1/3/2 → HV 1/2 → DCRC55(1)/DCRC50(3) | |
| 0008 | Tower = 1.65 K, Condenser = 268 mbar and falling, P1 >1000 mbar, 1K Pot = 1.67 K / 45 mbar. Open V6=13.6% |
| 0036 | Tower = 1.87 K, Condenser = 140 mbar and falling, P1 >1000 mbar, 1K Pot = 1.67 K / 45 mbar. Open V6=13.8% |
| 0046 | Tower = 1.92 K, Condenser = 118 mbar and falling, P1 >1000 mbar, 1K Pot = 1.67 K / 45 mbar. Open V6=14.2% |
| 0052 | Tower = 1.92 K, Condenser = 108 mbar and falling, P1 >1000 mbar, 1K Pot = 1.67 K / 45 mbar. Open V6=14.4% |
| 0100 | Tower = 1.95 K, Condenser = 97 mbar and flattening out, P1 >1000 mbar, 1K Pot = 1.67 K / 45 mbar. Open V6=14.8% |
| 0108 | Tower = 1.92 K, Condenser = 89 mbar, P1 >1000 mbar. Open V6=15.0% |
| 0112 | Tower = 1.94 K, Condenser = 84 mbar, P1 >1000 mbar. Open V6=15.2% |
| 0115 | Tower = 1.93 K, Condenser = 81 mbar, P1 >1000 mbar. Open V6=15.6% |
| 0120 | Tower = 1.93 K, Condenser = 79 mbar, P1 >1000 mbar, 1K Pot = 1.67 K / 45 mbar. Open V6=16.0% |
| 0125 | Tower = 1.94 K, Condenser = 76 mbar, P1 >1000 mbar. Open V6=16.4% |
| 0127 | Tower = 1.93 K, Condenser = 73 mbar, P1 >1000 mbar. Open V6=17.0% |
| 0130 | Tower = 1.94 K, Condenser = 72 mbar, P1 >1000 mbar. Open V6=17.6% |
| 0135 | Have been slowly opening V6 further while watching G1/G2. Currently at V6=23% but G1 = 68mbar and falling. P1>1000 mbar. |
| 0138 | Finally, at V6=26% G1 rises to 283 mbar. Tower drops to 1.7 K, 1K pot = 1.67 K / 45 mbar |
| 0609 | Tower = 1.91 K, Condenser = 95 mbar, P1 >1000 mbar, 1K pot = 1.67 K / 40 mbar. Open V6=27.0% |
| 0630 | Tower = 1.62 K and falling, Condenser = 340 mbar and rising, P1 = 200 mbar and falling, 1K pot = 1.67 K / 40 mbar. |
| 0700 | Tower = 1.60 K, Condenser = 347 mbar, Still(P1) = 80 mbar, all flat. 1K pot = 1.67 K / 40 mbar (stable). Open V6→37%. Still pressure falls quickly. |
| 0735 | Tower = 0.99 K with small oscillations, Condenser = 436 mbar, Still(P1) = 0.140 mbar. |
| not good. Tower can't get below 1 K. Still pressure low, probably poor throughput. | |
| 0740 | Close V13a, open V9, to start pumping to dump. G1/G2 fall. |
| 0745 | Increase Still heater 0.1→10 mW. No increase in Still pressure, actually goes down a little. |
| 0745 | Open V3. Now pumping both sides of fridge to dump. |
| 0750 | Set MC and Still heaters both to 20 mW |
| 0900 | Close 1K Pot needle valve and speedivalve |
| 0950 | Tower = 7.5 K and warming. G2 (dump) = 717.5 mbar, flat. 1K Pot = 4.3 K. Still = 0.014 mbar, Condenser = 1.3 mbar. Still thermometer = 60 K(?). Calibration must be no good around 4K. |
| 1000 | Fill LN2 cold trap bath |
| 1000 | Top off HPGe dewar. |
| 1000 | Close V6. Close V12a+speedi, open V13. This lets mix cool in cold trap before letting it back in the fridge. |
| 1005 | Open V12a+speedi. This lets mix into fridge (both sides, V3 is open). |
| 1010 | Tower = 8.5 K. Turn off MC and Still heaters. |
| 1015 | Tower = 8.3 K, Condenser = 579 mbar. Turbo on IVC = 3.0 nbar |
| 1015 | Open 1K Pot speedivalve to start pumping on pot. Pot → 1.76 K / 100 mbar, Condenser → 495 mbar, Tower → 8.1 K. |
| 1020 | 1K Pot → 3.4 K / 0.45 mbar |
| 1020 | Open 1K Pot needle valve → 10°. 1K Pot → 2.9 K / 3.2 mbar Open 1K Pot needle valve → 20°. 1K Pot → 2.6 K / 5.5 mbar |
| 1025 | Open 1K Pot needle valve → 30°. 1K Pot → 2.4 K / 11 mbar |
| 1030 | Tower = 6.7 K, Condenser = 490 mbar, both falling |
| 1030 | Open 1K Pot needle valve → 40°. 1K Pot → 2.1 K / 16 mbar |
| 1035 | Open 1K Pot needle valve → 50°. 1K Pot → 1.8 K / 24 mbar |
| 1035 | Tower = 3.5 K, Condenser = 433 mbar, both falling |
| 1040 | Open 1K Pot needle valve → 60°. 1K Pot → 1.7 K / 36 mbar |
| 1055 | Tower = 2.6 K, Condenser = 198 mbar, both falling. 1K Pot = 1.66 K / 34 mbar |
| 1115 | Tower = 2.3 K, Condenser = 106 mbar, flattish. 1K Pot = 1.55 K / 9.5 mbar |
| 1115 | Close V3 and V9, leaving 82 mbar in the dump. |
| 1120 | Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 235 mbar. Tower falls to 2.15 K. Small response in 1K Pot |
| 1125 | 1K Pot spontaneously jumps → 1.65 K / 34 mbar |
