UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20170325

2017/03/25 (Saturday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
Took data at 20 V overnight
1015 Tower = 33.2 mK, Main bath = 28.7%.
1015 Fill LN2 cold trap bath.
1017 Turn off HV.
1020 Commence 6 hours of baking (DCRC 1,2 with 30 sec on/5 sec off).
1135 Tower temperature stable within 400-455 mK (average around 426 mK)while baking.
1427 Bake* Tower/MC/Coldplate = 408/112/221 mK, Still = 1.3 K / 0.130 mbar, G1/G2 = 292/265 mbar, 1K Pot = 1.694 K / 28 mbar, Main bath = 20.4%. Boil off rate: close to slow mode (predicted level: 21.6%).
1440 Start LHe transfer (MB = 20.0%)
1552 End LHe transfer (MB = 95.4%)
1555 Fill LN2 cold trap full.

Next Day