start › cdms › k100 › run_diary › 20230417

2023/04/17 (Monday)

Previous Day

  • Lab in-personnel: Himangshu
Refer to this run diary for 1D source mover radioactive source setup information.
Source is at center of the detector. Distance from near endstop: 15.98 mm. Adjusted the position of the carriage to be right at the center, using the centerline scratch marks.
The chosen setup:
Accessories at Face 6, meaning 2D source mover striplines at Face 6.
This puts 2D source mover 0-direction at Face 4.
DIB1 at Face 4 (so source 0-direction is detector -y direction
Phonon channel wiring same as last run, so DIB1 channels connect to Face 5
Accessory interposer will connect at Face 5
1443 Put S104 in Position 1 on tower. DIB1-Face4.
1445 Place source mover housing in Position 2 above detector (facing Side 1). The far endstop is in the +y direction.
Run 78 Tower Electrical Checkout
Device Label Unit Value
Face5pA Side1pE Ω 326.2
Face5pB Side1pF Ω 405.4
Face5pC Side1pA Ω 97.5
Face5pD Side2pE Ω 318.7
Face5pA-pB same side inter kΩ 0.850
Face5pA-pC same side inter kΩ 0.672
Face5pA-pD opp side inter kΩ 2.107
Face5pB-pC same side inter kΩ 0.510
Face5pB-pD opp side inter kΩ 2.032
Face5pC-pD opp side inter kΩ 1.869
Face5LED1 pin-to-chassis V 1.068
Face5LED2 pin-to-chassis V 1.60
Face3pA Side2pC Ω 375.5
Face3pB Side2pF Ω 514.0
Face3pC Side1pC Ω 437.7
Face3pD Side1pD Ω 403.2
Face3pA-pB same side inter kΩ 1.707
Face3pA-pC opp side inter kΩ 2.561
Face3pA-pD opp side inter kΩ 2.293
Face3pB-pC opp side inter kΩ 2.643
Face3pB-pD opp side inter kΩ 2.363
Face3pC-pD same side inter kΩ 1.380
Face2pA Side2pB Ω 140.5
Face2pB Side2pA Ω 87.0
Face2pC Side1pB Ω 143.4
Face2pD Side2pD Ω 386.0
Face2pA-pB same side inter kΩ 0.734
Face2pA-pC opp side inter kΩ 1.654
Face2pA-pD same side inter kΩ 1.184
Face2pB-pC opp side inter kΩ 1.789
Face2pB-pD same side inter kΩ 1.466
Face2pC-pD opp side inter kΩ 1.927
Face2pA-chassis detector shorts Ω
Face6 1-2 Motor Coil1 1-2 kΩ 7.49
Face6 3-4 Motor Coil2 3-4 kΩ 7.44
Face6 6-chassis Near Endstop kΩ 5.685
Face6 7-chassis Far Endstop kΩ 22.48
Face6 12-13 Rox kΩ 1.072