2018/03/06 (Tuesday)

Previous Day

Status: First LHe transfer yesterday. Baking detector with all 4 LEDs overnight, at 4K. DIB1/2 connected to DCRC50(3)/DCRC55(1).
0856 Fill LN2 cold trap bath
1015 MC(Cernox) = 4.34 K, Sorb = 5.9 K, 1K Pot = 4.68 K, Turbo (on IVC) = 14 nbar, G1 (fridge) = 539 mbar, Main bath = 44.7%
1020 Turn off all 4 LEDs at the ODB.
1030 Set a +5V bias on all 4 channels. Look for reasonable threshold settings. DCRC3qI (DIB1qI) is clearly the noisiest channel, needs a much higher threshold.
1040 Let's take some data here at 4K. We can compare noise to what we get at base temperature, see if there's a difference. Do we even have to cool to base?
1042 Start 30min series of 5V data (DCRC bias), with 1000 beginning randoms, and a 700/100/150/150 bin threshold trigger on DIB1qI/qO/DIB2qI/qO, with 1.6384 ms pulses (8192 length) and 400 us prepulse (500 length).
1105 Pump out line leading to 1K Pot, then open 1K Pot speedivalve. (Note that 1K Pot was evacuated before LHe transfer and needle valve was closed.) About 0.23 bar in 1K Pot. Pot temperature dips to 3.6 K when valve opened, then rises again. Pump out for a few seconds with scroll pump, then switch to 4He pump. Open the two valves to connect 4He exhaust to recovery.
1115 Set detector bias to 0 V after ending data series
1115 1K Pot = 4.70 K / 1.9 mbar
Open 1K Pot Needle Valve to ~10 degrees. 1K Pot → 3.3 K / 40 mbar. After a few minutes, pressure falls to 2.4 mbar.
1120 Open Needle Valve a few more degrees. 1K Pot → 3.3 K / 3.0 mbar
Open Needle Valve a few more degrees. 1K Pot → 3.2 K / 4.0 mbar
Open Needle Valve a few more degrees. 1K Pot → 3.10K / 5.0 mbar
Open Needle Valve a few more degrees. 1K Pot → 2.9 K / 6.0 mbar
1125 Open Needle Valve a few more degrees. 1K Pot → 2.6 K / 10 mbar, some oscillations
1130 Tower temp and Fridge pressure falling. Tower = 4.20 K, G1(fridge) = 490 mbar
1130 Open Needle Valve a few more degrees. 1K Pot → 2.4 K / 16 mbar
Open Needle Valve a few more degrees. 1K Pot → 2.1 K / 22 mbar
1135 Tower temp falling faster, Tower = 3.76 K and falling. (Tower temperature drop coincides with flat period in 1KPot pressure, oscillations gone)
1135 Open Needle Valve a few more degrees. 1K Pot → 1.9 K / 26 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.85 K / 36 mbar
1140 Open Needle Valve a few more degrees. 1K Pot → 1.8 K / 40 mbar
1145 Tower = 3.67 K (falling), 1K Pot = 1.81 K, Fridge(G1) = 400 mbar (falling)
1210 Tower = 3.60 K (flat), 1K Pot = 1.77 K, Fridge(G1) = 341 mbar (rising a little)
1210 1K Pot pressure has fallen to 15 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.77 K / 26 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.79 K / 36 mbar
1235 Had lunch, and again pot pressure is falling, Pot & Tower temps & fridge pressure rising
1235 Tower = 3.62 K, 1K Pot = 2.2 K, Fridge(G1) = 357 mbar
Open Needle Valve several more degrees. 1K Pot → 1.8 K / 45 mbar
1245 1K Pot pressure is falling again, down to 30 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.79 K / 40 mbar
1305 1K Pot pressure is falling again, down to 22 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.78 K / 34 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.79 K / 38 mbar
1315 Tower = 3.52 K, 1K Pot = 2.8 K, Fridge(G1) = 271 mbar
1325 1K Pot pressure is falling again, down to 24 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.78 K / 34 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.81 K / 55 mbar
1400 1K Pot pressure is falling again, down to 22 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.80 K / 45 mbar
1445 1K Pot pressure is falling again, down to 20 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.79 K / 40 mbar
1445 Tower = 3.47 K, 1K Pot = 1.8 K, Fridge(G1) = 237 mbar
1505 1K Pot pressure is falling again, down to 24 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.78 K / 34 mbar
Open Needle Valve several more degrees. 1K Pot → 1.83 K / 120 mbar
1507 Start LHe transfer (MB = 37.5%)
1510 1K Pot pressure spiked off scale, but it's coming down again, currently 40 mbar
1540 1K Pot pressure is falling again, down to 18 mbar
Open Needle Valve a few more degrees. 1K Pot → 1.78 K / 45 mbar
1605 End LHe transfer (MB = 88.2%)
1650 I opened the 1K Pot Needle valve wider several more times, got tired of writing down the numbers every time, same pattern of behavior, keeps running dry. Cooldown going slowly.
1655 Tower = 3.42 K, 1K Pot = 1.75 K, Fridge(G1) = 220 mbar
1656 Fill LN2 cold trap full.