2023/10/03 (Tuesday)

Previous Day

  • Lab in-personnel: Zach, Himangshu
1210 Check contents of warm LN2 Trap 1: 15.2 mbar. Pump it out with the scroll pump.
1213 G1 = 2.7 mbar and stable, stop pumping on trap 1.
1215 Insert cold trap 1 into the LN2 bath, fill bath
1216 Turn on the 3He pump
1217 Begin cryocleaning through cold trap 1 following procedure Cryo-Clean Mixture
1218 Open V6 to 11%
1222 Open V6 to 11.9% to restart flow
1226 Open V6 to 12.0% to restart flow
1231 Open V6 to 12.1% to restart flow
1254 Open V6 to 12.2% to restart flow
1330 Stop cryocleaning through cold trap 1, pump mixture to dump
1400 G1 = 1.7, P1 =.045 mbar, close all valves, leaving 714.5 mbar in the dump
1400 Turn off 3He pump and pull cold trap 1 out of the LN2 bath