2018/03/07 (Wednesday)

Previous Day

Status: started 1K Pot yesterday, it kept running dry. DIB1/2 connected to DCRC50(3)/DCRC55(1).
0845 MC(Cernox) = 3.44 K, Sorb = 4.0 K, Turbo (on IVC) = 3.4 nbar, G1 (fridge) = 330 mbar, Main bath = 70.7%
0845 1K Pot = 2.26 K / 5.5 mbar
overnight, there was one period a couple hours long where the 1K Pot and Tower temperatures and the fridge pressure were all elevated. Then they all came back down.
0850 Open Needle Valve a few more degrees. 1K Pot → 1.8 K / 200 mbar
0855 Fill LN2 cold trap bath
0855 Tower = 3.25 K, Fridge (G1) = 236 mbar. Both are falling. Tower is lower than at any time yesterday.
0905 Tower = 3.15 K, Fridge (G1) = 186 mbar
0905 1K Pot = 1.69 K / 26 mbar
Open Needle Valve a few more degrees, couple times. 1K Pot → 1.7 K / 36 mbar
0940 Start 5 min series of 5V data (DCRC bias), with 1000 beginning randoms, and a 700/100/150/150 bin threshold trigger on DIB1qI/qO/DIB2qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
0944 Logger seems not be to be running. Stopped run. Restart midas.
0947 Start 30 min series of 5V data (DCRC bias), with 1000 beginning randoms, and a 700/100/150/150 bin threshold trigger on DIB1qI/qO/DIB2qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
0950 Tower = 3.14 K, Fridge (G1) = 162 mbar
0955 1K Pot = 1.61 K / 8.5 mbar
Open Needle Valve a few more degrees, couple times. 1K Pot pressure rises off scale, but comes back down again.
1023 Start 30 min series of 10V data (DCRC bias), with 1000 beginning randoms, and a 700/100/150/150 bin threshold trigger on DIB1qI/qO/DIB2qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
1033 1K Pot = 1.61 K / 8.0 mbar
Open Needle Valve a few more degrees. 1K Pot pressure rises to 30mbar.
1110 Tower = 3.11 K, Fridge (G1) = 158 mbar
1110 1K Pot = 1.6 K / 7.5 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.74 K / 50 mbar.
1130 1K Pot = 1.61 K / 9.0 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.74 K / 50 mbar.
1200 1K Pot = 1.58 K / 6.5 mbar. Fridge pressure has been decreasing but now leveling out.
Open Needle Valve a few more degrees. 1K Pot → 1.69 K / 24 mbar.
We got the best decrease in fridge pressure at 11:30. At that time the needle valve was opened enough to cause the 1K Pot pressure to go off scale temporarily
1220 1K Pot = 1.59 K / 7.0 mbar
Open Needle Valve further several times, until 1K Pot pressure goes off scale. 1K Pot → 1.80 K.
OK, we'll start trying to circulate. Will open needle valve occasionally to keep 1K Pot pressure above 20 mbar.
1330 Close V9 leaving 143 mbar in dump. Close V3. Turn on 3He pump, which will push a little more mix into condenser.
1335 Start circulation through Trap 2. V6 → 12.8%. Condenser peaks at 300 mbar. 1K Pot pressure spikes off scale then falls. Tower temperature falling quickly, <2.6 K.
1340 V6 → 13.2%. Condenser peaks at 170 mbar. Tower temperature seems to be varying with 1K Pot - hit a low of 2.1 K but came back up to 3.4 K
1350 Condenser pressure increasing apparently because 1K Pot ran dry. Open needle valve more. Wait for condenser to come down, currently at 345 mbar
1355 Spike in 1K Pot causes Condenser pressure and Tower temperature to start falling again. Condenser → 304 mbar.
1355 V6 → 13.5%. Condenser → 500 mbar. Tower → 3.05 K. Condenser plateaus.
1355 V6 → 13.8%. Condenser → 630 mbar. Tower → 3.10 K.
1405 While I was away from the fridge, Still pressure fell as low as 50 mbar and Tower temperature fell to 1.5 K. But then the 1K Pot ran dry and became hot, then Still & Condenser pressures rose and Tower temperature rose.
1410 Open 1K Pot needle valve some more to get it going again
1415 Condenser is pretty high, 856 mbar, but 1K Pot is working and Tower is cooling again. Hopefully condenser will start falling.
1420 Condenser = 823 mbar and falling.
1420 V6 → 14.0%. Condenser peaks at 875 mbar.
1430 V6 → 14.2%. Condenser peaks at 920 mbar. Still pressure falls, → 200 mbar
1435 V6 → 14.4%. Condenser peaks at 930 mbar. Still pressure falls, → 60 mbar

