2023/10/02 (Monday)
- Lab in-personnel: Himangshu, Zach
| Run 78 Tower Electrical Checkout | |||
|---|---|---|---|
| Device | Label | Unit | Value |
| Face5pA | Side1pE | Ω | 326.1 |
| Face5pB | Side1pF | Ω | 404.9 |
| Face5pC | Side1pA | Ω | 97.5 |
| Face5pD | Side2pE | Ω | 318.7 |
| Face5pA-pB | same side inter | kΩ | 0.855 |
| Face5pA-pC | same side inter | kΩ | 0.675 |
| Face5pA-pD | opp side inter | kΩ | 2.112 |
| Face5pB-pC | same side inter | Ω | 510 |
| Face5pB-pD | opp side inter | kΩ | 2.035 |
| Face5pC-pD | opp side inter | kΩ | 1.872 |
| Face5LED1 | pin-to-chassis | V | 1.068 |
| Face5LED2 | pin-to-chassis | V | 1.639 |
| Face3pA | Side2pC | Ω | 375.4 |
| Face3pB | Side2pF | Ω | 513.6 |
| Face3pC | Side1pC | Ω | 436.6 |
| Face3pD | Side1pD | Ω | 401.9 |
| Face3pA-pB | same side inter | kΩ | 1.707 |
| Face3pA-pC | opp side inter | kΩ | 2.552 |
| Face3pA-pD | opp side inter | kΩ | 2.279 |
| Face3pB-pC | opp side inter | kΩ | 2.626 |
| Face3pB-pD | opp side inter | kΩ | 2.341 |
| Face3pC-pD | same side inter | kΩ | 1.383 |
| Face2pA | Side2pB | Ω | 140.5 |
| Face2pB | Side2pA | Ω | 87.0 |
| Face2pC | Side1pB | Ω | 144.4 |
| Face2pD | Side2pD | Ω | 386.7 |
| Face2pA-pB | same side inter | kΩ | 0.734 |
| Face2pA-pC | opp side inter | kΩ | 1.654 |
| Face2pA-pD | same side inter | kΩ | 1.179 |
| Face2pB-pC | opp side inter | kΩ | 1.789 |
| Face2pB-pD | same side inter | kΩ | 1.462 |
| Face2pC-pD | opp side inter | kΩ | 1.927 |
| Face2pA-chassis | detector shorts | Ω | |
| Face6 1-2 | Motor CoilY 1-2 | kΩ | 113 |
| Face6 3-4 | Motor CoilY 3-4 | kΩ | 112.7 |
| Face66-7 | Motor CoilX 3-4 | kΩ | 111.8 |
| Face6 8_9 | Motor CoilX 3-4 | kΩ | 110.5 |
| Face6 6-chassis | Near Endstop | kΩ | 5.685 |
| Face6 7-chassis | Far Endstop | kΩ | 22.48 |
| Face6 12-13 | Rox | kΩ | 1.072 |
| 1146 | Tower assembly complete |
|---|---|
| 1425 | Tower orientation: Face 6 towards East(Stripline) |
| Finish bolts attaching 30mK adapter to 30mK can and torque them. | |
| Put cap on bottom hole of Tower. | |
| Attach Outer first, then inner shield piece. We use medium-sized screws for the heatsinks outer screw-holes, longer screws for inner screwholes of heatsinks and short screws for other screwholes to complete the 100mK shield. Completed installing shields for the 100mK Can. |
| 1607 | Check contents of warm LN2 Trap 1: 20.8 mbar. Pump it out with the scroll pump. |
|---|---|
| 1610 | G1 = 2.6 mbar and stable, stop pumping on trap 1. |
| 1612 | Insert cold trap 1 into the LN2 bath, fill bath |
| 1613 | Begin cryocleaning through cold trap 1 following procedure Cryo-Clean Mixture |
| 1613 | Turn on the 3He pump |
| I think the 3He pump is still relatively noisy, should look into adding oil? or whatever needs to be done to maintain it | |
| 1616 | Open V6 to 11% |
| 1620 | Open V6 to 11.9% to restart flow |
| 1624 | Open V6 to 12.1% to restart flow |
| 1649 | Open V6 to 12.2% to restart flow |
| 1725 | Open V6 to 12.3% to restart flow |
| 1745 | Stop cryocleaning through cold trap 1, pump mixture to dump |
| 1800 | G1 = 1.6, P1 =.022 mbar, close all valves, leaving 713.0 mbar in the dump |
| 1805 | Turn off 3He pump and pull cold trap 1 out of the LN2 bath |