Table of Contents
Run 5
Summary
- Placed two Am-241 sources of activity 37 kBq in source holder mounted in tower.
- Collimation performed by means of a lead disk of thickness 1/16“ and outer diameter 1”, with a hole of diameter 0.008“ drilled through the its centre.
- Hole covered with Cu tape to prevent alphas from contaminating the data.
- Unable to see peaks at energies < 30 keV due to copper's high atomic number (Z = 29) and density (8.92 g/cm³).
- Lesson: use Al tape in future (Z = 13, density = 2.7 g/cm³).
- Tower assembled using threaded rods and side coax 4.
- One connector found to have a damaged pin.
- One connector gave strange results when an electrical check was performed.
- Both sides have been labelled so as not to be used again.
- Base temperature is slightly higher than before at 40 mK.
- Could not get DCRCv2 to work for phonon channels, so switched to DCRCv1.
- Known problem in DCRCv2, which will be fixed in DCRCv3.
- 8 keV line from Cu fluorescence not visible due to higher energy threshold of DCRCv1.
- Performed overnight (approximately 11-hour) Am-241 run on 28/29 July 2010.
- See 60 keV line clearly and position of sources through reconstruction.
- No degredation in neutralisation.
- Ambient neutron rate is 2.5 ± 0.3 mHz.
- Performed 1-hour Cf-252 (& Am-241) runs to test shielding effects of polyethylene sheets on 29 July 2010.
- Tested shielding thicknesses of 0”, 2“, 4” & 8“.
- Neutrons selected with Total Phonon Energy (Pt) between 20 and 60 keV, and Yield 0 to 0.6.
- 8” of polyethylene reduces neutron flux by a factor of 6.
- Drilled holes in polyehthylene shield and line-of-sight only coverage of shielding between detector and source means these values are fairly conservative.
- Performed overnight (approximately 8-hour) Ba-133 (& Am-241) run on 30/31 July 2010.
- Ideal position of Ba-133 source is 13“ below top of metal shield.
- Am-241 peak visible at 60 keV.
- Ba-133 peaks visible at 303 keV, 356 keV and 384 keV.
- No degredation in neutralisation.
- Ambient neutron rate is 2.1 ± 0.2 mHz, which agrees within errors with that obtained from the overnight Am-241 run.
Setup
| Detector | G9 |
|---|---|
| Tower | T1 |
| Face | 6 (ZB-7), SQUID gusset holes are stripped |
| Side coax | 4 |
| SQUET | ZC15 |
| Stripline | #20 |
| Readout | DCRCv1 |
Data Summary & Cuts
The eps, matlab fig and png files of all the following plots can be found here
Cuts
% position cut cQin = 0.93 <= Qpart & Qpart <=1.07; cDel = (-50 < Xdel & Xdel <50) & (-50 < Ydel & Ydel <50 ); cAmDel1 = ((-7.86 < Xdel & Xdel < -1) & (-29.1 < Ydel & Ydel < -20.5)); cAmDel2 = ((3.4 < Xdel & Xdel < 10.5) & (15.7 < Ydel & Ydel < 25.5)); cAmDel = cAmDel1 | cAmDel2; % fitting quality cut cQchi2 = QSOFchisq(detnum,0) < 9000 + Qsum.^2*0.21; % energy cut cE = 5 < Qsum & Qsum < 100; cPt = Pt > 5; cQsum = Qsum >5;
Plot cut summary
| Plot | Cut |
|---|---|
| Qpart logstep | cDel & cPt & cQsum |
| Qsum linstep | cQin & cDel & cQchi2 & cPt |
| (Qsum, QSOFchisq) | cDel & cQin |
| (Qi, Qo) | cDel & cPt |
| (Xdel, Qi) | cDel & cQin & cQchi2 & cPt |
| Pt overall calibration | cDel & cQin & cQchi2 & cQsum & cPt |
| (Pt, Qsum) | cDel & cQin & cQchi2 |
| Phonon relative calibration | No cut |
| (Xdel, Ydel) | cQin & cQchi2 & cE(55<Qsum<65)', 'cQin & cQchi2 & cE & cAmDel' |
| (Xpart, Ypart) | 'cc = cDel & cQin & cQchi2 & cE', 'cc = cc & cAmDel' |
