| 1313 | HV = 74.96V, V_leak = 33.06 mV Note: These are ~10x higher than yesterday. We probably need to control the humidity at the E-box. It's rainy today and the humidity in the lab is around 40%. |
| 1426 | Begin LHe transfer @ 28.5% |
| 1516 | End transfer @ 89.7% |
| 1525 | 1K Pot = 1.60 K / 24 mbar. Tweak needle valve open a little further. 1K Pot → 1.61 K / 26 mbar, only a slight difference. |
| 1535 | As in Run 62 we are sealing a bag around the ebox and running a hose from the LN2 Dewar vent to the bag, with the vent valve open just enough to feel some flow. |
| 1550 | Apply the new QET bias settings from the table above. Set all the Driver Gains to x8. Adjust Driver Offsets to place baselines around -0.2 - -0.3 V. |
| 1628 | Set HV=70V to select reasonable trigger threshold w/HV. Change Driver gains back to x1 to make this easier. Triggering on DCRC1PA with an ADC threshold of 4 just starts to trigger on noise. Let's use a threshold of 10 to be safe. |
| 1637 | HV = 69.93 V, V_leak = 4.64 mV The N2 purge bag seems to be helping! |
| 1639 | Set HV = 0 V |
| 1640 | Fill LN2 cold trap full. |
| 1642 | Flash for 30s |
| 1804 | Start HV scan sequence HVscan-0_30_65_70_75V-15_5PreBias-30Flash-30wait-1hrData.msl Sequence takes data at 0, 30, 65, 70, 75 V HV biases with 15% prebias for 5 min and 30s flash with 30 min wait between each 1 hr series. Trigger = DCRC1PA = 10 ADC, 10000 BOR randoms. DriverGains=1, QETbias = (DCRC1PA/PB/PC)/(DCRC3PB/PC/PD) = table above. Taking extra randoms to get clean PSDs. |
| 1954 | HV = 64.92 V, V_leak = 2.63 mV |