UMN CDMS Wiki archive
start › cdms › k100 › run_diary › 20170321

2017/03/21 (Tuesday)

Previous Day

Setup: DIB1/3/4 → 3.2 MHz filters 3/1/2 → cable 3/1/2 →gender changer 3/1/2 → HVConn 1/3/2 → HV 1/2 → DCRC55/DCRC33
0904 Tower/MC/Coldplate = 33.5/30.7/191 mK, Still = 1.2 K / 0.110 mbar, G1/G2 = 256/224 mbar, 1K Pot = 1.676 K / 24 mbar, Main bath = 71.7%. Boil off rate: faster than slow mode (predicted level: 76.0%).
0907 Fill LN2 cold trap bath.
1058 Set HV = 60 V
1100 Set QET biases to 0 uA
Channel Driver Offset
DCRC1 PA -1
DCRC1 PB -0
DCRC1 PC -0
DCRC2 PC 1
DCRC2 PD -0.4
1104 Start 30 min 60V HV superconducting long trace random data set with 26.2144ms trace length, Frontend Gain = (20/20/20)/(-/4/5), OG = 4/2, QET biases=(0/0/0)/(-,0,0), 20Hz in-run randoms with no starting/ending randoms.
1107 HV = 60V, V_leak = 4.4 mV, V_crystal = 58.1 V
1150 Ground and flash detector
1337 Set HV = 90V
Channel Driver Offset
DCRC1 PA -1
DCRC1 PB -0
DCRC1 PC -0
DCRC2 PC 1.15
DCRC2 PD -0.15
1342 Start 30 min 90V HV superconducting long trace random data set with 26.2144ms trace length, Frontend Gain = (20/20/20)/(-/4/5), OG = 4/2, QET biases=(0/0/0)/(-,0,0), 20Hz in-run randoms with no starting/ending randoms.
1343 HV = 90V, V_leak = 9.6 mV, V_crystal = 85.8 V
1419 Ground and flash
1720 Tweak 1K Pot needle valve open a few more degrees. Pressure peaks at 36 mbar then falls back to 28 mbar, oscillates a little.
1738 Fill LN2 cold trap bath.
1745 Set HV to 10.01 V. After about 5 minutes Vtest is down t0 0.6 mV, so Vdetector = 9.75 V
1756 Start sequence of 14, 1-hour data series of 1-sided 10V HV data with Frontend Gain = (20/20/20)/(-/4/4), OG = 4/2, QET biases = table from Friday, with 1000 beginning randoms, no flashing, and a 15ADC bin threshold trigger on DCRC1 pA

Next Day