This is beginning to feel like a losing battle. We have to keep opening the needle valve more and more - the pot cools to 1.8 K, stays there for a few minutes, then rises immediately to around 2.5 K accompanied by a falling pressure. When the pot goes dry the condenser pressure goes up too, not surprisingly. Tower has been pretty flat at 2.6 K.

Is there too much helium in the IVC? The temperatures are bouncing aroudn the lambda point, roughly. Maybe superfluid He climbs up to the pot, warms it, then gets heated above superfluid and goes away again.

1450 Still = 3.3 K, Coldplate = 3.1 K. Not good…
1450 Condenser = 971 mbar. Too high, can't keep it down. Close V6 to let mixture expand into Still.
The Turbo pump on the IVC has generally been around 3 nbar. But watching it while the 1K Pot warmed up again, it rose to 5.4 nbar. Should close the Turbo valve, then turn off the pump and let things warm to 4K, see how high the IVC pressure goes.
1455 Open needle valve more to get pot working again, 1K Pot → 1.8 K as usual
1455 Turbo on IVC → 4.6 nbar. Close valve to turbo pump.
1500 IVC = 86 nbar. 1K Pot = 1.71 K. Close V4a to stop pumping on 1K Pot.
1505 1K Pot has risen to 3.1 K. Tower = 3.5 K. IVC = 120 nbar, not a dramatic increase.
1505 1K Pot = 3.7 K. Tower = 3.76 K. IVC = 160 nbar.

can compare with Run 59. In that run we were repeatedly warming the tower to test the mu-metal shield, and we also had a leak from bath to IVC, which we could see when we warmed up - accumulated He in the IVC would be seen at the pressure gauge when we were warm. We have a similar amount of gas in the IVC now. One difference with Run 59 is that then we were leaving the 1K Pot on when we warmed up.

1515 1K Pot = 3.9 K. Tower = 3.90 K. IVC = 220 nbar. Pump out IVC with turbo pump.
1520 Close 1K Pot needle valve. We don't want the pot too full if/when we start pumping on it again.
1520 Turbo = 10 nbar. Close turbopump valve to check IVC pressure. IVC = 140 nbar. A bit lower. Open turbopump again to continue pumping on IVC.
1540 1K Pot = 4.1 K. Tower = 4.25 K. Turbo = 5.4 nbar.
Open V4a to pump on 1K Pot. (Needle valve is closed.) 1K Pot falls quickly to 1.8 K then rises when it empties - Tower and Fridge pressures fall too.
1545 Open needle valve enough to get 1K Pot pressure up to 50 mbar. Then adjust to keep temperature ⇐ 1.8 K.
1610 Have been increasing the needle valve periodically to keep the 1K Pot above 10 mbar. Meanwhile mixture in fridge is condensing again and Tower temperature is coming down. But it doesn't look like any improvement over this morning.
1635 Situation is similar to earlier before we started circulation: Fridge (G1) = 165 mbar, Tower = 3.12 K, Pot pressure keeps running low. I don't think things are getting better…
… so let's stop the 1K Pot and let things settle to 4K, and also stop pumping on the IVC to see how high the pressure goes up overnight.
1640 Close turbo pump valve. IVC = 74 nbar.
1645 Close 1K Pot needle valve and speedivalve. Close V4a and 4He pump valve, turn off 4He pump.
1650 Fill LN2 cold trap bath
1655 1K Pot = 3.0 K, Tower = 3.5 K, IVC = 120 nbar