| Yield plot (Pt, Y=Qsum/Pt) | cDel & cQin & cQchi2 |
| Neutralization plot (Time, Y) | cAmDel & cQin & cQchi2 & cE(30<Qsum<90) |
| Neutralization plot (Time, Qsum) | cAmDel & cQin & cQchi2 |
Deposited Energy Spectra
| Figure 1: Deposited energy spectrum from overnight Am-241 run |
|---|
![]() |
| Matlab fig, eps, data set 07100728_2249: link to all plots |
| Figure 2: Position reconstruction of Am-241 sources |
|---|
![]() |
| Matlab fig, eps, data set 07100728_2249: link to all plots |
![]() |
| Matlab fig, eps, data set 07100728_2249: link to all plots |
| Figure 3: Deposited energy spectrum after cut on position of Am-241 sources |
|---|
![]() |
| Matlab fig, eps, data set 07100728_2249: link to all plots |
| Figure 4: Deposited energy spectrum from overnight Ba-133 & Am-241 run |
|---|
![]() |
| Matlab fig, eps, data set 07100731_0111: link to all plots |
Neutralisation Tests
| Figure 5: Qsum & Yield as a function of time over 11-hour Am-241 run |
|---|
![]() |
| Matlab fig, eps, data set 07100728_2249: link to all plots |
![]() |
| Matlab fig, eps, data set 07100728_2249: link to all plots |
| Figure 6: Qsum & Yield as a function of time over 8-hour Ba-133 & Am-241 run |
|---|
![]() |
| Matlab fig, eps, data set 07100731_0111: link to all plots |
![]() |
| Matlab fig, eps, data set 07100731_0111: link to all plots |
Polyethylene Neutron Shielding Tests
Neutron rates and fluxes in Figure 7 obtained with data taken in 1 hour Cf-252 runs with 0", 2", 4", & 8" of polyethylene shielding. Ambient neutron rate and flux obtained with data from the overnight Am-241 run.
| Figure 7: Neutron rate and flux as a function of polyethylene neutron shield thickness |
|---|
![]() |
| Matlab fig, eps |
![]() |
| Matlab fig, eps |
Neutrons are selected as events with Total Phonon Energy (Pt) between 20 and 60 keV, and Yield 0 to 0.6. An ambient neutron rate of 2.5 ± 0.3 mHz is found from the overnight Am-241 run:
| Figure 8: Yield Y as a function of Total Phonon Energy Pt from overnight Am-241 run |
|---|
![]() |
| Matlab fig, eps, data set 07100728_2249: link to all plots |
The ambient neutron rate obtained from the overnight Ba-133 and Am-241 run of 2.1 ± 0.2 mHz lies within errors:
| Figure 9: Yield Y as a function of Total Phonon Energy Pt from overnight Ba-133 & Am-241 run |
|---|
![]() |
| Matlab fig, eps, data set 07100731_0111: link to all plots |
Polyethylene sheets of approximate height 6” and width 2' were placed between the source and detector such that the line-of-sight was covered. However, neutrons, unlike gamma-rays, do not necessarily follow a straight path to the detector. Additionally, drilled holes in the polyethylene also means that shielding does not provide a uniform coverage. Therefore, the reduction in neutron rate by a factor of 6 with 8“ of polyethylene shielding is fairly conservative, and hence should be assumed to be a lower limit. The neutron castle will better cover the detector, and hence will improve this result.
| Figure 10: Yield Y as a function of Total Phonon Energy Pt from 1 hour Cf-252 & Am-241 runs with 0”, 2“, 4” & 8“ of polyethylene shielding |
|---|
![]() |
| Matlab fig, eps, data set 07100729_2305: link to all plots |
![]() |
| Matlab fig, eps, data set 07100729_2138: link to all plots |
![]() |
| Matlab fig, eps, data set 07100729_1856: link to all plots |
![]() |
| Matlab fig, eps, data set 07100729_2305: link to all plots |