| 1130 | V6 → 13.2%. Condenser peaks at 225 mbar. Tower flat at 2.2 K |
| 1145 | Stop DAQ programs and turn off DCRCs. Tower temperature doesn't change much. |
| 1150 | V6 → 13.5%. Condenser peaks at 460 mbar. Tower → 1.65 K |
| 1215 | V6 → 13.8%. Condenser peaks at 505 mbar. Still → 6.5 mbar. Tower → 1.34 K |
| 1240 | V6 → 14.2%. Condenser flattens at 495 mbar. Still → 3.0 mbar. Tower → 1.28 K |
| 1305 | V6 → 15.0%. Condenser flattens at 515 mbar. Still → 0.75 mbar. Tower → 1.31 K |
| 1320 | V6 → 100%. Condenser flattens at 55 mbar. Still → 0.13 mbar. Tower → 0.97 K |
| 1330 | Still = 1.40 K / 0.14 mbar. Turn on Still heater to 2 mW. Still → 1.42 K / 0.15 mbar. |
| 1345 | Increase Still heater to 5 mW. Still → 1.45 K / 0.17 mbar. |
| 1350 | Switch back to Tower thermometer: Tower has fallen to 202 mK! that's a surprise… But it's rising, reaches 210 mK in a few minutes. |
| 1405 | Tower = 255 mbar and rising very steadily, strange. |
| 1415 | Try lowering Still heat: temperature goes up faster. Try raising Still heat: temperature goes up more slowly. |
| 1420 | At Still = 20 mW, Still → 1.55 K / 0.20 mbar. Tower continues to get warmer, now 314 mK |
| 1430 | Tower 430 mbar and rising quickly. Modifying still heat doesn't seem to matter much. Condenser = 740 mbar, goes up a bit with still heat applied. |
| 1430 | Close V13a, open V9. Now pumping to dump. |
| 1435 | Open V5. Now pumping both sides of fridge to dump. Set MC heater to 20 mW. Tower → 1.5 K. |
| 1440 | Close 1K Pot needle valve and speedivalve. |
| 1555 | Mixture mostly pumped out to dump. Check contents of warm LN2 Trap 2: 13 mbar. Pump out with 4He pump. |
| Throughput test procedure, Cold Trap 1: |
- All mixture in dump, LN2 traps cold, condenser & still valves open (3He pump on - optional)
- Open V9, V13a, V12a+speedivalve, V3. Wait for G1 to be close to G2.
- Close V12a+speedivalve and V3, open V1
- monitor G1
| 1610 | Perform mixture throughput test at MC = 9.8 K using above procedure with 3He pump on. Let G1 rise to 560 mbar before Step 3, then it falls to 240 mbar and starts to rise. Throughput = 116 mbar/min. Is this good? Certainly is not zero. It's more than twice a typical 77K rate. Not sure what it should be at 10K. |
|---|---|
| 1630 | Turn off MC and Still heaters |
| 1635 | Open 1K Pot speedivalve to start pumping on pot. Pot → 1.9 K / 400 mbar, then → 4.0 K / 0.95 mbar Open 1K Pot needle valve → 15°. 1K Pot → 3.2 K / 4.5 mbar |
| 1640 | Open 1K Pot needle valve → 30°. 1K Pot → 2.6 K / 12 mbar Open 1K Pot needle valve → 45°. 1K Pot → 2.1 K / 24 mbar |
| 1645 | Open 1K Pot needle valve → 60°. 1K Pot → 1.7 K / 50 mbar |
| 1700 | Tower = 2.95 K, Condenser = 300 mbar, both falling |
| 1705 | Started to pump mixture to dump to try another throughput test. But mixture is partly condensed so this won't work. Open mixture to fridge again. |
| 1710 | Fill LN2 cold trap bath. |
| 1720 | Tower = 2.53 K, Condenser = 131 mbar, both falling. 1K Pot = 1.67 K / 40 mbar, falling ever since returning the mixture (which caused a spike) |
| 1730 | Tower = 2.3 K, Condenser = 84 mbar, both falling. 1K Pot = 1.61 K / 22 mbar Open 1K Pot needle valve → 70°. 1K Pot → 1.62 K / 24 mbar |
| 1735 | Open 1K Pot needle valve → 80°. 1K Pot → 1.62 K / 24 mbar shortly after it spikes → 1.74 K / off scale |
| 1740 | Close 1K Pot needle valve → 70°. 1K Pot → 1.72 K / 200 mbar |
| 1745 | Close 1K Pot needle valve → 65°. 1K Pot → 1.71 K / 100 mbar |
| 1730 | Tower = 2.1 K, Condenser = 81 mbar |
| 1820 | Tower = 2.0 K, Condenser = 68 mbar, 1K Pot = 1.71 K / 100 mbar Close V3. Close V9 leaving 47 mbar of mixture in dump. |
| 1820 | Start circulation through Trap 1. V6 → 12.8%. Condenser peaks at 145 mbar. |
| 1830 | V6 → 13.2%. Condenser peaks at 195 mbar. Tower → 1.8 K |
| 1930 | Tower = 1.7 K. Condenser = 127 mbar and falling. |
| 2140 | Tower = 1.7 K. Condenser = 67 mbar and falling. |
| 2300 | Tower = 1.8 K. Condenser = 60 mbar. V6 → 13.6%. Looks like the condenser will peak around 350 mbar. |