| 1310 | Take LN2 dewar to be filled. |
| 1420 | LN2 dewar is filled. |
| 1431 | Check contents of dump by opening V9, 704.7 mbar, turn on 3He pump to warm up pump, close V9. |
| 1433 | G1 is pretty high 72mbar, pump out lines with scroll pump. P1 is 7mbar but leave it |
| 1435 | Check contents of warm LN2 Trap 1: 11.9 mbar. Pump it out with the scroll pump. |
| 1439 | G1 = 2.5 mbar and stable, stop pumping on trap 1. |
| 1440 | Check contents of warm LN2 Trap 2: 7.8 mbar. Pump it out with the scroll pump. |
| 1443 | G1 = 1.6 mbar and stable, stop pumping on trap 2. |
| 1447 | 3He pump became quite noisy, maybe it doesn't like pumping on nothing for this long? Turn it off |
| 1448 | Insert cold trap 1 into the LN2 bath, fill bath |
| 1250 | Begin cryocleaning through cold trap 1 following procedure Cryo-Clean Mixture |
| 1253 | Turn on the 3He pump |
| 1255 | Open V6 to 12% to start flow, G2 rising too fast, reduce V6 to 11% |
| 1258 | Open V6 to 12% to restart flow, then reduce to 11.9% |
| 1311 | Open V6 to 12.2% to restart flow |
| 1608 | Stop cryocleaning through cold trap 1, pump mixture to dump |
| 1620 | G1 = 1.6, P1 =.19 mbar, close all valves, leaving 710.9 mbar in the dump |
| | Barometric pressure = 29.92“ (source is https://w1.weather.gov/data/obhistory/KMSP.html.) G2 = 710.9 mbar. Corrected dump pressure = 710.9 mbar |