2017/03/24 (Friday)
| Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33 | |
| Took data at 20 V overnight | |
| 0930 | Tower = 33.0 mK, Main bath = 65.4%. Boil off rate: faster than slow mode (predicted level: 68.2%). |
| 0935 | Fill LN2 cold trap bath. |
| 0935 | HV = 20.0 V, Vtest = 0.7 mV, Vdetector = 19.7 V |
| 0935 | Dial HV down to 0V, ground detector, flash for 30 seconds |
| 1015 | Set HV=90 V. |
| 1020 | Adjust trace lengths in ODB for DCRC1, DCRC2 to 26.2144ms (ChargeWaveFormLength = 131072, PhononWaveFormLength = 65536) |
| 1021 | Set QET biases to values used for earlier noise data: here |
| Channel | Driver Offset |
|---|---|
| DCRC1 PA | -0.59 |
| DCRC1 PB | 0.85 |
| DCRC1 PC | -0.24 |
| DCRC2 PC | 0.66 |
| DCRC2 PD | 0.15 |
| 1023 | Start 30 min 90V HV transition long trace random data set with 26.2144ms trace length, Frontend Gain = (20/20/20)/(-/4/5), OG = 4/2, QET biases=table here, 5Hz in-run randoms with no starting/ending randoms. |
|---|---|
| 1027 | HV = 90V, V_leak = 6.2 mV, V_crystal = 87.3 V |
| 1114 | Ground and flash detector |
| 1200 | Set HV=90 V. |
| Set QET biases to 300uA |
| Channel | Driver Offset |
|---|---|
| DCRC1 PA | -1 |
| DCRC1 PB | 0.5 |
| DCRC1 PC | -0.7 |
| DCRC2 PC | 1.3 |
| DCRC2 PD | 0.9 |
| 1208 | Start 30 min 90V HV normal long trace random data set with 26.2144ms trace length, Frontend Gain = (20/20/20)/(-/4/5), OG = 4/2, QET biases=(300/300/300)/(-/-300/-300), 20Hz in-run randoms with no starting/ending randoms. |
|---|---|
| 1210 | HV = 90V, V_leak = 6.6 mV, V_crystal = 87.1 V |
| 1400 | HV = 90V, V_leak = 6.2 mV, V_crystal = 87.3 V |
| 1400 | Dial HV down to zero and ground detector. |
| 1415 | Flash detector for 30 seconds |
| 1540 | Set HV to 20.00 V |
| 1545 | HV = 20.0 V, Vtest = 0.3 mV → Vdetector = 19.9 V |
| 1548 | Start sequence of 16, 1-hour data series of 1-sided 20V HV data with Frontend Gain = (20/20/20)/(-/4/5), OG = 4/2, QET biases = table from Mar 15, with 1000 beginning randoms, no flashing, and a 15ADC bin threshold trigger on DCRC1 pA |
| 1550 | Tower/MC/Coldplate = 33.1/31.5/189 mK, Still = 1.2 K / 0.110 mbar, G1/G2 = 260/240 mbar, 1K Pot = 1.695 K / 30 mbar, Main bath = 56.1%. |