2017/07/11 (Tuesday)

Previous Day

Setup: DIB 1/3/4 → 3.2 MHz filters 1/2/3 → cable 1/2/3 → gender changer 1/3/4 → DCRC55(1)/33(2)/X(4)
0855 Fill LN2 cold trap full.
0850 Tower = 27.3 mK, Still = 0.060 mbar, G1/G2 = 166/159 mbar, 1K Pot = 1.66 K / 20 mbar. MB = 71.8%
0930 Discovered that on DCRC4, Connect Phonon to LED was enabled. Switch this off - now pA is responsive! (Thanks Nick)
1325 Finish erecting North and South wall of lead shield
1329 Fill LN2 cold trap full.
1330 Tower/MC/Coldplate = 28.7/30.8/170 mK, Still = 1.42 K / 0.060 mbar, G1/G2 = 166/162 mbar, 1K Pot = 1.66 K / 20 mbar. MB = 64.7% (Predicted level: 66.0%)
1335 Lab = 23.9 C, Humidity = 42%
1345 Estimate event rate: count 102 events in 1 second → 102 Hz. This is about 46 Hz less than with no shield. In Run 62 it was a 65 Hz reduction, although that also included the top corner lead, which seemed to account for maybe an additional 3-9 Hz reduction.
It's possible that the discrepancy is a matter of threshold: we count pulses that rise noticeably above background, but the HV detector had much peakier events so the threshold would be lower. If we had as low a threshold now maybe all the rates would be higher, and the reduction would also be greater.
If we increase all this run's rates by 28% we get 190/131 Hz without/with shield.
DIB chan Side chan nominal QET bias [uA] DIB chan Side chan nominal QET bias [uA] DIB chan Side chan nominal QET bias [uA]
DIB1 pA Side1 pD 50 DIB3 pA Side2 pA * DIB4 pA Side2 pE 60
DIB1 pB Side1 pF DIB3 pB Side2 pC 75 DIB4 pB Side2 pF 67
DIB1 pC Side1 pE 60 DIB3 pC Side1 pA DIB4 pC Side2 pD 60
DIB1 pD mistake: Side2 pA DIB3 pD Side1 pC 55 DIB4 pD Side1 pB 65
*this is the shorted one, currently can't get any pulses in it
1555 Finish IbIs on DCRC1pA / DIB1pA / Side1pD
1718 Fill LN2 cold trap full.